User Manual
16 Pages
Preview
Page 1
AKRON® BOBATH Models 4411, 4412, 4421, 4422
USER MANUAL
AKRON 4411, 4412, 4421, 4422
Contents 1.
Introduction
2
Warnings, Cautions General Warnings
2 2
2.
Installation
3
3.
Operation
4 5
4.
Accessories – Operating Instructions
5
Breathing Hole and Plug
5
5.
Cleaning and Decontamination
6
6.
Troubleshooting
6
7.
Maintenance
7
8.
Warranty and Service
8 9
9.
Revision Status
9
10.
Technical Specification
10
"The
mark on this product demonstrates conformity with the EC Directive 93/42/EEC."
1
AKRON 4411, 4412, 4421, 4422
1. Introduction Always keep this Operating Manual available for reference.
Warnings & Cautions WARNINGS Warnings identify possible hazards in procedures or conditions which, if not correctly followed, could result in death, injury or other serious adverse reactions.
CAUTIONS Cautions identify conditions or procedures which, if not correctly followed, could result in equipment failure or damage.
General Warnings WARNINGS Use only accessories that have been designed or approved for use with this couch. Do not use electric couches in the presence of flammable gasses such as an anaesthetic agent. Maximum Load 225 Kg (495 lbs) distributed evenly. Do not concentrate weight on either side of the couch. Plug acts as a disconnect device. Do not allow children to play with the couch at any time. Adequately supervise children in the proximity of the couch.
2
AKRON 4411, 4412, 4421, 4422
2. Installation Adjustable Foot WARNING Lock castors before adjusting.
A)
Lock castors first.
B)
Adjust foot.
C)
Lock foot by tightening half nut.
WARNING Do not insert the 2-pin mains connector directly into a 3-pin 13A mains outlet. Always use the adapter provided. Plug acts as a disconnect device. Electric Models Use adapter if 3 pin 13A plug is required. (Order reference 5S0409Z)
Ensure mains supply cable is not stretched when connected. Ensure, during couch use, no obstacles are in its path. Ensure hand / foot switch tubes or leads are not trapped by moving parts of the couch.
CAUTION When not in use, switch couch off.
3
AKRON 4411, 4412, 4421, 4422
3. Operation Elevation Controls WARNING Safely position patient before elevating couch. Position all electrical cables away from the moving parts. Position Hand/Foot switch where accidental operation is not possible. Hydraulic Models
PUMP PEDAL TO RAISE COUCH
Hydraulic Pedal
Electric Models
Press é to raise couch. Press ê to lower couch.
Handswitch
AUF
Press é to raise couch AB
Press ê to lower couch. Footswitch
4
LIFT PEDAL TO LOWER COUCH
AKRON 4411, 4412, 4421, 4422
3. Operation Wheel Locking WARNING Lock couch wheels/castors prior to patient transfer or treatment. The retractable castors are not designed for transportation of patients. Disconnect and coil electrical cord away from moving parts prior to repositioning couch. Castors
Castors Unlocked
Castors Locked
Head Section CAUTION Support upholstery section while manually altering angle.
Head Section Operating Lever
A)
Press lever UP to alter section angle.
B)
Support section when altering angle.
C)
Release lever to lock section in position.
Angles = +85° - 0°
4. Accessories – Operating Instructions Breathing Hole and Plug
Remove plug for patient prone position Press out from underneath
5
AKRON 4411, 4412, 4421, 4422
5. Cleaning and Decontamination WARNING Do not use Alkalis or other organic solvents for routine cleaning. Replace torn or damaged vinyl immediately. Always cover vinyl with fresh paper or sheet between patients. If contamination is known or suspected always decontaminate couch before use.
Frequency of routine couch cleaning will depend on exact application. CLEANING / DECONTAMINATION Wear suitable protective clothing Disconnect mains supply Locate couch in a suitable cleaning area with castors disengaged or wheels locked. A) B) C)
Using a clean damp cloth, warm water and soap carefully wipe couch from top to bottom, taking care around electrical components. Wipe couch carefully from top to bottom using a damp cloth and clean water. Wipe couch carefully dry with clean paper towels from top to bottom.
If DECONTAMINATING then repeat steps a) to c) using a dilute Hypochlorite / warm water mixture (10,000ppm) instead of warm water and soap. Safely discard cleaning / decontaminating materials. Wash hands thoroughly. Put couch back into service.
6. Troubleshooting – Electric Actuator Symptom Actuator will not work
Possible cause Power disconnected
Action Ensure mains supply is connected
Fuse blown
Replace fuse in socket
Hand/footswitch disconnected
Check that the hand/foot switch is fully inserted into the actuator
If the problem persists consult your local Akron Supplier.
6
AKRON 4411, 4412, 4421, 4422
7. Maintenance Check couch visually before use. Carry out routine servicing every six months or more frequently if in heavy use. Routine servicing or repair must only be performed by suitably qualified person.
WARNING Before commencing any maintenance activity, disconnect the couch from the electrical supply. Plug acts as disconnect device. Examine the couch for any signs of damage. Make sure that all nuts, bolts and other fasteners are tight. Do not over tighten as this may cause binding. Upholstery fasteners should be checked, tightened or replaced if necessary. Check section hinge pins to ensure that no pin protrudes. Check gas strut controls, throughout the full range of movement for locking and assistance. Check hydraulic couches for smooth operation and signs of leakage.
WARNING All electrical checks should be performed by a suitably qualified and licensed electrician. Neither BLUE or BROWN wires are to be connected to the earth terminal of a three pin plug. Examine the flexible cables for cuts, abrasions, kinking or other deterioration and replace if necessary. Ensure that the power supply connector plug is securely attached. Check mains plug connections are tight and fuse is correctly rated (5A).
7
AKRON 4411, 4412, 4421, 4422
8. Warranty and Service (UK only) Huntleigh Akron’s standard terms and conditions apply to this couch. A copy is available upon request. These contain full details of warranty terms and do not limit the statutory rights of the consumer. For service or maintenance please contact: Service Department Huntleigh Healthcare 310-312 Dallow Road Luton Beds LU1 1TD
Telephone: 01582 745700 Fax: 01582 745869
For information on this product, or any other Huntleigh Akron product, please contact: Huntleigh Akron 1 Farthing Road Ipswich IP1 5AP
Telephone: 01473 461042 Fax: 01473 462924
Huntleigh Akron T/A a Division of Huntleigh Nesbit Evans Limited A Member of the Huntleigh Technology PLC Group of Companies © Huntleigh Technology PLC 2001. If for any reason your couch is being returned, please inform Huntleigh Akron prior to sending, and: 1. 2. 3. 4.
Clean the product, as described in the cleaning section. Pack it in suitable packing. Attach the decontamination certificate (or other written statement declaring that the product has been cleaned) to the outside of the package. Mark the package “Service Department”.
Manufactured in the UK by Huntleigh Akron. As part of the ongoing development programme this company reserves the right to modify specifications and materials of this couch without notice.
8
AKRON 4411, 4412, 4421, 4422
8. Warranty and Service (UK only) Huntleigh Akron’s standard terms and conditions apply to this couch. A copy is available upon request. These contain full details of warranty terms and do not limit the statutory rights of the consumer. For service, maintenance and any questions regarding this, or any other Huntleigh Akron Product, please contact: Huntleigh Akron 1 Farthing Road Ipswich Suffolk IP1 5AP. England Telephone: +44 (0) 1473 461042 Fax: +44 (0) 1473 462924 Huntleigh Akron T/A a Division of Huntleigh Nesbit Evans Limited A Member of the Huntleigh Technology PLC Group of Companies © Huntleigh Technology PLC 2001. If for any reason your couch is being returned, please inform Huntleigh Akron prior to sending and: 1. 2. 3. 4.
Clean the product, as described in the cleaning section. Pack it in suitable packing. Attach the decontamination certificate (or other written statement declaring that the product has been cleaned) to the outside of the package. Mark the package “Service Department”.
Manufactured in the UK by Huntleigh Akron. As part of the ongoing development programme this company reserves the right to modify specifications and materials of this couch without notice.
9. Revision Status
4411, 4412, 4421, 4422 PAGES ALL PAGES ALL PAGES PAGES 8 & 9 PAGE 2
ISSUE UM - 01 UM - 02 UM - 03 UM - 04 (DCN958)
WRITTEN S.Rees S.Rees J. Austin J. Barker
9
DATE 26/05/98 07/01/99 18/11/99 30/10/00
AKRON 4411, 4412, 4421, 4422
10. Technical Specification
Head Length (cm) Foot Length (cm) Width (cm) Overall Length (cm) Height Range (cm) Max. Load Weight Elevation Control
4411
4412
4421
4422
71±1.5 124±1.5 106±1.5 196±3 46±1.5 91±1.5 225 Kg Hydraulic
71±1.5 124±1.5 106±1.5 196±3 46±1.5 91±1.5 225 Kg Electric
71±1.5 124±1.5 125±1.5 196±3 46±1.5 91±1.5 225 Kg Hydraulic
71±1.5 124±1.5 125±1.5 196±3 46±1.5 91±1.5 225 Kg Electric
Electrical Data 230v Actuator Internal Fuse Rt Current Power Mains Plug Fuse Rt (UK)
220-240v T1.6A x 1 1.6A 480W 5.0A x 1
220-240v T1.6A x 1 1.6A 480W 5.0A x 1
Electrical Data 115v Actuator Internal Fuse Rt Current Power
115v F4.0A x 1 3.3A 380W
115v F4.0A x 1 3.3A 380W
IEC 601-1 Classification Check serial number label for classification 230v
115v
Double Insulated
Double Insulated
Class 1
Type B equipment
Type B equipment
Type B equipment
Attention consult this manual
Attention consult this manual
Attention consult this manual
10
115v
AKRON 4411, 4412, 4421, 4422
11
AKRON 4411, 4412, 4421, 4422
12
Huntleigh Akron, 1 Farthing Road, Ipswich, Suffolk, IP1 5AP, England Tel: +44 (0) 1473 461042 Fax: +44 (0) 1473 462924
792-077-04