Alaris
Alaris Pump Module Improper Set Loading Guide Oct 2013
Guide
1 Page
Preview
Page 1
Improper set loading guide for the Alaris Pump module ®
Alaris System ®
Practices that may lead to set misloading: Do not place the safety clamp into the Alaris Pump module prior to inserting the set upper fitment.
Illustrations of misloaded sets
Do not hold the tubing while closing and latching the door.
Do not touch the upper fitment while closing and latching the door.
Examples of improper loading
If you observe any of these conditions, close the roller clamp and reload the set upper fitment into the receptacle. In A and B, the upper fitment is tilted from the vertical position. This tilting was likely caused by lateral force on the upper fitment while the door was closed and latched. This condition may cause infusion rate inaccuracy.
A Front view
B Side view
If you observe any of these conditions, close the roller clamp and reload the set upper fitment into the receptacle. In C and D, the upper fitment is elevated above the receptacle on the Alaris Pump module. This elevation was likely caused by maintained traction on the set while the door was latched and closed. This condition may cause infusion rate inaccuracy.
C
Warning: Failure to follow the Alaris System user manual may result in infusion rate inaccuracy. For product support, contact Customer Advocacy at 888.812.3266 or [email protected]. For technical support, contact Instrument Technical Support at 888.812.3229. For product orders, contact Customer Order Management at 800.482.4822. © 2013 CareFusion Corporation or one of its subsidiaries. All rights reserved. Alaris, CareFusion and the CareFusion logo are trademarks or registered trademarks of CareFusion Corporation or one of its subsidiaries. IF2410 (1013/2000)
CareFusion San Diego, CA
carefusion.com
D