Ackermann
Ultra D-Grip clip applier, straight,280 mm, for medium-large clips
Reusable Surgical Instruments
12 Pages
Preview
Page 1
STENT PLACEMENT KITS
en
Reusable Surgical Instruments Important All products covered by these reprocessing instructions can be found in the appendix. Before using these products, please read the following information thoroughly! Warnings •
Reusable instruments from Ackermann Instrumente GmbH are supplied non-sterile, which is indicated on the product label by the following symbol:
Not for use in patients allergic to NI, CR, brass or aluminum. The products are contraindicated for all applications other than the intended use described. If there are any product-specific contraindications, these are listed in the instructions for use of the respective product.
Limitations on Reprocessing The reusable surgical instruments are designed for repeated use. The service life of the instruments results from reprocessing and handling. Therefore, a limit for the number of reprocessing cycles cannot be specified. Before each use, theproduct must be checked for function and damage. Damaged products must be disposed of. Warning
•
•
•
•
•
These products must be cleaned and sterilised before first use in accordance with the instructions in this reprocessing manual. The instruments may only be used by trained and, if applicable, instructed physicians. Disinfection, cleaning, and sterilization may only be performed by specially trained medical personal. The following instructions only apply to reusable Ackermann products. Please note that any deviation from these instructions, including the use of cleaners / detergents not specifically indicated in these instructions will require an evaluation of device specific efficacy and suitability of the actually performed cycle. Respective evaluation usually requires equipment qualification and device specific performance qualification / validation. In case of any serious incident that has occurred in relation to the respective device shall be reported to the manufacturer and the competent authority of the member state in which the user and/or patient is established. You can inform us using the form provided on our homepage. If product-specific instructions for use are enclosed with the products, these must also be followed.
Intended Use Reusable surgical instruments manufactured by Ackermann Instrument GmbH are intended to perform general surgical procedures. It is the responsibility of the attending physician to determine the use of each product for a specific operation. The intended use of the respective product group can be found in the appendix.
Contraindication(s) The instruments are contraindicated for use on the central circulatory system and central nervous system.
If the instruments are used on patients infected with prion-based diseases (e.g. CJD) or HIV, the instrument must not be reused. Even after reprocessing and sterilization, the risk of crosscontamination cannot be eliminated and the instrument must be destroyed. Inspection before Use Warning Before operating the instruments, a visual inspection and a functional test must be carried out in accordance with section “Functional control and maintenance of the products in the reprocessing cycle according to DIN 96298-4”. Non-functional instruments must be discarded. Reprocessing Instructions
Preparation before Use Warning Ackermann reusable products are supplied nonsterile and must be cleaned and sterilized before first use and before each subsequent use. The packaging cannot withstand the high temperatures of autoclaving and should be disposed of before sterilization.
Rinse
Cleaning Warning
Thermal disinfection
Improper cleaning, rinsing, and drying can result in potentially hazardous residues or inadequate sterilization.
Drying
Always ensure that the products are used and reprocessed by qualified personnel with appropriate experience in hospital hygiene and sterilization techniques. To ensure safe and effective reprocessing of the products, the following instructions have been validated by the manufacturer for effectiveness and compatibility with the instruments. It is the responsibility of the end user to ensure that cleaning and sterilization is indeed performed with appropriate equipment, materials and personnel to achieve the desired result. Any deviation from these instructions should be evaluated for effectiveness and possible adverse consequences. Warning All reusable Ackermann products must be subjected to reprocessing as described in the following sections before being used for the first time. Follow the instructions and warnings of the manufacturers of the decontamination, disinfection and cleaning agents to be used.
Cold DI water
times and wipe off excess oil with a lint-free cloth.
Exposure time
60 secs. / 1 min.
Visual and Fuctional Inspection
Neutralizing agent
acidic (e.g. Neodisher® Z)
Concentration
0.10%
Rinsing temperature
Cold DI water
Visual Inspection
Exposure time
120 secs. / 2 mins.
Disinfection temperature
90 °C (A0 3000)
Visual inspection for cleanliness • The instruments should be free of macroscopic contaminants.
Exposure time
300 secs. / 5 mins.
Drying
60 mins. / 1 hour
Pre-Cleaning Instruments, or fully disassembled instruments if applicable, must be brushed under cold water until all visible contamination is removed. If the instrument has a flushing port, the lumen of the shaft must be flushed for at least 10 seconds through the flushing port with a water jet gun (static pressure greater than 4 bar) after manual brushing. The instruments are then treated in a 0.5% alkaline enzymatic (e.g. Neodisher® Mediclean forte) ultrasonic bath for 10 minutes. Finally, rinse again for 30 seconds with a water jet gun.
Automated Cleaning 1 The automated cleaning was validated with Neodisher® Mediclean forte (alkaline-enzymatic: pH-value 10,4 – 10,8) for cleaning and Neodisher® Z (acidic: pH-value 3,0 – 2,6) for neutralization.
Steam Sterilization 2 Procedure
Triple fractionated pre-vacuum
Temperature
min. 132°C
Duration
min. 3 mins.
Drying time
min. 20 mins.
The instructions provided above have been validated as being capable of preparing a medical device for reuse. It remains the responsibility of the processor to ensure that the processing, as actually performed using equipment, materials and personnel in the processing facility, achieves the desired result. This requires verification and/or validation and routine monitoring of the process. 1
Warning
Rinsing temperature
Step
Parameter
Pre-rinse
Rinsing temperature
Cold tap water
Exposure time
180 secs. / 3 mins.
Cleaning temperature
55°C
Exposure time
600 secs. / 10 mins.
Detergent
Alkalineenzymatic (e.g. Neodisher® Mediclean forte)
Cleaning
Concentration
2
0.50%
Products marked as nonfunctional must be discarded. The following tests and maintenance must be performed by qualified personnel.
Visual control for missing parts • All parts such as screws, pins, springs, valves, hooks, etc. must be present. If this is not the case, the whereabouts of the missing parts must be clarified. Visual control for breakage • If parts (such as tips, teeth, springs, working ends, TC inserts) are missing due to material breakage, the whereabouts of the missing parts must be clarified. Check for deformation • Instruments with deformed working ends, cutting edges, tips, etc. need to be sorted out. Visual inspection for corrosion • Instruments with visible corrosion in the form of pitting, fretting, stress or crevice corrosion must be sorted out. Visual inspection for discoloration • Instruments with a discoloration must be sorted out.
Validated with washer-disinfector Miele PG 8535 Validated with steam autoclave Lautenschläger ZentraCert
Visual inspection for cracks • Instruments with a crack must be sorted out.
Function Control and Maintenance of the Products in the Reprocessing Cycle acc. to DIN 96298-4
Visual inspection of coatings • Coatings must be free of imperfections and notches. If damage to a coating is detected, this instrument must be sorted out.
Warning If the product is damaged or suspected of being damaged, do not attempt to repair the product yourself. Do not reuse damaged instruments. Warning If a product-specific inspection or maintenance is required, the relevant information can be found in the instruction of use for the respective product.
Maintenance Maintenance is always performed prior to functional testing. Oils can be applied by squeeze, spray, or pipe. Oil the joints and sliding surfaces in a directed way, then move the joint and sliding surfaces several
Visual inspection of markings • The marking must be legible. If the instrument is marked with a machine-readable marking, this must be readable with an appropriate scanner. If the marking is no longer legible, the instrument must be sorted out.
Fuctional Check Function check for clamps • The clamps must have a uniform, backlashfree movement. • The lock must hold securely in any position. • The working ends must be congruent with each other. • The teeth, if any, must mesh without hooking.
Function check for scissors • When closing the scissors, there must be minimal and continuous friction between the scissor blades, but no noticeable snagging. A cutting test can be carried out to eliminate any doubts. Function check for needleholders • The needleholder must have a smooth, backlash-free movement. • The catch must hold securely in any position. • The working ends must be congruent. • The working ends must close completely; if a gap is detected, the tooth profile must be checked for wear and replaced if necessary. Function check for tweezers • The teeth must engage without hooking. This is checked by pressing the tweezer legs together. The tips of the forceps (grasping surfaces) must be congruent with each other and must not slide sideways from each other when the forceps legs are pressed together. Function check for retractors • The retractor must hold securely in any position during opening. This is checked by opening and simultaneously applying pressure to the arms. Valves must be easy to insert and still fit securely. Functional check for chisels and osteotomes with handles made of plastics or aluminum • The components of the instrument must be firmly connected to each other without wobbling. The working ends must be checked for the presence of nicks. Impact caps with severe deformation or splintering must be discarded. Function check for bone splitters and roungeurs • The rongeurs with spring must open independently. The cutting edges should not have any major nicks, otherwise the function according to the intended purpose is no longer guaranteed. Rongeurs with severely damaged cutting edges must be discarded. Labeling and Storage Labeling Warning Please pay attention to the symbols on the product label and in this instruction for use. Care must be taken to ensure that the instruments are fully traceable at all times. Storage Conditions Storage conditions can be found on the product label.
IFU_00156_Reusable_Surgical_Instruments_INT_08/24
0483
STENT PLACEMENT KITS
Warning The storage area must be dust-free, low in microbiological contamination, dark and free from temperature fluctuations.
•
Prepare the instrument before returning it to the manufacturer. Return the reconditioned instrument in its original packaging to the manufacturer.
•
Symbols used on labeling (acc. ISO 15223-1)
Contact Ackermann Instrumente GmbH Eisenbahnstrasse 65-67 78604 Rietheim-Weilheim Germany
Disposal Environmentally sound disposal enables valuable raw materials to be recycled. Dispose of the device in an environmentally friendly manner in accordance with the valid hospital guidelines.
Phone: Fax: E-mail: Web:
Legal manufacturer
+49 (0)7461 966 17 - 0 +49 (0)7461 966 17 - 70 [email protected] www.ackermannsurgical.com
Batch code End of document
Repairs
Product number
Despite being used as intended, medical devices are subject to varying degrees of wear depending on the intensity of use. •
•
Non-sterile
Please note that repairs may only be carried out by the manufacturer or authorised personnel. The medical devices must be cleaned, disinfected and sterilised before being sent in for repair. Soiled or contaminated medical devices must not be sent.
Do not use if package is damaged
Keep dry
Medical device
Unique device identifier
Additional Information Do not exceed maximum loading capacity of the sterilizer when processing multiple instruments in one sterilization cycle.
Consult instruction for use
Caution
Packaging Packaging suitable for steam sterilization must comply with the requirements according to ISO 11607 / ANSI / AAMI ST79 / AAMI TIR12, for example, disposable sterilization packs (single or double packs) temperature resistant up to at least 137°C (279°F) and sufficient steam permeability, which provide sufficient protection against mechanical damage, or sterilization containers which need to be maintained according to the manufacturer‘s instructions.
Keep away from sunlight
ONLY
Only by prespcription from medical doctor
Quantity
0483
mdc medical device certification GmbH Kriegerstraße 6 70191 Stuttgart, Germany
Warranty A guarantee of two years from handover to the end customer is given for material and manufacturing defects. Transport costs and shipping risk are not assumed. In addition, the warranty stated in the General Terms and Conditions applies. This warranty does not apply to instruments that have been repaired, modified or improperly handled by unauthorised persons. In this case, no liability is assumed for the operational safety of the instrument either. In addition, the warranty becomes null and void.
Further Information is a trademark Instrumente GmbH.
of
Ackermann
Warning Risk of infection due to non-sterile instruments.
IFU_00156_Reusable_Surgical_Instruments_INT_08/24
0483
STENT PLACEMENT KITS
Appendix
Product Group
Intended Use
Basic UDI
Cannula, reusable, injection, non-sterile, laparoscopy
The instrument is intended to be manually-operated to perform various needle functions (e.g., injection, aspiration, irrigation, puncture).
42503022034A0N4
Guiding instrument, reusable, non-sterile, arthroscopy
The instrument is intended to simultaneously place, angle, and guide a rotating surgical drill on hard tissue (e.g., bone).
42503022045A0NG
Endoscopic forceps, flexible, reusable, biopsy, non-sterile, gynecology
The instrument is intended to obtain tissue specimens for examination to determine the presence or extent of a disease.
42503022046A0NM
Endoscopic forceps, rigid, reusable biopsy, non-sterile, arthroscopy
The instrument is intended to obtain tissue specimens for examination to determine the presence or extent of a disease.
42503022047A0NS
Endoscopic forceps, rigid, reusable biopsy, non-sterile, bronchoscopy
The instrument is intended to obtain tissue specimens for examination to determine the presence or extent of a disease.
42503022048A0NX
Endoscopic forceps, rigid, reusable biopsy, non-sterile, laparoscopy
The instrument is intended to obtain tissue specimens for examination to determine the presence or extent of a disease.
42503022049A0P4
Endoscopic forceps, rigid, reusable biopsy, non-sterile, urology
The instrument is intended to obtain tissue specimens for examination to determine the presence or extent of a disease.
42503022050A0MW
Endoscopic forceps, rigid, reusable cutting/punching, non-sterile, ENT
The instrument is intended to punch/cut/excise soft-tissue.
42503022051A0N3
Endoscopic forceps, rigid, reusable cutting/punching, non-sterile, arthroscopy
The instrument is intended to punch/cut/excise soft-tissue.
42503022052A0N8
Dilator, reusable, non-sterile, arthroscopy
The instrument is intended to dilate a body lumen or access channel.
42503022053A0ND
Drilling instrument, reusable, non-sterile, arthroscopy
The instrument is intended to cut and remove circular bone sections by rotary motions.
42503022054A0NJ
Endoscopic retractor, reusable, non-sterile, laparoscopy
The instrument is intended to lift urogenital structures.
42503022055A0NP
Endoscopic forceps, flexible, reusable, grasping, non-sterile, gynecology
The instrument is intended to grasp and manipulate tissue or additionally foreign bodies.
42503022056A0NU
Endoscopic forceps, flexible, reusable, grasping, non-sterile, urology
The instrument is intended to grasp and manipulate tissue or additionally foreign bodies.
42503022057A0NZ
Endoscopic forceps, flexible, reusable, grasping, non-sterile, arthroscopy
The instrument is intended to grasp and manipulate tissue or additionally foreign bodies.
42503022058A0P6
Endoscopic forceps, rigid, reusable grasping, non-sterile, bronchoscopy
The instrument is intended to grasp and manipulate tissue or additionally foreign bodies.
42503022059A0PB
Endoscopic forceps, rigid, reusable grasping, non-sterile, ENT
The instrument is intended to grasp and manipulate tissue or additionally foreign bodies.
42503022060A0N5
Endoscopic forceps, rigid, reusable grasping, non-sterile, laparoscopy
The instrument is intended to grasp and manipulate tissue or additionally foreign bodies.
42503022061A0NA
Endoscopic forceps, rigid, reusable grasping, non-sterile, urology
The instrument is intended to grasp and manipulate tissue or additionally foreign bodies.
42503022062A0NF
Impact instrument, reusable, non-sterile, arthroscopy
The instrument is intended to position an orthopaedic implant within the body by transmission of an impact force that drives the implant into its final position.
42503022063A0NL
Knife, reusable, non-sterile, arthroscopy
The Instrument is intended to manually cut or dissect tissue.
42503022066A0P3
Knife, reusable, non-sterile, ENT
The Instrument is intended to manually cut or dissect tissue.
42503022067A0P8
Knife, reusable, urethrotome system, non-sterile, urology
The Instrument is intended to manually cut or dissect tissue.
42503022068A0PD
Endoscopic forceps, rigid, reusable optical forceps, non-sterile, bronchoscopy
The instrument is intended to grasp and manipulate tissue or additionally foreign bodies.
42503022069A0PJ
Probe, reusable, non-sterile, arthroscopy
The instrument is intended to explore, measure, and/or perform therapeutic procedures.
42503022070A0NC
Ring applicator, reusable, non-sterile, laparoscopy
The instrument is intended to apply small elastic rings to block the fallopian tubes.
42503022071A0NH
Endoscopic scissors, flexible, reusable, non-sterile, gynecology
The instrument is intended to cut tissue or sutures.
42503022072A0NN
Endoscopic scissors, flexible, reusable, non-sterile, urology
The instrument is intended to cut tissue or sutures.
42503022073A0NT
Endoscopic scissors, rigid, reusable, non-sterile, urology
The instrument is intended to cut tissue or sutures.
42503022074A0NY
Endoscopic scissors, flexible, reusable, non-sterile, Bronchoscopy
The instrument is intended to cut tissue or sutures.
42503022075A0P5
Endoscopic scissors, rigid, reusable non-sterile, laparoscopy
The instrument is intended to cut tissue or sutures.
42503022076A0PA
Hammer, reusable, surgical, non-sterile, arthroscopy
The Instrument is intended to strike a surface or another device.
42503022077A0PF
Surgical procedure kit, reusable, non-sterile, arthroscopy
The instrument kit is intended to perform a specific surgery.
42503022078A0PL
Screwdriver, reusable, non-sterile, arthroscopy
The instrument is intended to fit into a screw head to introduce/remove a screw.
42503022079A0PR
Endoscopic forceps, rigid, reusable suture passer, non-sterile, arthroscopy
The instrument is intended to pass sutures or a needle/suture assembly through tissues/prosthetic materials to facilitate suturing.
42503022080A0NK
Endoscopic forceps, rigid, reusable suture passer, non-sterile, laparoscopy
The instrument is intended to pass sutures or a needle/suture assembly through tissues/prosthetic materials to facilitate suturing.
42503022081A0NQ
Knot pusher, reusable, non-sterile, arthroscopy
The instrument is intended to advance extracorporeal suture knots to the point of appropriate suture tension.
42503022082A0NV
Knot pusher, reusable, non-sterile, laparoscopy
The instrument is intended to advance extracorporeal suture knots to the point of appropriate suture tension.
42503022083A0P2
Myoma screw, reusable, non-sterile, gynecology
The instrument is intended to engage, stabilize, and manipulate a myoma (i.e., a uterine fibroid) and other gynaecological tissues during their surgical removal.
42503022084A0P7
Awl, reusable, non-sterile, arthroscopy
The instrument is intended to bore holes in bone.
42503022085A0PC
Curette, reusable, non-sterile, arthroscopy
The Instrument is intended for scraping out or off bones.
42503022086A0PH
Curette, reusable, non-sterile, ENT
The Instrument is intended for scraping out or off bones.
42503022087A0PN
Tendon stripper, reusable, non-sterile, arthroscopy
The Instrument is intended to excise a length of ligament, tendon or fascia for use as living graft.
42503022088A0PT
Rasp, reusable, bone, non-sterile, arthroscopy
The Instrument is intended to be used to scrape and smooth the surface of bone.
42503022090A0NS
Endoscopic forceps, rigid, reusable clip applier, non-sterile, ligation system
The instrument is intended to apply small atraumatic clips for the ligation of blood vessels, or similar tubular anatomy.
42503022091A0NX
Forceps, open-surgery, reusable, clip applier, non-sterile, ligation system
The instrument is intended to apply small atraumatic clips for the ligation of blood vessels, or similar tubular anatomy.
42503022092A0P4
IFU_00156_Reusable_Surgical_Instruments_INT_08/24
STENT PLACEMENT KITS
Endoscopic forceps, rigid, reusable clip remover, non-sterile, ligation system
The instrument is intended to grasp, open, and remove a tissue clip.
42503022093A0P9
Forceps, open surgery, reusable, clip remover, non-sterile, ligation system
The instrument is intended to grasp, open, and remove a tissue clip.
42503022094A0PE
Probe, reusable, non-sterile, laparoscopy
The instrument is intended to explore, measure, and/or perform therapeutic procedures.
42503022095A0PK
Retractor, reusable, hand-held, non-sterile, laparoscopy
The instrument is intended to separate/draw aside the margins of a wound/incision to allow access to tissues/organs.
42503022096A0PQ
Retractor, reusable, system component, non-sterile, retraction and positioning
The Instrument is intended to separate the margins of a wound/incision.
42503022097A0PV
Elevator/lever, reusable, non-sterile, ENT
The instrument is intended to lift, position or pry anatomical structures or surgical material.
42503022098A0Q2
Syringe, reusable, bladder, non-sterile, urology
The product is intended for the injection or withdrawal of liquids or gases into or out of the body.
42503022099A0Q7
Syringe adapter, reusable, non-sterile, urology
The adapter is intended to enable the interconnection of compatible components.
42503022099A1Q9
Bladder evacuator, reusable, non-sterile, urology
The Product is intended to suction tissue and body fluids.
42503022102A0MG
Endoscopic forceps, rigid, reusable grasping, non-sterile, VATS
The instrument is intended to grasp and manipulate tissue or additionally foreign bodies.
42503022112A0MP
Endoscopic scissors, rigid, reusable non-sterile, VATS
The instrument is intended to cut tissue or sutures.
42503022113A0MU
Dilator, reusable, non-sterile, urology
The instrument is intended to dilate a body lumen or access channel.
42503022116A0NB
Knot pusher, reusable, non-sterile, VATS
The instrument is intended to advance extracorporeal suture knots to the point of appropriate suture tension.
42503022118A0NM
Endoscopic forceps, rigid, reusable biopsy, non-sterile, VATS
The instrument is intended to obtain tissue specimens for examination to determine the presence or extent of a disease.
42503022119A0NS
Endoscopic forceps, rigid, reusable needle holder, non-sterile, laparoscopy
The Instrument is intended to grasp a suture needle when pushing/pulling the needle and attached suture through tissue during suturing.
42503022120A0ML
Endoscopic forceps, rigid, reusable needle holder, non-sterile, VATS
The Instrument is intended to grasp a suture needle when pushing/pulling the needle and attached suture through tissue during suturing.
42503022121A0MR
Suction/irrigation cannula, rigid, reusable non-sterile, VATS
The instrument is intended for irrigation and/or aspiration of body cavities/wounds.
42503022122A0MW
Endoscopic forceps, rigid, reusable suture cutter, non-sterile, VATS
The instrument is intended for cutting suture.
42503022123A0N3
Speculum, reusable, nasal, non-sterile, ENT
The instrument is intended to create a channel for examination, suction, irrigation, or the insertion of another surgical device in a body opening.
42503022130A0MT
Knife, reusable, non-sterile, urology
The Instrument is intended to manually cut or dissect tissue.
42503022131A0MY
Endoscopic forceps, flexible, reusable, biopsy, non-sterile, urology
The instrument is intended to obtain tissue specimens for examination to determine the presence or extent of a disease.
42503022133A0NA
Gall bladder extractor, reusable, non-sterile, laparoscopy
The instrument is intended to remove the gallbladder.
42503022134A0NF
Clamp, reusable, penis clamp, non-sterile, urology
The instrument is intended to grasp the penis for manipulation during surgery.
42503022135A0NL
Suture cannula, reusable, non-sterile, laparoscopy
The instrument is intended to pass sutures or a needle/suture assembly through tissues/prosthetic materials to facilitate suturing.
42503022136A0NR
Awl, reusable, non-sterile, general surgery
The instrument is intended to bore holes in bone.
42503022137A0NW
Drill, reusable, orthopaedics, non-sterile, general surgery
The instrument is intended to cut and remove circular bone sections by rotary motions.
42503022138A0P3
Drilling instrument, reusable, non-sterile, general surgery
The instrument is intended to cut and remove circular bone sections by rotary motions.
42503022139A0P8
Hammer, reusable, surgical, non-sterile, general surgery
The Instrument is intended to strike a surface or another device.
42503022140A0N2
Dermatome, reusable, non-sterile, general surgery
The Instrument is intended to manually cut or dissect tissue.
42503022141A0N7
Dilator, reusable, bile, non-sterile, general surgery
The instrument is intended to dilate a body lumen or access channel.
42503022142A0NC
Dilator, reusable, vascular, non-sterile, general surgery
The instrument is intended to dilate a body lumen or access channel.
42503022143A0NH
Dissector / Enucleator, reusable, non-sterile, general surgery
The instrument is intended to separate a soft tissue or body structure from another.
42503022144A0NN
Elevator/lever, reusable, ENT, non-sterile, general surgery
The instrument is intended to lift, position or pry anatomical structures or surgical material.
42503022145A0NT
Elevator/lever, reusable, non-sterile, general surgery
The instrument is intended to lift, position or pry anatomical structures or surgical material.
42503022146A0NY
Guide, reusable, suture, non-sterile, general surgery
The Instrument is intended for passing a length of a flexible material/device, though a surgical site.
42503022147A0P5
File, reusable, orthopaedic, non-sterile, general surgery
The Instrument is intended to scrape and smooth the surface of bone.
42503022148A0PA
Vascular approximator, reusable, non-sterile, general surgery
The instrument is intened to grasp the torn or resected portions of a vessel to hold the portions in apposition for the purpose of reattaching one portion to the other in order to repair the structure.
42503022149A0PF
Implantation instrument, reusable, non-sterile, general surgery
The instrument is intended to aid navigation/placement of an implant/prosthesis or other instruments during implantation.
42503022150A0N9
Cannula, reusable, injection, non-sterile, general surgery
The instrument is intended to be inserted beneath the skin of patient while connected to a noninvasive device (e.g., syringe, secondary medication set) to administer and/or withdraw (aspirate) fluids/drugs.
42503022151A0NE
Intubation tube, reusable, non-sterile, general surgery
The instrument is intended to be inserted into an artificial opening made in the trachea (windpipe) during tracheotomy to provide a patent airway.
42503022152A0NK
Clamp, reusable, arterial, non-sterile, general surgery
The instrument is intended to grasp, manipulate, dissect, and/or clamp soft-tissues.
42503022153A0NQ
Clamp, reusable, clip applier, non-sterile, general surgery
The instrument is intended to apply small atraumatic clips for the ligation of blood vessels, or similar tubular anatomy.
42503022154A0NV
Clamp, reusable, vascular, non-sterile, general surgery
The instrument is intended to grasp, manipulate, dissect, and/or clamp soft-tissues.
42503022155A0P2
Clamp, reusable, grasping, non-sterile, general surgery
The instrument is intended to grasp and manipulate tissue or additionally foreign bodies.
42503022156A0P7
Clamp, reusable, penis clamp, non-sterile, general surgery
The instrument is intended to grasp the penis for manipulation during surgery.
42503022157A0PC
Clamp, reusable, sponge/bandage/cloth, non-sterile, general surgery
The instrument is intended to be used for non-dedicated grasping of devices, sponges and/or dressings during a procedure.
42503022158A0PH
Clamp, reusable, bone holding, non-sterile, general surgery
The instrument is intended to grasp and hold a bone.
42503022159A0PN
Curette, reusable, non-sterile, general surgery
The Instrument is intended for scraping out or off bones.
42503022160A0NG
Marker, reusable, corneal, non-sterile, general surgery
The instrument is intended to imprint, indent, and/or incise corneal tissue prior to an ophthalmic surgical procedure.
42503022161A0NM
IFU_00156_Reusable_Surgical_Instruments_INT_08/24
STENT PLACEMENT KITS
Chisel, reusable, non-sterile, general surgery
The instrument is intended to remove or cut through bone.
42503022162A0NS
Knife, reusable, non-sterile, general surgery
The Instrument is intended to manually cut or dissect tissue.
42503022163A0NX
Clamp, reusable, umbilical cord, non-sterile, general surgery
The instrument is intended to temporarily compress the umbilical cord immediately after birth.
42503022165A0P9
Needle, reusable, general-purpose, non-sterile, general surgery
The instrument is intended to be manually-operated to perform various needle functions (e.g., injection, aspiration, irrigation, puncture).
42503022166A0PE
Guide needle, reusable, non-sterile, general surgery
The instrument is intended to drain fluids.
42503022166A0PE
Clamp, reusable, needle holder, non-sterile, general surgery
The Instrument is intended to grasp a suture needle when pushing/pulling the needle and attached suture through tissue during suturing.
42503022168A0PQ
Needle, reusable, suture, non-sterile, general surgery
The instrument is intended to insert and pull suture material through tissue.
42503022169A0PV
Osteotome, reusable, non-sterile, general surgery
The Instrument is intended to cut and/or shape bone by percussion during an orthopaedic procedure.
42503022170A0NP
Tweezers, reusable, grasping, non-sterile, general surgery
The instrument is intended to grasp and manipulate tissue or additionally foreign bodies.
42503022171A0NU
Tweezers, reusable, grasping, ophthalmology, non-sterile, general surgery
The instrument is intended to grasp and manipulate tissue or additionally foreign bodies.
42503022172A0NZ
Rasp, reusable, bone, non-sterile, general surgery
The instrument is intended to scrape and smooth the surface of bone.
42503022173A0P6
Rasp, reusable, nose, non-sterile, general surgery
The instrument is intended to remove and profile the cartilage/bone of the nose.
42503022174A0PB
Retractor, reusable, self-retaining, non-sterile, general surgery
The instrument is intended to separate/draw aside the margins of a wound/incision to allow access to tissues/organs.
42503022175A0PG
Retractor, reusable, hand-held, non-sterile, general surgery
The instrument is intended to separate/draw aside the margins of a wound/incision to allow access to tissues/organs.
42503022176A0PM
Retractor, reusable, ophthalmology, hand-held, non-sterile, general surgery
The instrument is intended to separate/draw aside the margins of a wound/incision to allow access to tissues/organs.
42503022178A0PX
Retractor, reusable, system component, self-retaining, non-sterile, general surgery
The Instrument is intended to separate the margins of a wound/incision.
42503022179A0Q4
Retractor, reusable, ophthalmology, self-retaining, non-sterile, general surgery
The instrument is intended to separate/draw aside the margins of a wound/incision to allow access to tissues/organs.
42503022180A0NW
Rongeur, reusable, non-sterile, general surgery
The instrument is intended for cutting tough tissues (intervertebral disk) or bone, through a cutting/biting action.
42503022181A0P3
Saw, reusable surgical, manual, non-sterile, general surgery
The instrument is intended for cutting bones or other body tissue by means of a sawing action performed by the user.
42503022182A0P8
Saw, reusable, nose, non-sterile, general surgery
The instrument is intended for cutting bone and/or cartilage in the nose.
42503022183A0PD
Suction/irrigation cannula, rigid, reusable non-sterile, general surgery
The instrument is intended for irrigation and/or aspiration of body cavities/wounds.
42503022184A0PJ
Scissors, open-surgery, reusable, non-sterile, general surgery
The instrument is intended to cut tissue or sutures.
42503022185A0PP
Snare, reusable, nasal, non-sterile, general surgery
The instrument is intended to be inserted into the naris for the removal of tissue, typically polyps, tumours, and other abnormal tissue from the nasal cavity during ear/nose/throat (ENT) surgery.
42503022186A0PU
Snare, reusable, ear, non-sterile, general surgery
The instrument is intended to be inserted through the outer ear canal or through a surgical incision in the ear to remove tissue, typically tumorous or damaged tissue, from the ear during ear/nose/throat (ENT) surgery.
42503022187A0PZ
Snare, reusable, tonsil, non-sterile, general surgery
The Instrument is intended to remove the tonsils during ear/nose/throat (ENT) surgery.
42503022188A0Q6
Screwdriver, reusable, non-sterile, general surgery
The instrument is intended to fit into a screw head to introduce/remove a screw.
42503022189A0QB
Tendon stripper, reusable, non-sterile, general surgery
The Instrument is intended to excise a length of ligament, tendon or fascia for use as living graft.
42503022190A0P5
Probe, reusable, non-sterile, general surgery
The instrument is intended to explore, measure, and/or perform therapeutic procedures.
42503022191A0PA
Spatula/needle, reusable, ophthalmology, non-sterile, general surgery
The instrument is intended to remove a foreign body/object embedded in or adhering to the surface of the eye globe.
42503022192A0PF
Spatula, open-surgery, reusable, non-sterile, general surgery
The instrument is intended to manipulate tissue during various types of surgical interventions.
42503022193A0PL
Forceps, open-sugery, reusable, skin punching-, general surgery
The instrument is intended to punch/cut/excise soft-tissue.
42503022194A0PR
Syndesmotome, reusable, non-sterile, general surgery
The instrument is intended for cutting desmodontal fibres and for detaching the alveolodental ligaments.
42503022195A0PW
Marker, reusable, ophthalmologic, non-sterile, general surgery
The instrument is intended to measure the diameter, length, angles, and thicknesses of eye. This instrument is not calibrated.
42503022196A0Q3
Trephine, reusable, corneal, non-sterile, general surgery
The instrument is intended to cut and remove circular pieces of corneal tissue.
42503022197A0Q8
Trocar, reusable, general-purpose, non-sterile, general surgery
The instrument is intended to create a body cavity by puncture to facilitate the introduction of an access cannula.
42503022198A0QD
Trocar, reusable, tracheal, non-sterile, general surgery
The instrument is intended to create a body cavity by puncture to facilitate the introduction of an access cannula.
42503022199A0QJ
Catheter, reusable, ear, non-sterile, general surgery
The instrument is intended for middle ear drainage and/or temporary equalization of the pressure on both sides of the eardrum via insufflation.
42503022200A0MH
Forceps, open surgery, reusable, biopsy, non-sterile, general surgery
The instrument is intended to obtain tissue specimens for examination to determine the presence or extent of a disease.
42503022201A0MN
Forceps, open surgery, reusable, punch, ENT, non-sterile, general surgery
The instrument is intended to punch/cut/excise soft-tissue.
42503022202A0MT
Forceps, open surgery, reusable, grasping, non-sterile, general surgery
The instrument is intended to grasp and manipulate tissue or additionally foreign bodies.
42503022203A0MY
Forceps, open surgery, reusable, bone holding, non-sterile, general surgery
The instrument is intended to grasp and hold a bone.
42503022204A0N5
Forceps, open surgery, reusable, bone punch, non-sterile, general surgery
The instrument is intended to punch/cut/excise posterior rib ends and bone tissue.
42503022205A0NA
Forceps, reusable, tooth extraction, non-sterile, general surgery
The instrument is intended for the extraction of teeth.
42503022206A0NF
Forceps, open surgery, reusable, wire forceps, non-sterile, general surgery
The instrument is intended to cut, grip, tighten and/or twist wires that are being applied to the patient.
42503022207A0NL
Intubation laryngoscope rigid, reusable non-sterile, general surgery
The instrument is intended to manipulate the tongue, preventing it from obstructing the oropharynx and enabling a clear view of the trachea.
42503022216A0NN
Speculum, reusable, nasal, non-sterile, general surgery
The instrument is intended to create a channel for examination, suction, irrigation, or the insertion of another surgical device in a body opening.
42503022218A0NY
Speculum, vaginal, reusable, general surgery
The instrument is intended to create a channel for examination, suction, irrigation, or the insertion of another surgical device in a body opening.
42503022219A0P5
Dilator, reusable, urethral, non-sterile, general surgery
The instrument is intended to dilate a body lumen or access channel.
42503022220A0MX
IFU_00156_Reusable_Surgical_Instruments_INT_08/24
STENT PLACEMENT KITS
Speculum, reusable, ear, non-sterile, general surgery
The instrument is intended to create a channel for examination, suction, irrigation, or the insertion of another surgical device in a body opening.
42503022221A0N4
Mouth gag, reusable, non-sterile, general surgery
The instrument is intended to be placed between the teeth of the upper and lower jaws of a patient to maintain an open oral cavity.
42503022222A0N9
Catheter, reusable, ureteral, non-sterile, general surgery
The device is intended for ureter dilation, bypass of partial obstructions, urinary tract irrigation/drainage during surgery, stone removal, retrograde urography, guidewire placement, and/or brush biopsy.
42503022223A0NE
Dilator, uterine, reusable, general surgery
The instrument is intended to dilate a body lumen or access channel.
42503022224A0NK
Dental hand instrument, reusable, non-sterile, general surgery
The dental hand instruments are intended in dentistry to perform various diagnostic, preventive and therapeutic procedures.
42503022225A0NQ
Dilator, reusable, lacrimal, non-sterile, general surgery
The instrument is intended to dilate a body lumen or access channel.
42503022228A0P7
Dilator, reusable, tracheal surgery, non-sterile, general surgery
The instrument is intended to dilate a body lumen or access channel.
42503022229A0PC
Speculum, reusable, rectal, non-sterile, general surgery
The instrument is intended to create a channel for examination, suction, irrigation, or the insertion of another surgical device in a body opening.
42503022234A0NS
Syringe, reusable, general-purpose, non-sterile, general surgery
The product is intended for the injection or withdrawal of liquids or gases into or out of the body.
42503022235A0NX
Endoscopic forceps, rigid, reusable cutting, ENT, non-sterile, general surgery
The instrument is intended to be used to punch/cut/excise soft-tissue.
42503022237A0P9
Surgical procedure kit, reusable, non-sterile, general surgery
The instrument kit is intended to perform a specific surgery.
42503022239A0PK
Optical tube, reusable, rectal, non-sterile, general surgery
The cannula is intended to provide access from a small body opening (surgical/non-surgical) to the surgical site to allow access for endoscopes, instruments and/or fluids.
42503022240A0ND
Spatula, non-surgical, reusable, non-sterile, general surgery
The instrument is intended to manipulate tissue. The instrument is intended to be non-surgical.
42503022253A0P3
IFU_00156_Reusable_Surgical_Instruments_INT_08/24