AKRON BOBATH 5 Models 4632 User Manual June 2002
User Manual
16 Pages
Preview
Page 1
AKRON 4632
Contents 1.
Introduction
2
Warnings, Cautions General Warnings
2 2
2.
Installation
3
3.
Operation
4
4.
Optional Accessories - Operating Instructions
5
5.
Cleaning and Decontamination
6
6.
Troubleshooting
6
7.
Maintenance
7
8.
Warranty and Service
8 9
9.
Revision Status
9
10.
Technical Specification
10
"The
mark on this product demonstrates conformity with the EC Directive 93/42/EEC."
1
AKRON 4632
1. Introduction Always keep this Operating Manual available for reference.
Warnings & Cautions WARNINGS Warnings identify possible hazards in procedures or conditions which, if not correctly followed, could result in death, injury or other serious adverse reactions.
CAUTIONS Cautions identify conditions or procedures which, if not correctly followed, could result in equipment failure or damage.
General Warnings WARNINGS Use only accessories that have been designed or approved for use with this table. Do not use electric tables in the presence of flammable gasses such as an anaesthetic agent. Maximum Load 325 Kg (715 lbs) distributed evenly. Do not concentrate weight on either side of the table. Plug acts as a disconnect device. Do not allow children to play with the table at any time. Adequately supervise children in the proximity of the table.
2
AKRON 4632
2. Installation WARNING Do not insert the 2-pin mains connector directly into a 3-pin 13A mains outlet. Always use the adapter provided. Plug acts as a disconnect device.
Use adapter if 3 pin 13A plug is required. (Order reference 5S0409Z)
Ensure mains supply cable is not stretched when connected. Ensure, during table use, no obstacles are in its path. Ensure hand / foot switch tubes or leads are not trapped by moving parts of the table.
CAUTION When not in use, switch table off.
Battery Backup CAUTION Battery Backup is for the unlikely event of mains power failure / not normal use. The table should always be used with mains power connected. The light on the control box indicates mains power is on. Whilst 'on' batteries are trickle charged – in case of power failure. To charge the accumulators, the control unit must be activated by shortly pressing one of the switch keys.
3
AKRON 4632
3. Operation Elevation Controls WARNING Safely position patient before elevating table. Position all electrical cables away from the moving parts. Position Hand/Foot switch where accidental operation is not possible.
Press up to raise table Press down to lower table
Press up to raise backrest Press down to lower backrest
Handswitch
Backrest Angles = +85º -0º
Wheel Locking WARNING Lock table wheels/castors prior to patient transfer or treatment. The retractable castors are not designed for transportation of patients. Disconnect and coil electrical cord away from moving parts prior to repositioning table.
Wheels
Wheels Locked
Directional Lock
4
Wheels Unlocked
AKRON 4632
4. Optional Accessories – Operating Instructions CAUTION Ensure accessories are secure prior to use.
Fold Down Safety Sides a)
To raise the safety side when in the down position, lift and rotate towards you through 180°. Once upright, push the safety side downwards into the moulded block until located fully.
b)
To lower the safety side from the upright position, lift out of the moulded block and rotate through 180° into the down position.
Patient Assist Support Handles a)
To insert support handles, position pin into socket.
b)
Secure by tightening locking knob.
c)
To remove support handles, loosen locking knob and lift support handles out of the socket.
Patient Assist Safety Support Loops a)
To insert arm loop, position pin into socket.
b)
Secure by tightening locking knob.
c)
To remove arm loop, loosen locking knob and lift arm loop out of the socket.
Breathing Hole and Plug
Remove plug for patient prone position Press out from underneath
5
AKRON 4632
5. Cleaning and Decontamination WARNING Do not use Alkalis or other organic solvents for routine cleaning. Replace torn or damaged vinyl immediately. Always cover vinyl with fresh paper or sheet between patients. If contamination is known or suspected always decontaminate table before use.
Frequency of routine table cleaning will depend on exact application. CLEANING / DECONTAMINATION Wear suitable protective clothing Disconnect mains supply Locate couch in a suitable cleaning area with castors disengaged or wheels locked. A) B) C)
Using a clean damp cloth, warm water and soap carefully wipe table from top to bottom, taking care around electrical components. Wipe table carefully from top to bottom using a damp cloth and clean water. Wipe table carefully dry with clean paper towels from top to bottom.
If DECONTAMINATING then repeat steps a) to c) using a dilute Hypochlorite / warm water mixture (10,000ppm) instead of warm water and soap. Safely discard cleaning / decontaminating materials. Wash hands thoroughly. Put table back into service.
6. Troubleshooting – Electric Actuator Symptom Actuator will not work
Possible cause Power disconnected
Action Ensure mains supply is connected
Fuse blown
Replace fuse in socket
Hand/footswitch disconnected
Check that the hand/foot switch is fully inserted into the actuator
Thermal protection
Disconnect power cord from mains supply. Leave unit for 1 minute then re-connect power cord.
If the problem persists consult your local Akron Supplier.
6
AKRON 4632
7. Maintenance Check table visually before use. Carry out routine servicing every six months or more frequently if in heavy use. Routine servicing or repair must only be performed by suitably qualified person.
WARNING Before commencing any maintenance activity, disconnect the table from the electrical supply. Plug acts as disconnect device. Examine the table for any signs of damage. Make sure that all nuts, bolts and other fasteners are tight. Do not over tighten as this may cause binding. Upholstery fasteners should be checked, tightened or replaced if necessary. Check section hinge pins to ensure that no pin protrudes. Check gas strut controls, throughout the full range of movement for locking and assistance. Check hydraulic tables for smooth operation and signs of leakage.
WARNING All electrical checks should be performed by a suitably qualified and licensed electrician. Neither BLUE or BROWN wires are to be connected to the earth terminal of a three pin plug. Examine the flexible cables for cuts, abrasions, kinking or other deterioration and replace if necessary. Ensure that the power supply connector plug is securely attached. Check mains plug connections are tight and fuse is correctly rated (5A).
7
AKRON 4632
8. Warranty and Service (UK only) Huntleigh Akron’s standard terms and conditions apply to this couch. A copy is available upon request. These contain full details of warranty terms and do not limit the statutory rights of the consumer. For service or maintenance please contact: Service Department Huntleigh Healthcare 310-312 Dallow Road Luton Beds LU1 1TD
Telephone: 01582 745700 Fax: 01582 745869
For information on this product, or any other Huntleigh Akron product, please contact: Huntleigh Akron 1 Farthing Road Ipswich IP1 5AP
Telephone: 01473 461042 Fax: 01473 462924
Huntleigh Akron T/A a Division of Huntleigh Nesbit Evans Limited A Member of the Huntleigh Technology PLC Group of Companies © Huntleigh Technology PLC 2002. If for any reason your couch is being returned, please inform Huntleigh Akron prior to sending, and: 1. 2. 3. 4.
Clean the product, as described in the cleaning section. Pack it in suitable packing. Attach the decontamination certificate (or other written statement declaring that the product has been cleaned) to the outside of the package. Mark the package “Service Department”.
Manufactured in the UK by Huntleigh Akron. As part of the ongoing development programme this company reserves the right to modify specifications and materials of this couch without notice.
8
AKRON 4632
8. Warranty and Service (UK only) Huntleigh Akron’s standard terms and conditions apply to this couch. A copy is available upon request. These contain full details of warranty terms and do not limit the statutory rights of the consumer. For service, maintenance and any questions regarding this, or any other Huntleigh Akron Product, please contact: Huntleigh Akron 1 Farthing Road Ipswich Suffolk IP1 5AP England Telephone: +44 (0) 1473 461042 Fax: +44 (0) 1473 462924 Huntleigh Akron T/A a Division of Huntleigh Nesbit Evans Limited A Member of the Huntleigh Technology PLC Group of Companies © Huntleigh Technology PLC 2002. If for any reason your couch is being returned, please inform Huntleigh Akron prior to sending and: 1. 2. 3. 4.
Clean the product, as described in the cleaning section. Pack it in suitable packing. Attach the decontamination certificate (or other written statement declaring that the product has been cleaned) to the outside of the package. Mark the package “Service Department”.
Manufactured in the UK by Huntleigh Akron. As part of the ongoing development programme this company reserves the right to modify specifications and materials of this couch without notice.
9. Revision Status
4632 PAGES ALL PAGES
ISSUE UM - 01
WRITTEN J. Barker
9
DATE 28/06/02
AKRON 4632
10. Technical Specification 4632 Head Length (cm) Foot Length (cm) Width (cm) Overall Length (cm) Height Range (cm)
71±1.5 124±1.5 107±1.5 196±3 50±1.5 97±1.5 325 Kg Electric
Max. Load Weight Elevation Control Electrical Data 230v Voltage Current Power
220-240v / 50Hz 0.8A 24v / 6A Electrical Data 115v
Voltage Current Power
115v / 60Hz 1.6A 24v / 6A
IEC 601-1 Classification
Class 1 Type B equipment Attention consult this manual