3M™ Attest™ Mini Auto-reader 490M System* Quick Reference Guide BI Preparation and Use for Tabletop Sterilisers
10
mins
1
Assemble PCD
Create a Process Challenge Device (PCD)by placing a BI** and a chemical integrator in representative packaging.
2 Place
Place PCD in a full load in the location recommended by steriliser manufacturer.
3 Process
4 Cool
3 Inspect
4 Incubate
Run sterilisation cycle.
After completion of cycle, retrieve PCD, remove BI and allow BI to cool 10 minutes prior to activation.
BI Activation and Incubation
1
Crush
Squeeze BI in activator to close cap and crush media ampoule.
2 Flick
Remove BI and flick wrist to ensure media flows to growth chamber.
Verify media has transferred to BI growth chamber at the vial bottom.
Place activated BI in any well. Display shows remaining minutes of incubation or test results. DO NOT REMOVE BI until the (+) or (-) symbol indicates the test is complete.
Interpretation of 3M™ Sterilisation Assurance Core Products 3M™ Attest™ Biological Indicator 1492V
3M™ Attest™ Steam Chemical Integrator 1243
3M™ Comply™ Indicator Tape 1322
3M™ Comply™ Indicator Tape 1355
Unexposed
Unexposed
Unprocessed
Acceptable Exposed
Acceptable Exposed
Acceptable Processed PASS
After processing, indicator stripes turn dark-brown/black.
After processing, indicator stripes turn dark-brown/black.
Unexposed
1492V
Acceptable Exposed
1492V After processing, indicator on cap turns light brown or darker.
Unacceptable Processed FAIL
3M™ Attest™ Mini Auto-reader 490M Positive Control BI Each day a processed BI is incubated, activate and incubate an unprocessed BI as a positive control. The control BI should be from the same lot code as the processed BI.
Results After Incubation If the color bar crosses into the ACCEPT region, it is considered a pass. If the color bar is in the REJECT region or on the line, it is considered a fail.
Control BI: Positive (+) Acceptable Processed BI: Negative (-) Acceptable Sterilisation Processed BI: Positive (+) Failed Sterilisation
Remaining Incubation Minutes
3M Australia Pty Ltd Medical Solutions Division Bldg A, 1 Rivett Road North Ryde NSW 2113
3M New Zealand Limited Medical Solutions Division 94 Apollo Drive Rosedale Auckland 0632
Phone Web
Phone Web
1300 363 878 www.3M.com.au/healthcare
0800 80 81 82 www.3M.co.nz/healthcare
3M, Attest and Comply are trademarks of 3M. Please recycle. Printed in Australia. © 3M 2020. All rights reserved. AV011483516
*Refer to 3M™ Attest™ Mini Auto-reader 490M Operator Manuals and 3M™ Attest™ Biological Indicator 1492V Instructions for Use for complete instructions and other important information. Refer to the Instructions for Use for complete instructions and other important information related to the use of all other 3M products included in this Guide. **Refer to the indications for use provided in the biological indicator Instructions for Use. The indications for use should align with the sterilisation cycle to be monitored.