Guide
5 Pages
Preview
Page 1
VENTILATION BEYOND LIMITS
fabian HFO with Masimo SET®
Pulse Oximetry from Masimo SET® is now available as new feature in the fabian HFO The integration of Masimo SET® with ACUTRONIC’s ventilator fabian HFO is offering new options for caregivers. Masimo SET® is widely used and has set new standards in the world of SpO 2 measurement. ® Due to its excellent built-in algorithms, Masimo SET is able to provide accurate SpO 2 values, pulse rates and Perfusion Index (PI) even under the most challenging patient conditions, such as patient motion, low perfusion, low signal amplitude, or intense ambient light. Conditions commonly found in the neonatology arena. It is the main reason, why ACUTRONIC has selected Masimo as their preferred provider of SpO 2 . The fabian HFO is the first dedicated neonatology ventilator in the world to offer an intelligent integration of the SpO 2 feature. In a first step it will directly improve the workflow for caregivers. Instead of using two units, they now can simply attach the Pulse Oximetry sensor via USB interface to the fabian HFO and display the SpO 2 values on the ventilator screen. This will save the clinicians more for the patient and allows an even better treatment. ACUTRONIC is aiming to provide our customers additional innovative advantages, by introducing closed-loop, or near-closed-loop features in our products in the near future. ACUTRONIC – Ventilation Beyond Limits. More information: www.acutronic-medical.com or www.masimo.com
Rev. 1.0
TDS – fabian HFO with Masimo SET®
1/5
VENTILATION BEYOND LIMITS
Technical Data Measuring Capabilities Displayed parameters
SpO 2 , pulse rate and PI (Perfusion Index)
Measuring method
Absorption spectrophotometry
Measuring capabilities
Pediatric and Neonatal SpO 2 : 1% to 100%
Measuring range
Pulse rate: 25 bpm to 240 bpm PI: 0,02 to 20
Measuring Accuracy Saturation (%SpO 2 )
0% to 69% not specified
During No Motion Conditions
70% to 100% Infants / Pediatrics: ± 2% Neonates: ± 3%
Saturation (%SpO 2 )
0% to 69% not specified
During Motion Conditions
70% to 100% Infants / Pediatrics / Neonates: ± 3%
Pulse Rate (bpm)
Infants / Pediatrics / Neonates: ± 3 bpm
During No Motion Conditions Pulse Rate (bpm)
Infants / Pediatrics / Neonates: ± 5 bpm
During Motion Conditions Pulse Rate (bpm)
Infants / Pediatrics / Neonates: ± 3 bpm
During Low Perfusion Conditions
Rev. 1.0
TDS – fabian HFO with Masimo SET®
2/5
VENTILATION BEYOND LIMITS
Measuring Accuracy Alarms
User-adjustable upper and lower alarm limit settings for SpO 2 , Pulse Rate Lower alarm limit for PI (Perfusion Index)
Module Connector
USB connection to fabian HFO USB port
Power Source
Powered from the ventilator
Product Specifications Length of the uSpO 2 cable
170 cm
Weight
125 gr
Environmental Requirements Temperature Range Operating:
5º C to +45º C
Storage:
-20º C to +60º C Relative Humidity Range
Operating:
10% to 95% non-condensing
Storage:
10% to 95% with packaging Atmospheric Pressure Range
Operating:
485 mmHg to 795 mmHg (57 kPa to 106 kPa)
Storage:
375 mmHg to 795 mmHg (57 kPa to 106 kPa)
Standards IEC 60601 – 1: 2001 EN 60601 – 1 EN 10993 The ACUTRONIC Masimo SET® uSpO 2 module complies with the Medical Devices Directive (MDD) 93/42/EEC and bears the CE mark.
Rev. 1.0
TDS – fabian HFO with Masimo SET®
3/5
VENTILATION BEYOND LIMITS
ACUTRONIC Masimo uSpO 2 USB Option Ordering Information Part Number
Description
7381 Masimo SET High Performance/ Low Power Module - USB with USB connector comes with: LNCS® Neo Sample Pack, consisting of 1 LNCS® Inf,1 LNCS® Neo, 1 LNCS® Inf
Sensor with Starter-Kit (only be ordered together with P/N 7381) Part Number 7390 P/N 7381
P/N 7382
Box of 20 pcs. P/N 7383
P/N 7384
Box of 20 pcs.
Rev. 1.0
TDS – fabian HFO with Masimo SET®
Box of 20 pcs.
4/5
VENTILATION BEYOND LIMITS
ACUTRONIC Masimo uSpO 2 USB Option Ordering Information Part Number
Description
7382 LNCS® Inf Infant SpO 2 Adhesive Sensors, 18 in • • • •
Weight 3 kg – 20 kg Single Patient Use Latex free Box of 20 pcs.
7383 LNCS® Neo Neonatal SpO 2 Adhesive Sensors, 18 in. • • • •
Weight < 3 kg or > 40 kg Single Patient Use Latex free Box of 20 pcs.
7384 ®
LNCS NeoPt Neonatal SpO 2 Adhesive Sensors, 18 in. • • • •
Weight < 1 kg Single Patient Use Latex free Box of 20 pcs.
7385
9-pin Masimo SET Tester (Simulator) Connects to LNCS patient cables
Rev. 1.0
TDS – fabian HFO with Masimo SET®
5/5