ABBOTT LABORATORIES

i-STAT 1 User Guide Feb 2020

User Guide

8 Pages

Quality Check Messages and Codes (continued) Message  Cause  Action  Analyzer Error, Use Electronic Simulator  Handheld detects problem from which it is likely to recover.  Insert the Electronic Simulator. If PASS, continue to use handheld.  Analyzer Error, See Manual  Handheld detects problem from which it may not recover.  Insert Electronic Simulator. If PASS, insert a cartridge with sample or control. If the code does not reappear, continue to use handheld.  No display  Disposable batteries dead or rechargeable battery fully discharged. Keypad not responding. Start switch broken.  Change disposable batteries or recharge battery. If still no display, call Support Services.  Cartridge Locked does not disappear after test cycle completed  Dead battery(s). Mechanical problem.  Wait until handheld turns off. Turn handheld on. If resets, remove cartridge. If not, change or recharge battery(s) and turn handheld on.  i-STAT1 User Guide  06F16-57  (01)00054749002139 Abbott Point of Care Inc. 100 and 200 Abbott Park Road Abbott Park, IL 60064 • USA  Emergo Europe Prinsessegracht 20 2514 AP The Hague The Netherlands  For information related to Article 33 of the EU REACH regulation (EC No. 1907/2006), please refer to PMIS.abbott.com. If you have issues logging into the website, contact Abbott at: ... abbott.REACH@abbott.com. ©2020 Abbott Point of Care Inc. All rights reserved. Printed in USA. i-STAT is a trademark of the Abbott group of companies.  Art: 714254-01V  Rev. Date: 19-FEB-2020
File Type: PDF
File Size: 1.01 MB
File Name: ABBOTT LABORATORIES - 714254-01V - i-STAT 1 User Guide Feb 2020 - 2020-02.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions