Operators Manual
12 Pages
Preview
Page 1
3M Attest 290G Auto-reader ™
™
Operator’s Manual
Read and understand all safety information before using this product.
Health Care
INTENDED USE
3 2 1
9 4
7
5
8
6
1. 2. 3. 4. 5.
Cover Incubation/Reader Wells Crusher Well Red Light Positive (+) Indicator Yellow Light (•) Test In Progress Indicator
6. 7. 8. 9.
Green Light Negative (-) Indicator Buzzer Alarm OFF Button Digital Display Panel Time Remaining Button
Intended Use The 3MTM AttestTM 290G Auto-reader is designed to incubate and automatically read the 3MTM AttestTM 1294 Rapid Readout Biological Indicators (RRBI) for Ethylene Oxide (EO) at 37oC for a final negative fluorescence reading at 4 hours. The Attest 290G Auto-reader is also designed to allow for further incubation of the Attest 1294 RRBI for EO for visual pH color change of the growth media at 7 days. The 7-day incubation is not intended for routine use and may be used for special studies to confirm the validity of the fluorescent results.
3M © 2003 78-8083-0136-6
Page 1 of 9
A positive fluorescence reading indicates an EO gas sterilization process failure. The visual pH color change also indicates a sterilization process failure. However, due to the high sensitivity of the 4-hour fluorescent results, there is no advantage to incubate the Attest 1294 RRBI for EO beyond 4 hours. Other biological indicators (BIs), are not compatible with this device and cannot be used. The unit is designed to be used only with the Power Supply module supplied by 3M.
Attest 290G Auto-reader
SAFETY INFORMATION
CAUTION
WARNING •
•
•
To avoid the risk of electric shock, which if not avoided could result in serious injury or death: - Do not immerse the Attest 290G Auto-reader in any liquid. - Use indoors only. - Disconnect power cord before cleaning. - To clean, use only a mild detergent and wipe with a damp cloth. - There are no user serviceable parts. - The Attest 290G Auto-reader must be returned to the manufacturer for repair. - Make sure a properly rated power cord is installed on the Attest 290G Auto-reader power supply. To reduce the risk of using incompletely sterilized loads: - Match the color on the exterior label of the incubator with the same color cap on the RRBI. - Do not remove RRBI until incubation cycle is complete. - Do not use on a vibrating surface. There is a glass ampule inside the plastic vial of the RRBI. To avoid the risk of injury from flying debris due to a ruptured RRBI, which if not avoided could result in serious injury or death: - Wear safety glasses and gloves when removing the RRBI from the sterilizer. - Wear safety glasses when crushing the RRBI. - Handle the RRBI by the cap when crushing and tapping. - Do not use your fingers to crush the glass ampule. - Do not roll the RRBI between your fingers to wet the spore strip.
3M © 2003 78-8083-0136-6
Page 2 of 9
•
To avoid possible environmental contamination, dispose of the Attest 290G Auto-reader according to applicable governmental regulations.
•
To avoid RRBI vial from absorbing fluorescent residue from a chemical indicator or tape, place the RRBI vial so it does not come in direct contact with chemical indicators or tape.
Explanation of symbols:
Caution – Risk of electric shock
Attention – Refer to accompanying documents
Attest 290G Auto-reader
SAFETY INFORMATION Power Supply Specifications Power supply
Operating Condition
Units
Voltage Range Frequency Current
100-240 50/60 1.2
Volts AC Hertz Amps
Environmental Condition
Operating Condition
Units
Altitude Operating Temperature Relative Humidity Voltage Range Frequency Current Installation/Over Voltage Pollution Degree
2500 (max) 16-29 20-80 (non-condensing) 12 DC 2.4 Category II 2
meters °C % Volts DC
Environmental Operating Conditions
Amps
Health and Safety Information Device Safety Compliance The Attest 290G Auto-reader complies with the standard IEC/EN 61010-1 as demonstrated by the CB Scheme Certificate and test report that Underwriters Laboratories Inc. (UL) generated. The Attest 290G Auto-reader is listed as Laboratory equipment and carries the UL and c-UL marks based on compliance to the standards UL 3101-1 and CSA 22.2 No. 1010-1.92. The Attest 290G Auto-reader complies with the CE mark related to the Low Voltage Directive 73/23/EEC as confirmed in the Declaration of Conformity. EMC Compliance The Attest 290G Auto-reader complies with IEC/EN 61326-1 as confirmed in the Certificate of Compliance generated by 3M. The Attest 290G Auto-reader complies with the EMC requirements of the CE mark EMC Directive 89/336/EEC.
Note: This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide a reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at their own expense. This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations. Cet appareil numerique de la classe A respecte toutes les exigences due Reglement sur le materiel brouilleur du Canada.
The Attest 290G Auto-reader complies with the Australian EMC requirements as confirmed in the Supplier’s Declaration of Conformity that is linked to the C-tick mark.
3M © 2003 78-8083-0136-6
Page 3 of 9
Attest 290G Auto-reader
INSTRUCTIONS FOR USE Operating Instructions Start-up
Readout Procedure
Place unit on a firm level surface. Plug unit into an appropriate AC grounded outlet.
Follow this procedure for every RRBI to be read:
Allow a 30 minute warm-up period before placing RRBIs into the incubation wells. The C1 caution code will disappear when proper incubation temperature is reached. It is recommended that the unit be left on to eliminate warm-up periods. The unit may be turned off when it is anticipated that it will not be used for an extended period.
Step 1. While wearing gloves and safety glasses, CLOSE RRBI cap by pressing down.
Step 2. CRUSH the glass ampule (containing growth media) in the designated crushing well built into the incubator.
Step 3. Hold the RRBI by the cap and TAP bottom of the vial on a tabletop until media wets spore strip at bottom of vial. Do not tap vial on unit.
WARNING
Step 4. Open the cover and PLACE the RRBI into an incubation/reader well.
3M © 2003 78-8083-0136-6
Step 5. CLOSE the cover on the Attest 290G Auto-reader and wait for either the red or green indicator light to signal the result.
Page 4 of 9
Do not remove or change placement of RRBI once it is placed into a well. Doing so may result in loss of results and the RRBI test may be invalid.
Attest 290G Auto-reader
INSTRUCTIONS FOR USE
Wells are color coded to match the cap of each type of RRBI; i.e. the Attest 1294 RRBI with a green cap goes into a green well.
All red, yellow and green lights will flash for one second when the cover is opened demonstrating they are functional.
A yellow light will come on to indicate the RRBI is incubating in the reader well and should not be touched. A yellow light indicates the results are not available and incubation is in progress.
The cover should only be opened when placing additional RRBIs into incubation wells or when removing RRBIs at the completion of the designated incubation period.
Do not remove RRBIs until the red (+) or green (-) light indicates the test is complete.
A positive result (+) will automatically be shown as soon as it is detected. The full incubation period is required before a negative result (-) is shown. The Attest 1294 RRBI for EO requires 4 hours of incubation before a negative (-) result will be shown. The negative result will be shown automatically after completion of the required incubation time.
Note: If RRBIs are inadvertently removed before the incubation period is complete, an alarm will sound and all three lights (+, •, and -), for that incubation well will light. C2 Caution code will also be displayed. You must return the RRBI to the reader well within 10 seconds or the results will be lost.
WARNING Do not remove or change placement of RRBI once it is placed into a well. Doing so may result in loss of results and the RRBI test may be invalid.
3M © 2003 78-8083-0136-6
Page 5 of 9
The processed RRBI and the positive control may also be further incubated for 7 days to obtain a visual pH color change. The 7-day incubation is not intended for routine use and may be used for special studies to confirm the validity of the fluorescent results. This visual pH color change also indicates a sterilization process failure however, due to the high sensitivity of the 4-hour fluorescent results, there is no advantage to incubate the Attest 1294 RRBI for EO beyond 4 hours. The final negative reading (media remains green) for a visual color change is made at 7 days. The positive control should show a yellow color change of the growth media.
Attest 290G Auto-reader
INSTRUCTIONS FOR USE Positive Control Crush, tap and incubate at least one non-processed Attest 1294 RRBI for EO to use as the positive control each day that a processed RRBI is incubated and read by the unit. Write a “C” (for “control”) and the date on the label. The positive control can be placed in any incubator well of the unit. The positive control should be the same part number, from the same manufacturing date and lot number as the processed RRBI in the unit.
It is good practice to use a positive control each day a sterilized RRBI is processed. This helps ensure: - correct incubation temperatures are met - viability of spores has not been altered due to improper storage temperature, humidity, or proximity to chemicals - capability of media to promote rapid growth, and - proper functioning of Attest 290G Auto-reader components.
Interpretation of Results Each incubation/reader well has three corresponding indicator lights: a red (+), yellow (•) and green (-) light. After placing a RRBI into a corresponding color-coded well, a yellow light will illuminate to indicate that incubation/reading is in progress.
The final negative processed RRBI reading for a fluorescence change (green light -) after 4 hours of incubation indicates an acceptable sterilization process. The positive control should read positive (red light +).
A red light (+), will illuminate and an alarm will sound (if the sound feature is turned on), as soon as a positive RRBI result is detected.
If the processed RRBI is further incubated for a visual pH color change, when special studies are desired, the media color change from green to yellow will also indicate a sterilization failure. The positive control should show a yellow color change at 7 days. A final negative color change result (no color change), media remains green , is made at 7 days.
At the end of the specified incubation time, if a negative RRBI result is detected, the green light (-) will illuminate indicating an acceptable sterilization process. After either the green or red light comes on, the RRBI can be removed from the well. The RRBI should not be removed or touched when the yellow light is on. The positive (unprocessed) control RRBI must provide a positive fluorescent result (red light +). Processed RRBI results are not valid until the positive control reads fluorescent positive (red light +). With the processed RRBI, a positive (red light +) result means a sterilization process failure may have occurred.
Act on any positive result immediately. The positive result indicates a sterilization process failure may have occurred. Determine the cause of the positive RRBI following current facility policies and procedures. Always retest the sterilizer and do not use the sterilizer for processing loads until 3 consecutive RRBI results are negative.
Disposal Dispose of the used Attest 1294 RRBI for EO according to your healthcare facility policy.
3M © 2003 78-8083-0136-6
You may wish to sterilize any positive RRBI prior to disposal.
Page 6 of 9
Attest 290G Auto-reader
FEATURES, CARE AND CLEANING Features Audible Alarm An audible alarm will sound and the positive red (+) light will illuminate for the designated incubation/reader well when a positive result is detected. This provides results in the least amount of time without the need to manually take frequent readings. An alarm will also sound if a RRBI is removed from the Attest 290G Auto-reader well before a final result has been detected. Data will be lost if the RRBI is not replaced in the reader/incubation well within 10 seconds. Alarm On/Off The audible alarm for positive results can be turned off or on for a particular RRBI. This is accomplished by pressing the "alarm off" button after the alarm has sounded. This will turn off the alarm for that RRBI only. Any positive RRBI results detected in other wells will result in the alarm sounding. Removing the RRBI will reactivate the alarm for that well. Pressing the “alarm off” button will not silence the alarm for error or caution code alarm warnings.
To turn the alarm function back on, press the "alarm off" and "time remaining" buttons at the same time. The yellow period on the display panel will turn off, indicating that the alarm is again active. Time Remaining When the “time remaining” button is pressed, the incubation time remaining for each individual incubation/reader well that contains a RRBI will be displayed. All three lights for each individual incubation/reader well will illuminate as the time remaining is displayed. The process begins with the incubation/reader well containing a RRBI on the left and proceeds to the right. The time remaining appears in hours until there is less than an hour remaining when it displays minutes remaining. Temperature Monitoring The Attest 290G Auto-reader contains an internal temperature self-diagnostic program. Should the temperature fall outside the specified range (37ºC +/-2ºC) an error code will be displayed. Power Cord Retaining Clip
To turn alarm function off (disable), press both the "alarm off" and "time remaining" buttons at the same time. A yellow period will appear on the lower right corner of the display panel to indicate the alarm is disabled. The alarm will be disabled for all positive RRBIs, error codes, and caution code alarm warnings.
Place the power cord through the power cordretaining clip on the rear of the Attest Model 290G Auto-reader near the plug in. This will help to ensure that the power connection does not become inadvertently disconnected.
Care and Cleaning
WARNING Always unplug the Attest 290G Autoreader and allow to cool before cleaning. Do not immerse the unit in liquid.
3M © 2003 78-8083-0136-6
Page 7 of 9
Clean the exterior surface of the Attest 290G Autoreader, with the dust cover closed, by wiping it with a damp cloth moistened with a mild detergent. Do not open the dust cover and attempt to clean interior components. If additional cleaning of the wells is required, call for repair or replacement. Note: The Attest 290G Auto-reader does not contain any user serviceable parts or require any routine maintenance. Any malfunctions will be indicated by an error code. Refer to the Trouble Shooting Guide section for error code explanations.
Attest 290G Auto-reader
TROUBLESHOOTING GUIDE Trouble Shooting Guide In the event of an error code, unplug the Attest 290G Auto-reader and plug it back in to initiate the selfdiagnostic test. If error code re-occurs, call for repair or replacement. The display panel will indicate malfunctions detected by a specific error code. Refer to the following: Code
Problem
Solution
C1
Unit Warm-up Caution Incubation block not at temperature (RRBI readings are disabled until incubator block reaches 37ºC +/-2ºC).
Allow a 30 minute warm-up after Attest 290G Auto-reader is plugged in.
C2
RRBI has been removed before incubation complete. All three lights for affected RRBI will light and alarm will sound if active.
Replace RRBI in proper well within 10 seconds.
E1
Temperature Control Error This error will occur if system is no longer able to control the temperature of the incubator block (37ºC +/-2ºC).
Call for repair or replacement.
E3
Memory Error.
Call for repair or replacement.
E4
Analog-Digital Conversion Error.
Call for repair or replacement.
E7
Communications Error This error will occur if internal communications have failed.
Call for repair or replacement.
E8
Bulb Error This error occurs if the system detects that the bulb is too dim to properly read RRBI results.
Call for repair or replacement.
E9
Transport Error This error will occur if the system detects a motor failure or positioning problem.
Call for repair or replacement.
3M © 2003 78-8083-0136-6
Page 8 of 9
Attest 290G Auto-reader
TROUBLESHOOTING GUIDE Trouble Shooting Guide (Continued) Code
Problem
Solution
Other
Unwanted alarm sounds when positive detected.
Turn alarm off, refer to features Section.
Other
All three lights blink and buzzer sounds. RRBI has been removed from incubation/reader well prematurely.
Replace RRBI in incubation/reader well within 10 seconds to prevent loss of data collected.
Other
Red, yellow, or green lights do not illuminate when expected such as following insertion of a RRBI or when time remaining button is pressed.
Specific light burned out. Call for repair or replacement.
Other
Negative Control Vial
Other
Vial not crushed.
Crush vial before inserting into well.
Media did not wet the spore strip.
Tap vial on table until media wets spore strip.
Control vial was sterilized.
Use a non-sterilized vial as a control. Check chemical indicator on label.
Unexplained positive Adhesive residue on vial.
Do not place tape or labels on vial.
Test vial is absorbing fluorescent residue from a chemical indicator or tape.
Place the vial so it does not come in direct contact with chemical indicators or tape.
Repair and Replacement Attest 290G Auto-readers must be repaired at: 3M Health Care Service Center Suite 200, Bldg. 502 3350 Granada Avenue North Oakdale, MN 55128 1-800-292-6298 Fax: 1-800-770-8016 Outside of the USA, contact your local 3M Subsidiary. If you have questions, call our Customer Service number: 1-800-292-6298.
3M © 2003 78-8083-0136-6
Page 9 of 9
Attest 290G Auto-reader
3M Health Care 3M Center Building 275-4E-01 St. Paul, MN 55144-1000 1-800-228-3957
3M © 2003 78-8083-0136-6
3M Health Care D-41460 Neuss, Germany
Page 10 of 9
Recycled paper 40% pre-consumer 10% post-consumer Litho in U.S.A. with 3M film, proofing systems and offset plates. Attest is a registered trademark of 3M. 3M 2003 78-8083-0136-6 Rev. B Lit. Code: 70-2009-4687-2
Attest 290G Auto-reader