System Manual
241 Pages
Preview
Page 1
System Manual IVD
Distributed by Epocal Inc.
Alere North America, LLC
Emergo Europe
2060 Walkley Road
30 S. Keller Road, Suite 100,
Molenstraat 15, 2513 BH
Ottawa, Ontario,
Orlando, FL 32810 USA
The Hague, Netherlands
Canada K1G 3P5
Tel (US): (877) 441-7440
Tel: +31.70.345.8570
Tel: (613) 738-6192
Tel (outside US): (321) 441-7200 Fax: +31.70.346.7299
Fax: (613) 738-6195
Email: [email protected]
www.epocal.com
© 2015 Alere. All rights reserved. The Alere Logo, Alere, epoc, and Care-Fill are trademarks of the Alere group of companies. All other trademarks referenced are trademarks of their respective owners. 51000623 Rev: 12
Patents US D639956; US 6845327; US 6896778; US 7094330; US 7767068; US 7824529; US 7842234; US 8506778; additional patents pending.
51000623 Rev: 12
Table of Contents Section 1
2 3
4
Title
Page
Introduction
1.1 epoc System Manual ... 1-1 1.2 Cautions ... 1-2 1.3 Warranty ... 1-2 1.4 Warranty Limitations ... 1-3 1.5 WEEE Compliance ... 1-3
epoc Blood Analysis System
2.1 System Overview... 2-1 2.2 Operation Overview ... 2-2
epoc System Operation
3.1 System Operation Overview ... 3-1 3.2 Turning ”ON” the epoc Reader ... 3-1 3.3 Turning ”ON” the epoc Host ... 3-2 3.4 Logging in to epoc Host Software Application ... 3-2 3.5 Running a Test on a Dedicated Reader ... 3-2 3.6 Alternate Means to Run a Test ... 3-3 3.7 Reader Electronic Internal QC Test ... 3-3 3.8 Reader Screen ... 3-4 3.9 Obtaining the Test Card ... 3-4 3.10 Inserting the Test Card ... 3-5 3.11 Calibration Sequence ... 3-6 3.12 Entering Patient Information (or Lot Number) and Test Selection ... 3-6 3.13 Using Barcode Scanner to Enter Patient ID ... 3-7 3.14 Collecting a Blood Sample ... 3-7 3.15 Timing of Sample Introduction ... 3-7 3.16 Sample Introduction ... 3-8 3.17 Test Completion... 3-9 3.18 Running Another Test ... 3-9 3.19 Closing the Test and Disconnecting the Reader ... 3-10 3.20 EDM Synchronization ... 3-10 3.21 Logging Out and Turning the Power ”OFF” ... 3-10 3.22 Multiple Reader Testing ... 3-11
epoc Test Cards
4.1 General Test Card Information ... 4-1 4.2 Test Card Physical Characteristics ... 4-1 4.3 Test Card Packaging, Storage, and Shelf Life... 4-2
51000624 Rev: 23
Table of Contents
Page 1 of 6
Section 5
6
Title
Page
epoc Reader
5.1 Overview ... 5-1 5.2 Power Requirements ... 5-2 5.3 Rechargeable Battery ... 5-3 5.4 Reader Status and Firmware Version ... 5-3 5.5 Test Status ... 5-4 5.6 Audible Signals ... 5-4 5.7 Card Insertion Slot... 5-5 5.8 Docking Pivot ... 5-5 5.9 USB Maintenance Port ... 5-6 5.10 Motorized Mechanism ... 5-6 5.11 Wireless Module ... 5-6 5.12 Barcode Scanner ... 5-6 5.13 Thermal System ... 5-7 5.14 Operating Environment ... 5-7
epoc Host
6.1 Overview ... 6-1 6.2 Rechargeable Battery ... 6-1 6.3 Barcode Scanner ... 6-1 6.4 Soft (Warm) and Hard (Cold) Reset ... 6-2 6.5 Navigation ... 6-2 6.6 Entering Text ... 6-3 6.7 Date/Time Clock ... 6-4 6.8 Status Indicators ... 6-4 6.9 User Accounts ... 6-5 6.10 epoc Host Application User Interface ... 6-5 6.11 Startup Screen ... 6-6 6.12 Login Screen ... 6-6 6.13 Operational Modes ... 6-7 6.14 Testing Mode ... 6-7 6.15 Main Reader Screen ... 6-8 6.16 Reader Screen ... 6-9 6.17 Reader Screen Pages ... 6-9 6.18 Test Results Page... 6-10 6.19 Critical Values Actions ... 6-12 6.20 Test Information Page ... 6-13 6.21 Test Selection Page ... 6-13 6.22 Respiratory Therapy Parameters ... 6-14 6.23 Age and Gender ... 6-17 6.24 EDM Synchronization ... 6-17 6.25 Viewing Tests Mode ... 6-18 6.26 Viewing Electronic QC Mode ... 6-20 6.27 Viewing Thermal QA Mode... 6-21 6.28 Personal Options Page ... 6-22
51000624 Rev: 23
Table of Contents
Page 2 of 6
Section Title 7 epoc Host Administration
Page
7.1 General ... 7-1 7.2 First Time Administrator Login ... 7-1 7.3 Limitations For Use ... 7-1 7.4 Administrator Access ... 7-2 7.5 Windows Mobile Operating System... 7-2 7.6 Administrator User Interface ... 7-3 7.7 Administrator Options ... 7-4 7.8 Barcode Options Page ... 7-10 7.9 EDM Options Page ... 7-11 7.10 Card Options 1 ... 7-11 7.11 Card Options 2 ... 7-14 7.12 Personal Options Page ... 7-15 7.13 Software Upgrades ... 7-15 7.14 Set Date, Time, and Time Zone ... 7-19 7.15 Reader Identifiers ... 7-19 7.16 Reader Configuration and Reader Software Upgrade ... 7-20 7.17 Dedicating Readers ... 7-21 7.18 Test Record Administration ... 7-22
8
epoc Enterprise Data Manager
8.1 Introduction ... 8-1 8.2 Deployment ... 8-1 8.3 Hardware and Software Requirements ... 8-1 8.4 Installation ... 8-2 8.5 epoc Database ... 8-2 8.6 epoc Link Application... 8-2 8.7 Using EDM Web Application ... 8-3 8.8 Blood Tests Page ... 8-4 8.9 Blood Tests: Filtering Results ... 8-6 8.10 Blood Tests: Printing Results ... 8-6 8.11 Blood Tests: Exporting List of Tests to CSV File ... 8-8 8.12 QA Tests ... 8-8 8.13 Reports: Electronic QC ... 8-9 8.14 Reports: Thermal QA ... 8-9 8.15 Reports: Usage Statistics ... 8-10 8.16 Reports: Patients List (ADT) ... 8-11 8.17 Inventory: Hosts ... 8-11 8.18 Inventory: Readers ... 8-12 8.19 Inventory: Card Lots ... 8-13 8.20 Inventory: Printers ... 8-14 8.21 Users Page ... 8-14 8.22 Groups Page... 8-16 8.23 Host Settings ... 8-17 8.24 Host Settings: Units ... 8-18 8.25 Host Settings: Ranges ... 8-19 8.26 Host Settings: Software Update ... 8-20 8.27 Departments ... 8-21 8.28 Configurations ... 8-22 8.29 EDM Settings... 8-24
51000624 Rev: 23
Table of Contents
Page 3 of 6
Section 9
10
11
12
13
Title
Page
Quality Assurance
9.1 Overview ... 9-1 9.2 Recommended Quality Control for epoc System ... 9-1 9.3 Optional Quality Control for epoc System ... 9-3 9.4 Aqueous Fluid Handling ... 9-4 9.5 Value Assignment Datasheets ... 9-6
Care and Maintenance of epoc System
10.1 General Information ... 10-1 10.2 Care of the epoc System ... 10-1 10.3 Cleaning ... 10-1 10.4 Maintenance ... 10-2
Theory of Operation
11.1 epoc Host ... 11-1 11.2 epoc Reader ... 11-2 11.3 epoc Test Card ... 11-3 11.4 Sensor Module ... 11-4 11.5 Sensor Electrodes ... 11-6 11.6 Quality Control and epoc System ... 11-10 11.7 References ... 11-14
BGEM Test Card Specifications
12.1 General BGEM Test Card Specifications ... ..12-1 12.2 Test Card Configuration and Use ... 12-2 12.3 pH ... 12-8 12.4 pCO2 ... 12-12 12.5 pO2 ... 12-16 12.6 Sodium (Na+) ... 12-20 12.7 Potassium (K+) ... 12-25 12.8 Ionized Calcium (Ca++) ... 12-29 12.9 Chloride (Cl-) ... 12-34 12.10 Glucose (Glu) ... 12-38 12.11 Lactate (Lac) ... 12-47 12.11 Creatinine (Crea) ... 12-52 12.13 Hematocrit (Hct) ... 12-57 12.14 Calculated Values ... 12-61
epoc Reader and Host Specifications
13.1 epoc Reader ... 13-1 13.2 epoc Reader Components ... 13-2 13.3 epoc Host (Socket) ... 13-3 13.4 epoc Host Accessories (for Socket) ... 13-4 13.5 epoc Host2 (Motorola) ... 13-5 13.6 epoc System Accessories ... 13-7 13.7 epoc System Compliance ... 13-7
51000624 Rev: 23
Table of Contents
Page 4 of 6
Section
Title
Page
14
Troubleshooting and Error Messages
15
Glossary
14.1 General ... 14-1 14.2 Out-of-Range Results on the Test Card for Liquid Quality Control or Calibration Verification ... 14-1 14.3 Failed Reader Electronic QC ... 14-1 14.4 Failed Thermal QA ... 14-2 14.5 epoc Host Application Messages ... 14-2
15.1 Symbols ... 15-1 15.2 Terminology and Abbreviations ... 15-2
Appendix A.
epoc Host Quick Start Guide (Socket SoMo 650)
Appendix B.
epoc Host2 Quick Start Guide (Motorola MC55A0)
Appendix C.
epoc Reader Quick Start Guide
Appendix D.
epoc Care-Fill™ Capillary Tube Package Insert
51000624 Rev: 23
Table of Contents
Page 5 of 6
Document / Revision List Title
Document Number
Revision
Cover Page
51000623
12
Table of Contents
51000624
23
01 - Introduction
51000625
08
02 - epoc Blood Analysis System
51000627
06
03 - epoc System Operation
51000635
13
04 - epoc Test Cards
51000629
11
05 - epoc Reader
51000631
08
06 - epoc Host
51000632
14
07 - epoc Host Administration
51000634
15
08 - epoc Enterprise Data Manager
51001364
09
09 - Quality Assurance
51000636
15
10 - Care and Maintenance of the epoc System
51000639
10
11 - Theory of Operation
51000628
10
12 - BGEM Test Card Specifications
51000630
17
13 - epoc Reader and Host Specifications
51000633
11
14 - Troubleshooting and Error Messages
51000638
12
15 - Glossary
51000626
10
Appendix A. epoc Host Quick Start Guide (Socket SoMo 650)
51001684
07
Appendix B. epoc Host2 Quick Start Guide (Motorola MC55A0) 51005527
07
Appendix C. epoc Reader Quick Start Guide
51000848
04
Appendix D. epoc Care-Fill™ Capillary Tube Package Insert
51003903
07
epoc System Manual, Master PDF
51004969
11
51000624 Rev: 23
Table of Contents
Page 6 of 6
01
Introduction
1.1 epoc System Manual General The epoc System Manual describes the proper use and operation of the epoc Blood Analysis System. System operators and the system administrator should familiarize themselves with the applicable sections in the manual prior to conducting testing. All operators and the system administrator require training by Epocal authorized personnel prior to conducting patient testing. This training is based upon the information contained in this manual. The System Manual should be made available to the administrator and operators at all times while the epoc Blood Analysis System is in use. Organization of System Manual The System Manual is organized into sections corresponding with the Table of Contents at the front of the manual. Each section is controlled and maintained separately by the document and revision number at the bottom of each page. Section 2 “epoc Blood Analysis System” provides a brief description of the epoc System and its major components. Section 3 “epoc System Operation” describes the steps and necessary information for performing a test, including sample collection and interpretation of test results. Sections 4, 5 and 6 “epoc Test Cards”, “epoc Reader” and “epoc Host” describe the function and construction details of the major system components. Section 7 “epoc Host Administration” describes to the system administrator how to set up and maintain key settings in the epoc Host application prior to releasing the system for patient use. Requirements for ongoing administrative maintenance are also described. Section 8 “epoc Enterprise Data Manager” describes how to set up and operate the optional “epoc Data Manager”. Section 9 “Quality Assurance” describes recommended and additional Quality Control and Calibration Verification procedures used to verify the performance of the epoc System. These procedures include internal, liquid, whole blood, calibration verification, and proficiency testing quality control procedures. The rationale for the epoc System quality control is described in the “Theory of Operation” section of this manual Section 10 “Care and Maintenance of the epoc System” section describes the cleaning and maintenance procedures. Section 11 “Theory of Operation” contains the methodology, principles of operation and the special merits and limitations of the epoc System. Section 12 “BGEM Test Card Specifications” contains performance characteristics and specifications for the test cards, including measurement ranges, limitations and interferences for each analyte. Section 13 “epoc Reader and Host Specifications” provides specifications for the entire system and each of the major components as well as compliance information. Section 14 “Troubleshooting and Error Messages” should be referenced if problems arise during the use of the epoc System. Section 15 “Glossary” defines terms and acronyms used in the system manual.
51000625 Rev: 08
Introduction
1-1
1.2 Cautions Cautions are identified throughout the manual using the “Caution risk of danger” or “Caution, consult accompanying documents” symbol. The operator and administrator need to pay special attention to the instructions accompanying this symbol to ensure that the epoc Blood Analysis System is used properly, reliably and safely.
The “Biological risks” symbol is used in the manual to identify potential biological risks associated with the handling of blood samples. Precautions, as stipulated by the facility where the epoc System is utilized, must be taken to ensure that the risk of transmitting blood borne pathogens is minimized. Other symbols may also be used in the Operator’s Manual. The correct interpretation of these symbols is located in the Glossary Section.
1.3 Warranty Epocal Inc. (“Epocal”) warrants to the original customer that the medical equipment manufactured by Epocal is free from defects in materials and workmanship, and under normal and proper use conditions, for a period of one (1) full year from the date of shipment. Upon receiving notice from the customer of any defects within this warranty period, Epocal shall, at its option and sole discretion, either repair, replace or modify this medical equipment or part thereof, which proves to be defective. Epocal shall repair or replace software media and firmware which does not execute as intended due to such defects. These replacements, repairs, or alterations will in no case extend the specified herein warranty period. Epocal does not warrant that the operating of the software, firmware or hardware shall be uninterrupted or error free. If Epocal is unable, within a reasonable time, to repair, replace or modify any product to a condition as warranted, the customer shall be entitled to a refund of the purchase price upon return of the product to Epocal, together with a copy of the dated itemized purchase receipt and the original packaging. The test card warranty is limited to, and prorated in accordance with the “use by” date indicated on the test card labeling. The customer must inform Epocal immediately upon receipt if it is evident that the test cards were improperly stored or handled during shipment. Epocal is not responsible for test cards that have been handled and stored outside the specified requirements stated in the System Manual after delivery to the customer. The warranty does not cover these parts that deteriorate, or which are in any case considered consumables, or those parts or items, which by their nature are normally required to be replaced periodically consistent with normal maintenance . Note: Warranty rights may vary from state to state, province to province and country to country.
51000625 Rev: 08
Introduction
1-2
1.4 Warranty Limitations The foregoing warranty shall not apply to defects resulting from: 1
6 7 8
Improper, inadequate, insufficient or negligent storage, care or maintenance by the customer or an unauthorized person, Misuse due to carelessness, negligence or inexperience, Using accessories and/or consumables that are not approved by Epocal, Unauthorized use of buyer-supplied hardware, software or interfacing, Unauthorized repairs, modifications, misuse, or damage caused by disposable batteries, or rechargeable batteries not supplied by Epocal, Failure to use the device and accessories in accordance with operating instructions, Operating outside of the environmental specifications of the product, Improper site preparation or maintenance, or
9
Expiration of test card “use by” date.
2 3 4 5
NO OTHER WARRANTIES, EXPRESS OR IMPLIED, ARE MADE. EPOCAL WILL NOT BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES RESULTING DIRECTLY OR INDIRECTLY FROM THE FAILURE OF THE PRODUCT TO PERFORM IN ACCORDANCE WITH SPECIFICATIONS. Some states do not allow the exclusion or limitation of other express or implied warranties or incidental or consequential damages, so the above limitations or exclusions may not apply. No agent or employee of Epocal is authorized to extend any other warranty or to assume for Epocal any liability except as set forth above.
1.5 WEEE Compliance Epocal Inc. complies with Directive 2002/96/EC of the European Parliament and of the Council of 27 January 2003 on waste electrical and electronic equipment (WEEE).
Compliance is indicated on the epoc hardware using the Wheelie Bin symbol.
Customers should contact their epoc distributer or Epocal Inc., the manufacturer, to arrange for disposal of their epoc electrical and electronic hardware at the end of product life. Contact information is provided on the cover page of this instruction manual.
51000625 Rev: 08
Introduction
1-3
02
epoc Blood Analysis System
2.1 System Overview The epoc Blood Analysis System o
Is a Portable Blood Analyzer comprised of 3 components:
epoc Reader
epoc Host mobile computer
epoc Test Card
The epoc Reader o
Is a Battery powered portable device
o
Has an internal barcode scanner
o
Has card slot for accepting test card
o
Reads epoc test cards during blood test
o
Has status indicators to inform user of test progress
o
Measures electrical signals from test card sensors
o
Transmits test results wirelessly via Bluetooth to the epoc Host
The epoc Host o
Is a Dedicated use mobile computer with epoc Host software application installed
o
Communicates wirelessly via Bluetooth with epoc Reader
o
Calculates analytical values from sensor data sent by epoc Reader
o
Displays test results
The epoc Test Card o
Is a single-use device with port for blood sample introduction
o
Contains array of sensors on a Sensor Module
o
Contains calibration fluid within sealed reservoir
o
Generates electrical signals proportional to analyte concentrations in sample
o
Uses Barcode to identify Card Type, “Use By” Date, Serial and Lot Number
51000627 Rev: 06
epoc Blood Analysis System
2-1
2.2 Operation Overview
A single blood test is performed in the following way: 1
Use epoc Host to establish wireless connection with an epoc Reader.
2
Obtain a new Test Card and remove from pouch.
3
Insert Test Card into Reader. The Reader reads the barcode on the card. “Use By” date is checked and card lot number is linked to test result and any related patient data that was entered.
4
Internal motor of the Reader is actuated to start the calibration process. This process releases a calibration fluid in the card that flows across sensors within the card. User has time to prepare the patient and draw the blood sample.
5
User introduces the sample into the Test Card upon completion of calibration.
6
Reader sends test data to the Host. Results are calculated and displayed on the Host in about half a minute.
51000627 Rev: 06
epoc Blood Analysis System
2-2
03
epoc System Operation
3.1 System Operation Overview Follow set up instructions for the epoc Reader and epoc Host before conducting a test. Only use the Test Cards that have been properly stored. To complete a test, the following steps are required: 1. Turn “ON” the epoc Reader and epoc Host. 2. Log in to epoc Host software application. 3. Discover the epoc Reader by connecting wirelessly from the epoc Host. 4. Begin the test sequence. 5. Insert a new Test Card into the epoc Reader. 6. Enter patient information, select tests and sample type (if applicable). 7. Introduce a blood sample into the Test Card. 8. Observe and possibly print the Test Results. 9. Remove the Test Card and discard it as biohazard waste. Once logged in and connected to an epoc Reader, steps 5 through 9 above are required to perform another test.
3.2 Turning “ON” the epoc Reader Depress the Power Switch to turn “ON” the epoc Reader. The Power LED will turn green indicating the epoc Reader is “ON” and ready for use. epoc Readers that are switched “ON” will be located by the epoc Host and display on its main screen. Dedicated Readers display on the screen regardless of whether or not they are turned on. However, connection will be successful only if they are ”ON.”
Power Switch
Turn “OFF” any epoc Readers not in active use to conserve battery power.
51000635 Rev: 13
epoc System Operation
3-1
3.3 Turning “ON” the epoc Host The Power button of the epoc Host mobile computer starts and stops the device. Press the Power button to wake up the epoc Host if the screen is blank. The Soft Reset (Warm Reset) button reboots epoc Host application and displays the Login page. Note: Refer to the epoc Host Quick Start Guide to locate these buttons on your Host.
3.4 Logging in to epoc Host Software Application After a soft (warm) reset or logging out, epoc Host software application displays the Login Page. Enter a valid User ID and Password and press the Login button. Note: User ID and Password requirements may vary depending upon how the Administrator has set up the login requirements.
3.5 Running a Test on a Dedicated Reader If the epoc Host is already configured to connect with a single dedicated epoc Reader, the epoc Host will automatically connect to that epoc Reader to run a blood test and begin the Electronic QC Test. A connection is cancelled by tapping the Cancel button.
51000635 Rev: 13
epoc System Operation
3-2
3.6 Alternate Means to Run a Test If the system administrator has configured the system for multiple dedicated epoc Readers, all epoc Readers available for connection are listed. The Reader Icon displays the Reader name and the serial number below.
can also be tapped to find (“discover”) The Discovery icon more Readers. Tapping the discovery icon when inactive will initiate this discovery process. Tapping the discovery icon while discovering will end the discovery process.
Once the desired Reader is displayed, press and hold its Reader Icon to select it for testing. A dropdown menu will be displayed. For a blood test, select Run blood test. For a QA test (if authorized), select Run QA test. You can also start a blood test by double-tapping the Reader Icon
.
3.7 Reader Electronic Internal QC Test Connecting to an epoc Reader displays a Reader Screen unique to the epoc Reader’s serial number. The Reader name is displayed on the bottom tab, with the serial number in parentheses. Every time a Host and Reader connect, the Reader begins a two-level Electronic QC Test. Configuration data is sent by the Host to the Reader, and preparations begin for a test. After completing the electronic QC test and configuration, the Reader Screen displays Insert card to begin test, and the Test Status Indicator of the Reader turns and stays green. 8-Hour Electronic QC Check: The epoc Host checks that an Electronic QC Test has been run within the past 8 hours. If a new Test Card is inserted after the epoc Host has been connected to the Reader continuously for 8 hours or more, the Host will disconnect from the Reader and inform the user they must reconnect to the Host to run another Electronic QC Test.
51000635 Rev: 13
epoc System Operation
3-3
3.8 Reader Screen The epoc Host and epoc Reader are ready to begin a test by inserting a Test Card. The Reader Screen displays: 1. Type of test:
Blood test or
QA test.
2. Battery charge level of the Reader
.
3. Current Date and Time. 4. Patient ID or Lot Number field. 5. Additional tabs for accessing other test information entries that may be used to ensure a complete test record. Requirements may vary according to the policy of the healthcare institution. Always verify that current date and time are correct before running a test. The date and time displayed become part of the test record. Contact the administrator prior to running a test if adjustment of the date and/or time is required. The policy of the healthcare institution may require the user to select the Analytes, Sample Type, and/or Hemodilution Correction Factor for testing. If running Test Cards on the Use By date identified by the Hourglass Icon on the bottom of the Test Card, allow sufficient time to complete the test before midnight. Test results do not display after midnight of the Use By date.
3.9 Obtaining the Test Card 1. Select a properly stored Test Card. 2. Starting at the Notch, tear open the Test Card pouch as shown. 3. Carefully (read the cautions below) remove the Test Card from the Test Card pouch. 4. Place the Test Card directly into the epoc Reader’s Card Insertion Slot. 5. Discard the empty pouch.
51000635 Rev: 13
epoc System Operation
Notch
3-4
Sensor Module’s contact surface
Never touch the Sensor Module’s contact surface or Blood Sample Entry Port. Never place the Test Card on any surface before running a test. Always take the Test Card directly from the pouch before inserting into the Reader. The Test Card pouch should be opened only when conducting a blood or QA test to assure a low humidity environment for the Test Card. Blood Sample Entry Port
3.10 Inserting the Test Card The epoc Reader must be placed on a stable horizontal surface, such as a tabletop, prior to inserting the Test Card. Never insert anything except a Test Card into the Reader’s Card Insertion Slot.
This side up
Position the Test Card with the Blue Label side facing upwards and the Sensor Module towards the Reader. Test Cards are “keyed” using a notch in the corner to ensure correct card orientation during insertion. Insertion of a Test Card causes the Barcode Reader in the Reader to turn “ON.” Insert the Test Card into the Reader’s Card Insertion Slot at the front of the Reader with a smooth single motion to enable the Barcode on the Card Label to be correctly read by the Reader during insertion.
Continue inserting the Test Card until slight resistance is felt. Push the Test Card past this point to “lock” it into place. This is the final Test Card position. Avoid abrupt stops or unevenness in speed during Test Card insertion. Upon correct Test Card insertion, the Reader is configured for the card type indicated by the Test Card Barcode. The Reader performs a series of card integrity checks. The Reader beeps once, and the Test Status Indicator turns solid green to notify the user that the Test Card has been successfully inserted. 51000635 Rev: 13
epoc System Operation
3-5
Any problem reading the Barcode (or any other error) causes the Test Status Indicator to turn solid red. Check the Host for an error message and completely remove the Test Card from the Reader. Reinsert the card for successful insertion confirmed by a solid green Test Status Indicator.
3.11 Calibration Sequence Once the Test Card is successfully inserted, a motorized mechanism in the Reader can be heard as the calibration fluid is released over the sensors within the Test Card. The Test Status Indicator on the Reader flashes green to indicate the start of the test calibration sequence. The Host confirms the start of the test by entering the calibration mode and displaying the calibration progress. The calibration process takes approximately 165 seconds to complete. During this time, the user can prepare the patient and obtain a blood sample. The Reader must rest on a flat horizontal surface without movement for the duration of the test.
3.12 Entering Patient Information (or Lot Number) and Test Selection The Patient ID and related information can be entered at any time during the test. For a Blood Test, the Patient ID number is entered to identify the test results for the Test Card used for the test. For a QA Test (not shown), the QA fluid Lot Number is entered instead of the Patient ID. on the Reader Screen to Select the Test Information Tab enter the patient information. The Administrator may require Sample Type or Hemodilution entries. Using the arrow , additional settings related to respiratory therapy, age, and gender may be entered. Patient information entered prior to completion of the test is saved automatically with the test results when the test is complete. Patient information entered after the test is complete, but before the next test starts, must be saved by tapping the Save button. The Test Information Page for performing QA test contains only the Comments field (not shown). If the Patient ID is not entered prior to completion of the test, the user is prompted to enter the Patient ID when the test results are displayed. 51000635 Rev: 13
epoc System Operation
3-6
Exercise care when entering Patient ID and other information. Ensure the correct Reader is selected by verifying that the Reader name corresponds with the Reader used to conduct the test.
on the Reader Screen to Select the Test Selection Tab select or deselect analytes to display in the test results. The system administrator settings may require analyte selection before results are displayed. Additional analytes may be selected after test completion. Once test results are displayed, analytes can no longer be de-selected.
3.13 Using Barcode Scanner to Enter Patient ID Press the stylus in the Patient ID field. A cursor appears. Activate the Barcode Scanner by pressing any of the Scan buttons on the epoc Host. The Barcode Icon at the top of the screen indicates when the barcode scanner is ready to scan. Point the light coming from the top of the Barcode Scanner towards the desired barcode until a beep is heard. The scanner turns off. The scanned text appears in the field where the cursor was left. Warning: Never look directly at the laser light. Point the laser only at a barcode and away from the eyes at all times. The Patient ID may also be entered using the stylus and the Onscreen Keyboard accessible from bottom of screen via the Text Input button.
3.14 Collecting a Blood Sample Read information on Sample Collection in the BGEM Test Card Specifications section of this Manual to ensure that blood samples are properly collected and handled for testing.
3.15 Timing of Sample Introduction After about 165 seconds of calibration the Test Status Indicator stops flashing green indicating that the Test Card is ready to receive a test sample. The epoc Host displays the message, “Inject sample….” The screen has a bar indicating the time remaining to introduce a sample. The blood sample must be introduced into the Test Card during this 450-second (or 7.5-minute) period. 51000635 Rev: 13
epoc System Operation
3-7