Advantage Medical
Model mme 1003 OBSTETRIC TENS Instruction Book
Instruction Book
10 Pages
Preview
Page 1
__________________________________
Manufactured in Australia by: Advantage Health Care Pty Ltd
CONTENTS __________________________________
incorporating Meyer Medical Electronics
SECTION
Sold and distributed by:
Page 2
PAGE
1
WARNING
4
2
DESCRIPTION
5
3
SPECIFICATIONS
9
4
OPERATION
11
5
GUARANTEE
17
Page 3
__________________________________
__________________________________
1 . . . WARNING
2 . . . DESCRIPTION
__________________________________
__________________________________
mme 1003 OBSTETRIC TENS should
The mme 1003 OBSTETRIC TENS has been
only be used under the supervision of medical
specifically developed for the relief of pain during
staff.
pregnancy and childbirth. The effectiveness of
The
TENS On NO ACCOUNT should the
mme 1003
OBSTETRIC TENS be connected to a mains
(Transcutaneous
Electrical
Nerve
Stimulation) for the relief of pain has long been known to the medical profession.
powered battery eliminator or similar power supply, as under certain conditions this could
HOW TENS WORKS
prove hazardous to the user.
TENS is a method of applying electrical
TENS should not be used on patients who are
impulses
fitted with cardiac pace-makers, or on those who
underlying peripheral nerves.
have a history of heart disorders.
activity affects the large diameter nerve fibres
through
the
skin
to
effect
This electrical
and facilitates the operation of a “Gating
Page 4
Page 5
the
and
The results of these studies have shown that a
effectively overrides the transmission of pain
significant reduction of pain may be achieved,
signals in the small diameter nerve fibres
with possible reduction in use of medication and
(Melzack & Wall 1965).
therefore lessening the risk of side effects to
Mechanism”
within
the
spinal
cord
both mother and child. It is also suggested that, TENS stimulates the production of naturally occurring pain relieving substances, called endorphins and encephalins, which act at segmental levels in the spinal cord, and also within the central nervous system to relieve pain and promote a feeling of well-being (Salar et al 1981).
The use of TENS in labour has been studied by many researchers, among them Augustinsson et al (1977), Robson (1979) and more recently by G. Bortoluzzi (1989) at Box Hill Hospital.
Page 6
Page 7
__________________________________
10
3 . . . SPECIFICATIONS
Obstetric mme
__________________________________
1003
9 8 7
5
4
6
6
3
5
2
8
4
1
9
7
0
3
OUTPUT
Biphasic Pulse (No net DC component)
1 Frequency
2
PULSE RATE
1 - 100 Hz
PULSE WIDTH
250 micro-seconds
1
Power On
Ready
Of Am plitude
AMPLITUDE
150 Volts (1K ohm load)
POWER SOURCE
9 Volt Battery
(Alkaline battery recommended) HAND CONTROL
2:1 Range (Amplitude)
DIMENSIONS
Width
Height
30 mm
Length
120 mm
MME 1003 Page 8
Page 9
65 mm
Each mme 1003 OBSTETRIC TENS SYSTEM is supplied in a carry case containing the following items:
__________________________________
4 . . .OPERATIONS __________________________________
1 Stimulator unit 1 Lead set 1 Hand control unit 1 9 volt battery 1 Instruction manual
It is strongly recommended that the use of the TENS unit is practised for some time before the expected birth, so that its operation is fully understood and the mother-to-be becomes familiar with the sensations produced. Use the smaller (50 mm x 50 mm) electrodes for this purpose.
Electrodes are available separately To prepare these electrodes for use, first insert the pins on the ends of the four way lead into the receiving lead attached to the self-adhesive electrodes.
Page 10
Page 11
Attach all four electrodes to the skin and press
on the unit by rotating the AMPLITUDE knob
them firmly against the skin (see diagram page
clockwise until it clicks. The red POWER ON
16).
and the green READY lights should both illuminate. If the READY light does not come on
IMPORTANT when practising using the TENS,
or is dim, the battery needs to be replaced. To
place the electrodes on your arm or leg, NOT in
replace the battery, remove the cover on the
the position that is used for labour and delivery.
back of the istrument, withdraw the battery and remove the battery connector. Install new
Having attached the electrodes as described
bayttery and replace cover.
above, make sure that the TENS unit is switched off (AMPLITUDE knob fully anti-clockwise), then
Now SLOWLY increase the AMPLITUDE setting
insert the plug on the four way lead into the
until a tingling sensation is felt under the
socket on the top right hand side of the TENS
electrodes.
unit.
Next, connect the HAND CONTROL by
AMPLITUDE a little at a time to become used to
inserting the three pin plug into the socket on the
the sensation. Do not increase it to point where
top left of the unit.
it becomes uncomfortable. As the AMPLITUDE
You are now ready to
Continue
to
increase
the
is increased, the red lights (bar graph) on the left
commence treatment.
of the machine will progressively illuminate. Set
the
FREQUENCY
knob
to
10
(this
corresponds to a frequency of 100 Hz), switch Page 12
These lights switch on and off at a rate set by the FREQUENCY knob but at a higher Page 13
FREQUENCY above (20Hz) they appear to be on continuously.
When you have finished using the TENS, switch it off and remove the electrodes from the skin. It
Next take the hand controller and GENTLY
is not necessary to remove the electrodes, but
press the plunger down - this boosts the output -
the leads should be disconnected from the
the more you press the plunger down the
electrodes until next used.
stronger the stimulation. This boost will be used to control your pain during contractions. When
Practise using your TENS from time to time, so
you release the plunger, the stimulation returns
that when you go into labour you will be
to the level set by the AMPLITUDE knob - this is called the background level. Note that as the plunger is depressed the reading on the bar graph will increase.
thoroughly conversant with its use. During labour and for delivery, the large self adhesive electrodes are used.
These are
applied to the lower back as shown in the diagram.
After using the TENS for some time, you will notice that the tingling sensation diminishes
To enhance adhesion, the area to
which the electrodes are to be applied should first be moistened with a damp sponge.
somewhat - this is called accommodation, so the AMPLITUDE should be increased a little to maintain a satisfactory background level. Page 14
Page 15
As was done during the practise sessions, with the frequency set on ten, set the background stimulation (AMPLITUDE) to a comfortable level (strong but not painful).
During contractions
increase the stimulation by using the hand control - this can be operated by the expectant mother herself or her partner - to relieve the pain.
As mentioned before, the background
level may need to be increased from time to time as accommodation occurs.
Page 16
ELECTRODE PLACEMENT
__________________________________
CHART
5 . . . GUARANTEE __________________________________ The
mme 1003 OBSTETRIC TENS UNIT is
covered by a full two year guarantee, covering the
materials
(except
batteries),
workmanship used in its construction.
and Should
any fault develop within the guarantee period, meyer medical electronics undertake to repair or replace the instrument free of charge. Simply return the defective equipment along with evidence of purchase date, and a description of the fault, to meyer medical electronics or its agent for service. This guarantee does not cover damage caused by misuse, accident or battery leakage and is not transferable. Page 17
Page 18