AKRON
AKRON Tables and Couches
AKRON STANDARD HI-LI TILT TABLE Models 8622, 8631, 8632 User Manual Feb 2001
User Manual
20 Pages
Preview
Page 1
AKRON® STANDARD HI/LO TILT TABLE Models 8622, 8631, 8632
USER MANUAL
AKRON 8622, 8631, 8632
Contents 1.
Introduction
2
Warnings, Cautions General Warnings
2 2
2.
Installation
3
3.
Operation
4
Elevation Controls Wheel Locking, Tilting Control Angle Indicator, Foot Board Work Table, Hand Grips Fixation Belts & Harnesses
4 5-6 7 8 9
Accessories
10
Emergency Quick Release
10
Cleaning and Decontamination
11
4. 5.
12
6.
Troubleshooting
12
7.
Maintenance
13
8.
Warranty and Service
14 15
9.
Revision Status
15
10.
Technical Specification
16
"The
mark on this product demonstrates conformity with the EC Directive 93/42/EEC."
1
AKRON 8622, 8631, 8632
1. Introduction Always keep this Operating Manual available for reference.
Warnings & Cautions WARNINGS Warnings identify possible hazards in procedures or conditions which, if not correctly followed, could result in death, injury or other serious adverse reactions.
CAUTIONS Cautions identify conditions or procedures which, if not correctly followed, could result in equipment failure or damage.
General Warnings WARNINGS Use only accessories that have been designed or approved for use with this table. Do not use electric couches in the presence of flammable gasses such as an anaesthetic agent. Maximum Load 180 Kg (396 lbs) distributed evenly. Do not concentrate weight on either side of the table. Plug acts as a disconnect device. Do not allow children to play with the table at any time. Adequately supervise children in the proximity of the table.
2
AKRON 8622, 8631, 8632
2. Installation WARNING Do not insert the 2-pin mains connector directly into a 3-pin 13A mains outlet always use the adapter provided. Electric Models
Use adapter if 3 pin 13A plug is required. (Order reference 5S0409Z)
Plug Adapter Ensure mains supply cable is not stretched when connected. Ensure, during table use, no obstacles are in its path. Ensure handswitch tubes or leads are not trapped by moving parts of the table.
CAUTION When not in use, switch table off. Plug acts as disconnect device.
CAUTION Ensure that the electricty supply is compatible with the table.
Serial No. Label Junction Box (Model 8622 only)
3
AKRON 8622, 8631, 8632
3. Operation Elevation Controls WARNING Safely position patient before elevating table. Position all electrical cables away from the moving parts. Position Hand/Footswitch where accidental operation is not possible. Hydraulic Model (8631) PUMP PEDAL TO RAISE TABLE
Electric Model (8622) Press é to raise table Press ê to lower table
Twin Hand Switch
Electric Model (8632)
Press é to raise table Press ê to lower table Handswitch AUF
Press é to raise table AB
Press ê to lower table Footswitch 4
LIFT PEDAL TO LOWER TABLE
AKRON 8622, 8631, 8632
3. Operation Wheel Locking WARNING Lock table wheels/castors prior to patient transfer or treatment. The retractable castors are not designed for transportation of patients. Disconnect and coil electrical cord away from moving parts prior to repositioning table. Castors
Castors Unlocked
Castors Locked
5
AKRON 8622, 8631, 8632
3. Operation Tilting Controls WARNING Ensure that the patient is safely positioned prior to tilting the table. Always secure the patient using the straps provided prior to tilting the table. Under no circumstances should the table be moved with the patent in situ and in tilt mode. In the interest of patient safety, always ensure that a member of staff (or other competent individual) is close at hand when the patient is in tilt. Ensure that all electrical cables are positioned away from the moving parts of the table. Care should be taken when tilting the tilt table into the vertical position. Ensure table is elevated sufficiently prior to tilting vertical as table may lift off wheels. Ensure no object is underneath footboard when tilting as this could cause entrapment. Models 8631 & 8632
Operating Lever
A)
Press lever UP to alter section angle.
B)
Support section when altering angle.
C)
Release lever to lock section in position.
Angles = 0° - 90°
Model 8622 Press é to raise table Press ê to lower table
6
AKRON 8622, 8631, 8632
3. Operation Angle Indicator
Tilt table as desired and view angle indicator.
Adjustable Foot Board WARNING For reasons of safety, always adjust the footboard to the negative position when the table is not in use. CAUTION When returning the table from tilt to horizontal it may be necessary to support the footboard to prevent it falling towards the upholstery.
A)
Rotate the locking screws located on the end of the chrome rails clockwise, until the foot board can swing into a positive position.
B)
Once in a positive position, the foot board can then be positioned as required by rotating the locking screws in an anti-clockwise direction.
7
AKRON 8622, 8631, 8632
3. Operation Accessory Work Table WARNING The Accessory Work Table has a safe working load of 20 Kg.
To fit the Work Table :A) B)
Insert Work Table Frame into sliding fixation brackets. Tighten locking knobs on fixation brackets to secure Work Table.
To adjust the height of the Work Table :Work Table
A) B) C)
Loosen locking knobs on sliding fixation brackets on both sides of the table. Slide Work Table Assembly along chrome rails to desired height Re-tighten locking knobs to secure Work Table.
To alter the angle of the Work Table :A) B) C)
Unscrew hand wheel on the underside of the Work Table. Adjust Work Table to the required angle. Tighten hand wheel to secure Work Table in the new position.
Hand wheel
Hand Grips To fit Hand Grips :A) B) C)
Remove Work Table (if fitted). Slide Hand Grips into fixation brackets on both sides of the table. Tighten locking knobs on fixation brackets to secure.
To adjust height of Hand Grips :A) B) C)
Loosen locking knobs on sliding fixation brackets on both sides of the table. Slide Hand Grips along chrome rails to desired height. Re-tighten locking knobs to secure Hand Grips.
8
AKRON 8622, 8631, 8632
3. Operation Fixation Belts and Harnesses WARNING Check condition of straps and harnesses. If worn out or damaged, replace immediately. Always ensure that the straps and harnesses are in the correct position and that the patient is fully secure before operating the tilt facility. Only use Huntleigh Akron straps and harnesses. The Standard Tilt Table is supplied with the following Fixation Harnesses and Belts (TT3)
2 x Narrow Belts (TT1)
1 x Wide Belt (TT2) When the Standard Tilt Table is supplied with Split Leg Facility a Split Leg Harness is also included.
Split Leg Harness (TT4) The Wide Belt is used across the patient’s knees and fastened using the Velcro strips which wrap around the chrome rails. One Narrow Belt is fitted across the chest with the other across the waist. The Split Leg Harness is an optional accessory and can only be used in conjunction with the optional split leg upholstery models (TTSL). This allows each leg to be secured individually by using the special harness. The centre of the harness is inserted into the split in the upholstery and is held in position by the plywood section fixed to the harness. The harness is then fitted around the legs individually and fixed to the chrome rails as previously described.
9
AKRON 8622, 8631, 8632
4. Accessories Emergency quick release mechanism (model 8622) CAUTION The Quick Release Mechanism is for EMERGENCY USE ONLY in the event of a power failure. It must not be used for routine lowering of the tilted section. Always support the upholstery section when changing the tilt angle.
Red Handle (A)
Red Handle and Release Lever together (B)
Plastic Retaining Clips (C)
To lower from tilt position in an emergency situation, grasp and pull Red Handle (A) downwards until table is at approximately 45°. Do not stand on any part of table frame to reach this handle. From this position, quickly transfer grip by grasping Red Handle and Release Lever together (B). Pull down until horizontal to allow removal of the patient from the table. Please note that from 45° angle, weight of patient will be positive and will need support as the table is lowered. Releasing the lever will immediately cause the table to lock into position. After use of ‘Emergency Quick Release’ function, the normal controls cannot be operated by the hand/foot switch until the table is returned to the same tilt angle prior to using the gas strut release lever. Use gas strut release lever grasped to handle to return table to tilt angle prior to emergency use. It can then be operated as normal with the hand/foot switch. The Red Handle, when not in use, can be secured to the frame by ‘twisting’ the Plastic Retaining Clips (C).
WARNING Do not stand on any part of table frame to reach the Red Handle. Do not re-position patient using the Quick Release Mechanism.
10
AKRON 8622, 8631, 8632
5. Cleaning and Decontamination WARNING Do not use Alkalis or other organic solvents for routine cleaning. Replace torn or damaged vinyl immediately. Always cover vinyl with fresh paper or sheet between patients. If contamination is known or suspected always decontaminate table before use.
Frequency of routine table cleaning will depend on exact application. CLEANING / DECONTAMINATION Wear suitable protective clothing. Disconnect mains supply. Locate couch in a suitable cleaning area with castors disengaged or wheels locked. A) B) C)
Using a clean damp cloth, warm water and soap carefully wipe table from top to bottom, taking care around electrical components. Wipe table carefully from top to bottom using a damp cloth and clean water. Wipe table carefully dry with clean paper towels from top to bottom.
If DECONTAMINATING then repeat steps a) to c) using a dilute Hypochlorite / warm water mixture (10,000ppm) instead of warm water and soap. Safely discard cleaning / decontaminating materials. Wash hands thoroughly. Put table back into service.
11
AKRON 8622, 8631, 8632
5. Cleaning and Decontamination Decontamination of Straps It is not recommended that the straps are laundered on a regular or routine basis as this will eventually compromise the integrity of the material and associated Tac-Flex. However, under certain conditions, ie when the straps become heavily soiled or when they are suspected or known to have been in contact with an infectious organism, they can be laundered or decontaminated using one of the following methods.
General Wash The straps should be washed separately, as the Tac-Flex will attract debris from other material which will ultimately reduce its effectiveness. The Tac-Flex should be joined prior to washing. The straps may then be laundered in the normal fashion, at a temperature of approximately 71°C and allowed to dry in ambient conditions. After such wash, the straps should be thoroughly inspected prior to patient use. If deterioration is present or the Tac-Flex fails to maintain integrity, the straps must be replaced.
Decontamination Eye protection and suitable clothing should be worn. Where contamination is known or suspected, the surface of the straps should be sprayed liberally using a hypochlorite solution of 1,000 ppm or a 1,000 ppm solution of sodium dichloroisocyanurate (NaDCC), after removing excessive organic material. The straps should then be air dried at ambient temperature and inspected (as above) prior to patient use.
6. Troubleshooting – Electric Actuator Symptom Actuator will not work
Possible cause Power disconnected
Action Ensure mains supply is connected
Fuse blown
Replace fuse in socket
Hand/footswitch disconnected
Check that the hand/foot switch is fully inserted into the actuator
If the problem persists consult your local Akron Supplier.
12
AKRON 8622, 8631, 8632
7. Maintenance Check table visually before use. Carry out routine servicing every six months or more frequently if in heavy use. Routine servicing or repair must only be performed by suitably qualified person.
WARNING Before commencing any maintenance activity, disconnect the table from the electrical supply. Plug acts as disconnect device. Examine the table for any signs of damage. Make sure that all nuts, bolts and other fasteners are tight. Do not over tighten as this may cause binding. Upholstery fasteners should be checked, tightened or replaced if necessary. Check section hinge pins to ensure that no pin protrudes. Check gas strut controls, throughout the full range of movement for locking and assistance. Check hydraulic tables for smooth operation and signs of leakage.
WARNING All electrical checks should be performed by a suitably qualified and licensed electrician. Neither BLUE or BROWN wires are to be connected to the earth terminal of a three pin plug. Examine the flexible cables for cuts, abrasions, kinking or other deterioration and replace if necessary. Ensure that the power supply connector plug is securely attached. Check mains plug connections are tight and fuse is correctly rated (5A).
13
AKRON 8622, 8631, 8632
8. Warranty and Service (UK only) Huntleigh Akron’s standard terms and conditions apply to this couch. A copy is available upon request. These contain full details of warranty terms and do not limit the statutory rights of the consumer. For service or maintenance please contact: Service Department Huntleigh Healthcare 310-312 Dallow Road Luton Beds LU1 1TD
Telephone: 01582 745700 Fax: 01582 745869
For information on this product, or any other Huntleigh Akron product, please contact: Huntleigh Akron 1 Farthing Road Ipswich IP1 5AP
Telephone: 01473 461042 Fax: 01473 462924
Huntleigh Akron T/A a Division of Huntleigh Nesbit Evans Limited A Member of the Huntleigh Technology PLC Group of Companies © Huntleigh Technology PLC 2001. If for any reason your couch is being returned, please inform Huntleigh Akron prior to sending, and: 1. 2. 3. 4.
Clean the product, as described in the cleaning section. Pack it in suitable packing. Attach the decontamination certificate (or other written statement declaring that the product has been cleaned) to the outside of the package. Mark the package “Service Department”.
Manufactured in the UK by Huntleigh Akron. As part of the ongoing development programme this company reserves the right to modify specifications and materials of this couch without notice.
14
AKRON 8622, 8631, 8632
8. Warranty and Service (UK only) Huntleigh Akron’s standard terms and conditions apply to this couch. A copy is available upon request. These contain full details of warranty terms and do not limit the statutory rights of the consumer. For service, maintenance and any questions regarding this, or any other Huntleigh Akron Product, please contact: Huntleigh Akron 1 Farthing Road Ipswich Suffolk IP1 5AP. England Telephone: +44 (0) 1473 461042 Fax: +44 (0) 1473 462924 Huntleigh Akron T/A a Division of Huntleigh Nesbit Evans Limited A Member of the Huntleigh Technology PLC Group of Companies © Huntleigh Technology PLC 2001. If for any reason your couch is being returned, please inform Huntleigh Akron prior to sending and: 1. 2. 3. 4.
Clean the product, as described in the cleaning section. Pack it in suitable packing. Attach the decontamination certificate (or other written statement declaring that the product has been cleaned) to the outside of the package. Mark the package “Service Department”.
Manufactured in the UK by Huntleigh Akron. As part of the ongoing development programme this company reserves the right to modify specifications and materials of this couch without notice.
9. Revision Status
8622, 8631, 8632 PAGES ALL PAGES PAGES 12 & 13 PAGES 2 & 9 PAGE 9 PAGE 6
ISSUE UM - 03 UM - 04 UM - 05 (DCN958/943) UM - 06 (DCN966) UM - 07 (DCN967)
WRITTEN P. Hammond J. Austin J. Austin J. Barker J. Austin
15
DATE 25/01/99 18/11/99 10/10/00 09/02/01 27/02/01
AKRON 8622, 8631, 8632
10. Technical Specification
Width (cm) Overall Length (cm) Height Range (cm) Max. Load Weight Elevation Control Tilting Control
8622
8631
8632
63±1.5 186±3 46±1.5 81±1.5 180 Kg Electric Electric
63±1.5 186±3 46±1.5 81±1.5 180 Kg Hydraulic Manual
63±1.5 186±3 46±1.5 81±1.5 180 Kg Electric Manual
Electrical Data 230v Actuator 220-240v 220-240v Internal Fuse Rt T1.6A x 1 T1.6A x 1 Current 1.6A 1.6A Power 480W 480W Mains Plug Fuse Rt Junction Box Mains Plug Fuse Rt (UK) 5.0A x 1 Junction Box T 3.15 x 1 Fuse Rating Output 2 x (230v – 480W) Electrical Data 115v
220-24v T1.6A x 1 1.6A 480W 5.0A x 1
Actuator Internal Fuse Rt Current Power
115v F4.0A x 1 3.3A 380W
Junction Box Fuse Rating Output
115v 115v F4.0A x 1 F4.0A x 1 3.3A 3.3A 380W 380W Junction Box T 8.0 x 1 2 x (115v – 380W)
IEC 601-1 Classification Check serial number label for classification 230v
115v
Double Insulated
Double Insulated
Class 1
Type B equipment
Type B equipment
Type B equipment
Attention consult this manual
Attention consult this manual
Attention consult this manual
16
115v
Huntleigh Akron, 1 Farthing Road, Ipswich, Suffolk, IP1 5AP, England Tel: +44 (0) 1473 461042 Fax: +44 (0) 1473 462924
792-220-07