ABBOTT LABORATORIES

i-STAT System

i-STAT User Guide Rev Date 4 Feb 2016

User Guide

8 Pages

Quality Check Messages and Codes (continued) Message  Cause  Action  Analyzer Error, Use Electronic Simulator  Handheld detects problem from which it is likely to recover.  Insert the Electronic Simulator. If PASS, continue to use handheld.  Analyzer Error, See Manual  Handheld detects problem from which it may not recover.  Insert Electronic Simulator. If PASS, insert a cartridge with sample or control. If the code does not reappear, continue to use handheld.  No display  Disposable batteries dead or rechargeable battery fully discharged. Keypad not responding. Start switch broken.  Change disposable batteries or recharge battery. If still no display, call Support Services.  Cartridge Locked does not disappear after test cycle completed  Dead battery(s). Mechanical problem.  Wait until handheld turns off. Turn handheld on. If resets, remove cartridge. If not, change or recharge battery(s) and turn handheld on.  i-STAT  06F16-57  (01)00054749002139 ®  i-STAT 1 User Guide  Abbott Point of Care Inc. 100 and 200 Abbott Park Road Abbott Park, IL 60064 • USA  Emergo Europe Molenstraat 15 2513 BH The Hague The Netherlands  For information related to Article 33 of the EU REACH regulation (EC No. 1907/2006), please refer to PMIS.abbott.com. If you have issues logging into the website, contact Abbott at: ... abbott.REACH@abbott.com. ©2016 Abbott Point of Care Inc. All rights reserved. Printed in USA. i-STAT is a registered trademark of the Abbott Group of Companies in various jurisdictions.  Art: 714254-01Q  Rev. Date: 04-Feb-16  ®
File Type: PDF
File Size: 965 KB
File Name: ABBOTT LABORATORIES - i-STAT User Guide Rev Date 4 Feb 2016 - 2016-02.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions