Aesculap
285 mm x 5 mm, KAIRISON Pneumatic Bone Punch, Standard Footplate
KAIRISON Pneumatic Bone Punch Instructions for Use/Technical Description
6 Pages
Preview
Page 1
Aesculap® KAIRISON pneumatic bone punch
► Prior to each use, inspect the product for loose, bent, broken, cracked, worn, or fractured components. ► Do not use the product if it is damaged or defective. Set aside the product if it is damaged. ► Replace any damaged components immediately with original spare parts. ► To avoid damage to the working end: Carefully insert the product through the working channel (e.g. trocar). ► Only combine Aesculap products with each other. ► Operate the product only with compressed air or compressed nitrogen.
Legend
Single-use products
1 KAIRISON pneumatic bone punch 2 Punch slider part 3 Punch lower part 4 Sprung ball bearing 5 Bone ejector pin 6 Punch shaft complete 7 Shaft release/locking slider 8 Handle 9 Compressed-air connector (air inlet) 10 Compressed-air connector (air exhaust) 11 Safety button (to lock the trigger) 12 Trigger 13 Locking lever (for locking the slider part) 14 Hook 15 Actuator lever 16 Single-use tube set 17 Pressure reducer 18 Aesculap compressed-air tube 19 Compressed-air wall connector
Symbols on product and packages Sterilization using ethylene oxide
Infection hazard for patients and/or users and impairment of product functionality due to reuse. Risk of injury, illness or death due to contamination and/or impaired functionality of the product! ► Do not reprocess the product.
DANGER
Art. no.
Designation
FK888SU
Single-use tube set
The product is EO sterilized and sterile packed. The product must not be reused. ► Ensure that the product and its accessories are operated and used only by persons with the requisite training, knowledge, or experience. ► Read, follow, and keep the instructions for use. ► Use the product only in accordance with its intended use, see Intended use. ► Do not use products from open or damaged sterile packaging. ► Do not re-sterilize the product. ► Prior to each use, inspect the product for loose, bent, broken, cracked or fractured components. ► Do not use the product if it is damaged or defective. Set aside the product if it is damaged. ► Do not use the product after its use-by date.
Safe operation Risk of injury and/or malfunction! ► Always carry out a function check prior to using the product.
Not for reuse in intended applications as defined by the manufacturer
WARNING Use by
Risk of injury when using the product beyond the field of view! ► Apply the product only under visual control. WARNING
Caution, general warning symbol Caution, see documentation supplied with the product
Applicable to ► For item-specific instructions for use and information on material compatibility, see also the Aesculap Extranet
at www.extranet.bbraun.com
Components required for operation Art. no.
Designation
FK899R
Connect the handpiece
see brochure, KAIRISON pneumatic bone punch
Shafts
FK898R
Pressure reducer
Intended use
FK888SU
Single-use tube set
The KAIRISON pneumatic bone punch is used for removing bone, cartilage and tissue in various surgical disciplines (e.g. neurosurgery, orthopedics and traumatology).
see brochure, KAIRISON pneumatic bone punch
Aesculap compressed-air tube*
Available sizes
* Regarding the operation and function of this product, follow instructions for use TA011106.
For detailed information on available sizes/ variants, see KAIRISON pneumatic bone punch brochure.
Operating principle
Safe handling and preparation CAUTION Federal law restricts this device to sale by, or on order of a physician!
WARNING
CAUTION
Injury to the patient due to uncontrolled movement of the foot plate/sticking of the slider part in the bone! ► While punching, apply slight pull away force from the patient. ► After punching, keep the KAIRISON pneumatic bone punch closed and remove it from the bone before opening it. ► If the bone cannot be punched through, remove the closed KAIRISON pneumatic bone punch from the bone by slightly turning and pulling it away from the bone. ► Do not open the KAIRISON pneumatic bone punch while it is in contact with the bone. ► Note that only a limited volume of bone material can be punched (slider aperture larger than the bone volume to be taken). ► Do not use blunt shafts. ► To ensure proper operation of the KAIRISON pneumatic bone punch, make certain that the operating air pressure is at least 7 bar. Damage to, or destruction of the KAIRISON pneumatic bone punch caused by incorrect handling! ► Use the KAIRISON pneumatic bone punch only for its intended applications, see intended purpose. ► Avoid overstrain to the product due to turning or levering during the punching movement.
► Ensure that the product and its accessories are operated and used only by persons with the requisite training,
knowledge, or experience. ► Read, follow, and keep the instructions for use. ► Use the product only in accordance with its intended use, see Intended use. ► Remove the transport packaging and clean the new product, either manually or mechanically, prior to its initial sterilization. ► Store any new or unused products in a dry, clean, and safe place.
The KAIRISON pneumatic bone punch is powered by compressed air taken in through the compressed-air connector (supply air) 9 and passed out through the compressed-air connector (exhaust air) 10. The compressed air power is transmitted into the working direction via the accuator lever 15. Moveable slider part 2 is moved forward and the punching action is carried out (in two steps).
Preparation Non-compliance with the following rules will result in complete exclusion of liability on the part of Aesculap.
Start-up Note Use the KAIRISON pneumatic bone punch with authentic Aesculap products only, see Accessories/Spare parts. Observe the operating pressure of at least 7 bar to ensure full cutting performance. Note To insert the shaft, compressed air must be connected and the actuator lever must be in the neutral shaft-insertion position. Note If there is only one compressed-air supply, compressed-air bifurcation adaptor RT035 can be used. In this way the user can avoid having to reconnect the system repeatedly (when a pneumatic motor is used as well). ► Connect Aesculap pneumatic tube 18 to the pressure reducer 17, see Fig. 1. ► Connect Aesculap pneumatic tube 18 to compressed-air wall connector 19. ► Turn safety switch 11 to its OFF position, see Fig. 3. ► Connect single-use tube set 16 to handle 8 and pressure reducer 17. ► Connect compressed-air connector (exhaust air) 10. ► Connect compressed-air connector (supply air) 9. ► Insert the shaft, see Inserting the shaft in the handle. ► Turn safety switch 11 to its ON position, see Fig. 3.
The KAIRISON pneumatic bone punch 1 is ready for operation.
Safety devices
Single-use products
Shaft release/locking slider
WARNING
Risk of injury due to inadvertent activation of the KAIRISON pneumatic bone punch! ► When cleaning the punch reservoir, the O.R. assistant carrying out this task must keep locking lever 13 pressed down until the cleaning procedure is completed.
► Press shaft release/locking slider 13 (for locking the slider part 2), see Fig. 2. ► Clean the punch aperture. ► Release shaft release/locking slider 13.
Slider part 2 can be moved again. Safety switch Note Hand over or put down KAIRISON pneumatic bone punch 1 in safe mode only! ► Turn safety switch 11 to its ON position, see Fig. 3.
Trigger 12 is in operating position. ► Turn safety switch 11 to its OFF position, see Fig. 3.
Trigger 12 is in safe position.
Inserting the shaft in the handle ► Turn safety switch 11 to its OFF position, see Fig. 3. ► Prior to inserting shaft 6 in handle 8, verify that slider part 2 is in its correct position, see Fig. 14. ► Connect compressed air so that angle lever 15 moves to neutral insertion position, see Fig. 4. ► Engage hook 14 of shaft 6 in the handle location pin of handle 8 in the direction indicated by the arrow, see
Fig. 5. ► Insert shaft 6 as far as it will go in the direction of the arrow and push-rotate it down until it engages, see Fig. 6.
The KAIRISON pneumatic bone punch 1 is ready for operation. ► Make certain that shaft 6 is secured by locking button 7.
WARNING
Infection hazard for patients and/or users and impairment of product functionality due to reuse. Risk of injury, illness or death due to contamination and/or impaired functionality of the product! ► Do not reprocess the product!
Art. no.
Designation
FK888SU
Single-use tube set
General information Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore the time interval between application and processing should not exceed 6 h; also, neither fixating pre-cleaning temperatures >45 °C nor fixating disinfecting agents (active ingredient: aldehydes/alcohols) should be used. Excessive measures of neutralizing agents or basic cleaners may result in a chemical attack and/or to fading and the laser marking becoming unreadable visually or by machine for stainless steel. Residues containing chlorine or chlorides e.g. in surgical residues, medicines, saline solutions and in the service water used for cleaning, disinfection and sterilization will cause corrosion damage (pitting, stress corrosion) and result in the destruction of stainless steel products. These must be removed by rinsing thoroughly with demineralized water and then drying. Additional drying, if necessary. Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems: ■ Optical changes of materials, e.g. fading or discoloration of titanium or aluminum. For aluminum, the application/process solution only needs to be of pH >8 to cause visible surface changes. ■ Material damage such as corrosion, cracks, fracturing, premature aging or swelling. ► Do not use metal cleaning brushes or other abrasives that would damage the product surfaces and could cause corrosion. ► Further detailed advice on hygienically safe and material-/value-preserving reprocessing can be found at www.ak-i.org, link to Publications, Red Brochure – Proper maintenance of instruments.
Triggering the punch action The tissue/bone can be punched either in two steps or, by skipping the positioning step, in a single punching action: ■ Positioning: Careful application of the cutting edge at the intended punching position, applying only mild force with minimal cutting effect. ■ Punching/Cutting: Forceful punching/cutting of the tissue. ► For the positioning step, gently press actuating lever 12, see Fig. 7 (short arrow). ► For punching/cutting, firmly press actuating lever 12, see Fig. 7 (long arrow).
Detaching the shaft from the handle ► Turn safety switch 11 to its OFF position, see Fig. 3. ► Pull back locking button 7 in the direction of the arrow, see Fig. 8. ► Press down shaft 6 in the direction of the arrow, see Fig. 9. ► Disengage hook 14 of shaft 6 from handle adapter of handle 8 in the direction indicated by the arrow, see
Fig. 10. This completes detaching shaft 6 from the handle.
Disassembling the product before carrying out the reprocessing procedure ► Disassemble system components (tubes, shafts and pressure reducer) immediately after use.
Preparations at the place of use ► If applicable, rinse non-visible surfaces preferably with deionized water, with a disposable syringe for example. ► Remove any visible surgical residues to the extent possible with a damp, lint-free cloth. ► Transport the dry product in a sealed waste container for cleaning and disinfection within 6 hours.
Preparation before cleaning ► Check for debris on the sealing caps; clean if necessary. ► Seal compressed air connector (exhaust air) 10 and compressed-air connector (supply air) 9 at handpiece and
pressure reducer 17 with the sealing caps. ► Dismantle the product prior to cleaning, see Disassembling.
Cleaning/disinfection
Disassembling ► Push slider part 2 as far as it will go in the direction of the arrow, see Fig. 11. ► Remove slider part 2 from main part 3 in the direction of the arrow, see Fig. 12.
Product-specific safety notes on the reprocessing procedure
This completes disassembling of shaft 6.
Assembling
CAUTION
CAUTION Damage to the slider part caused by knocking the cutting edge! ► Verify that the guides are fitting properly. ► Note the pairing code (XXXX), see Fig. 13. ► Make certain the slider part is in its correct position (note the marking, see Fig. 14).
► Place the slider part 2 on the main part 3 in such a way that the ejector pin 5, bore and guides fit together, see
Risk of corrosion/functional failure due to incorrect processing! ► Do not apply ultrasonic cleaning to the handpiece and pressure reducer. ► Allow any fluid that has penetrated the device to run out immediately.
Fig. 13. ► Push forward slider2 in the direction indicated by the arrow until ball thrust piece 4 engages in its correct posi-
tion, see Fig. 14. Shaft 6 is assembled now and can be inserted in handle 8.
CAUTION
Validated reprocessing procedure General safety instructions Note Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for sterile processing. Note For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possible variants of CJD, observe the relevant national regulations concerning the reprocessing of products. Note Mechanical reprocessing should be favored over manual cleaning as it gives better and more reliable results. Note Successful processing of this medical device can only be ensured if the processing method is first validated. The operator/sterile processing technician is responsible for this. The recommended chemistry was used for validation. Note If there is no final sterilization, then a virucidal disinfectant must be used. Note For the latest information on reprocessing and material compatibility see also the Aesculap extranet at www.extranet.bbraun.com The validated steam sterilization procedure was carried out in the Aesculap sterile container system.
Damage to the product due to inappropriate cleaning/disinfecting agents and/or excessive temperatures! ► Use cleaning and disinfecting agents according to the manufacturer’s instructions which – are approved for plastics (handle), surgical steel (shafts) and aluminum (pressure reducer), – do not attack softeners (e.g. in silicone). ► Observe specifications regarding concentration, temperature and exposure time. ► Do not exceed the maximum permitted cleaning temperature of 55 °C.
CAUTION
Damage to the product/functional failure caused by fluids entering compressedair connector (exhaust air)10/ compressed-air connector (supply air) 9 at the handle and pressure reducer! ► Apply sealing caps.
► Use suitable cleaning/disinfecting agents if the product is put away in a wet condition. To prevent foam forma-
tion and reduced effectiveness of the process chemicals: Prior to mechanical cleaning and disinfection, rinse the product thoroughly with running water. ► Carry out ultrasound cleaning: – as an effective mechanical supplement to manual cleaning/disinfecting. – as a pre-cleaning procedure for products with encrusted residues, in preparation for mechanical cleaning/ disinfecting. – as an integrated mechanical support measure for mechanical cleaning/disinfecting. – for additional cleaning of products with residues left after mechanical cleaning/disinfecting.
Validated cleaning and disinfection procedure
Mechanical cleaning/disinfecting
For mechanical cleaning: Note Use appropriate (neutral, enzymatic and mild-alkaline) detergents on this product with its aluminum components.
Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883).
Validated procedure
Specific requirements
Reference
Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
Manual cleaning with immersion disinfection
■ Suitable cleaning brush ■ Disposable syringe, 20 ml ■ Drying phase: Use a lint-free cloth
Chapter Manual cleaning/disinfection and sub-chapter:
or medical compressed air Mechanical alkaline cleaning and thermal disinfection
■ Place the product in a tray that is suitable for cleaning (avoiding rinsing blind spots).
■ Shafts only
■ Place products in the tray with their hinges open. Manual pre-cleaning with brush and subsequent mechanical alkaline cleaning and thermal disinfection
■ Suitable cleaning brush ■ 20 ml disposable syringe ■ Place the product in a tray that is
■ Handpiece only
suitable for cleaning (avoiding rinsing blind spots).
■ Position the product in such a way that fluids can flow out.
■ Place the product in the correct
■ Chapter Manual cleaning with immersion disinfection
Mechanical alkaline cleaning and thermal disinfecting Machine type: single-chamber cleaning/disinfection device without ultrasound Phase
Step
T [°C/°F]
t [min]
Water quality
Chemical/Note
Chapter Mechanical cleaning/disinfecting and sub-chapter:
I
Prerinse
<25/77
3
D–W
-
■ Chapter Mechanical alkaline
II
Cleaning
55/131
10
FD-W
■ Concentrate, alkaline:
cleaning and thermal disinfecting
– pH = 13 – <5 % anionic surfactant
■ 0.5 % working solution
Chapter Mechanical cleaning/disinfection with manual pre-cleaning and sub-chapter:
■ Chapter Manual pre-cleaning
– pH = 11* III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD-W
-
V
Drying
-
-
-
According to the program for cleaning and disinfection device
with a brush
■ Chapter Mechanical alkaline cleaning and thermal disinfecting
position in a suitable wire basket. Manual pre-cleaning with brush and subsequent mechanical neutral or mild alkaline cleaning and thermal disinfection
■ Pressure reducer only
■ Suitable cleaning brush ■ 20 ml disposable syringe ■ Place the product in a tray that is suitable for cleaning (avoiding rinsing blind spots).
■ Position the product in such a way that fluids can flow out.
■ Place the product in the correct
Chapter Mechanical cleaning/disinfection with manual pre-cleaning and sub-chapter:
■ Chapter Manual pre-cleaning
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alkaline
with a brush
■ Chapter Mechanical neutral or mild alkaline cleaning and thermal disinfecting
► Check visible surfaces for residues after mechanical cleaning/disinfecting.
Mechanical cleaning/disinfection with manual pre-cleaning Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883).
position in a suitable wire basket.
Manual cleaning/disinfection ► Prior to manual disinfecting, allow water to drip off for a sufficient length of time to prevent dilution of the dis-
Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
infecting solution. ► After manual cleaning/disinfection, check visible surfaces visually for residues. ► Repeat the cleaning /disinfection process if necessary.
Manual pre-cleaning with a brush Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
Manual cleaning with immersion disinfection Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Disinfectant cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
I
Disinfecting cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Rinsing
RT (cold)
1
-
D–W
-
II
Intermediate rinse
RT (cold)
1
-
D–W
-
D–W: Drinking water RT: Room temperature *Recommended: BBraun Stabimed
III
Disinfection
RT (cold)
15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
IV
Final rinse
RT (cold)
1
-
FD-W
-
V
Drying
RT
-
-
-
-
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) RT: Room temperature *Recommended: BBraun Stabimed ► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure. Phase I ► Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are moistened. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe.
infection procedure. Phase I ► Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are moistened. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
Mechanical alkaline cleaning and thermal disinfecting Machine type: single-chamber cleaning/disinfection device without ultrasound Phase
Step
T [°C/°F]
t [min]
Water quality
Chemical
I
Prerinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD-W
■ Concentrate, alkaline: – pH = 13 – <5 % anionic surfactant
Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Drain any remaining water fully.
■ 0.5 % working solution – pH = 11*
Phase III ► Fully immerse the product in the disinfectant solution. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Rinse lumens at least 5 times at the beginning of the exposure time using an appropriate disposable syringe. Ensure that all accessible surfaces are moistened.
III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD-W
-
Phase IV ► Rinse/flush the product thoroughly (all accessible surfaces). ► Mobilize non-rigid components, such as set screws, joints, etc. during final rinse. ► Rinse lumens with an appropriate disposable syringe at least five times. ► Drain any remaining water fully.
V
Drying
-
-
-
According to the program for cleaning and disinfection device
Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure.
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alkaline ► Check visible surfaces for residues after mechanical cleaning/disinfecting.
Mechanical neutral or mild alkaline cleaning and thermal disinfecting
Storage
Machine type: single-chamber cleaning/disinfection device without ultrasound
► Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area. ► Store sterile packed single-use products dust-protected in a dry, dark and temperature-controlled room.
Phase
Step
T [°C/°F]
t [min]
Water quality
Chemical
I
Prerinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD-W
Neutral:
Maintenance To ensure reliable operation, the product must be maintained at least once a year. For technical service, please contact your national B. Braun/Aesculap agency, see Technical Service.
■ Concentrate: – pH neutral – <5 % anionic surfactant
■ 0.5 %* working solution
Troubleshooting list Malfunction
Detection
Cause
Remedy
Slider part not moving
Product does not work
No compressed air
Check/connect compressed-air supply
Product does not work
Insufficient compressed air
Check/connect compressed-air supply
No material removed on the bone
Blunt cutting edge
Repair by manufacturer
Mildly alkaline:
■ Concentrate: – pH = 9.5 – <5 % anionic surfactant
■ 0.5 % solution Punching is not carried out
III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD-W
-
Slider part does not return to its rest position
Compressed-air failure/ system defect
Dismount/remove shaft slider part from handle
V
Drying
-
-
-
According to the program for cleaning and disinfection device
Compressed-air leak (hissing)
Defective seals in handle
Repair by manufacturer
Product does not work
Shaft guides installed incorrectly
Install shaft guides correctly
Product does not work (air pressure drop)
Insufficient pressure
Check compressed-air supply (min. 7 bar operating pressure)
Incorrect slider part position
Components not centered correctly Correct positioning of the slider part and main part of the shaft.
Check positioning Sprung ball bearing must be engaged, see Assembling
Incorrect position of the actuator
Actuator not in insertion position
Connect compressed air at the handle
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner neutral
Repair by manufacturer
► Check visible surfaces for residues after mechanical cleaning/disinfecting. ► Repeat the cleaning/disinfecting process if necessary.
Shaft cannot be installed in the handle
Inspection, maintenance and checks
CAUTION
Damage (metal seizure/friction corrosion) to the product caused by insufficient lubrication! ► Prior to function checks, lubricate moving parts (e.g. joints, pusher components and threaded rods) with maintenance oil suitable for the respective sterilization process (e.g. for steam sterilization: Aesculap STERILIT® I oil spray JG600 or STERILIT® I drip lubricator JG598).
Technical Service ► Allow the product to cool down to room temperature. ► After each complete cleaning, disinfecting and drying cycle, check that the product is dry, clean, operational, and
free of damage (e.g. broken insulation or corroded, loose, bent, broken, cracked, worn, or fractured components). ► Dry the product if it is wet or damp. ► Repeat cleaning and disinfection of products that still show impurities or contamination. ► Check that the product functions correctly. ► Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service, see Technical Service. ► Assemble dismountable products, see Assembling. ► Check for compatibility with associated products.
Packaging
WARNING
The sterility of the KAIRISON pneumatic bone punch can not be assured if the connectors at the handle and pressure reducer are sealed! ► Prior to packaging, remove the sealing caps from the connectors at the handle and at the pressure reducer.
► Appropriately protect products with fine working tips. ► Place the product in its holder or on a suitable tray. Ensure that all cutting edges are protected. ► Pack trays appropriately for the intended sterilization process (e.g. in Aesculap sterile containers). ► Ensure that the packaging provides sufficient protection against recontamination of the product during storage.
Steam sterilization Note The shafts can be sterilized either in disassembled condition or in assembled condition. ► Check to ensure that the sterilizing agent will come into contact with all external and internal surfaces (e.g. by
opening any valves and faucets). ► Validated sterilization process – Steam sterilization using fractional vacuum process – Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665 – Sterilization using fractional vacuum process at 134 °C/holding time 5 min ► When sterilizing several products at the same time in a steam sterilizer, ensure that the maximum load capacity of the steam sterilizer specified by the manufacturer is not exceeded.
Sterilization for the US market ■ Aesculap advises against sterilizing the device by flash sterilization or chemical sterilization. ■ Sterilization may be accomplished by a standard prevacuum cycle in a steam autoclave. To achieve a sterility assurance level of 10-6, Aesculap recommends the following parameters: Aesculap Orga Tray/Sterile container (perforated bottom) Minimum cycle parameters* Sterilization method Prevacuum
Temp. 270 °F/275 °F
Time 4 min
Minimum drying time 20 min
*Aesculap has validated the above sterilization cycle and has the data on file. The validation was accomplished in an Aesculap sterile container cleared by FDA for the sterilization and storage of these products. Other sterilization cycles may also be suitable, however individuals or hospitals not using the recommended method are advised to validate any alternative method using appropriate laboratory techniques. Use an FDA cleared accessory to maintain sterility after processing, such as a wrap, pouch, etc. WARNING for the US market If this device is/was used in a patient with, or suspected of having Creutzfeldt-Jakob Disease (CJD), the device cannot be reused and must be destroyed due to the inability to reprocess or sterilize to eliminate the risk of crosscontamination.
Risk of injury and/or malfunction! ► Do not modify the product. WARNING ► For service and repairs, please contact your national B. Braun/Aesculap agency.
Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and forfeiture of applicable licenses. Service addresses Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 (7461) 95-1602 Fax: +49 (7461) 16-5621 E-Mail: [email protected] Or in the US: Attn. Aesculap Technical Services 615 Lambert Pointe Drive Hazelwood MO, 63042 Aesculap Repair Hotline Phone: +1 (800) 214-3392 Fax: +1 (314) 895-4420 Other service addresses can be obtained from the address indicated above.
Accessories/Spare parts Art. no.
Designation
FK881R
Storage for handle and pressure reducer
FK888SU
Single-use tube set
FK891R
Storage for shafts
FK898R
Pressure reducer
FK898214
Sealing cap for compressed-air connector (pressure reducer)
FK898215
Sealing cap
FK899R
Connect the handpiece
RT035
Compressed-air bifurcation
see brochure, KAIRISON pneumatic bone punch
Shafts
see brochure, KAIRISON pneumatic bone punch
Aesculap compressed-air tube
Technical Data Nominal operating pressure
8 bar
Weight (complete system)
450 g
Dimensions incl. shaft
330 mm x 150 mm x 35 mm
Disposal ► Adhere to national regulations when disposing of or recycling the product, its components and its packaging!
Distributor in the US/Contact in Canada for product information and complaints 3773 Corporate Parkway Center Valley, PA, 18034, USA TA-Nr. 011795
11/12
V6
Änd.-Nr. 45542