ULTRASOUND DIAGNOSTIC INSTRUMENT
Instruction Manual Measurement (volume 2/2)
(volume 1/2)
Instruction manuals consist of Safety Instruction, How to Use and this manual. Before using this instrument, please read Safety Instruction.
MN1-5370 rev.8
ProSound logo is registered mark of Hitachi Aloka Medical, Ltd. in Japan and other countries. Copyright©Hitachi Aloka Medical, Ltd. All rights reserved. Microsoft and Windows Media player is registered trademark of Microsoft Corporations in United States and/or other countries. All brand name and product name are trademarks or registered trademarks of their respective companies. In this manual, ® and ™ are omitted. VS-FlexGrid Pro copyright©1999-2000 Videosoft Corporation. Portions of this software are based in part on the work of the Independent JPEG Group. Real-time Tissue Elastography is a registered trademark of Hitachi Medical Corporation.
MN1-5370 rev.8
Introduction
Introduction This is an instruction manual for model ProSound α7, an ultrasound diagnostic instrument. Read the manual carefully before using the instrument. Take special note of the items in Chapter 1, "Safety Precautions" of Safety Instruction manual. Keep this manual securely for future reference.
Symbols Used in this Document The following items are important in preventing harm or injury to equipment operator or patient. There are four levels of harm/damage that can be caused by ignoring instructions/displays or using the equipment incorrectly: "Danger," "Warning," "Caution," and "Note." These types are indicated by the following symbols. Indicates an imminently hazardous situation that will result in the death of or serious injury to the equipment operator. Indicates a hazardous situation that could result in death or serious injury. Indicates a hazardous situation that may result in slight or moderate injury, or property damage. Indicates a request concerning an item that must be observed in order to prevent damage or deterioration to the equipment and also to ensure effective use.
Cautions use the following graphics.
This mark indicates and alert, additional information.
This mark indicates that the action is not allowed.
This mark indicates that the action is required.
Conventions used in this manual. NOTE: Notes containing additional information. IMPORTANT: Information that is considered especially important.
Input, output and screen-messages are presented in the following font: message. Menus and switches are written as Menu. Submenus are indicated by the use of angle brackets: Menu > sub-menu > sub-menu.
MN1-5370 rev.8
3
Introduction
About the model “ProSound α7” The ProSound α7 is intended to be used by doctors and other qualified personnel in fracture diagnostics and hemodynamic diagnostics. However, this equipment is not designed to be used in ophthalmic ultrasound diagnosis, as its sound intensity is not compliant with ophthalmic restrictions established by the FDA. Only physicians and other qualified personnel should operate this equipment for diagnostic purposes. Read section 1-1 of the Safety Instruction. 1)
PRECAUTIONS Concerning the Use/Management of the ProSound α7 •
Do not disassemble, repair or remodel this equipment or optional features without our consent. NOTE: Disassemble means to remove the parts or options from the equipment
using tools. NOTE: Remodel means to install or connect unauthorized parts or equipment
including the power cord. •
Assembly of the equipment or optional accessories shall be performed by our third party certified. Please contact one of our offices listed on back cover. NOTE: Assemble means to install or connect parts or optional accessories in/to the
device using tools.
2)
•
Transporting this equipment (via automobile/ship) shall be performed by a third party certified by the manufacturer. Please contact one of our offices listed on back cover.
•
Please conduct routine cleaning and inspection of the equipment. Refer to Chapter 5 of the Safety Instruction for details. .
•
Ensure that the output level of the scan conforms to the required duration of diagnosis.
•
If any malfunction or abnormality is discovered during operation of the equipment, remove the probe from the patient immediately and discontinue use. If any abnormality is observed in the patient, provide proper care as quickly as possible. Refer to Chapter 4 of the Safety Instructions for more information on dealing with the equipment appropriately. If the malfunction is not listed in Chapter 4 of the Safety Instructions, contact one of our offices listed on back cover.
PRECAUTIONS for the ProSound α7 Installation This equipment is a electrical medical device that intended for use in hospitals and research facilities. The device should be installed in accordance with the following guidelines. •
Install in accordance with Chapter 3 of the Safety Instructions.
•
Install in an environment that conforms to the operating environments indicated in section 2-2-2 of the Safety Instruction manual.
•
Install in an environment that ensures electromagnetic compatibility, in accordance with section 1-2-6 of the Safety Instruction manual, "Precautions Concerning the Maintenance of Electromagnetic Compatibility," and Item 1-3, "Guidelines for Electromagnetic compatibility." NOTE: The electromagnetic compatibility (EMC) is the ability of device to function
satisfactorily in its electromagnetic environment without introducing intolerable electromagnetic disturbance to anything in that environment.
4
MN1-5370 rev.8
Introduction
Classification ProSound α7 • Protection against electric shock (ME equipment): class I • ME equipment • Protection Against Electric Shock (Applied Parts): Type BF Applied Parts –
Probe/scanner applied parts and parts treated as applied parts: Refer to the following diagram (Probe/Scanner Pattern Diagram) and table.
Figure: Probe/Scanner Pattern Diagram Above illustrates a surface/intraoperative probe. Below shows a coelomic probe. B
C
A
connector
D
connector
C
A
Applicable part of body
Applied part
parts treated as applied parts
B - C length
surface of body
Ultrasonic irradiation area (D)
A to B
100 cm
Intraoperative
Ultrasonic irradiation area (D)
A to B
20 cm
Endocavity
A to C
A to C
N/A
–
Physiological signal applied part: ECG electrodes Part treated as applied part: 2m from the ECG electrode of the ECG patient cable (consult following diagram) 2 meters ECG electrodes
connector ECG patient lead
• Protection against electric shock (Defibrillation-proof applied parts): Not suitable • Protection against harmful ingress of water or particulate matter –
equipment: IPX0 (Ordinary equipment)
–
Probe applied part: IPX7 (Watertight equipment)
• Suitability for use in an oxygen rich environment: Not suitable • Method(s) of sterilization: Not suitable for sterilization/disinfection with medicinal solution, gas or radiation. • Mode of operation: Continuous operation
MN1-5370 rev.8
5
CONTENTS This book consists of two separate volumes. These two volumes have the same table of contents and the index respectively.
1. Measurement Functions 1-1. 1-2. 1-3. 1-4. 1-5.
1-6. 1-7.
1-8. 1-9.
1-10.
1-11. 1-12.
Preface... 1-1 Flow of Measurement Operations... 1-2 Switches Used for Measurement... 1-3 1-3-1. The function of the panel switch used by the measurement operation ... 1-3 Basic Operation Procedure for Measurement ... 1-4 1-4-1. Method of starting measurement... 1-4 Explanation of the Measurement Menus ... 1-9 1-5-1. When the measurement Study is changed... 1-9 1-5-2. When the measurement application is changed ... 1-10 Executing the Application Measurement Using the Transfer Function... 1-11 Measurement Mark and Measurement Method ... 1-13 1-7-1. Basic types of marks ... 1-13 1-7-2. Auxiliary line type marks... 1-13 1-7-3. Display mark ... 1-14 1-7-4. The basic operating method for each mark type ... 1-15 Basic Measurement Functional Outline... 1-22 1-8-1. Basic Measurement Functional List... 1-22 Measurement operation procedure... 1-26 1-9-1. B mode ... 1-26 1-9-2. M mode ... 1-45 1-9-3. D mode ... 1-50 1-9-4. B/D mode ... 1-66 1-9-5. B (Flow) mode ... 1-73 1-9-6. Calibration... 1-77 Preset Function... 1-82 1-10-1. Composition of the preset function ... 1-82 1-10-2. Basic operations and function of each button ... 1-83 1-10-3. Initializing Preset ... 1-84 1-10-4. Create Measurement Tools ... 1-85 1-10-5. Application Measurement Tools ... 1-91 1-10-6. Study Assignment ... 1-100 1-10-7. SW Assignment... 1-104 Calculation Formula & Reference ... 1-108 1-11-1. Calculation ... 1-108 Abbreviation... 1-111
2. Cardiac Measurement 2-1. 2-2.
Preface... 2-1 Cardiac Measurement Functional Outline ... 2-2 2-2-1. Cardiac Measurement Functional List ... 2-2 6
MN1-5370 Rev.8
2-3.
2-4.
2-5.
2-6.
2-7.
2-2-2. Items of Special Note... 2-7 2-2-3. Measurement Views for Measuring Cardiac Functions ... 2-8 Measurement operation procedure ... 2-10 2-3-1. B mode... 2-12 2-3-2. M mode... 2-34 2-3-3. D mode... 2-43 2-3-4. Flow mode ... 2-75 Report function ... 2-83 2-4-1. Basic Operation of a Report ... 2-83 2-4-2. Report Block...2-84 2-4-3. Description of Various Data Displayed in a Report ... 2-88 2-4-4. Function that Attaches an Ultrasound Image to a Report... 2-90 2-4-5. Printing Function ... 2-93 2-4-6. Output to a Personal Computer... 2-95 2-4-7. Output to a CSV file ... 2-96 Preset function ... 2-97 2-5-1. Preset Settings(Cardiac Measurement)... 2-97 2-5-2. PRESET list ... 2-98 Calculation Formula & Reference...2-106 2-6-1. Calculation... 2-106 2-6-2. Clinical References... 2-115 Abbreviation ... 2-123
3. Vascular Measurement 3-1. 3-2.
3-3.
3-4.
3-5.
3-6.
MN1-5370 Rev.8
Preface ... 3-1 Vascular Measurement Functional Outline ... 3-2 3-2-1. Vascular Measurement Functional List ... 3-2 3-2-2. Items of Special Note... 3-3 Measurement operation procedure ... 3-4 3-3-1. B mode... 3-6 3-3-2. D mode... 3-16 Report function ... 3-27 3-4-1. Basic Operation of a Report ... 3-27 3-4-2. Report Block...3-28 3-4-3. Description of Various Data Displayed in a Report ... 3-32 3-4-4. W. Trace Function ... 3-39 3-4-5. Function that Attaches an Ultrasound Image to a Report... 3-41 3-4-6. Printing Function ... 3-44 3-4-7. Output to a Personal Computer... 3-46 3-4-8. Output to a CSV file ... 3-47 Preset function ... 3-48 3-5-1. Preset Settings... 3-48 3-5-2. PRESET list ... 3-49 Calculation Formula & Reference...3-56 3-6-1. Calculation... 3-56 3-6-2. Anatomy Check List ... 3-57 7
3-7.
3-6-3. Clinical References ... 3-58 Abbreviation... 3-60
4. Abdominal measurement 4-1. 4-2.
4-3.
4-4.
4-5.
4-6.
4-7.
Preface... 4-1 Abdominal Measurement Functional Outline... 4-2 4-2-1. Abdominal Measurement Functional List... 4-2 4-2-2. Items of Particular Note ... 4-4 Measurement operation procedure... 4-5 4-3-1. B mode ... 4-6 4-3-2. D mode ... 4-14 Report function ... 4-24 4-4-1. Basic Operation of a Report... 4-24 4-4-2. Report Block ... 4-25 4-4-3. Description of Various Data Displayed in a Report... 4-29 4-4-4. Graph... 4-32 4-4-5. W. Trace... 4-33 4-4-6. Function that Attaches an Ultrasound Image to a Report ... 4-35 4-4-7. Printing Function... 4-38 4-4-8. Output to a Personal Computer ... 4-40 4-4-9. Output to a CSV file... 4-41 Preset function... 4-42 4-5-1. Preset Settings ... 4-42 4-5-2. PRESET list ... 4-43 Calculation Formula & Reference ... 4-49 4-6-1. Calculation ... 4-49 4-6-2. Clinical References ... 4-49 Abbreviation... 4-50
5. Obstetrical Measurement 5-1. 5-2.
5-3.
5-4.
Preface... 5-1 Obstetrical Measurement Functional Outline ... 5-2 5-2-1. Obstetrical Measurement Functional List ... 5-2 5-2-2. List of obstetrical measurement name built into system... 5-5 5-2-3. Items of Special Note ... 5-10 Measurement operation procedure... 5-11 5-3-1. B mode ... 5-12 5-3-2. M mode ... 5-28 5-3-3. D mode ... 5-30 5-3-4. Multiple pregnancies... 5-35 5-3-5. Interval Growth Rate... 5-37 Report function ... 5-38 5-4-1. Basic Operation of a Report... 5-38 5-4-2. Report Block ... 5-39 5-4-3. Description of Various Data Displayed in a Report... 5-43
8
MN1-5370 Rev.8
5-5.
5-6.
5-7.
5-4-4. Function that Attaches an Ultrasound Image to a Report... 5-58 5-4-5. Printing Function ... 5-61 5-4-6. Output to a Personal Computer... 5-63 5-4-7. Output to a CSV file ... 5-64 Preset function ... 5-65 5-5-1. Preset Settings... 5-65 5-5-2. PRESET list ... 5-67 Calculation Formula & Reference & Table...5-76 5-6-1. Calculation... 5-76 5-6-2. Anatomy Check List ... 5-80 5-6-3. BPP Scoring... 5-82 5-6-4. References... 5-84 5-6-5. Data in the fetal growth table inside the system ... 5-98 Abbreviation ... 5-139
6. Gynecological Measurement 6-1. 6-2.
6-3.
6-4.
6-5.
6-6.
6-7.
Preface ... 6-1 Gynecological Measurement Functional Outline ... 6-2 6-2-1. Gynecological Measurement Functional List... 6-2 6-2-2. Items of Special Note... 6-3 Measurement operation procedure ... 6-4 6-3-1. B mode... 6-6 6-3-2. D mode... 6-13 Report function ... 6-15 6-4-1. Basic Operation of a Report ... 6-15 6-4-2. Report Block...6-16 6-4-3. Description of Various Data Displayed in a Report ... 6-20 6-4-4. Function that Attaches an Ultrasound Image to a Report... 6-25 6-4-5. Printing Function ... 6-28 6-4-6. Output to a Personal Computer... 6-30 6-4-7. Output to a CSV file ... 6-31 Preset function ... 6-32 6-5-1. Preset Settings... 6-32 6-5-2. PRESET list ... 6-33 Calculation Formula & Reference...6-38 6-6-1. Calculation... 6-38 6-6-2. Anatomy Check List ... 6-38 6-6-3. Clinical References... 6-39 Abbreviation ... 6-41
7. Urological Measurement 7-1. 7-2.
MN1-5370 Rev.8
Preface ... 7-1 Urological Measurement Functional Outline ... 7-2 7-2-1. Urological Measurement Functional List ... 7-2 7-2-2. Items of Particular Note... 7-3
9
7-3.
7-4.
7-5.
7-6.
7-7.
Measurement operation procedure... 7-4 7-3-1. B mode ... 7-6 7-3-2. D mode ... 7-18 Report function ... 7-20 7-4-1. Basic Operation of a Report... 7-20 7-4-2. Report Block ... 7-21 7-4-3. Description of Various Data Displayed in a Report... 7-25 7-4-4. Graph function ... 7-30 7-4-5. Function that Attaches an Ultrasound Image to a Report ... 7-31 7-4-6. Printing Function... 7-34 7-4-7. Output to a Personal Computer ... 7-36 7-4-8. Output to a CSV file... 7-37 Preset function... 7-38 7-5-1. Preset Settings ... 7-38 7-5-2. PRESET list ... 7-39 Calculation Formula & Reference ... 7-43 7-6-1. Calculation ... 7-43 7-6-2. References ... 7-43 Abbreviation... 7-45
8. eTRACKING Measurement 8-1. 8-2.
8-3.
8-4.
Preface... 8-1 eTRACKING Measurement Functional Outline ... 8-1 8-2-1. Outline of Function ... 8-1 8-2-2. Measurement operation procedure... 8-1 Report function ... 8-2 8-3-1. Basic Operation of a Report... 8-2 8-3-2. Report Block ... 8-3 8-3-3. Description of Various Data Displayed in a Report... 8-5 8-3-4. Function that Attaches an Ultrasound Image to a Report ... 8-6 8-3-5. Printing Function... 8-10 8-3-6. Output to a Personal Computer ... 8-12 8-3-7. Output to a CSV file... 8-13 Preset function... 8-14 8-4-1. Preset Settings ... 8-14 8-4-2. PRESET list ... 8-15
9. Small Parts Measurement 9-1. 9-2.
9-3.
Introduction... 9-1 Overview of the Small Parts Measurement Function ... 9-2 9-2-1. Small Parts Measurement Function List ... 9-2 9-2-2. Items of Special Note ... 9-3 Measurement Procedure... 9-4 9-3-1. B mode ... 9-6 9-3-2. D mode ... 9-11
10
MN1-5370 Rev.8
9-4.
9-5.
9-6.
9-7.
MN1-5370 Rev.8
Report Function ... 9-13 9-4-1. Basic Operation for Reports ... 9-13 9-4-2. Report Block...9-14 9-4-3. Description of Various Data Displayed in a Report ... 9-18 9-4-4. Function to Attach Ultrasound Images to a Report ... 9-27 9-4-5. Print function ... 9-30 9-4-6. Output to a Personal Computer... 9-32 9-4-7. Output to a CSV file ... 9-33 Preset function ... 9-34 9-5-1. Preset Settings... 9-34 9-5-2. Preset list... 9-35 Calculation Formula & References ... 9-39 9-6-1. Calculation... 9-39 9-6-2. Clinical References... 9-39 Abbreviation ... 9-40
11
12
MN1-5370 Rev.8
4. Abdominal measurement 4-1.
Preface The description concerning the Abdominal measurement functions is divided into the following six sub-sections. 4-1.
Preface
4-2.
Abdominal Measurement Functional Outline
4-3.
Measurement operation procedure
4-4.
Report Function
4-5.
Preset Function
4-6.
Calculation Formula & Reference
This section describes the procedure for carrying out abdominal measurements, based on the assumption that products are on the factory default. Descriptions of the basic operations of the measurement functions and each measurement method (mark type = Caliper, Trace, etc.) are given in the Section 1. "MEASUREMENT FUNCTIONS".
This section consists of 52 pages. MN1-5370 Rev.8
4-1
4.Abdominal measurement 4-2.Abdominal Measurement Functional Outline
4-2.
Abdominal Measurement Functional Outline
4-2-1. Abdominal Measurement Functional List Abdominal measurements use studies consisting of various combinations of the measurement menu, report display, and so on, depending which part is being examined and the purpose of the examination. [Remark] On the factory default, it contains Basic Study. The abdominal measurement is divided as shown in the figure below for each mode. : Items that are displayed on the factory default.
4-2-1-1.
B mode
Measurement function Gallbladder measurement
Measurement menu
Display items
Gallbladder
GB-L
GB Wall-T
GB Wall-T
Remark
GB-W
Measured at one cross-section.
CBD measurement
CBD
CBD
Liver measurement
Liver (Left Lobe) Liver (Right Lobe)
L1
L2
Pancreas measurement
Pancreas
Head
Body
P-Duct
P-Duct
Renal measurement
Renal Volume (Right, Left)
Vol.
Rnl-L
Rnl-W
Spleen measurement Spleen
a
b
SI
Space-occupying lesion measurement
SOL1 SOL2
d1
d2
d3
Blood vessel diameter measurement
Aorta Diam
Aorta Diam
PV Diam
PV Diam
% stenosis measurement
%STENO Diam %STENO Area
Measured at one cross-section. Measurement of left lobe and right lobe. Tail
Measured at multiple cross-sections. Measured at one cross-section. Rnl-AP
Measured at two cross-sections.
SI: Spleen Index(axb)
Area
Measured at two cross-sections.
Volume
Vessel
Measured at one cross-section.
Resid
4-2
%STENO
Diagnosis for the rate of stenosis.
MN1-5370 Rev.8
4.Abdominal measurement 4-2-1.Abdominal Measurement Functional List
4-2-1-2.
D mode
Measurement function Artery measurement
Renal Artery measurement
Measurement menu A-Ao, CA, CHA, SA, SMA, IMA, CIA, HA, Prandial SMA, Artery1, Artery2, Artery3 *Artery1- 3: It is possible to freely define and use a name according to the particular purpose and application. Renal-A
Display items
Remark
PI
RI
S/D
PSV
EDV
MnV
ACC
AccT
PI
RI,
S/D
PSV
EDV
MnV
ACC
AccT
FlowT
AccT/FT
Diagnosis for the rate of stenosis CIA: Consists of Right/ Left Prandial SMA: Consists of Pre/Pst
Vm
Diagnosis of degree of stenosis Consists of Right/Left
Vm
Portal vein measurement
Main PV Rt.PV Lt.PV
pV
For portal vein blood flow.
Shunt blood vessel measurement
Pre Shunt PV Prox Shunt Mid Shunt Distal Shunt
pV
For shunt blood flow measurement.
4-2-1-3.
B mode, D mode
Measurement function Flow quantity measurement
Measurement menu FV(Artery) Abdom
Mean V
VTI
Display items
Remark
MnV
PSV
EDV
Δv
Δt
VTI
PI
RI
FT
AccT
ACC
AccT/FT
S/D
Vm
FV(MnV)
FVm(MnV)
CSD
CSA
MnV
PSV
EDV
Δv
Δt
VTI
PI
RI
FT
AccT
ACC
AccT/FT
S/D
Vm
FV(beat)
FV(min)
CSD
CSA
HR
For arterial blood flow.
*For peripheral blood vessels. FV(Vein) Abdom
MN1-5370 Rev.8
MnV
pV
VTI
AccT
ACC
FV
CSD
CSA
4-3
For steady state blood flow.
4.Abdominal measurement 4-2.Abdominal Measurement Functional Outline
4-2-2. Items of Particular Note For blood flow velocity measurement: When recording an arterial flow Doppler waveform (pulse method), if the Doppler incident angle exceeds 60× the measurement error becomes large. The measured values of the blood flow values obtained using this equipment are the absolute values displayed on the observation monitor. They are controlled as positive and negative values for the purpose of calculating the arithmetic index. If the display of each measured value in a report is set to “Average” in a preset, the positive and negative values are added together and displayed as a mean value. Consequently, when performing multiple measurements of blood flow on the blood flow waveform drawn using the color Doppler method as a guide, use identical recording conditions (forward and reverse flow directions) for all of the blood flow waveforms in order to correctly display each of the arithmetic values arranged in the report.
4-4
MN1-5370 Rev.8
4.Abdominal measurement 4-3.Measurement operation procedure
4-3.
Measurement operation procedure Abdominal measurement has the following study. Basic Study When multiple Studies are set by the preset function, it can be possible to switch the Study & Application by the touch panel.
<Displaying marks of registered reports > When the registration of the report is made after a measurement is made for each item, the number of registration times is displayed on the touch panel.
The number of registration times [Remark] The display examples of the measurement results, in this chapter, are displayed with a vertical display layout. [Remark] On the factory default, SEND switch is not assigned to the operation panel. To use the switch, assign it using CustomSW2 in Preset Setup. When the SEND switch is assigned to the USER2 switch, select Enter/Send for assigning switches. For details, refer to the How to Use manual.
MN1-5370 Rev.8
4-5
4.Abdominal measurement 4-3.Measurement operation procedure
4-3-1. B mode 4-3-1-1.
Gallbladder measurement
This is the measurement of the length and width of the vertical cross-section of the gallbladder.
<Operation method> (1)
Record a slice image of the gall bladder in the B mode.
(2)
Select the Gallbladder on the touch panel. The + mark is displayed, so measure the Length.
(3)
→ (4)
Length
Press the + switch. Measure the Width.
Width
Press the SEND switch. →
GB Wall-T
Measurement is finalized.
<Example of Gallbladder results display> Gallbladder
4-3-1-2.
GB-L:
.
cm
← Length of the gallbladder
GB-W:
.
cm
← Width of the gallbladder
GB Wall-T measurement
This is the measurement of the thickness of the gallbladder wall.
<Operation method> (1)
Record a slice image of the gallbladder in the B mode.
(2)
Select the GB Wall-T on the touch panel. →
(3)
The + mark is displayed, so measure the GB wall thickness.
Press the SEND switch. →
Measurement is finalized.
<Example of GB Wall-T results display> GB-Wall-T .
cm
← Thickness of gallbladder wall
4-6
MN1-5370 Rev.8
4.Abdominal measurement 4-3-1.B mode
4-3-1-3.
CBD measurement
This is the measurement of the diameter of the common bile duct.
<Operation method> (1)
Record a slice image of the common bile duct in the B mode.
(2)
Select the CBD on the touch panel. A + mark appears, so measure the diameter of the common bile duct.
(3)
Press the SEND switch. →
Measurement is finalized. CBD
<Example of CBD results display> ← Common Bile Duct
CBD: .
4-3-1-4.
← Diameter of the common bile duct
cm
Liver measurement
This is the measurement of the left and right lobes of the liver. The method of measurement for both the left and right lobes is the same.
<Operation method> (1)
Record a slice image of the right (or left) lobe in the B mode.
(2)
Select Right or Left of Liver on the touch panel. →
(3)
Length1
The + mark is displayed, so measure the Length1.
Press the + switch. →
Measure the Length2. Left Lobe
(4)
Press the SEND switch. →
Measurement is finalized.
<Example of Liver results display> Liver (Right)
Liver (Left)
L1:
.
cm
L1:
.
cm
← Length 1 of the right or left lobe of the liver
L2:
.
cm
L2:
.
cm
← Length 2 of the right or left lobe of the liver
MN1-5370 Rev.8
Length1
4-7
Right Lobe
4.Abdominal measurement 4-3.Measurement operation procedure
4-3-1-5.
Pancreas measurement
This is the measurement of the head, body and tail of the pancreas from its cross-section.
<Operation method> (1)
Record a slice image showing the head of the pancreas.
(2)
Select the Pancreas on the touch panel, and then select Head. →
(3)
Tail Body Head
Measure the Body.
Record a slice image showing the tail of the pancreas, and then press the + switch. →
(5)
The + mark is displayed, so measure the Head.
Record a slice image showing the body of the pancreas, and then press the + switch. →
(4)
P-Duct
Measure the Tail.
Press the SEND switch. →
Measurement is finalized.
<Example of Pancreas results display> Pancreas
4-3-1-6.
Head:
.
cm
← Diameter of head of pancreas
Body:
.
cm
← Diameter of body of pancreas
Tail:
.
cm
← Diameter of tail of pancreas
P-Duct measurement
This is the measurement of the diameter of the pancreatic duct.
<Operation method> (1)
Record a slice image of the pancreas showing the pancreatic duct in the B mode.
(2)
Select the P-Duct on the touch panel. →
(3)
The + mark is displayed, so measure the pancreatic duct.
Press the SEND switch. →
Measurement is finalized.
<Example of P-Duct results display> P-Duct: .
cm
← Diameter of pancreatic duct
4-8