3M
3M Temperature Monitoring Systems
3M Bair Hugger Model 370 Operators Manual Rev 34-8719-5759-2 Aug 2016
Operators Manual
22 Pages
Preview
Page 1
Revision History Revision
Reason for Change
Pages Affected
Date
370096 A
New Revision - First Release
All
May 2012
370096 B
Languages Added
25-155
May 2012
370096 C
Added specifications and added languages
All
Feb 2013
202052 A
Update company address
All
May 2013
34-8714-3113-5
Update product description
34-8719-5759-2
Update to new Bair Hugger Logo and Circular font layout
Mar 2014
All
August 2016
Temperature Monitoring System
Temperature Monitoring System Model 370 Operator’s Manual
1 English... 1
Q Lietuvių Kalba... 101
= Ελληνικά... 21
M Русский... 119
H Polski... 43
d Türkçe... 141
J Magyar... 63
I 傈劥铃161
F Česky... 83
U
TOC / 34-8719-5759-2
... 179
iii
Revision History Revision
Reason for Change
Pages Affected
Date
370096 A
New Revision - First Release
All
May 2012
370096 B
Languages Added
25-155
May 2012
370096 C
Added specifications and added languages
All
Feb 2013
202052 A
Update company address
All
May 2013
34-8714-3113-5
Update product description
34-8719-5759-2
Update to new Bair Hugger Logo and Circular font layout
Mar 2014
All
August 2016
Temperature Monitoring System
Temperature Monitoring System Model 370 Operator’s Manual
1 English... 1
Q Lietuvių Kalba... 101
= Ελληνικά... 21
M Русский... 119
H Polski... 43
d Türkçe... 141
J Magyar... 63
I 傈劥铃161
F Česky... 83
U
TOC / 34-8719-5759-2
... 179
iii
1 Table of Contents Section 1: Introduction ... 3 About this manual ... 3 Intended use ... 3 Safety information and symbols... 3 Explanation of signal word consequences ... 4 Proper use and maintenance ...5 Read before servicing equipment ...5 Section 2: Overview and Operation ... 6 Description of the 3M™ Bair Hugger™ temperature monitoring system ... 6 Back of control unit ... 8 Front of control unit ... 8 Setting up the control unit ... 8 Instructions for use ... 9 To change temperature unit display (Celsius or Fahrenheit) ...10 Shutting down the control unit ...10 Section 3: Display Panel ... 11 Display screens ... 11 Start-up... 11 Standby ... 11 Ready ... 11 Equilibration ... 11 Running ... 12 Control unit error... 12 Sensor error ... 12 Section 4: Troubleshooting ... 12 On Mode ... 12 Displays ... 13 Errors... 14 Section 5: General Maintenance, Storage and Service... 14 Cleaning Procedure ... 14 Storage ... 14 Service ... 14 Section 6: Technical Service and Order Placement... 15 When you call for technical support ... 15 Technical service and order placement ... 15 Section 7: Specifications ... 15 Physical characteristics...18 Electrical characteristics ...18 Temperature characteristics ...18 Performance characteristics ...18 Environmental conditions ...18 List of Bair Hugger Accessories and Cables ...19
GB / 34-8719-5759-2
1
1 3M™ Bair Hugger™ Temperature Monitoring System Model 370 Operator’s Manual
Section 1: Introduction About this manual ... 3 Intended use ... 3 Safety information and symbols... 3 Explanation of signal word consequences ... 4 Proper use and maintenance ...5 Read before servicing equipment ...5
About this manual This Operator’s Manual describes the setup, use, and maintenance of the temperature monitoring system. The temperature monitoring system and the manual are to be used by healthcare professionals in clinical environments only. Read and follow all instructions, labeling and accompanying documents supplied with this temperature monitoring system. Failure to follow instructions could lead to misuse of the device, device malfunction, or patient injury.
This system is subject to the European WEEE Directive 2002/96/EC. This product contains electrical and electronic components and must not be disposed of using standard refuse collection. Please consult local directives for disposal of electrical and electronic equipment.
Intended use Measure, monitor, and trend body temperature of adult and pediatric patients.
Defibrillation-proof type CF applied part
Safety information and symbols Date of manufacture
Class II equipment
Manufacturer
Authorized Representative in the European Community
Single use only CAUTION: Recycle to avoid environmental contamination This product contains recyclable parts. For information on recycling please contact your nearest 3M Service Center for advice.
Container quantity
CAUTION Direct current
See accompanying documents
3
GB / 34-8719-5759-2
4395459 Description: 3M TM Bair Hugger™ Temperature Monitoring System Model 370 Operator’s Manual - Part 2 Base File Name: PA: 34871957592.indb Book GA: Software: InDesign CC 2014 SGS Contact: Email: Printer: Printer Location: Supplier:
PROCESS BLACK
PROCESS CYAN
PROCESS MAGENTA
Supersedes#:XX-XXXX-XXXX-X Structure#:XX-XXXXX Cat/Product#: Model 370
08.07.16 NEW - PU previous file, convert to Circular fonts.
kmh
09.26.16 09.30.16 10.05.16 10.17.16
DJ kmh kmh kmh
Reference: #### Requester: Debra Staack Die # / Doc. Size: 5.5" x 8.5"
PROCESS YELLOW
PANTONE 280
4
Item Spec#: 34-8719-5759-2
PANTONE 267
PDF Scaled to 85%
Increase FC margin to .625 & Final release. Alts per Proof report PEWR_16698 PW BH Rebranding_MD370_Operator’s Manual Add supplied translations and Final Release Alts per Proof Report and Final Release
1 3M™ Bair Hugger™ Temperature Monitoring System Model 370 Operator’s Manual
Section 1: Introduction About this manual ... 3 Intended use ... 3 Safety information and symbols... 3 Explanation of signal word consequences ... 4 Proper use and maintenance ...5 Read before servicing equipment ...5
About this manual This Operator’s Manual describes the setup, use, and maintenance of the temperature monitoring system. The temperature monitoring system and the manual are to be used by healthcare professionals in clinical environments only. Read and follow all instructions, labeling and accompanying documents supplied with this temperature monitoring system. Failure to follow instructions could lead to misuse of the device, device malfunction, or patient injury.
This system is subject to the European WEEE Directive 2002/96/EC. This product contains electrical and electronic components and must not be disposed of using standard refuse collection. Please consult local directives for disposal of electrical and electronic equipment.
Intended use Measure, monitor, and trend body temperature of adult and pediatric patients.
Defibrillation-proof type CF applied part
Safety information and symbols Date of manufacture
Class II equipment
Manufacturer
Authorized Representative in the European Community
Single use only CAUTION: Recycle to avoid environmental contamination This product contains recyclable parts. For information on recycling please contact your nearest 3M Service Center for advice.
Container quantity
CAUTION Direct current
See accompanying documents
3
GB / 34-8719-5759-2
4395459 Description: 3M TM Bair Hugger™ Temperature Monitoring System Model 370 Operator’s Manual - Part 2 Base File Name: PA: 34871957592.indb Book GA: Software: InDesign CC 2014 SGS Contact: Email: Printer: Printer Location: Supplier:
PROCESS BLACK
PROCESS CYAN
PROCESS MAGENTA
Supersedes#:XX-XXXX-XXXX-X Structure#:XX-XXXXX Cat/Product#: Model 370
08.07.16 NEW - PU previous file, convert to Circular fonts.
kmh
09.26.16 09.30.16 10.05.16 10.17.16
DJ kmh kmh kmh
Reference: #### Requester: Debra Staack Die # / Doc. Size: 5.5" x 8.5"
PROCESS YELLOW
PANTONE 280
4
Item Spec#: 34-8719-5759-2
PANTONE 267
PDF Scaled to 85%
Increase FC margin to .625 & Final release. Alts per Proof report PEWR_16698 PW BH Rebranding_MD370_Operator’s Manual Add supplied translations and Final Release Alts per Proof Report and Final Release
Follow instructions for use
•
Use only the power supply specified for this product and certified for the country of use.
Keep dry
•
Use only a properly grounded power outlet; do not use extension cords or multiple portable socket outlets.
Hardware temperature limits
•
Do not allow the power supply to get wet.
•
Do not use the control unit when it appears the unit or system components are damaged. Contact 3M technical support at 1-800-228-3957.
•
Do not service or modify the power supply, control unit, cables, sensor, or any part of the temperature monitoring system. There are no user serviceable parts.
Sensor temperature limits
Explanation of signal word consequences
! !
WARNING: Indicates a hazardous situation which, if not avoided, could result in death or serious injury. CAUTION: Indicates a hazardous situation which, if not avoided, could result in minor or moderate injury.
NOTICE: Indicates a situation which, if not avoided, could result in property damage only.
!
!
•
The control unit is to be attached to other equipment by authorized service personnel only.
•
Do not re-install or re-locate the control unit.
WARNING: To reduce the risks associated with hazardous voltage, fire, and thermal energy hazards:
•
•
!
!
Do not connect the monitor output of the device to a non-CF rated monitor input.
Keep power supply visible and accessible at all times. The plug on the power supply serves as the disconnect device. The power outlet shall be as close as practical and shall be easily accessible.
The temperature monitoring system is not suitable for use in the presence of a flammable anesthetic mixture with air nitrous oxide.
WARNING: To reduce the risks associated with pressure, and crushed or entangled cords and connectors: •
Do not allow the patient to lie on any sensor cable or connector.
•
Always position cables and cords away from the patient’s body.
•
Do not use a headband or other device to secure the sensor to the patient.
WARNING: To reduce the risks associated with incorrect system or sensor use: •
Limit use of the sensor to 24 hours. Extended use may compromise skin, cause degradation of material or performance.
•
Avoid exposing the temperature monitoring sensor, sensor cable, power supply, and control unit to surgical skin surface preparation solutions or other fluids.
Do not use the Bair Hugger temperature monitoring system (sensor, control unit, cables, or cords) in an MRI environment.
WARNING: To reduce the risks associated with hazardous voltage and fire: •
4
Use ONLY Bair Hugger temperature monitoring system components (sensors, cables, and power supply). Do not substitute other devices for the control unit, sensor, cables, or power supply.
WARNING: To reduce the risks associated with fire: •
WARNING: To reduce the risks associated with access to critical patient information or equipment:
•
!
!
•
Use an additional independent thermometer to measure body temperature during intentional hyperthermia or hypothermia therapy.
•
Confirm unanticipated temperature reading with an independent thermometer, if needed.
•
Do not reposition the sensor; repositioning may weaken the sensor adhesive, damage the sensor, or compromise the device performance.
3M™ Bair Hugger™ Temperature Monitoring System Model 370 – Operator’s Manual
•
!
!
•
Follow facilities policies and procedures for disposal of contaminated materials.
•
Always perform the decontamination procedure prior to returning the temperature monitoring system for service and prior to disposal.
Do not leave pediatric patients unattended while using the temperature monitoring system.
Do not use the sensor on damaged or compromised skin.
CAUTION: To reduce the risks associated with cross-contamination: Clean the sensor cable before connecting to a new sensor.
CAUTION: To reduce the risks associated with impact and facility medical device damage: •
•
!
•
Do not store the control unit and system components in a wet or damp place.
•
Do not spray cleaning solutions onto the control unit or into the sensor cable connector.
•
Do not immerse the control unit or system components in any liquid or subject them to any sterilization process.
•
Do not use solvents such as acetone or thinner to clean the control unit; avoid abrasive cleaners.
•
Clean control unit exterior with a damp soft cloth and 70% mixture of isopropyl alcohol and water. For full cleaning procedures see “Section 5: General Maintenance, Storage and Service” on page 14.
•
Do not immerse any of the devices or use a dripping wet cloth for cleaning.
CAUTION: To reduce the risks associated with skin maceration:
•
!
To avoid temperature monitoring system damage that may impact performance:
WARNING: To reduce the risks associated with entanglement:
•
!
4.
WARNING: To reduce the risks associated with exposure to biohazards:
•
!
Avoid placing the sensor in the center of the forehead as this may affect sensor accuracy.
Do not use the control unit as a handle to transport or to move the device to which it is attached. Do not initiate temperature monitoring unless the control unit is safely placed on a hard, flat surface or is securely mounted.
CAUTION: To reduce the risks associated with environmental contamination: •
Follow applicable regulations when disposing of this device or any of its electronic components.
5.
The sensor is not made with natural rubber latex.
6.
To the full extent permitted by law, the manufacturer and/or importer declines all responsibility for injury resulting from the unit being used in conjunction with unapproved system components.
Proper use and maintenance 3M Health Care assumes no responsibility for the reliability, performance, or safety of the temperature monitoring system if the following events occur: •
Modifications or repairs are performed by unqualified personnel.
•
The control unit is used in a manner other than that described in the Operator’s Manual.
•
The control unit is installed in an environment that does not meet the appropriate electrical requirements.
NOTICES 1.
The temperature monitoring system meets medical electronic interference requirements. If radio frequency interference with other equipment should occur, connect the unit to a different power source.
2.
Protection against electrostatic discharge (ESD) is important when working with the temperature monitoring system. Always be mindful of the immediate work area and any potentially hazardous ESD situations that could impact temperature monitoring system performance.
3.
Federal law (USA) restricts this device to sale by or on the order of a licensed healthcare professional.
GB / 34-8719-5759-2
Read before servicing equipment •
All repair, calibration, and servicing of this equipment must be performed by 3M.
•
There are no user serviceable parts inside the equipment or power supply.
5
Follow instructions for use
•
Use only the power supply specified for this product and certified for the country of use.
Keep dry
•
Use only a properly grounded power outlet; do not use extension cords or multiple portable socket outlets.
Hardware temperature limits
•
Do not allow the power supply to get wet.
•
Do not use the control unit when it appears the unit or system components are damaged. Contact 3M technical support at 1-800-228-3957.
•
Do not service or modify the power supply, control unit, cables, sensor, or any part of the temperature monitoring system. There are no user serviceable parts.
Sensor temperature limits
Explanation of signal word consequences
! !
WARNING: Indicates a hazardous situation which, if not avoided, could result in death or serious injury. CAUTION: Indicates a hazardous situation which, if not avoided, could result in minor or moderate injury.
NOTICE: Indicates a situation which, if not avoided, could result in property damage only.
!
!
•
The control unit is to be attached to other equipment by authorized service personnel only.
•
Do not re-install or re-locate the control unit.
WARNING: To reduce the risks associated with hazardous voltage, fire, and thermal energy hazards:
•
•
!
!
Do not connect the monitor output of the device to a non-CF rated monitor input.
Keep power supply visible and accessible at all times. The plug on the power supply serves as the disconnect device. The power outlet shall be as close as practical and shall be easily accessible.
The temperature monitoring system is not suitable for use in the presence of a flammable anesthetic mixture with air nitrous oxide.
WARNING: To reduce the risks associated with pressure, and crushed or entangled cords and connectors: •
Do not allow the patient to lie on any sensor cable or connector.
•
Always position cables and cords away from the patient’s body.
•
Do not use a headband or other device to secure the sensor to the patient.
WARNING: To reduce the risks associated with incorrect system or sensor use: •
Limit use of the sensor to 24 hours. Extended use may compromise skin, cause degradation of material or performance.
•
Avoid exposing the temperature monitoring sensor, sensor cable, power supply, and control unit to surgical skin surface preparation solutions or other fluids.
Do not use the Bair Hugger temperature monitoring system (sensor, control unit, cables, or cords) in an MRI environment.
WARNING: To reduce the risks associated with hazardous voltage and fire: •
4
Use ONLY Bair Hugger temperature monitoring system components (sensors, cables, and power supply). Do not substitute other devices for the control unit, sensor, cables, or power supply.
WARNING: To reduce the risks associated with fire: •
WARNING: To reduce the risks associated with access to critical patient information or equipment:
•
!
!
•
Use an additional independent thermometer to measure body temperature during intentional hyperthermia or hypothermia therapy.
•
Confirm unanticipated temperature reading with an independent thermometer, if needed.
•
Do not reposition the sensor; repositioning may weaken the sensor adhesive, damage the sensor, or compromise the device performance.
3M™ Bair Hugger™ Temperature Monitoring System Model 370 – Operator’s Manual
•
!
!
•
Follow facilities policies and procedures for disposal of contaminated materials.
•
Always perform the decontamination procedure prior to returning the temperature monitoring system for service and prior to disposal.
Do not leave pediatric patients unattended while using the temperature monitoring system.
Do not use the sensor on damaged or compromised skin.
CAUTION: To reduce the risks associated with cross-contamination: Clean the sensor cable before connecting to a new sensor.
CAUTION: To reduce the risks associated with impact and facility medical device damage: •
•
!
•
Do not store the control unit and system components in a wet or damp place.
•
Do not spray cleaning solutions onto the control unit or into the sensor cable connector.
•
Do not immerse the control unit or system components in any liquid or subject them to any sterilization process.
•
Do not use solvents such as acetone or thinner to clean the control unit; avoid abrasive cleaners.
•
Clean control unit exterior with a damp soft cloth and 70% mixture of isopropyl alcohol and water. For full cleaning procedures see “Section 5: General Maintenance, Storage and Service” on page 14.
•
Do not immerse any of the devices or use a dripping wet cloth for cleaning.
CAUTION: To reduce the risks associated with skin maceration:
•
!
To avoid temperature monitoring system damage that may impact performance:
WARNING: To reduce the risks associated with entanglement:
•
!
4.
WARNING: To reduce the risks associated with exposure to biohazards:
•
!
Avoid placing the sensor in the center of the forehead as this may affect sensor accuracy.
Do not use the control unit as a handle to transport or to move the device to which it is attached. Do not initiate temperature monitoring unless the control unit is safely placed on a hard, flat surface or is securely mounted.
CAUTION: To reduce the risks associated with environmental contamination: •
Follow applicable regulations when disposing of this device or any of its electronic components.
5.
The sensor is not made with natural rubber latex.
6.
To the full extent permitted by law, the manufacturer and/or importer declines all responsibility for injury resulting from the unit being used in conjunction with unapproved system components.
Proper use and maintenance 3M Health Care assumes no responsibility for the reliability, performance, or safety of the temperature monitoring system if the following events occur: •
Modifications or repairs are performed by unqualified personnel.
•
The control unit is used in a manner other than that described in the Operator’s Manual.
•
The control unit is installed in an environment that does not meet the appropriate electrical requirements.
NOTICES 1.
The temperature monitoring system meets medical electronic interference requirements. If radio frequency interference with other equipment should occur, connect the unit to a different power source.
2.
Protection against electrostatic discharge (ESD) is important when working with the temperature monitoring system. Always be mindful of the immediate work area and any potentially hazardous ESD situations that could impact temperature monitoring system performance.
3.
Federal law (USA) restricts this device to sale by or on the order of a licensed healthcare professional.
GB / 34-8719-5759-2
Read before servicing equipment •
All repair, calibration, and servicing of this equipment must be performed by 3M.
•
There are no user serviceable parts inside the equipment or power supply.
5
Section 2: Overview and Operation
Skin Surface
Temperature Sensor
Description of the 3M™ Bair Hugger™ temperature monitoring system ... 6 Back of control unit ... 8
Deep Tissue
1) Placement and connection
Front of control unit ... 8 Setting up the control unit ... 8 Instructions for use ... 9 To change temperature unit display (Celsius or Fahrenheit) ...10 Shutting down the control unit ...10
2) Equilibration - isothermal zone development
Description of the 3M™ Bair Hugger™ temperature monitoring system The 3M Bair Hugger temperature monitoring system measures the core temperature of the patient. The system is comprised of a sensor, which connects to the control unit through the sensor cable, a power supply, and an optional monitor cable. For additional sensors, sensor cable, power supply, control unit stand, monitor cable, hook-and-loop, or foam adhesive, contact 3M.
How does it work? The 3M Bair Hugger temperature monitoring system uses zero-heat-flux thermometry to accurately measure a person’s core temperature as characterized in Figure 2-2 below. 1) The 3M Bair Hugger temperature monitoring system gently warms the sensor creating an isothermal zone under the sensor. 2) Once equilibrated to the core temperature, skin-surface heat loss to the environment is prevented and a zero-heat-flux condition is established. 3) When the temperature sensor reaches equilibrium with the patient’s core temperature, the control unit displays an accurate, noninvasive measurement of the patient’s core temperature.
3) Equilibrated - isothermal zone established Figure 2-2: Isothermal tunnel formation with zero-heat-flux technology
The 3M™ Bair Hugger™ control unit The control unit displays noninvasive core temperature measurements obtained from the temperature sensor. The current temperature is displayed numerically, and previous temperature data is displayed graphically as a temperature trend graph. When the sensor is plugged into the sensor cable, the control unit reads up to the last two hours of stored temperature data and displays the temperature trend graph. Disconnections of the sensor from the sensor cable are indicated by
Monitor Cable
gaps in the temperature trend graph. The current temperature measurement can be continuously transmitted to a patient vital signs monitor through the optional monitor cable using a standard YSI-400 type input. To transmit data from the temperature monitoring system to the patient’s monitor, simply connect the appropriate monitor cable (see Table 2-1: Monitor Cables) to the back of the control unit and to the input on the patient monitor.
Patient Monitor Connector Cables
Power Supply
SO R
3M™ Bair Hugger™ Control Unit
Table 2-1: Monitor Cables
PAT T SEN IEN
Sensor Cable tem
ring nito re Mo atu per Tem
Sys
Temperature Sensor Sensor Connector Figure 2-1: Bair Hugger Temperature Monitoring System
6
3M™ Bair Hugger™ Temperature Monitoring System Model 370 – Operator’s Manual
The 3M™ Bair Hugger™ sensor The sensor is constructed of two layers of medical grade foam and a flexible circuit which contains a resistive warming circuit, two calibrated thermistors, and nonvolatile memory. The information that generates the temperature trend graph is stored on the sensor. If the patient is transferred from one location to another, the sensor is disconnected from the sensor cable so the sensor is left in place on the patient. The sensor may be reconnected
GB / 34-8719-5759-2
to a different sensor cable and control unit when the patient arrives at the next point of care. Once the sensor is reconnected to the sensor cable, the system will begin to reequilibrate and will regenerate the stored temperature trend graph and display the patient’s current temperature. The act of disconnecting and reconnecting the sensor from the sensor cable is indicated by a gap in the temperature trend graph.
7
Section 2: Overview and Operation
Skin Surface
Temperature Sensor
Description of the 3M™ Bair Hugger™ temperature monitoring system ... 6 Back of control unit ... 8
Deep Tissue
1) Placement and connection
Front of control unit ... 8 Setting up the control unit ... 8 Instructions for use ... 9 To change temperature unit display (Celsius or Fahrenheit) ...10 Shutting down the control unit ...10
2) Equilibration - isothermal zone development
Description of the 3M™ Bair Hugger™ temperature monitoring system The 3M Bair Hugger temperature monitoring system measures the core temperature of the patient. The system is comprised of a sensor, which connects to the control unit through the sensor cable, a power supply, and an optional monitor cable. For additional sensors, sensor cable, power supply, control unit stand, monitor cable, hook-and-loop, or foam adhesive, contact 3M.
How does it work? The 3M Bair Hugger temperature monitoring system uses zero-heat-flux thermometry to accurately measure a person’s core temperature as characterized in Figure 2-2 below. 1) The 3M Bair Hugger temperature monitoring system gently warms the sensor creating an isothermal zone under the sensor. 2) Once equilibrated to the core temperature, skin-surface heat loss to the environment is prevented and a zero-heat-flux condition is established. 3) When the temperature sensor reaches equilibrium with the patient’s core temperature, the control unit displays an accurate, noninvasive measurement of the patient’s core temperature.
3) Equilibrated - isothermal zone established Figure 2-2: Isothermal tunnel formation with zero-heat-flux technology
The 3M™ Bair Hugger™ control unit The control unit displays noninvasive core temperature measurements obtained from the temperature sensor. The current temperature is displayed numerically, and previous temperature data is displayed graphically as a temperature trend graph. When the sensor is plugged into the sensor cable, the control unit reads up to the last two hours of stored temperature data and displays the temperature trend graph. Disconnections of the sensor from the sensor cable are indicated by
Monitor Cable
gaps in the temperature trend graph. The current temperature measurement can be continuously transmitted to a patient vital signs monitor through the optional monitor cable using a standard YSI-400 type input. To transmit data from the temperature monitoring system to the patient’s monitor, simply connect the appropriate monitor cable (see Table 2-1: Monitor Cables) to the back of the control unit and to the input on the patient monitor.
Patient Monitor Connector Cables
Power Supply
SO R
3M™ Bair Hugger™ Control Unit
Table 2-1: Monitor Cables
PAT T SEN IEN
Sensor Cable tem
ring nito re Mo atu per Tem
Sys
Temperature Sensor Sensor Connector Figure 2-1: Bair Hugger Temperature Monitoring System
6
3M™ Bair Hugger™ Temperature Monitoring System Model 370 – Operator’s Manual
The 3M™ Bair Hugger™ sensor The sensor is constructed of two layers of medical grade foam and a flexible circuit which contains a resistive warming circuit, two calibrated thermistors, and nonvolatile memory. The information that generates the temperature trend graph is stored on the sensor. If the patient is transferred from one location to another, the sensor is disconnected from the sensor cable so the sensor is left in place on the patient. The sensor may be reconnected
GB / 34-8719-5759-2
to a different sensor cable and control unit when the patient arrives at the next point of care. Once the sensor is reconnected to the sensor cable, the system will begin to reequilibrate and will regenerate the stored temperature trend graph and display the patient’s current temperature. The act of disconnecting and reconnecting the sensor from the sensor cable is indicated by a gap in the temperature trend graph.
7
2.
Use an alcohol wipe to clean and disinfect the patient’s forehead above the orbital ridge. Allow the patient’s forehead to completely dry.
3.
Connect the sensor to the sensor cable connector. Ensure the sensor tab is properly connected and is fully inserted into the sensor cable (Figure 2-5). The display on the control unit will change to the ready screen (See “Ready” on page 11) once it confirms the sensor connection.
Stand Knob Digital Port (used by 3M personnel only)
°C/°F Temperature Display Button
Monitor Cable Port
Power Input Port Figure 2-3: Back of Control Unit
The digital port is used only by 3M Health Care personnel for testing purposes and to access variables within the control unit.
Figure 2-7: Apply the sensor to the patient’s forehead
6.
Gently press the edge of the sensor to ensure good adhesion with the skin. OPTIONAL: Support the sensor cable with the provided sensor cable clip.
7.
The control unit will sense patient attachment after several seconds and the control unit display will change to the equilibration screen (see “Equilibration” on page 11).
8.
The patient’s temperature will display on the patient monitor once equilibrium is reached (approximately three minutes).
Temp eratu re
The monitor cable port provides a resistance that corresponds to a YSI-400 thermistor at the displayed temperature. It is electrically isolated from the control unit.
Moni toring Syste m
Front of control unit
m
oring e Monit eratur Temp
Syste
Figure 2-5: Connect the sensor to the sensor cable
4.
Transferring the patient and sensor 1.
Remove backing liners on the sensor to expose adhesive (Figure 2-6).
2.
Carefully reconnect a sensor cable and control unit to the sensor once transfer is complete.
3.
The control unit will sense patient attachment after several seconds and the control unit display will change to the equilibration screen (see “Equilibration” on page 11).
4.
The patient’s temperature will display on the patient monitor once equilibrium is reached (approximately three minutes).
Setting up the control unit The control unit must be properly installed and mounted by qualified personnel. NOTE: The sensor and control unit are suitable for use in the patient environment. 2.
Connect the power supply to the back of the control unit (see Figure 2-3: Back of Control Unit). Plug the power supply into an appropriate outlet. The standby screen will illuminate.
3.
Connect the sensor cable to the front of the control unit. See Figure 2-4: Front of Control Unit for proper orientation and connection.
4.
If desired, use the provided monitor cable to connect the control unit to the patient monitor (see Figure 2-3: Back of Control Unit). NOTE: Temperature input on patient monitor typically labelled T1 or T2.
8
3M™ Bair Hugger™ Temperature Monitoring System Model 370 – Operator’s Manual
Prior to moving the patient, disconnect the sensor cable from the sensor. The sensor may be left on the patient and reconnected to another system once patient transfer is complete. NOTE: The 3M™ Bair Hugger™ sensor contains integrated memory that captures up to two hours of temperature data in five minute intervals. The sensor will retain these temperatures and history.
Figure 2-4: Front of Control Unit
1.
Position the sensor cable away from the patient’s face and apply the sensor to the patient’s forehead above the orbital ridge (Figure 2-7).
g System
Ensure the system power supply is connected to the back of the control unit and the power supply is plugged into an appropriate outlet. If desired, ensure the provided monitor cable is connected to the back of the control unit and to the YSI-400 type input on the patient monitor.
Monitorin
1.
ture
Instructions for use
To view information and connections on the back of the control unit, assure that controller is viewed from a distance no further than 15” (38 cm) from the operator.
Tempera
5.
Back of control unit
Figure 2-6: Remove the backing liners
GB / 34-8719-5759-2
9
2.
Use an alcohol wipe to clean and disinfect the patient’s forehead above the orbital ridge. Allow the patient’s forehead to completely dry.
3.
Connect the sensor to the sensor cable connector. Ensure the sensor tab is properly connected and is fully inserted into the sensor cable (Figure 2-5). The display on the control unit will change to the ready screen (See “Ready” on page 11) once it confirms the sensor connection.
Stand Knob Digital Port (used by 3M personnel only)
°C/°F Temperature Display Button
Monitor Cable Port
Power Input Port Figure 2-3: Back of Control Unit
The digital port is used only by 3M Health Care personnel for testing purposes and to access variables within the control unit.
Figure 2-7: Apply the sensor to the patient’s forehead
6.
Gently press the edge of the sensor to ensure good adhesion with the skin. OPTIONAL: Support the sensor cable with the provided sensor cable clip.
7.
The control unit will sense patient attachment after several seconds and the control unit display will change to the equilibration screen (see “Equilibration” on page 11).
8.
The patient’s temperature will display on the patient monitor once equilibrium is reached (approximately three minutes).
Temp eratu re
The monitor cable port provides a resistance that corresponds to a YSI-400 thermistor at the displayed temperature. It is electrically isolated from the control unit.
Moni toring Syste m
Front of control unit
m
oring e Monit eratur Temp
Syste
Figure 2-5: Connect the sensor to the sensor cable
4.
Transferring the patient and sensor 1.
Remove backing liners on the sensor to expose adhesive (Figure 2-6).
2.
Carefully reconnect a sensor cable and control unit to the sensor once transfer is complete.
3.
The control unit will sense patient attachment after several seconds and the control unit display will change to the equilibration screen (see “Equilibration” on page 11).
4.
The patient’s temperature will display on the patient monitor once equilibrium is reached (approximately three minutes).
Setting up the control unit The control unit must be properly installed and mounted by qualified personnel. NOTE: The sensor and control unit are suitable for use in the patient environment. 2.
Connect the power supply to the back of the control unit (see Figure 2-3: Back of Control Unit). Plug the power supply into an appropriate outlet. The standby screen will illuminate.
3.
Connect the sensor cable to the front of the control unit. See Figure 2-4: Front of Control Unit for proper orientation and connection.
4.
If desired, use the provided monitor cable to connect the control unit to the patient monitor (see Figure 2-3: Back of Control Unit). NOTE: Temperature input on patient monitor typically labelled T1 or T2.
8
3M™ Bair Hugger™ Temperature Monitoring System Model 370 – Operator’s Manual
Prior to moving the patient, disconnect the sensor cable from the sensor. The sensor may be left on the patient and reconnected to another system once patient transfer is complete. NOTE: The 3M™ Bair Hugger™ sensor contains integrated memory that captures up to two hours of temperature data in five minute intervals. The sensor will retain these temperatures and history.
Figure 2-4: Front of Control Unit
1.
Position the sensor cable away from the patient’s face and apply the sensor to the patient’s forehead above the orbital ridge (Figure 2-7).
g System
Ensure the system power supply is connected to the back of the control unit and the power supply is plugged into an appropriate outlet. If desired, ensure the provided monitor cable is connected to the back of the control unit and to the YSI-400 type input on the patient monitor.
Monitorin
1.
ture
Instructions for use
To view information and connections on the back of the control unit, assure that controller is viewed from a distance no further than 15” (38 cm) from the operator.
Tempera
5.
Back of control unit
Figure 2-6: Remove the backing liners
GB / 34-8719-5759-2
9
Removing the sensor
3.
1.
Disconnect the sensor cable from the sensor. Disconnect the sensor cable clip from the bed or sheet, if used.
2.
Starting at the edge of the sensor, gently remove the sensor from the patient (Figure 2-8). NOTE: Do not use the sensor tab to remove the sensor from the patient.
Discard the sensor and clean the sensor cable. (See Section 5: General Maintenance, Storage and Service for details).
Section 3: Display Panel Display screens ... 11 Start-up ... 11 Standby ... 11
Sensor Tab
NOTE: If needed, use an alcohol swab along the sensor edge to help remove the sensor from the patient.
Ready ... 11 Equilibration ... 11 Running... 12 Control unit error... 12 Sensor error ... 12
Figure 2-8: Remove the sensor
To change temperature unit display (Celsius or Fahrenheit) The control unit can display temperature measurements in either Celsius or Fahrenheit. The unit will be factory preset to Celsius. 1.
Plug the control unit into an appropriate outlet.
2.
Push the temperature display button on the back of the control unit to change to the desired measurement scale. NOTE: The trend display is always shown in Celsius.
Display screens Start-up The start-up screen displays for approximately five seconds when the control unit is initially powered ON. The revision number of the control unit software is also displayed. FW-x.x.xxx
Standby The system graphic is displayed when the unit is not in use.
Temperature Monitoring System
Ready °C/°F Temperature Display Button
The ready screen displays the serial number of the sensor and indicates the sensor cable and sensor are properly attached to the control unit. The sensor is now ready to be attached to the patient. SN - 000000000
Figure 2-9: Temperature Display Button
Shutting down the control unit The plug on the power supply serves as the disconnect device. Turn the control unit on and off by plugging or unplugging the power supply from the wall outlet. The power outlet shall be as close as practical and shall be easily accessible.
Equilibration A flashing yellow temperature display indicates the sensor has been connected to the patient and to the control unit and is in the process of equilibration. The graph below the temperature output indicates the progress of equilibration. This process will take approximately three minutes. If you see a temperature in the upper left hand corner of the screen that is the last recorded temperature that has been captured by the sensor on the patient’s head. The sensor captures and records a temperature every five minutes.
36 9 C 39 37 35
36 0
C
After the equilibration is complete, the patient’s body temperature is displayed on the control unit in bold white numbers and can be automatically transmitted to the patient monitor via the monitor cable.
10
3M™ Bair Hugger™ Temperature Monitoring System Model 370 – Operator’s Manual
GB / 34-8719-5759-2
11
Removing the sensor
3.
1.
Disconnect the sensor cable from the sensor. Disconnect the sensor cable clip from the bed or sheet, if used.
2.
Starting at the edge of the sensor, gently remove the sensor from the patient (Figure 2-8). NOTE: Do not use the sensor tab to remove the sensor from the patient.
Discard the sensor and clean the sensor cable. (See Section 5: General Maintenance, Storage and Service for details).
Section 3: Display Panel Display screens ... 11 Start-up ... 11 Standby ... 11
Sensor Tab
NOTE: If needed, use an alcohol swab along the sensor edge to help remove the sensor from the patient.
Ready ... 11 Equilibration ... 11 Running... 12 Control unit error... 12 Sensor error ... 12
Figure 2-8: Remove the sensor
To change temperature unit display (Celsius or Fahrenheit) The control unit can display temperature measurements in either Celsius or Fahrenheit. The unit will be factory preset to Celsius. 1.
Plug the control unit into an appropriate outlet.
2.
Push the temperature display button on the back of the control unit to change to the desired measurement scale. NOTE: The trend display is always shown in Celsius.
Display screens Start-up The start-up screen displays for approximately five seconds when the control unit is initially powered ON. The revision number of the control unit software is also displayed. FW-x.x.xxx
Standby The system graphic is displayed when the unit is not in use.
Temperature Monitoring System
Ready °C/°F Temperature Display Button
The ready screen displays the serial number of the sensor and indicates the sensor cable and sensor are properly attached to the control unit. The sensor is now ready to be attached to the patient. SN - 000000000
Figure 2-9: Temperature Display Button
Shutting down the control unit The plug on the power supply serves as the disconnect device. Turn the control unit on and off by plugging or unplugging the power supply from the wall outlet. The power outlet shall be as close as practical and shall be easily accessible.
Equilibration A flashing yellow temperature display indicates the sensor has been connected to the patient and to the control unit and is in the process of equilibration. The graph below the temperature output indicates the progress of equilibration. This process will take approximately three minutes. If you see a temperature in the upper left hand corner of the screen that is the last recorded temperature that has been captured by the sensor on the patient’s head. The sensor captures and records a temperature every five minutes.
36 9 C 39 37 35
36 0
C
After the equilibration is complete, the patient’s body temperature is displayed on the control unit in bold white numbers and can be automatically transmitted to the patient monitor via the monitor cable.
10
3M™ Bair Hugger™ Temperature Monitoring System Model 370 – Operator’s Manual
GB / 34-8719-5759-2
11
Displays
Running The running screen displays the deep tissue temperature of the patient in white. The trend graph at the bottom of the screen displays the last two hours of the patient’s temperature in five minute increments. The bars on the graph will begin filling from the right with the 36°C always visible. The blue bars indicate that the temperature dropped below 36°C (white indicates the temperature is above 36°C).
36 7
C
36°
Control unit error
Condition
Cause
Action
“Ready” screen continues to display while sensor is adhered to the patient.
Sensor not properly adhered to the patient.
Make sure the sensor is connected to the patient.
Low patient temperature (below 30°C).
Press and hold the °C/°F button on the back of the control unit for five seconds to force the control unit into equilibration mode. The patient’s temperature will then display.
Control unit failure.
Discontinue use of unit. Contact 3M technical service.
Control unit may still be equilibrating.
Wait until the control unit has finished equilibration. Continue use.
Control unit failure.
Contact 3M technical service.
Control unit may still be equilibrating.
Wait until the control unit has finished equilibration. Continue use.
Monitor cable is not plugged in.
Make sure the monitor cable is plugged into the control unit and patient monitor.
Cable failure.
Replace monitor cable.
Control unit failure.
Contact 3M technical service after use.
Cable failure.
Make sure the monitor cable is securely plugged into the control unit and patient monitor.
The control unit error screen displays when there is a system error. Discontinue use of unit. Contact a biomedical technician. SN - 000000000
E-xx
Sensor error The sensor error screen displays when there is a temperature sensor or cable error. See “Section 4: Troubleshooting” on page 12 for details.
“Running” screen does not appear on the control unit.
36 7 E-xx
C
36°
Patient temperature does not appear on the patient monitor.
Section 4: Troubleshooting
36 9 C
On Mode ... 12 Displays ... 13
39 37 35
36 0
C
Errors... 14
The following conditions are listed in the order of which troubleshooting actions should be performed.
On Mode Condition
Cause
Action
Unit does not power up.
Unit is not plugged in, or power supply is not plugged into an appropriate outlet.
Make sure the power supply is plugged into the control unit. Make sure the control unit is plugged into an appropriate power outlet.
Unit failure.
Contact 3M technical service.
12
3M™ Bair Hugger™ Temperature Monitoring System Model 370 – Operator’s Manual
The patient’s temperature output on the control unit and patient monitor is > +/-0.2°C.
Contact 3M technical service.
Replace monitor cable. Verify monitor output (see Section 5 of the Installation and Service Manual). Contact 3M technical service.
GB / 34-8719-5759-2
13
Displays
Running The running screen displays the deep tissue temperature of the patient in white. The trend graph at the bottom of the screen displays the last two hours of the patient’s temperature in five minute increments. The bars on the graph will begin filling from the right with the 36°C always visible. The blue bars indicate that the temperature dropped below 36°C (white indicates the temperature is above 36°C).
36 7
C
36°
Control unit error
Condition
Cause
Action
“Ready” screen continues to display while sensor is adhered to the patient.
Sensor not properly adhered to the patient.
Make sure the sensor is connected to the patient.
Low patient temperature (below 30°C).
Press and hold the °C/°F button on the back of the control unit for five seconds to force the control unit into equilibration mode. The patient’s temperature will then display.
Control unit failure.
Discontinue use of unit. Contact 3M technical service.
Control unit may still be equilibrating.
Wait until the control unit has finished equilibration. Continue use.
Control unit failure.
Contact 3M technical service.
Control unit may still be equilibrating.
Wait until the control unit has finished equilibration. Continue use.
Monitor cable is not plugged in.
Make sure the monitor cable is plugged into the control unit and patient monitor.
Cable failure.
Replace monitor cable.
Control unit failure.
Contact 3M technical service after use.
Cable failure.
Make sure the monitor cable is securely plugged into the control unit and patient monitor.
The control unit error screen displays when there is a system error. Discontinue use of unit. Contact a biomedical technician. SN - 000000000
E-xx
Sensor error The sensor error screen displays when there is a temperature sensor or cable error. See “Section 4: Troubleshooting” on page 12 for details.
“Running” screen does not appear on the control unit.
36 7 E-xx
C
36°
Patient temperature does not appear on the patient monitor.
Section 4: Troubleshooting
36 9 C
On Mode ... 12 Displays ... 13
39 37 35
36 0
C
Errors... 14
The following conditions are listed in the order of which troubleshooting actions should be performed.
On Mode Condition
Cause
Action
Unit does not power up.
Unit is not plugged in, or power supply is not plugged into an appropriate outlet.
Make sure the power supply is plugged into the control unit. Make sure the control unit is plugged into an appropriate power outlet.
Unit failure.
Contact 3M technical service.
12
3M™ Bair Hugger™ Temperature Monitoring System Model 370 – Operator’s Manual
The patient’s temperature output on the control unit and patient monitor is > +/-0.2°C.
Contact 3M technical service.
Replace monitor cable. Verify monitor output (see Section 5 of the Installation and Service Manual). Contact 3M technical service.
GB / 34-8719-5759-2
13
Errors
Section 6: Technical Service and Order Placement
Error Screen
Condition
Cause
Action
E-1 through E-16, E-18 through E-99
Sensor not properly connected to the sensor cable.
Make sure the sensor is connected to the sensor cable.
Sensor failure.
Replace sensor.
E-xx
Cable failure.
E-17
Technical service and order placement ... 15
Replace sensor cable. Contact 3M technical service.
When you call for technical support
Technical service and order placement
Sensor failure.
Replace sensor.
3M Helpline at 1-800-228-3957 (US) or your local 3M representative (Worldwide).
Cable failure.
Replace sensor cable.
Remember, we will need to know the serial number, software number and error code of your control unit when you call us. The serial number label is located on the back of the control unit. (The software number displays on the control unit when the control unit is connected to a power outlet.)
E-xx
Contact 3M technical service.
E-100 and up
When you call for technical support ... 15
Control unit failure.
Discontinue use of the control unit. Contact a biomedical technician.
Control unit failure.
Discontinue use of the control unit. Contact a biomedical technician.
E-xx
Section 7: Specifications Physical characteristics ...18 Electrical characteristics ...18 Temperature characteristics ...18
Section 5: General Maintenance, Storage and Service
Performance characteristics ...18 Environmental conditions ...18
Cleaning Procedure ... 14
List of Bair Hugger Accessories and Cables ...19
Storage ... 14 Service ... 14 SPECIFICATIONS: Guidance and Manufacturer’s Declaration - Emissions
Cleaning Procedure
!
Storage
CAUTION: •
•
Do not immerse any of the devices or use a dripping wet cloth for cleaning. Moisture may seep inside the device and damage the electrical components and lead to incorrect temperature reporting. Do not spray cleaning solutions onto the control unit or into the sensor cable connector. Damage to the control unit or sensor cable connector may occur.
Cleaning the control unit and sensor cable Clean the control unit on an as-needed basis or per facility policies and procedures for cleaning electronic equipment. Clean the sensor cable between each use. 1.
Disconnect power supply from power outlet.
2.
Use a slightly damp soft cloth moistened with a mild, nonabrasive cleaning solution to clean the device surfaces, cords, and cables. Avoid getting liquid into electronic ports.
3.
Dry with a separate soft cloth.
14
Notice: Do not store the control unit and system components in a wet or damp place. Damage to the electrical components may occur. Store all components at room temperature and in a dry place when not in use.
Service All service must be performed by 3M Health Care or an authorized service technician. Call 3M Customer Service at 1-800-733-7775 or your local 3M representative for service information.
3M™ Bair Hugger™ Temperature Monitoring System Model 370 – Operator’s Manual
The 3M™ Bair Hugger™ Temperature Monitoring System is intended for use in the electromagnetic environment specified below. The customer or user of the Temperature Monitoring System should ensure that it is used in such an environment. Emissions Test
Compliance
Electromagnetic Environment – Guidance
RF Emissions
Group 1
The Temperature Monitoring System uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
Class B
The Temperature Monitoring System is suitable for use in all establishments, including domestic, and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
CISPR 11 RF Emissions CISPR 11 Harmonics
Class A
IEC 61000-3-2 Flicker
Complies
IEC 61000-3-3
GB / 34-8719-5759-2
15
Errors
Section 6: Technical Service and Order Placement
Error Screen
Condition
Cause
Action
E-1 through E-16, E-18 through E-99
Sensor not properly connected to the sensor cable.
Make sure the sensor is connected to the sensor cable.
Sensor failure.
Replace sensor.
E-xx
Cable failure.
E-17
Technical service and order placement ... 15
Replace sensor cable. Contact 3M technical service.
When you call for technical support
Technical service and order placement
Sensor failure.
Replace sensor.
3M Helpline at 1-800-228-3957 (US) or your local 3M representative (Worldwide).
Cable failure.
Replace sensor cable.
Remember, we will need to know the serial number, software number and error code of your control unit when you call us. The serial number label is located on the back of the control unit. (The software number displays on the control unit when the control unit is connected to a power outlet.)
E-xx
Contact 3M technical service.
E-100 and up
When you call for technical support ... 15
Control unit failure.
Discontinue use of the control unit. Contact a biomedical technician.
Control unit failure.
Discontinue use of the control unit. Contact a biomedical technician.
E-xx
Section 7: Specifications Physical characteristics ...18 Electrical characteristics ...18 Temperature characteristics ...18
Section 5: General Maintenance, Storage and Service
Performance characteristics ...18 Environmental conditions ...18
Cleaning Procedure ... 14
List of Bair Hugger Accessories and Cables ...19
Storage ... 14 Service ... 14 SPECIFICATIONS: Guidance and Manufacturer’s Declaration - Emissions
Cleaning Procedure
!
Storage
CAUTION: •
•
Do not immerse any of the devices or use a dripping wet cloth for cleaning. Moisture may seep inside the device and damage the electrical components and lead to incorrect temperature reporting. Do not spray cleaning solutions onto the control unit or into the sensor cable connector. Damage to the control unit or sensor cable connector may occur.
Cleaning the control unit and sensor cable Clean the control unit on an as-needed basis or per facility policies and procedures for cleaning electronic equipment. Clean the sensor cable between each use. 1.
Disconnect power supply from power outlet.
2.
Use a slightly damp soft cloth moistened with a mild, nonabrasive cleaning solution to clean the device surfaces, cords, and cables. Avoid getting liquid into electronic ports.
3.
Dry with a separate soft cloth.
14
Notice: Do not store the control unit and system components in a wet or damp place. Damage to the electrical components may occur. Store all components at room temperature and in a dry place when not in use.
Service All service must be performed by 3M Health Care or an authorized service technician. Call 3M Customer Service at 1-800-733-7775 or your local 3M representative for service information.
3M™ Bair Hugger™ Temperature Monitoring System Model 370 – Operator’s Manual
The 3M™ Bair Hugger™ Temperature Monitoring System is intended for use in the electromagnetic environment specified below. The customer or user of the Temperature Monitoring System should ensure that it is used in such an environment. Emissions Test
Compliance
Electromagnetic Environment – Guidance
RF Emissions
Group 1
The Temperature Monitoring System uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
Class B
The Temperature Monitoring System is suitable for use in all establishments, including domestic, and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
CISPR 11 RF Emissions CISPR 11 Harmonics
Class A
IEC 61000-3-2 Flicker
Complies
IEC 61000-3-3
GB / 34-8719-5759-2
15
Guidance and Manufacturer’s Declaration - Emissions
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment – Guidance
The 3M™ Bair Hugger™ Temperature Monitoring System is intended for use in the electromagnetic environment specified below. The customer or user of the Temperature Monitoring System should ensure that it is used in such an environment.
ESD
±6kV Contact
±6kV Contact
IEC 61000-4-2
±8kV Air
±8kV Air
Floors should be wood, concrete or ceramic tile. If floors are synthetic, the r/h should be at least 30%
EFT
±2kV Mains
±2kV Mains
IEC 61000-4-4
±1kV I/Os
±1kV I/Os
Surge
±1kV Differential
±1kV Differential
IEC 61000-4-5
±2kV Common
±2kV Common
Voltage Dips/ Dropout IEC 61000-4-11
>95% Dip for 0.5 Cycle
>95% Dip for 0.5 Cycle
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment – Guidance
Conducted RF IEC 61000-4-6
3 Vrms 150 kHz to 80 MHz
3Vrms
Portable and mobile communications equipment should be separated from the Temperature Monitoring System by no less than the distances calculated/listed below:
Radiated RF IEC 61000-4-3
3 V/m 80 MHz to 2.5 GHz
3V/m
D=1.2 D= 1.2
(80 to 800 MHz)
60% Dip for 5 Cycles
60% Dip for 5 Cycles
D= 2.3
(800 MHz to 2.5 GHz)
30% Dip for 25 Cycles
30% Dip for 25 Cycles
>95% Dip for 5 Seconds
>95% Dip for 5 Seconds
3A/m
3A/m
Where P is the max power in watts and D is the recommended separation distance in meters. Field strengths from fixed transmitters, as determined by an electromagnetic site survey, should be less than the compliance levels (V1 and E1).
Power Frequency 50/60Hz
Interference may occur in the vicinity of equipment containing a transmitter.
Magnetic Field IEC 61000-4-8
Guidance and Manufacturer’s Declaration –Immunity The 3M™ Bair Hugger™ Temperature Monitoring System is intended for use in the electromagnetic environment specified below. The customer or user of the Temperature Monitoring System should ensure that it is used in such an environment.
3M™ Bair Hugger™ Temperature Monitoring System Model 370 – Operator’s Manual
Mains power quality should be that of a typical commercial or hospital environment. Mains power quality should be that of a typical commercial or hospital environment. If the user of the Temperature Monitoring System requires continued operation during power mains interruptions, it is recommended that the Temperature Monitoring System be powered from an uninterruptible power supply or battery.
Power frequency magnetic fields should be that of a typical commercial or hospital environment.
Recommended Separations Distances The 3M™ Bair Hugger™ Temperature Monitoring System is intended for use in the electromagnetic environment in which radiated disturbances are controlled. The customer or user of the Temperature Monitoring System can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF Communications Equipment and the Temperature Monitoring System as recommended below, according to the maximum output power of the communications equipment.
Max Output Power (Watts)
16
Mains power quality should be that of a typical commercial or hospital environment.
Separation (m) 150kHz to 80MHz
Separation (m) 80 to 800MHz
Separation (m) 800MHz to 2.5GHz
D=1.2
D=1.2
D=2.3
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
GB / 34-8719-5759-2
17
Guidance and Manufacturer’s Declaration - Emissions
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment – Guidance
The 3M™ Bair Hugger™ Temperature Monitoring System is intended for use in the electromagnetic environment specified below. The customer or user of the Temperature Monitoring System should ensure that it is used in such an environment.
ESD
±6kV Contact
±6kV Contact
IEC 61000-4-2
±8kV Air
±8kV Air
Floors should be wood, concrete or ceramic tile. If floors are synthetic, the r/h should be at least 30%
EFT
±2kV Mains
±2kV Mains
IEC 61000-4-4
±1kV I/Os
±1kV I/Os
Surge
±1kV Differential
±1kV Differential
IEC 61000-4-5
±2kV Common
±2kV Common
Voltage Dips/ Dropout IEC 61000-4-11
>95% Dip for 0.5 Cycle
>95% Dip for 0.5 Cycle
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment – Guidance
Conducted RF IEC 61000-4-6
3 Vrms 150 kHz to 80 MHz
3Vrms
Portable and mobile communications equipment should be separated from the Temperature Monitoring System by no less than the distances calculated/listed below:
Radiated RF IEC 61000-4-3
3 V/m 80 MHz to 2.5 GHz
3V/m
D=1.2 D= 1.2
(80 to 800 MHz)
60% Dip for 5 Cycles
60% Dip for 5 Cycles
D= 2.3
(800 MHz to 2.5 GHz)
30% Dip for 25 Cycles
30% Dip for 25 Cycles
>95% Dip for 5 Seconds
>95% Dip for 5 Seconds
3A/m
3A/m
Where P is the max power in watts and D is the recommended separation distance in meters. Field strengths from fixed transmitters, as determined by an electromagnetic site survey, should be less than the compliance levels (V1 and E1).
Power Frequency 50/60Hz
Interference may occur in the vicinity of equipment containing a transmitter.
Magnetic Field IEC 61000-4-8
Guidance and Manufacturer’s Declaration –Immunity The 3M™ Bair Hugger™ Temperature Monitoring System is intended for use in the electromagnetic environment specified below. The customer or user of the Temperature Monitoring System should ensure that it is used in such an environment.
3M™ Bair Hugger™ Temperature Monitoring System Model 370 – Operator’s Manual
Mains power quality should be that of a typical commercial or hospital environment. Mains power quality should be that of a typical commercial or hospital environment. If the user of the Temperature Monitoring System requires continued operation during power mains interruptions, it is recommended that the Temperature Monitoring System be powered from an uninterruptible power supply or battery.
Power frequency magnetic fields should be that of a typical commercial or hospital environment.
Recommended Separations Distances The 3M™ Bair Hugger™ Temperature Monitoring System is intended for use in the electromagnetic environment in which radiated disturbances are controlled. The customer or user of the Temperature Monitoring System can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF Communications Equipment and the Temperature Monitoring System as recommended below, according to the maximum output power of the communications equipment.
Max Output Power (Watts)
16
Mains power quality should be that of a typical commercial or hospital environment.
Separation (m) 150kHz to 80MHz
Separation (m) 80 to 800MHz
Separation (m) 800MHz to 2.5GHz
D=1.2
D=1.2
D=2.3
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
GB / 34-8719-5759-2
17
Physical characteristics
Electrical characteristics
Dimensions of Control Unit 9.3 cm (3.7 in) high, extendable to 11.4 cm (4.5 in) high
External power supply 100-240 VAC, 50-60 Hz
7.1 cm (2.8 in) wide, 4.3 cm (1.7 in) deep Weight of Control Unit 128 g (4.5 oz)
Output; 5 VDC
Model Number
Cable Length (m)
70-2007-9192-2
90106
Affinity Medical
MD370
4.0 ± 0.08
MD370
2.0 ± 0.03
90107
Monitor cable, 3.5mm to 3.5mm
70-2007-9194-8
90108
TLC Electronics
MD370
2.0 ± 0.03
Temp. cable, Phillips to mini-phone
70-2007-9206-0
90130
AMC&E
MD370
2.0 ± 0.075
Temp. cable, GE to mini-phone
70-2007-9205-2
90131
AMC&E
MD370
2.0 ± 0.075
Temp. cable, Draeger to mini-phone
70-2007-9204-5
90132
AMC&E
MD370
2.0 ± 0.075
Temp. cable, Spacelabs to mini-phone
70-2007-9207-8
90133
AMC&E
MD370
2.0 ± 0.075
US Style Power Supply
70-2007-9197-1
90113
FRIWO
MD370
3.0 ± 0.1
Measurement locations Patient’s lateral forehead above the orbital ridge
EU style power supply
70-2007-9198-9
90115
FRIWO
MD370
3.0 ± 0.1
UK style power supply
70-2007-9199-7
90116
FRIWO
MD370
3.0 ± 0.1
Measurement range 25°C - 43°C
Bair Hugger Sensor Model 36000
70-2007-9188-0
36000
3M
MD360
-
Measurement readout Celsius or Fahrenheit
WARNING: The use of ACCESSORIES and CABLES other than those specified, with the exception of accessories and cables sold by 3M as replacement parts for internal components, may result in increased EMISSIONS or decreased IMMUNITY of the Bair Hugger system.
Over-temperature cutoff 43°C - Skin thermistor reading
Classification
44.5°C - Heater thermistor reading
Performance characteristics Equilibration time/Response time Approximately three minutes
Leakage current CF classification. Meets leakage current requirements in accordance with UL 60601-1 and EN60601-1.
Environmental conditions
Accuracy 25°C to 43°C ± 0.2°C
Ambient temperature range 10°C to 40°C (50°F to 104°F)
Sensor Material Medical grade foam and adhesive
Storage and transport temperature range -20°C to 60°C (-4°F to 140°F)
PET flexible circuit
Store all components at room temperature and in a dry place when not in use. Operating humidity 10 to 75% RH, noncondensing
Sensor Shelf Life 36 months
Atmospheric pressure range 80 kPa to 106 kPa (12 PSI to 15 PSI)
Digital Port Serial output 0 [-12 to +12] Monitor Cable Port Provides a resistance that corresponds to a YSI-400 thermistor at the displayed temperature. It is electrically isolated from the control unit.
18
Manufacturer
70-2007-9193-0
Length of the Sensor Cable 400 cm (158 in)
Recommended Calibration Check Every 12 months
3M Catalog #
TLC Electronics
Temperature characteristics
Classified by Underwriters Laboratories Inc. with respect to electric shock, fire, and mechanical hazards only, in accordance with UL 60601-1, IEC 60601-1, and Canadian/CSA C22.2 No. 601.1. ANSI/ AAMI ES 60601-1; 2005. EN60601-1.
Sensor Cable
Part Number
Monitor Cable, 3.5mm to 6.35mm
Dimensions of Sensor 4.1 cm (1.6 in) diameter, 0.5 cm (0.2 in) thick
Classified under IEC 60601-1 Guidelines (and other national versions of the Guidelines) as Class II, Type CF, Defibrillation-Proof, Ordinary Equipment. Conforms to EN12470. Not suitable for use in the presence of flammable anesthetic mixtures with air or nitrous oxide.
Component
Class II, double insulated medical grade Maximum heating power 2W
MEDICAL-GENERAL MEDICAL EQUIPMENT AS TO ELECTRICAL SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601-1; CAN/ CSA-C22.2, No. 60601-1; ANSI/ AAMI ES60601-1:2005 CSA-C22.2 No.601-1:08; EN60601-1; Control No.4HZ8
List of Bair Hugger Accessories and Cables
Oxygen environment The sensor only is suitable for use in the presence of an oxygen enriched environment.
3M™ Bair Hugger™ Temperature Monitoring System Model 370 – Operator’s Manual
GB / 34-8719-5759-2
19