3M
3M Ranger Blood-Fluid Warming System Model 245 Operators Manual May 2014
Operators Manual
22 Pages
Preview
Page 1
3M™ Ranger™ Blood/Fluid Warming System
English
1
Table of Contents Technical Service and Order Placement � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
2
Introduction � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
3
Indications for use � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
3
Definition of Symbols � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
3
Explanation of Signal Word Consequences � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
4
Warning: � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
5
Caution: � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
5
Notice: � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
6
Product Description � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
6
The Ranger blood/fluid warming unit � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
6
Ranger blood/fluid warming set � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
7
Product safety features � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
8
Instructions for Use � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
10
Preparation and setup of the Ranger blood/fluid warming unit� � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
10
Removing the warming set from the Ranger blood/fluid warming unit � � � � � � � � � � � � �
11
Transferring the warming set from one Ranger warming unit to another � � � � � � � � � � � �
11
Maintenance and Storage � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
14
Specifications � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � � �
16
PDF Scaled at 100.0%
English 34-8714-4406-2
English
3M™ Ranger™ Blood/Fluid Warming System
Technical Service and Order Placement USA TEL:
1-800-228-3957
Outside of the USA Contact your local 3M representative�
In-warranty repair and exchange To return a device to 3M for service, first obtain a Return Authorization (RA) number from a customer service representative� Please use the (RA) number on all correspondence when returning a device for service� A shipping carton will be delivered to you at no charge, if needed� Call your local supplier or sales representative to inquire about loaner devices while your device is being serviced�
When you call for technical support Remember, we will need to know the serial number of your unit when you call us� The serial number label is located on the bottom of the warming unit�
PDF Scaled at 100.0%
2
3M™ Ranger™ Blood/Fluid Warming System
English
3
Introduction The 3MTM RangerTM blood/fluid warming system includes a warming unit and a disposable warming set� The Ranger blood/fluid warming system is designed to warm blood, blood products, and liquids and deliver these at flow rates from KVO to 500 mL/min� At these flow rates, the device maintains fluid output temperatures ranging from 33°C to 41°C (Note: Output temperatures are dependant on input fluid temperatures and flow rate)� It takes less than 2 minutes to warm up to the 41°C set point temperature� Disposable blood/fluid warming sets are available in: pediatric flow, standard flow, and high flow applications� Warming sets are sterile, latex-free, single-use-only items and are designed to be used with the warming unit� The Ranger blood/fluid warming unit is designed to be mounted to an I�V� pole� A handle located on the top of the unit makes transport easy� When mounted to the I�V� pole, the unit fits easily above the 3MTM Bair HuggerTM warming unit� For additional information on the Ranger blood/ fluid warming sets visit us online at rangerfluidwarming�com� This manual includes operating instructions and unit specifications for the Ranger blood/fluid warming system� For information about using Ranger blood/fluid warming sets with the Ranger blood/fluid warming unit, Model 245, refer to the "Instructions for Use" included with each warming set� The Ranger blood/fluid warming system should only be used in healthcare facilities by trained medical professionals�
Indications for use The Ranger blood/fluid warming system is intended to warm blood, blood products, and liquids�
Definition of Symbols ON OFF Fuse CAUTION Type BF Applied Part Voltage, Alternating Current (AC) An equipotentiality plug (grounded) conductor other than a protective earth conductor or a neutral conductor, providing a direct connection between electrical equipment and the potential equalization busbar of the electrical installation� Please consult IEC 60601-1; 2005 for requirements�
PDF Scaled at 100.0%
English 34-8714-4406-2
English
3M™ Ranger™ Blood/Fluid Warming System
This system is subject to European WEEE Directive 2002/96/EC� This product contains electrical and electronic components and must not be disposed of using standard refuse collection� Please consult local directives for disposal of electrical and electronic equipment� Protective earth ground Consult instructions for use Follow instructions for use Recycle to avoid environmental contamination� This product contains recyclable parts� For information on recycling - please contact your nearest 3M Service Center for advice� Date of manufacture Manufacturer Sterile, ethylene oxide Store at room temperature DEHP-free Container quantity Single use Latex free Keep dry Temperature limits
Explanation of Signal Word Consequences WArNINg: Indicates a hazardous situation which, if not avoided, could result in death or serious injury.
cAUTION: Indicates a hazardous situation which, if not avoided, could result in minor or moderate injury.
NOTIcE: Indicates a situation which, if not avoided, could result in property damage only.
PDF Scaled at 100.0%
4
3M™ Ranger™ Blood/Fluid Warming System
English
5
WaRnIng: 1� To reduce the risks associated with hazardous voltage and fire and thermal energy hazards: • Do not substitute other devices (i.e. Model 247) for the Ranger blood/fluid warming unit or Ranger blood/fluid warming sets. • Do not continue use of the unit if the over-temperature alert continues to sound and the temperature does not return to the set point temperature. Immediately stop fluid flow and discard the warming set. Have the blood/fluid warming unit tested by a biomedical technician or call 3M Patient Warming technical service. 2� To reduce the risks associated with hazardous voltage and fire: • Do not modify or service this device, and do not open the warming unit case as there are no user-serviceable parts in the unit. • Connect power cord to receptacles marked “Hospital Only,” “Hospital Grade,” or a reliably grounded outlet. • Use only the power cord specified for this product and certified for the country of use. • Do not allow the power cord to get wet. • Do not use the Ranger blood/fluid warming system when it appears the unit power cord, or warming set is damaged. Use only 3M specified replacement parts. • Keep power cord visible and accessible at all times. The plug on the power cord serves as the disconnect device. The wall socket outlet shall be as close as practical and shall be easily accessible. 3� To reduce the risks associated with air embolism: • Never infuse fluids if air bubbles are present in the fluid line.
CauTIOn: 1� To reduce the risks associated with cross-contamination: • The cleaning tool provides only superficial cleaning, it does not disinfect or sterilize the interior of the unit. 2� To reduce the risks associated with impact and facility medical device damage: • Clamp the Ranger blood/fluid warming unit to an I.V. pole with a minimum 14” (35.6 cm) radius wheelbase and at a height no higher than 44” (112 cm). 3� To reduce the risks associated with environmental contamination: • Follow applicable regulations when disposing of this device or any of its electronic components. 4� Do not use for direct cardiac application� When using the Ranger blood/fluid warming system with a central venous catheter (CVC), ensure that the catheter tip does not have direct contact with the heart and ensure that all electrical devices connected to or near the patient have an appropriate leakage current rating for the application� If a CVC tip is found to be in direct contact with a patient’s heart, the blood/fluid warming unit, Model 245 should be immediately disconnected from the CVC until the CVC is safely repositioned� Failure to follow these precautions may cause cardiac disturbance and/or patient injury� 5� Follow the AABB Guidelines for the use of blood warming device which caution against warming when administering platelets, cryoprecipitates, or granulocyte suspensions�
PDF Scaled at 100.0%
English 34-8714-4406-2
English
3M™ Ranger™ Blood/Fluid Warming System
nOTICE: 1�
Federal law (USA) restricts this device to sale by or on the order of a licensed healthcare professional�
2�
To avoid device damage: • Do not clean the blood/fluid warming unit with solvents. Damage to the case, label, and internal components may result. • Do not immerse the blood/fluid warming unit in cleaning or sterilizing solutions. The unit is not liquid proof. • Do not insert metallic instruments in the blood/fluid warming unit. • Do not use abrasive materials or solutions to clean the heater plates. • Do not allow spills to dry inside the unit, as this may make it more difficult to clean the unit.
3�
The Ranger blood/fluid warming unit meets medical electronic interference requirements� If radio frequency interference with other equipment should occur, connect the unit to a different power source�
Product Description The Ranger blood/fluid warming system consists of a Model 245 warming unit and a sterile warming set�
The ranger blood/fluid warming unit The blood/fluid warming unit is a compact, lightweight, liquid-resistant device with a clamp located on the side for attachment to an I�V� pole� A carrying handle on the top of the unit makes it easy to transport� Located on the front panel you will find: • Alphanumeric display that indicates the heater temperature during normal operation� In an over-temperature condition, the display alternately flashes a temperature of 43°C or higher and the word “HI�” An audible alert also sounds� In an undertemperature condition, the display alternately flashes a temperature of 33°C or lower and the word “LO�”
Alert indicator light Slot for cassette
I.V. pole clamp Alphanumeric display
Power switch
Bubble trap holder
• Alert indicator light that comes on when either an over- or undertemperature condition occurs�
PDF Scaled at 100.0%
6
3M™ Ranger™ Blood/Fluid Warming System
English
7
ranger blood/fluid warming set Blood/fluid warming sets include a cassette, tubing and connectors compatible with hospital standard I�V� blood/fluid administration sets, bubble trap, and injection port� The high flow set also includes 2 blood/ fluid bag spikes and a 150 micron filter located in a central drip chamber�
Warming Set
Flow R ate
PRiming Volume
Patient line length
Standard Flow Set
150 mL/min
39-44 mL
76 cm 152 cm (76 cm patient line plus 76 cm extension)
High Flow Set
100-500 mL/min
150 mL
152 cm (76 cm patient line plus 76 cm extension)
Pediatric/Neonate
100 mL/min
20 mL
46 cm
Refer to the instructions provided with each warming set for information on use�
PDF Scaled at 100.0%
English 34-8714-4406-2
English
3M™ Ranger™ Blood/Fluid Warming System
Product safety features The following chart describes the safety alert features of the Ranger blood/fluid warming unit� aleRt tyPe
what to look FoR
descRiPtion / cause
action
Over-temperature alert - 43°C
Alert indicator light illuminates and alert sounds, alphanumeric display alternately flashes a temperature of 43°C or higher and the word “HI�”
Heater temperature rose to 43°C because of transient conditions�*
Observe alphanumeric display� If temperature does not drop to 41°C (may take a few minutes), discontinue use of unit� Call 3M customer service�
Under-temperature alert - 33°C
Alert indicator light illuminates and alert sounds, alphanumeric display alternately flashes a temperature of 33°C or lower and the word “LO�”
Heater temperature has fallen to 33°C�
Alerts stop when temperature rises above 33°C� Continue use of unit� If the temperature does not rise above 33°C, unplug the unit and call 3M customer service�
*Transient conditions may cause an over-temperature alert condition� These conditions include: • There was an extreme change in flow rates (e�g�, from 500 mL/min to stop flow)� • The unit was turned on and reached the set point temperature of 41°C before the warming cassette was inserted in the device� • Fluids were prewarmed to above 42°C before being infused�
PDF Scaled at 100.0%
8
3M™ Ranger™ Blood/Fluid Warming System
English
9
Warming units rev A to M aleRt tyPe
what to look FoR
Independent back-up • Alert indicator light safety alert - 46°C illuminates and alert sounds, alphanumeric display alternately flashes a temperature of 43°C or higher and the word “HI�” • Alphanumeric display is dark, alert sounds (backup safety alert still works even if display is dark)�
descRiPtion / cause
action
Heater temperature rose to 46°C� Safety backup system is activated at 46°C and unit shuts off power to the heating plates�
TURN OFF THE UNIT AND UNPLUG IT� Do not use the unit� Discard the disposable set� Call 3M customer service�
descRiPtion / cause
action
Heater temperature rose to 44°C� Safety backup system is activated at 44°C and unit shuts off power to the heating plates�
TURN OFF THE UNIT AND UNPLUG IT� Do not use the blood/fluid warming unit� Discard the set� Call 3M customer service�
Warming units rev N and newer aleRt tyPe
what to look FoR
Independent back-up • Alert indicator light safety alert - 44°C illuminates and alert sounds, alphanumeric display alternately flashes a temperature of 43°C or higher and the word “HI�” • Alphanumeric display is dark, alert sounds (backup safety alert still works even if display is dark)�
PDF Scaled at 100.0%
English 34-8714-4406-2
English
3M™ Ranger™ Blood/Fluid Warming System
Instructions for use Preparation and setup of the ranger blood/fluid warming unit 1� Attach the Ranger blood/fluid warming unit to the I�V� pole� Tighten the pole clamp securely� CauTIOn: To reduce the risks associated with impact and facility medical device damage:
44" (112 cm)
14" (35.6 cm)
• Clamp the Ranger blood/fluid warming unit to an I.V. pole with a minimum 14” (35.6 cm) radius wheelbase and at a height no higher than 44” (112 cm). 2� Slide the cassette into the slot in the warming unit� The cassette can only fit into the device one way� 3� Prime the warming set� For more information about priming the set, refer to instructions provided with the warming sets� 4� Place the bubble trap in the holder� 5� Plug the power cord into an appropriate outlet� Turn the unit ON� In a few seconds the alphanumeric display will illuminate� It takes less than two minutes to warm up to the 41°C set point temperature� 6� Begin infusion� When infusion is completed, remove the warming set and discard according to institutional protocol�
PDF Scaled at 100.0%
10
3M™ Ranger™ Blood/Fluid Warming System
English
11
removing the warming set from the ranger blood/fluid warming unit 1� Close the inlet clamp proximal to the cassette and open all clamps distal to the cassette� 2� Disconnect the warming set from the fluid source, if applicable� 3� Allow fluid to flow to the patient (this may take 2-3 seconds)� Close a distal clamp� 4� Remove the cassette from the warming unit and discard according to institutional protocol� 5� Reconnect the patient I�V� line to the fluid source to continue the infusion without warming�
Transferring the warming set from one ranger warming unit to another 1� Follow steps 1-3 above, then remove the warming set from the first warming unit� 2� During transport, keep the clamps closed and do not infuse fluids while the cassette is outside of the warming unit� 3� Slide the cassette into the second warming unit� 4� Ensure that air is removed from the tubing� 5� Open the clamps and continue the infusion�
PDF Scaled at 100.0%
English 34-8714-4406-2
12 English
3M™ Ranger™ Blood/Fluid Warming System
Troubleshooting condition
cause
solution
Nothing illuminates on the warming unit panel�
• Unit is not turned on, plugged in, or power cord is not plugged into an appropriate outlet�
• Turn unit on� Make sure the power cord is plugged into the power entry module of the warming unit� Make sure the warming unit is plugged into a properly grounded outlet�
• Unit failure�
• Call 3M customer service�
condition
cause
solution
Alert indicator illuminates and alert sounds, alphanumeric display alternately flashes a temperature of 43°C or higher and the word “HI�”
Temporary over-temperature condition because: • An extreme change in flow rates occurred (e�g�, from 500 mL/min to stop flow)� • Unit was turned on and reached set point temperature before cassette was inserted�
• Open flow to reduce temperature� Alerts will stop when the display reads 41°C� The unit is ready to use� • Alerts will stop when the display reads 41°C� The unit is ready to use�
condition
cause
solution
Alert sounds, alphanumeric display and alert indicator light go dark�
Primary controller failure� Unit will no longer operate�
Power to heating plates will shut off if temperature rises to 44°C (warming unit Rev N and newer) or 46°C (warming units Rev A to M)� Turn unit off and unplug it� Discontinue use of unit� Discard disposable set� Alert will continue to sound if you do not unplug unit� Call 3M customer service�
PDF Scaled at 100.0%
• Fluids were prewarmed to • Turn off unit and unplug it� above 42°C before being run Discontinue infusion of fluids� through the warming unit� Do not warm fluids before infusing them through the Ranger warming unit�
3M™ Ranger™ Blood/Fluid Warming System
condition
cause
Test screw on bottom of unit is • Unit alerts soon after plugging it in (unit does not loose or missing� have to be turned on for this condition to occur)� • Heater temperature rises to 44°C (warming units Rev N and newer) or 46°C (warming units Rev A to M) and unit shuts down soon after plugging it in (unit does not have to be turned on for this condition to occur)�
English
13
solution Make sure test screw is completely tightened� If it is missing, turn unit off and unplug it� Call 3M customer service�
condition
cause
solution
Alert sounds but unit has been turned off�
Independent backup safety system has been activated�
Unplug unit� Call 3M customer service�
condition
cause
solution
Cannot remove cassette from unit�
• Cassette is too full, fluids are still being infused, or clamp is open proximal to the cassette�
• Make sure fluid is drained from cassette before sliding out the cassette, that fluids are no longer infusing, and that clamp is closed proximal to the warming cassette�
• Warming unit is below patient level, creating excessive back pressure�
• Raise unit above patient level�
condition
cause
solution
Alert indicator light illuminates and alert sounds, alphanumeric display alternately flashes a temperature of 33°C or below and the word “LO�”
Under temperature condition caused by very high flow using very cold fluid, or defective heater/relay�
Alert should stop when temperature rises above 33°C� If alert continues, turn unit off, unplug unit and discontinue use� Call 3M customer service�
condition
cause
solution
Alphanumeric display reads “Er 4” or “Open�”
Open wire on temperature sensor�
Do not use unit� Call 3M customer service�
condition
cause
solution
Alphanumeric display reads “Er 5” or “Open�”
Electrical interference�
Remove the unit� Refer to biomedical technician or call 3M customer service�
PDF Scaled at 100.0%
English 34-8714-4406-2
English
3M™ Ranger™ Blood/Fluid Warming System
Maintenance and Storage cleaning the ranger blood/fluid warming unit Clean the Ranger warming unit on an as-needed basis� CauTIOn
1� To reduce the risks associated with cross-contamination: • The cleaning tool provides only superficial cleaning, it does not disinfect or sterilize the interior of the unit� nOTICE
1� To avoid device damage: • Do not immerse the warming unit in cleaning or sterilizing solutions� The unit is not liquid proof� • Do not clean the warming unit with solvents� Damage to the case, label, and internal components may result� • Do not insert metallic instruments in the warming unit� • Do not use abrasive materials or solutions to clean the heater plates� • Do not allow spills to dry inside the unit, as this may make it more difficult to clean the unit� to clean the exteRioR oF the waRming unit:
1� Disconnect the Ranger warming unit from the power source� 2� Wipe the outside of the unit with warm, soapy water, nonabrasive cleaning solutions, dilute bleach, or cold sterilants� Do not use abrasive materials� 3� Wipe with a dry, soft cloth� to clean the heating Plates:
The Ranger hardware cleaning tool is intended to clean both heating plates of the warming unit� It is not necessary to disassemble the warming unit to use the tool�
PDF Scaled at 100.0%
14
3M™ Ranger™ Blood/Fluid Warming System
English
15
Method 1� Unplug the warming unit� 2� Unfold the cleaning tool� Wet the foam pads with a nonabrasive solution, such as AlconoxTM brand detergent� 3� Insert the tool from the back of the unit and pull the tool all the way out from the front� 4� Rinse the tool with water and repeat 3 times� 5� Wipe off the unit to remove excess fluid� to clean Resistant, dRied-on Fluids:
1� Spray a nonabrasive solution inside the slot of the warming unit and let sit for 15-20 minutes� 2� Clean the unit by using the cleaning tool� Note: If the cleaning tool cannot be inserted through the slot of the warming unit because of excess dried-on fluids, send the unit to a biomedical technician�
Storage Store all components in a cool, dry place when not in use�
Servicing There are no user-serviceable parts in the Ranger blood/fluid warming unit� All service must be performed by 3M or an authorized service technician� Call 3M Technical Service at 800-228-3957 for service information� Outside of the USA contact your local 3M Patient Warming representative� Alconox is a trademark of Alconox, Inc�
PDF Scaled at 100.0%
English 34-8714-4406-2
English
3M™ Ranger™ Blood/Fluid Warming System
Specifications Physical characteristics warming unit
4�5 in� (11 cm) high x 7�5 in� (19 cm) wide x 10 in� (25 cm) long; wt�: 7 lb� 7 oz� (3�4 kg) certifications
IEC/EN 60601-1; CAN/CSA-C22�2, No�601�1 classification
MEDICAL - GENERAL MEDICAL EQUIPMENT AS TO ELECTRICAL SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601-1; CAN/CSA-C22.2, No.601.1; ANSI/AAMI ES60601-1:2005 CSA-C22.2 No. 60601-1:08; Control No.4HZ8 Classified under IEC 60601-1 Guidelines (and other national versions of the Guidelines) as Class I, Type BF, Ordinary equipment, Continuous operation� Classified by Underwriters Laboratories Inc� with respect to electric shock, fire and mechanical hazards only, in accordance with IEC/EN 60601-1 and in accordance with Canadian/CSA C22�2, No� 601�1� Classified under the Medical Device Directive as a Class IIb device�
Electrical characteristics
Temperature characteristics
device rating
set point temperature
100-120 VAC, 50/60 Hz 220-240 VAC, 50/60 Hz
41°C
maximum heating power
over-temperature alert
900 W
43°C
fuse
under-temperature alert 33°C
2 x T10A-H (250V) for 100-120 VAC 2 x T6�3A-H (250V) for 220-240 VAC fuse type
over-temperature cutoff
Time delay, high breaking
44°C (warming units Rev N and newer) 46°C (warming units Rev A to M)
Leakage current Meets leakage current requirements in accordance with UL/IEC 60601-1�
Environmental conditions operating temperature range
15° to 40°C (59° to 104°F) storage temperature range
-20° to 45°C (-4° to 113°F) operating humidity
10 to 85% RH, noncondensing atmospheric pressure range
50 kPa to 106 kPa
PDF Scaled at 100.0%
16
PDF Scaled at 100.0%
Made in the USA by 3M Health Care. 3M, RANGER, BAIR HUGGER, and the BAIR HUGGER logo are trademarks of 3M. Used under license in Canada. ©2014, 3M. All rights reserved. 3M Deutschland GmbH, Health Care Business Carl-Schurz-Str. 1, 41453 Neuss, Germany
PDF Scaled at 100.0%
3M Health Care, 2510 Conway Ave., St. Paul, MN 55144 USA TEL 800-228-3957 | www.bairhugger.com Issue Date: 2014-05 34-8714-4406-2