AKRON STREAMLINE TM DROP END Models 2281, 2282, 2283 2281A, 2282A, 2283A
USER MANUAL UM ENGLISH 792-702-01
AKRON STREAMLINE Model No.’s:- 2281, 2282, 2283, 2281A, 2282A, 2283A
Contents 1.
Introduction
2
2.
Warnings & Cautions
3
3.
Installation
4
4.
Operation
5
5.
Accessories – Operating Instructions
8
Paper Roll Holder, Arm Support, Moulded Arm Support, Head Support, Leg Supports/Stirrups, Phlebotomy Arm, Arm Loops, IV Pole
The
6.
Cleaning and Decontamination
11
7.
Troubleshooting
12
8.
Maintenance
12
9.
Warranty and Service
13
10.
Revision Status
14
11
Technical Specification
15
12.
Transport and Storage
17
13.
Environmental Protection
17
mark on this product demonstrates conformity with the EC Directive 93/42/EEC.
1
AKRON STREAMLINE Model No.’s:- 2281, 2282, 2283, 2281A, 2282A, 2283A
1. Introduction TM
The STREAMLINE product range of treatment & examination couches, have been designed to cover a wide range of disciplines, from simple examination to heavier manipulative and manual therapy techniques. TM
The STREAMLINE couches in this range provide ergonomic solutions to a variety of manual handling issues, benefiting both the patient and clinician. These products can be further enhanced with a wide range of additional accessories also available. However, these products are not designed to be used as a bed and are not suitable for sleeping on.
Always keep this Operating Manual available for reference.
2
AKRON STREAMLINE Model No.’s:- 2281, 2282, 2283, 2281A, 2282A, 2283A
2. Warnings & Cautions WARNINGS Warnings identify possible hazards in procedures or conditions which, if not correctly followed, could result in death, injury or other serious adverse reactions.
CAUTIONS Cautions identify conditions or procedures which, if not correctly followed, could result in equipment failure or damage.
General Warnings WARNINGS Electrical equipment can be hazardous. These products should only be operated by suitably qualified/trained personnel. This equipment should not be used adjacent to other electrical equipment. If such use is necessary, normal operation of the equipment should be verified by observation. Medical Electrical Equipment maybe affected by portable and/or mobile RF communications equipment (eg mobile telephones). Use only accessories that have been designed or approved for use with this couch. These products are not intended for Patient Transportation. Do not use electric couches in the presence of flammable gases such as an anaesthetic agent. Maximum Load 225 kg (495 lbs) distributed evenly. Do not concentrate weight on either side of the couch. Plug acts as a disconnect device. Do not allow children to play with the couch at any time. Adequately supervise children in the proximity of the couch. The patient should not be left on the Couch unattended at any time. CAUTIONS Electromagnetic Compatibility (EMC). This product complies with the requirements of applicable EMC standards. The use of accessories not specified by the manufacturer may result in increased emissions by, or decreased immunity of, the equipment, affecting its performance. Do not exceed the specified duty cycle as stated on the Product Identification Label USA only - To achieve grounding reliability, ensure that this equipment is connected to a receptacle marked “Hospital Only” or “Hospital Grade”.
3
AKRON STREAMLINE Model No.’s:- 2281, 2282, 2283, 2281A, 2282A, 2283A
3. Installation Adjustable Foot WARNING Lock castors before adjusting.
A)
Lock castors first.
B)
Adjust foot.
C)
Lock foot by tightening half nut.
Electric Models WARNING Do not insert the 2-pin mains connector directly into a UK style, BS1363, 3-pin 13A mains outlet. Always use the adapter provided. Plug acts as a disconnect device. Use adapter if UK style, BS1363, 3-pin 13A plug is required. (Order reference 5S0409Z)
Ensure mains supply cable is not stretched when connected. Ensure, during couch use, no obstacles are in its path. Ensure hand / foot switch tubes or leads are not trapped by moving parts of the couch.
CAUTION When not in use, switch couch off, by unplugging from the power socket/receptacle. Note that models with battery back-up will still operate.
CAUTION Do NOT Insert Hands or Feet into the Mechanism as this could result in serious injury!
4
AKRON STREAMLINE Model No.’s:- 2281, 2282, 2283, 2281A, 2282A, 2283A
4. Operation Elevation Controls WARNING Safely position patient before elevating couch. Position all electrical cables away from the moving parts. Position Hand/Foot switch where accidental operation is not possible. Hydraulic Models PUMP PEDAL TO RAISE COUCH
Hydraulic Pedal
Electric Models Press
to raise couch.
Press
to lower couch.
Press
to raise couch.
Press
to lower couch.
Hanning Handswitch Hanning Handswitch
Hanning Footswitch
Linak Handswitch (Model 2283 only)
5
LIFT PEDAL TO LOWER COUCH
AKRON STREAMLINE Model No.’s:- 2281, 2282, 2283, 2281A, 2282A, 2283A
4. Operation Wheel Locking WARNING Lock couch wheels/castors prior to patient transfer or treatment. These couches are not designed for the transportation of patients. Disconnect and coil electrical cord away from moving parts prior to repositioning couch.
Castors (RWB – Retractable Wheel Base)
Castors locked
Castors unlocked
WARNING Lock table wheels prior to patient transfer or treatment. Wheels (LWB - Large Wheel Base)
Wheels Locked
Directional Lock
6
Wheels Unlocked
AKRON STREAMLINE Model No.’s:- 2281, 2282, 2283, 2281A, 2282A, 2283A
4. Operation Backrest & Foot Section CAUTION Support upholstery section while manually altering the angle.
Backrest Operating Lever
A)
Press lever UP to alter section angle.
B)
Support section when altering angle.
C)
Release lever to lock section in position. Angles = 0° to +85 o
Foot Section Operating Lever A)
Press lever UP to alter section angle.
B)
Support section when altering angle.
C)
Release lever to lock section in position. Foot Section Angles = +35° to -90°
Breathing Hole and Plug
Remove plug for patient prone position. Press out from underneath. (Models 2281A, 2282A, 2283A only)
7
AKRON STREAMLINE Model No.’s:- 2281, 2282, 2283, 2281A, 2282A, 2283A
5. Accessories – Operating Instructions Paper Roll Holder Replace paper roll by:A)
Gently spring bracket out.
B)
Insert paper roll rod between brackets.
C)
Load paper roll and release brackets to secure.
Arm Support Adjust Arm Support to suit patient by:A)
Unlocking the locking knobs.
B)
Adjust position of arm support.
C)
Tighten locking knobs.
Moulded Arm Support Adjust Moulded Arm Support to suit patient by:A)
Unlocking the locking knobs.
B)
Adjust position of arm support.
C)
Tighten locking knobs.
Head Support Adjust Head Support to suit patient by: A)
Unfastening hook and loop strap.
B)
Position where required.
C)
Fasten hook and loop strap to secure.
8
AKRON STREAMLINE Model No.’s:- 2281, 2282, 2283, 2281A, 2282A, 2283A
5. Accessories – Operating Instructions Leg Supports CAUTION Ensure all locking knobs and levers have been sufficiently tightened before using leg supports.
To adjust leg support: A)
Locate leg support into bracket.
B)
Secure by tightening locking knob.
C)
To adjust angle of leg support pad, loosen lever on underside of pad, adjust to desired position and secure by tightening lever.
Leg Stirrups Adjust Leg Stirrups to suit patient by:A)
Unlocking the locking knobs.
B)
Adjust position of stirrup.
C)
Tighten locking knobs.
Phlebotomy Arm
To adjust arm support: Loosen knob and adjust sliding arm into required position.
9
AKRON STREAMLINE Model No.’s:- 2281, 2282, 2283, 2281A, 2282A, 2283A
5. Accessories – Operating Instructions Arm Loops CAUTION Ensure arm loop is secure prior to use.
A)
To insert arm loop, position pin into socket.
B)
Secure by tightening locking knob.
C)
To remove arm loop, loosen locking knob and lift arm loop out of socket.
IV Pole
Locate IV pole into socket and secure into position using locking knob.
CAUTION Ensure all accessories are fully tightened prior to use.
10
AKRON STREAMLINE Model No.’s:- 2281, 2282, 2283, 2281A, 2282A, 2283A
6. Cleaning and Decontamination WARNING Do not use Alkalis or other organic solvents for routine cleaning. Replace torn or damaged vinyl immediately. Always cover vinyl with fresh paper or sheet between patients. If contamination is known or suspected always decontaminate couch before use. Do not saturate hinge pin location and it is essential to ensure that they are dry before re-use. Disconnect any mains supply. Frequency of routine couch cleaning will depend on exact application. CLEANING / DECONTAMINATION Locate couch in a suitable cleaning area with castors disengaged or wheels locked. 1.
Wear protective clothing: plastic disposable apron and gloves.
2.
Disconnect any mains supply and lock couch wheels.
3.
Control handsets can be wiped over using a 70% Isopropranol alcohol wipe.
4.
Prepare a solution of warm water and neutral detergent.
5.
Start at the top of the couch and work your way downwards.
6.
Wipe over head end working towards mid torso.
7.
Wipe over remaining horizontal patient surface, working towards foot end.
8.
Wipe all horizontal parts under the patient surface.
9.
Wipe all vertical parts under the patient surface.
10.
Wipe over shroud (if fitted).
11.
Wipe over all parts of wheels.
12.
Ensure all parts of the couch are dry before re-use.
13.
Safely discard cleaning materials.
14.
Wash hands thoroughly.
If DECONTAMINATING then repeat steps 1) to 14) using a dilute Hypochlorite / warm water mixture (10,000ppm) instead of warm water and soap. Put couch back into service.
11
AKRON STREAMLINE Model No.’s:- 2281, 2282, 2283, 2281A, 2282A, 2283A
7. Troubleshooting – Electric Actuator Symptom
Possible cause
Action
Actuator will not work
Power disconnected Fuse blown Hand/footswitch disconnected
Ensure mains supply is connected Contact your local Akron supplier Check that the hand/foot switch is fully inserted into the actuator
If the problem persists consult your local AKRON Supplier.
8. Maintenance Check couch visually before use. Carry out routine servicing every six months or more frequently if in heavy use. Routine servicing or repair must only be performed by suitably qualified persons.
WARNING Before commencing any maintenance activity, disconnect the couch from the electrical supply. Plug acts as disconnect device. Examine the couch for any signs of damage. Make sure that all nuts, bolts and other fasteners are tight. Do not over tighten as this may cause binding. Upholstery fasteners should be checked, tightened or replaced if necessary. Check section hinge pins to ensure that no pin protrudes. Check gas strut controls, throughout the full range of movement for locking and assistance. Check hydraulic couches for smooth operation and signs of leakage.
WARNING All electrical checks should be performed by a suitably qualified and licensed electrician. Only the GREEN/YELLOW wire is to be connected to the earth/ground terminal of a three pin plug. Examine the flexible cables for cuts, abrasions, kinking or other deterioration and replace if necessary. Ensure that the power supply connector plug is securely attached. Check mains plug connections are tight and fuse (UK style, BS1363 plugs only) is correctly rated (5A).
12
AKRON STREAMLINE Model No.’s:- 2281, 2282, 2283, 2281A, 2282A, 2283A
9. Warranty and Service (UK only) ArjoHuntleigh AB standard terms and conditions apply to this couch. A copy is available upon request. These contain full details of warranty terms and do not limit the statutory rights of the consumer. For service or maintenance please contact: Service Department Huntleigh Healthcare 310-312 Dallow Road Luton Beds LU1 1TD Telephone: 01582 745700 Fax: 01582 745869 For information on this product, regarding warranty and spare parts, please contact: ArjoHuntleigh AB Verkstadsvägen 5 241 38 Eslöv SWEDEN www.arjohuntleigh.com If for any reason your couch is being returned, please inform ArjoHuntleigh AB prior to sending, and: 1. 2. 3. 4.
Clean the product, as described in the cleaning section. Pack it in suitable packing. Attach the decontamination certificate (or other written statement declaring that the product has been cleaned) to the outside of the package. Mark the package “Service Department”.
Manufactured in the UK by ArjoHuntleigh AB. As part of the ongoing development programme this company reserves the right to modify specifications and materials of this couch without notice.
13
AKRON STREAMLINE Model No.’s:- 2281, 2282, 2283, 2281A, 2282A, 2283A
9. Warranty and Service (Outside UK) ArjoHuntleigh AB standard terms and conditions apply to this couch. A copy is available upon request. These contain full details of warranty terms and do not limit the statutory rights of the consumer. For service, maintenance and any questions regarding this, or any other ArjoHuntleigh AB Product, please contact: ArjoHuntleigh AB Verkstadsvägen 5 241 38 Eslöv SWEDEN www.arjohuntleigh.com If for any reason your couch is being returned, please inform ArjoHuntleigh AB prior to sending and: 1. 2. 3. 4.
Clean the product, as described in the cleaning section. Pack it in suitable packing. Attach the decontamination certificate (or other written statement declaring that the product has been cleaned) to the outside of the package. Mark the package “Service Department”.
Manufactured in the UK by ArjoHuntleigh AB. As part of the ongoing development programme this company reserves the right to modify specifications and materials of this couch without notice.
10. Revision Status 2281, 2282, 2283, 2281A, 2282A, 2283A PAGES ALL PAGES
ISSUE UM – 01 (1253)
WRITTEN S. Ridley
14
DATE 05/07/2007
AKRON STREAMLINE Model No.’s:- 2281, 2282, 2283, 2281A, 2282A, 2283A
11. Technical Specification STREAMLINE DROP END Head Length (cm) Mid Length (cm) Foot Length (cm) Width (cm)
Overall Length (cm) Height Range (cm) Max. Load Weight Couch Mass Elevation Control Configuration
2281
2282
2281A
2282A
102±1.5 54±1.5 125±1.5 63±1.5 70±1.5 80±1.5 198±3 46±1.5 91±1.5 225 kg 82 kg Hydraulic
102±1.5 54±1.5 125±1.5 63±1.5 70±1.5 80±1.5 198±3 46±1.5 91±1.5 225 kg 82 kg Electric
102±1.5 54±1.5 125±1.5 63±1.5 70±1.5 80±1.5 198±3 46±1.5 91±1.5 225 kg 82 kg Hydraulic
102±1.5 54±1.5 125±1.5 63±1.5 70±1.5 80±1.5 198±3 46±1.5 91±1.5 225 kg 82 kg Electric
Electrical Data 230V (Hanning)
AC Actuator Internal Fuse Rt Current Power Mains Plug Fuse Rt (UK)
220-240V T3.15A x 1 2.0A 360W 5.0A x 1 Electrical Data 115V (Hanning)
Actuator Internal Fuse Rt Current Power
115V T5A x 1 3.5A 405W
STREAMLINE DROP END
2283 RWB ELEC/ELEC
Head Length (cm) Mid Length (cm) Foot Length (cm) Width (cm) Overall Length (cm) Height Range (cm) Max. Load Weight Elevation Control
102±1.5 54±1.5 125±1.5 63±1.5 / 70±1.5 / 80±1.5 198±3 46±1.5 - 91±1.5 225 kg Electric Electrical Data 230V (Linak) 230V 3A 700W 5Ax1 Electrical Data 120V (Linak) 120V 5.8A 700W
DC Control Box Current Power (Max) Mains Plug Fuse Rt (UK) DC Control Box Current (Max) Power (Max)
15
AKRON STREAMLINE Model No.’s:- 2281, 2282, 2283, 2281A, 2282A, 2283A
11. Technical Specification Complies with the following Electrical Safety Standards:IEC 60601-1:1988/A2:1995, EN 60601-1:1990/A13:1996, UL60601-1 1st Edition, CAN/CSA C22.2 No. 601.1- M90. EMC Standards: EN 60601-1-2:2001
Check Serial Number label for classification and duty cycle Model No.’s 2282, 2282A (Hanning) 230V ~ 50Hz 2.0A
~
Model No.’s 2282/UL, 2282A/UL 115V ~ 60Hz 3.6A
Double Insulated (Class II)
Class I
Type B equipment
Type B equipment
Attention consult accompanying documents
Attention consult accompanying documents
~
Alternating Current
25EA 82 kg
Couch Mass
82 kg
Alternating Current Model No.’s 2282, 2282A (Hanning). Classified with respect to Electric Shock, Fire and Mechanical hazards only in accordance with UL 606011 and CAN/CSA C.22.2 No. 601.1 (CAN/CSA C22.2 No. 601.2.38) Couch Mass
Check Serial Number label for classification and duty cycle Model No.’s 2283, 2283A (Linak)
Model No.’s 2283/9, 2283A/9 (Linak)
230V ~ 50Hz 3.0A
~ 82 kg
120V ~ 60Hz 5.8A
Class I
Class I
Type B equipment
Type B equipment
Attention consult accompanying documents
Attention consult accompanying documents
~
Alternating Current Couch Mass
82 kg
16
Alternating Current Couch Mass
AKRON STREAMLINE Model No.’s:- 2281, 2282, 2283, 2281A, 2282A, 2283A
11. Technical Specification CAUTION Electromagnetic compatibility (EMC). This product complies with the requirements of applicable EMC Standards. The use of accessories not specified by the manufacturer may result in increased emission by, or decreased immunity of, the equipment, affecting its performance. Do not exceed the specified duty cycle as stated on the Product Identification Label.
WARNING Medical Electrical Equipment needs special precautions regarding EMC (Electromagnetic Compatibility) and must be installed and put into service according to the EMC information provided in the documents accompanying this equipment.
12. Transportation and Storage Handle with care. Do not drop. Avoid shock or violent impact. The equipment should be stored in a clean, dry and well ventilated area. The following limits apply during transport and for a storage period of up to 15 weeks: Ambient temperature Relative humidity Air pressure
-25°C to +70°C 10% to 75% 50 kPa to 106 kPa
The following limits apply to operating conditions or longer periods of storage: Ambient temperature Relative humidity Air pressure
+10°C to +40°C 30% to 75% 70 kPa to 106 kPa
13. Environmental Protection Dispose of device in compliance with local regulations.