5. Product care
Do not use Alkalis or other organic solvents for routine cleaning. Replace torn or damaged vinyl immediately. Always cover vinyl with fresh paper or sheet between patients. If contamination is known or suspected always decontaminate couch before use. Do not saturate hinge pin location and it is essential to ensure that they are dry before re-use. Disconnect any mains supply. Do not allow the mains plug or power supply cord to get wet.
Do not use Phenol-based solutions or abrasive compounds or pads during the decontamination process as these will damage the surface coating. Avoid immersing electrical parts in water during the cleaning process.
Cleaning / Decontamination Frequency of routine couch cleaning will depend on exact application, local policy and contamination status.
To clean Clean all exposed surfaces and remove any organic debris by wiping with a cloth moistened with a simple (neutral) detergent and water. Chemical Disinfection We recommend a chlorine-releasing agent, such as sodium hypochlorite, at a strength of 1,000ppm available chlorine (this may vary from 250ppm to 10,000ppm depending on local policy and contamination status). Use low strength solution for upholstery. 17
Wipe all cleaned surfaces with the solution, then wipe with a cloth moistened with water and dry thoroughly. Alcohol based disinfectants (maximum strength 70%) may be used as an alternative. Ensure the product is dry before storage. If an alternative disinfectant is selected from the wide variety available we recommend that suitability for use is confirmed with the chemical supplier prior to use.
Preventive maintenance This product is subject to wear and tear during use. To ensure that it continues to perform within its original specification, preventive maintenance procedures should be carried out at the interval shown.
This list indicates the minimum recommended level of preventive maintenance. Failure to carry out these checks, or continuing to use the product if a fault is found, may compromise the safety of both the patient and carer. Preventive maintenance can help to prevent accidents.
Actions to be done by operative Check operation of side rails Check patient handset and cable Visually check castors Visually chleck power supply cord and mains plug Examine the accessories (if applicable) Visually check the upholstery Check gas strut controls, and moveable sections for their full range of movement for locking
18
Daily
Weekly