3M Quick Sterilization Assurance Reference Guide ™
For 3M™ Attest™ Auto-reader 390* and 3M™ Attest™ Rapid Readout Biological Indicators (BIs)**
1 2 3 4 5 6 7
Cool
Allow test BI to cool at least 10 minutes.
Close
While wearing safety glasses, close the cap of the cooled BI by pushing the cap down.
Crush
Hold BI by cap and use “crusher well” to crush the media ampoule.
Tap
Holding BI by cap, tap the bottom of the vial on the counter until media wets spore strip.
Place
Place the activated BI in color-coded incubation well.
Interpret Result
Observe LCD panel under incubating BIs for negative (–) or positive (+) result. The positive control should read positive (+). A positive result for a test (processed) BI, which indicates a sterilization process failure, should be acted upon immediately.
Record
Record result for both test BI (processed) and positive control BI (unprocessed).
Optional Visual pH Color Change Result
Negative • Media remains purple. • A negative pH color change result can be assessed at 24 hours for Attest™ biological indicators 1291 and at 48 hours for Attest™ biological indicators 1292.
Positive • A media color change from purple to yellow indicates a positive result. • The biological indicator may be discarded as soon as the positive result has been observed and documented.