Aeroneb Solo - Controller Checklist ®
1
Connect the control module to the AC/DC adapter and plug into an AC power source. Ensure AC/DC light is on.
2
Connect the nebulizer unit to the control module using the control module cable.
3
Press and release the on/off button and ensure the 30 Min mode light is on.
4
Verify that the green 30 Min indicator light illuminates. If using in continuous mode, ensure that continuous mode light is illuminated.
Aeroneb Solo - Nebulizer Orientation
www.aerogen.com
For optimum use of the Aeroneb Solo please ensure it is in the optimum orientation at all times, in both continuous and intermittent settings.
þ
ý þ
ý
Aeroneb Solo - Troubleshooting If you experience any difficulty using the Aeroneb Solo nebulizer please contact your local Aerogen representative. IF IT HAPPENS:
IT COULD MEAN:
TRY THIS:
No Aerosol being produced in intermittent mode.
Air is trapped between the medication and the aperture plate.
Tap the nebulizer to dislodge the air.
No Aerosol being produced in continuous mode.
The feed system is not operating.
Check the feed system. Check the feed rate does not exceed the maximum input feed rate of 12mL per hour.
Air is trapped between the medication and the aperture plate.
Tap the nebulizer to dislodge the air.
Nebulizer was not turned on or connected to power.
Ensure that nebulizer is connected to power and turned on.
Rechargeable battery is depleted.
Recharge battery.
A 30 minute cycle was selected when connected to the continuous feed system.
Run a continuous cycle.
The fault indicator light illuminates.
The control module cable is incorrectly connected to the nebulizer, or electronics are malfunctioning.
Verify that the control module cable is correctly connected to both the nebulizer unit and the control module.
Battery will not recharge or battery light flashes while connected to AC/DC adapter.
It may be time to replace the battery.
Contact your local Aerogen sales representative with the serial number of your controller, located on the back of the unit.
The 30 Min or Continuous light illuminates but aerosol is not visible.
No medication in the nebulizer unit.
Refill medication through filler cap in the nebulizer unit.
Air is trapped between the medication and the aperture plate.
Tap the nebulizer to dislodge the air.
There is no power to the system.
Verify that AC/DC adapter is securely attached the to the control module.
Rechargeable battery is depleted.
Recharge battery.
Medication is left in the nebulizer unit after nebulization.
30 Min or Continuous indicator does not light when on/off power button is pressed.