Cleaning your Alcon phaco handpiece CENTURION® ACTIVE SENTRY® Handpiece  1  CENTURION® OZIL® Handpiece  Remove irrigation and aspiration tubing from handpiece. Remove aspiration luer with hand and not using tools.  2 Unplug handpiece from the console by the metal handle and install protective cap.  3 Remove infusion sleeve and tip from handpiece using tip wrench and discard safely.  4 Pat away residue with a soft, clean, lint-free non-abrasive cloth and rinse handpiece. * If necessary, clean exterior with a soft-bristled brush. Submerge the nose cone (front part) in water. Using a syringe, PUSH (DO NOT DRAW) at least 120 cc of water through irrigation and aspiration paths.  5  Using a syringe, PUSH (DO NOT DRAW) at least 120 cc of water through irrigation and aspiration paths.  Always work with sterile, room-temperature deionised water 6 With the same syringe, flush both ports with at least 60 cc of air.