INSTRUCTIONS FOR USE ®
A-dec 500 Delivery Systems For A-dec 532, 533 and 542
A-DEC 500® DELIVERY SYSTEMS Instructions For Use
Copyright ©2006 by A-dec Inc. All Rights Reserved. 2601 Crestview Drive, Newberg, OR 97132, USA Printed in USA. A-dec Inc. makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. A-dec Inc. shall not be held liable for any errors contained herein or any consequential or other damages concerning the furnishing, performance or use of this material. The information in this document is subject to change without notice. If you find any problems in the documentation, please report them to us in writing. A-dec Inc. does not warrant that this document is error free. No part of this document may be copied, reproduced, altered or transmitted in any form or by any means, electronic or mechanical, including photocopying, recording or by any information storage and retrieval system, without prior written permission from A-dec Inc.
Publication Number: 86.0607.00 Revision Date: Rev C 2006-05
Trademarks A-dec logo, A-dec 500, Cascade, Cascade Master Series, Century Plus, Continental Decade, Performer, Preference, Preference Collection, and Radius are registered trademarks in the U.S. Patent and Trademarks office. A-dec, and ICX are also trademarks of A-dec Inc.
Comments and Feedback Thank you for taking the time to use the A-dec 500 Delivery Systems Instructions for Use. We would appreciate any feedback or comments you have about this document. Please mail, email, or phone us with your comments. You can reach us at: A-dec Inc. Technical Communications Department 2601 Crestview Drive Newberg, OR 97132 Reach us by phone at: 1.800.547.1883 email: [email protected] website: www.a-dec.com
A-dec 500 Delivery Systems Instructions for Use
CONTENTS 1 INTRODUCTION... 1 About A-dec 500 Delivery System... 1 Customer Service ... 3 Serial Numbers ... 3 2 FEATURES ... 5 Power On/Off Button ... 5 Delivery System Master On/Off Toggle ... 5 Handpiece Holder Positions ... 5 Activate Handpiece... 5 Handpiece Lockout ... 5 Accessory Holder ... 6 Foot Control ... 6 Chip Blower/Accessory Button ... 6 Wet/Dry Toggle ... 6 Flexarm Brake... 7 Touchpads... 7 Self-Contained Water System ... 8 3 OPERATION ... 9 System Operations ... 9 Master On/Off Toggle ... 9 Foot Control ... 10 Operate Touchpad ... 10 Status Icon ... 10 Chair Positions ... 11 Cuspidor Functions ... 12 Bowl Rinse ... 13 Dental Light... 13
86.0607.00 Rev C 2006-05
iii
A-dec 500 Delivery Systems Instructions for Use
Handpiece Settings (Deluxe Touchpad Only)... 14 Standard Mode ... 14 Endodontics Mode ... 17 Adjustments ... 19 Handpiece Control Adjustments ... 19 Positioning Adjustments... 23 4 MAINTENANCE ...27 Touch and Transfer Surfaces ... 27 Delivery System Maintenance ... 28 Handpiece Tubing Cleaning and Care ... 28 Heat Sterilization ... 28 Replace Bulb for Fiber-Optic Tubing ... 29 Oil Collector... 29 Waterline Treatment ... 30 Asepsis ... 30
iv
86.0607.00 Rev C 2006-05
List of Figures
List of Figures Figure 1 A-dec 532 Traditional Delivery System ... 2 Figure 2 A-dec 533 Continental Delivery System ... 2 Figure 3 A-dec 542 Side Delivery System ... 3 Figure 4 Serial Number Labeling ... 3 Figure 5 Optional Accessory Holder ... 6 Figure 6 Flexarm Brake ... 7 Figure 7 Standard and Deluxe Touchpads ... 7 Figure 8 Master On/Off Toggle ... 9 Figure 9 Foot Control ... 10 Figure 10 Status Icon ... 10 Figure 11 Chair Position Touchpad Buttons... 11 Figure 12 Cup Fill Button ... 12 Figure 13 Bowl Rinse Button... 13 Figure 14 Dental Light Button ... 13 Figure 15 Dental Light Composite Mode... 13 Figure 16 Operator Status... 15 Figure 17 Program Handpiece Standard Mode Settings ... 16 Figure 18 Forward/Reverse Button ... 16 Figure 19 Endodontics Mode Touchpad Screen ... 17 Figure 20 Handpiece Adjustment Keys... 19 Figure 21 Drive Air Adjustment... 21 Figure 22 Repositioned Holders on a Traditional Delivery System... 22 Figure 23 Tray Adjustment ... 22 Figure 24 Control Head Rotation Adjustment ... 23 Figure 25 Adjust Flexarm Rotation ... 23 Figure 26 Adjust Front-Mount Tension... 24 Figure 27 Flexarm Counterbalance Adjustment ... 24 Figure 28 Flexarm Adjustment ... 25 Figure 29 Flush Tubings ... 28 Figure 30 Bulb Replacement ... 29 Figure 31 Oil Collector Maintenance ... 29
86.0607.00 Rev C 2006-05
v
A-dec 500 Delivery Systems Instructions for Use
List of Tables Table 1 Symbols ...vii Table 2 Classification of Equipment ...vii Table 3 Environmental Specifications ...vii Table 4 Key to Month of Manufacture ... 3 Table 5 Chair Direction... 11 Table 6 Chair Position Factory Presets... 12 Table 7 Electric Motor, Air and Water Coolant Presets (Standard Mode)... 14 Table 8 Endodontics Mode Settings ... 18
vi
86.0607.00 Rev C 2006-05
Classification of Equipment (EN-60601-1) Identification of Symbols
Identification of Symbols Symbol
Description Recognized by Underwriters Laboratories Inc. with respect to electric shock, fire and mechanical hazards only in accordance with UL 60601-1 (2601-1) and under mutual recognition agreement with CAN/CSA C22.2, No. 601.1. Classified by Underwriters Laboratories Inc. with respect to electric shock, fire and mechanical hazards only in accordance with UL 60601-1 (2601-1) and under mutual recognition agreement with CAN/CSA C22.2, No. 601.1. UL listed to UL 61010A-1, BS EN 61010-2-010 and Canadian (CAN/CSA C22.2, No. 1010.1-92) safety standards. Conforms to applicable European Directives (refer to Declaration of Conformity). Protective earth (ground). Functional earth (ground). Attention, consult accompanying documents. No user serviceable parts. Attention, line voltage. Only licensed electrician should remove cover. Type B applied part. Class II equipment. Caution: Metal surfaces can be hot during and following the dry cycle.
Classification of Equipment (EN-60601-1) Type/Mode
Classification
Types of shock protection
CLASS I EQUIPMENT: Dental chairs, dental lights, and power supplies CLASS II EQUIPMENT: Chair, wall, and cart-mounted delivery systems
Degree of shock protection
TYPE B APPLIED PART: Delivery systems only
Degree of protection against water ingress
ORDINARY EQUIPMENT: All products
Mode of operation
CONTINUOUS OPERATION: All models except dental chairs CONTINUOUS OPERATION WITH INTERMITTENT LOADING: Dental chairs - 5% duty cycle
Flammable Gasses:
Not suitable for use in the presence of a flammable anesthetic mixture with air, oxygen, or nitrous oxide, where such gasses may accumulate in concentration (closed space).
Electrical Rating Type
Specification
Volts:
100/110-120/220-240 VAC
Frequency:
50-60 Hz
Current:
As configured and specified in equipment manual (products labeled 15A or greater require dedicated circuit, identified in distribution panel).
Environmental Specifications Temperature/Humidity Specification
86.0607.00 Rev C 2006-05
Storage/Transportation
Temperature: -40°C to 70°C (-40°F to 158°F) Relative humidity: 80% for up to 31°C, decreasing linearly to 50% at 40°C
Operating
Temperature: 10°C to 40°C (40°F to 104°F) Relative humidity: 80% for up to 31°C, decreasing linearly to 50% at 40°C.
Indoor Use:
Altitude up to 2,000M (6,563 ft.), installation category II, pollution degree 2. (UL 61010A-1 and CAN/CSA C22.2, No. 1010.1-92 only)
vii
A-dec 500 Delivery Systems Instructions for Use
Warranty A-dec warrants all products against defects in materials or workmanship for one year from time of delivery. A-dec’s sole obligation under the warranty is to provide parts for the repair, or at its option, to provide the replacement product (excluding labor). The buyer shall have no other remedy. All special, incidental, and coincidental damages are excluded. Written notice of breach of warranty must be given to A-dec within the warranty period. The warranty does not cover damage resulting from improper installation or maintenance, accident or misuse. The warranty does not cover damage resulting from the use of cleaning, disinfecting or sterilization chemicals and processes. The warranty also does not cover light bulbs. Failure to follow instructions provided in the A-dec Instructions For Use (operation and maintenance instructions) may void the warranty. A-dec warrants A-dec dental chair cylinders, both lift and tilt for ten years from the date of purchase of the chair or the cylinder. This warranty is retroactive to A-dec chair cylinders already in the field. The warranty covers chair cylinders A-dec finds to have manufacturing irregularities. Stool cylinders are covered under A-dec’s one-year warranty. No other warranties as to merchantability or otherwise are made. For service information, contact your local authorized A-dec dealer. Check with local codes and ADA (Americans with Disabilities Act) requirements for installation of this product. CAUTION Federal law restricts this device to sale by or on the order of a dentist, physician, or any other practitioner licensed by the law of the state in which he or she practices to use or order the use of this device.
Equipment Alterations Policy Certain modifications or alterations of A-dec equipment which expand the use of A-dec equipment beyond its design and intent, or which override any safety features of A-dec equipment may jeopardize doctor, patient or staff safety. Field modifications that alter the electrical and/or mechanical safety of A-dec dental devices are in conflict with Underwriters Laboratory (UL) construction file requirements and are not sanctioned by A-dec. Examples of field modifications that diminish safety design include, but are not limited to: rendering access to the line voltage without the use of tools, modification of supporting elements that increase or shift loading characteristics, and the addition of any powered device that exceeds the design limits of the dental system. The use of accessory equipment not complying with the equivalent safety requirements of A-dec equipment may lead to a reduced level of safety of the resulting
viii
86.0607.00 Rev C 2006-05
About This Document Equipment Alterations Policy
system. It is the responsibility of the equipment distributor and the installer to assure that the installation complies with all building code requirements. The responsibility to determine whether a modification or alteration of A-dec equipment falls within these constraints is with the person(s) who initiates, approves and/or performs such modification or alteration. A-dec will not respond to inquiries on an individual basis. This person(s) will be deemed to have assumed all associated risks with such alteration or modification and will hold A-dec harmless from resulting claims, including product liability claims. Additionally, such modification or alteration voids A-dec's warranty and may invalidate UL or other regulatory agency approval.
About This Document Welcome to the A-dec 500 Delivery Systems Instructions for Use. This guide provides an easy-to-use source of technical information for operating and maintaining your A-dec 500 Delivery Systems. Before using the A-dec 500 Delivery System, make note of all cautions and warnings. Read the component descriptions, and become familiar with the locations of features and controls. This document contains the following information: • Module descriptions • Feature descriptions • Adjustments • Maintenance tasks
Intended Audience This guide is intended for doctors and dental staff.
86.0607.00 Rev C 2006-05
ix
A-dec 500 Delivery Systems Instructions for Use
x
86.0607.00 Rev C 2006-05
A-dec 500 Delivery Systems Instructions for Use
1 INTRODUCTION This guide contains information on: • Features of the A-dec 500 delivery systems • Operating instructions • Adjustments • Programming • Maintenance • Specifications
About A-dec 500 Delivery Systems A-dec 500 Delivery Systems standard configuration has four handpiece control block positions and an A-dec autoclavable syringe. A-dec 532 and 542 are Traditional delivery systems with handpiece tubings located under the control head. The A-dec 533 is a Continental< delivery system with handpiece tubings located above the control head. The A-dec 532 and 533 are chair-mounted delivery systems that attach to the A-dec 511 dental chair (see Figure 1 and Figure 2). The A-dec 542 is a cabinet-mounted delivery system that attaches to a variety of Preference Collection® and Preference Slimline™ cabinets or mounts to a wall board (see Figure 3). Optional tray holders are available for the delivery systems. The trays mount above the control head on an A-dec 532 and 542 delivery system, or below the control head on an A-dec 533 Continental delivery system. Trays can be mounted on the right or the left side of the control head.
86.0607.00 Rev C 2006-05
1
A-dec 500 Delivery Systems Instructions for Use
Figure 1 A-dec 532 Traditional Delivery System
Figure 2 A-dec 533 Continental Delivery System
2
86.0607.00 Rev C 2006-05
INTRODUCTION Customer Service Figure 3 A-dec 542 Side Delivery System (Cabinet Mount Shown)
Customer Service For service information, contact your local authorized A-dec dealer.
Serial Numbers When you call customer service, please include the serial number for the product (see Figure 4) Figure 4 Serial Number Labeling
ICC REF:5590
A
S/N: F 4 4 3 3 0 8 B
C
(A) The REF number is the model number. (B) The first letter of the serial number is the month the product was manufactured. (C) The first digit of the serial number is the year of manufacture. For example, 4 is 2004. Table 1 Key to Month of Manufacture
86.0607.00 Rev C 2006-05
Letter
Month
Letter
Month
A
January
G
July
B
February
H
August
C
March
I
September
D
April
J
October
E
May
K
November
F
June
L
December
3
A-dec 500 Delivery Systems Instructions for Use
4
86.0607.00 Rev C 2006-05
A-dec 500 Delivery Systems Instructions for Use
2 F E AT U R E S This section describes the following A-dec 500 Delivery Systems features: • Power On/Off button • Handpiece holder positions • Accessory holder (optional) • Tray holders (optional) • Foot control • Flexarm brake • Touchpads
Power On/Off Button The delivery system has two power on/off locations: • The power on/off button, on the chair. • The master on/off toggle on the right side of the control head.
Delivery System Master On/Off Toggle The Master On/Off Toggle controls the power to the delivery system as well as other modules mounted to the chair.
Handpiece Holder Positions Activate Handpiece The handpieces automatically activate when you lift a handpiece from its holder or pull the Continental whip forward. Press the foot control to operate the activated handpiece.
Handpiece Lockout When a handpiece is removed from the holder and the foot control is pressed, the chair does not operate.
86.0607.00 Rev C 2006-05
5
A-dec 500 Delivery Systems Instructions for Use
Accessory Holder The optional accessory holder is a convenient holder position for additional accessories. The holder mounts on either the right or left side of the control head, in the unused touchpad position (see Figure 5). Figure 5 Optional Accessory Holder
A
(A) Accessory Holder
Foot Control The foot control regulates drive air to the active handpiece. It provides an air signal that activates air coolant and water coolant flow. The foot control is equipped with a wet/dry toggle and with a chip blower/ accessory button. NOTE The foot control has additional functionality when used with a deluxe touchpad or an A-dec Intraoral Camera.
Chip Blower/Accessory Button The chip blower/accessory button operates as either a chip blower or an accessory button. The chip blower function sends a jet of air through the handpiece without activating the handpiece bur. The accessory button function operates integrated accessory devices. For more details, see your authorized A-dec dealer.
Wet/Dry Toggle Use the wet/dry toggle to shut off the water coolant to a handpiece without moving your hands from the oral cavity.
6
86.0607.00 Rev C 2006-05
FEATURES Flexarm Brake
Flexarm Brake The flexarm brake allows you to place varying amounts of weight (8 lb. maximum) on the control head tray while maintaining the position. A thumb-actuated flexarm brake is located on each delivery system handle (see Figure 6). The brake restricts vertical movement but allows for side-to-side horizontal positioning of the control head. Figure 6 Flexarm Brake
Touchpads The A-dec 500 delivery system touchpads centralize treatment room controls into one touch surface (see Figure 7). Some touchpad buttons have indicators to alert you if the operation is functioning. Figure 7 Standard and Deluxe Touchpads
86.0607.00 Rev C 2006-05
7
A-dec 500 Delivery Systems Instructions for Use
Self-Contained Water System The self-contained water system supplies doctor’s syringes. It uses a 2 liter, quick disconnect water bottle, and mounts in a variety of locations.
8
86.0607.00 Rev C 2006-05
A-dec 500 Delivery Systems Instructions for Use
3 O P E R AT I O N This section describes how to: • Operate the A-dec 500 delivery system • Use the touchpad to operate the: • Chair • Cuspidor • Dental Light • Handpieces • Adjust the delivery system: • Height and positioning • Air pressure and water flow • Instrumentation holder adjustment and positioning
System Operations You can activate the system using the master on/off toggle. The foot control and optional touchpad (Standard or Deluxe) allow you to control delivery system functions with a minimum of touch surfaces.
Master On/Off Toggle • Flip the toggle to the right to turn the unit on. • Flip the toggle to the left to turn the unit off. NOTE To prolong the life of your equipment, turn the power off when it is not in use.
Figure 8 Master On/Off Toggle .
A
(A) Master On/Off Toggle
86.0607.00 Rev C 2006-05
9