fabian Therapy evolution
Preparation and Start-Up
INITIAL SET-UP CONNECTIONS: MAIN POWER CORD, GAS SUPPLY HOSES Connect power cord with respective power source. The device can be operated within a range of 100VAC to 240VAC at 50/60Hz. No manual switch-over is required. Connect medical gas hoses (air and Oxygen) to the corresponding wall/bottle plug connectors. Gas supply pressure required must be in the range from 2,0 up to to 6,5 bar. AIR & OXYGEN connectors are of NIST type.
INITIAL SET-UP FRONT CONNECTIONS
NIV with fabian Non Invasive Ventilation (NIV) in Open Valveless System : • nCPAP • duoPAP • nAPRV • HHFNC
Infant Flow® variable flow Clinical performance
Reduction in imposed work of breathing
Moa G, Nilsson K. Crit Care Med. 1988;16:1238–1242
PATIENT CIRCUIT SET-UP
Patient circuit for NIV (i.e. Infant Flow) • nasal CPAP • nasal DuoPAP
Patient circuit for HFNC – High Flow Nasal Cannula 38/51
Neonatal patient circuit for NIV i.e. by using Inspire/Infant Flow Generator and Circuit*
* P/N 7195: includes P/N 5292 (NIV breathing circuit) and P/N 7186 (generator + nasal prongs)
Neonatal patient circuit for NIV i.e. by using Inspire/Infant Flow Generator and Circuit
Generator and nasal prongs (P/N 7186)
Neonatal patient circuit for NIV i.e. by using Inspire/Infant Flow Generator and Circuit*
* P/N 5292, NIV breathing circuit, compatible with F&PMR850 humidifier * P/N 5291 same as 5292 but includes also humidity chamber
Neonatal patient circuit for NIV i.e. by using Inspire/Infant Flow Generator and Circuit*
* P/N 5292, NIV breathing circuit, compatible with F&PMR850 humidifier * P/N 5291 same as 5292 but includes also humidity chamber
Neonatal patient circuit for NIV i.e. by using Inspire/Infant Flow Generator and Circuit*
* P/N 5291 or P/N 5292, NIV breathing circuit, compatible with F&PMR850 humidifier
ASSEMBLY of NIV DISPOSABLE CIRCUIT*
connect Infant Flow connecting tube: from NIV port to the humidifier chamber inlet
* P/N 5291 or P/N 5292, NIV breathing circuit, compatible with F&PMR850 humidifier
ASSEMBLY of NIV DISPOSABLE CIRCUIT*
connect inspiratory heated tube onto the humidifier chamber outlet
* P/N 5291 or P/N 5292, NIV breathing circuit, compatible with F&PMR850 humidifier
ASSEMBLY of NIV DISPOSABLE CIRCUIT*
connect a green connector
connect the other end of
to one end of proximal
proximal pressure line to the
pressure line
Prox connector on fabian
* P/N 5291 or P/N 5292, NIV breathing circuit, compatible with F&PMR850 humidifier
ASSEMBLY of NIV DISPOSABLE CIRCUIT*
connect the HEATER WIRE cable for the HEATING of the RESISTANCE inside the breathing circuit * P/N 5291 or P/N 5292, NIV breathing circuit, compatible with F&PMR850 humidifier
ASSEMBLY of NIV DISPOSABLE CIRCUIT*
INSERT the TEMPERATURE SENSORS into the breathing circuit: the dual pronged sensor at the chamber outlet and the single pronged sensor at the patient end. The dual pronged sensor measures both flow and temperature as gas leaves the chamber. The patient sensor controls the heated wires to maintain gas temperature throughout the circuit. * P/N 5291 or P/N 5292, NIV breathing circuit, compatible with F&PMR850 humidifier
ASSEMBLY of NIV DISPOSABLE CIRCUIT* INSERT the TEMPERATURE SENSORS onto the breathing circuit Push the airway probe into the breathing circuit. Make sure the airway probe is correctly located in its hole and that it’s pushed home.
* P/N 5291 or P/N 5292, NIV breathing circuit, compatible with F&PMR850 humidifier
ASSEMBLY of NIV DISPOSABLE CIRCUIT*
If needed, CONNECT the EXTENSION tube onto the breathing circuit The extension tube is prepared also for inhaled NO (Nitric Oxide) with Luer-lock connection for sampling line) * P/N 5291 or P/N 5292, NIV breathing circuit, compatible with F&PMR850 humidifier
ASSEMBLY of NIV DISPOSABLE CIRCUIT*
The NIV breathing circuit is now ready to be connected to nCPAP/duoPAP generator (i.e Inspire nCPAP™) * P/N 5291 or P/N 5292, NIV breathing circuit, compatible with F&PMR850 humidifier
ASSEMBLY of NIV DISPOSABLE CIRCUIT*
INSERT the nCPAP GENERATOR onto the connectors of breathing circuit