Instruction Manual
362 Pages
Preview
Page 1
ULTRASOUND DIAGNOSTIC INSTRUMENT
SSD-α10 Instruction Manual How to Use (volume 1/2)
Instruction manuals consist of this manual, Safety Instruction and Measurement. Before using this instrument, please read Safety Instruction.
改 訂 出図 2012.3.14 MS 技 アロカ
MN1-5206 rev.20
ProSound logo is registered mark of Hitachi Aloka Medical, Ltd. in Japan and other countries. Copyright©Hitachi Aloka Medical, Ltd. All rights reserved. Microsoft and Windows Media player is registered trademark of Microsoft Corporations in United States and/or other countries. All brand name and product name are trademarks or registered trademarks of their respective companies. In this manual, ® and ™ are omitted. VS-FlexGrid Pro copyright©1999-2000 Videosoft Corporation. Portions of this software are based in part on the work of the Independent JPEG Group.
改 訂 出図 2012.3.14 MS 技 アロカ
MN1-5206 rev.20
Introduction
Introduction This is an instruction manual for model SSD-α10, an ultrasound diagnostic instrument. Read the manual carefully before using the instrument. Take special note of the items in Chapter 1, "Safety Precautions.". Keep this manual securely for future reference.
Symbols Used in this Document The following items are important in preventing harm or injury to equipment operator or patient. There are four levels of harm/damage that can be caused by ignoring instructions/displays or using the equipment incorrectly: "Danger," "Warning," "Caution," and "Note." These types are indicated by the following symbols. Indicates an imminently hazardous situation that will result in the death of or serious injury to the equipment operator. Indicates a hazardous situation that could result in death or serious injury. Indicates a hazardous situation that may result in slight or moderate injury, or property damage. Indicates a request concerning an item that must be observed in order to prevent damage or deterioration to the equipment and also to ensure effective use.
Cautions use the following graphics. This mark indicates and alert, additional information.
This mark indicates that the action is not allowed.
This mark indicates that the action is required.
Conventions used in this manual. NOTE: Notes containing additional information. IMPORTANT: Information that is considered especially important.
改 訂
Input, output and screen-messages are presented in the following font: message. Menus and switches are written as Menu. Submenus are indicated by the use of angle brackets: Menu > sub-menu > sub-menu.
出図 2012.3.14 MS 技 MN1-5206 rev.20 アロカ
3
Introduction
About the model “SSD-α10” The SSD-α10 is intended to be used by doctors and other qualified personnel in fracture diagnostics and hemodynamic diagnostics. However, this equipment is not designed to be used in ophthalmic ultrasound diagnosis, as its sound intensity is not compliant with ophthalmic restrictions established by the FDA. Only physicians and other qualified personnel should operate this equipment for diagnostic purposes. Read section 1-1 of the Safety Instruction. 1)
PRECAUTIONS Concerning the Use/Management of the SSD-α10 •
Do not disassemble, repair or remodel this equipment or optional features without our consent. NOTE: Disassemble means to remove the parts or options from the equipment
using tools. NOTE: Remodel means to install or connect unauthorized parts or equipment
including the power cord. •
Assembly of the equipment or optional accessories shall be performed by a third party certified by us. Please contact one of our offices listed on back cover. NOTE: Assemble means to install or connect parts or optional accessories in/to the
device using tools.
2)
•
Transporting this equipment (by vehicle/ship) shall be performed by a third party certified by the manufacturer. Please contact one of our offices listed on back cover.
•
Please conduct routine cleaning and inspection of the equipment. Refer to Chapter 5 of the Safety Instruction for details.
•
Ensure that the output level of the scan conforms to the required duration of diagnosis.
•
If any malfunction or abnormality is discovered during operation of the equipment, remove the probe from the patient immediately and discontinue use. If any abnormality is observed in the patient, provide proper care as quickly as possible. Refer to Chapter 4 of the Safety Instructions for more information on dealing with the equipment appropriately. If the malfunction is not listed in Chapter 4 of the Safety Instruction manual, contact one of our offices listed on back cover.
PRECAUTIONS for the SSD-α10 Installation This equipment is a electrical medical device that intended for use in hospitals and research facilities. The device should be installed in accordance with the following guidelines.
改 訂
•
Install in accordance with Chapter 3 of the Safety Instructions.
•
Install in an environment that conforms to the operating environments indicated in section 2-2-2 of the Safety Instruction manual.
•
Install in an environment that ensures electromagnetic compatibility, in accordance with section 1-2-6 of the Safety Instruction manual, "Precautions Concerning the Maintenance of Electromagnetic Compatibility," and Item 1-3, "Guidelines for Electromagnetic compatibility." NOTE: The electromagnetic compatibility (EMC) is the ability of device to function
satisfactorily in its electromagnetic environment without introducing intolerable electromagnetic disturbance to anything in that environment.
出図 2012.3.14 MS 技 4 アロカ
MN1-5206 rev.20
Introduction
Classification of SSD-α10 • Protection against electric shock (ME equipment): class I • ME equipment • Protection Against Electric Shock (Applied Parts): Type BF Applied Parts –
Probe/scanner applied parts and parts treated as applied parts: Refer to the following diagram (Probe/Scanner Pattern Diagram) and table.
Figure: Probe/Scanner Pattern Diagram Above illustrates a surface/intraoperative probe. Below shows a coelomic probe. B
C
A
connector
D
connector
C
A
Applicable part of body
Applied part
parts treated as applied parts
B - C length
surface of body
Ultrasonic irradiation area (D)
A to B
100 cm
Intraoperative
Ultrasonic irradiation area (D)
A to B
20 cm
Endocavity
A to C
A to C
N/A
–
Physiological signal applied part: ECG electrodes Part treated as applied part: 2m from the ECG electrode of the ECG patient cable (consult following diagram) 2 meters ECG electrodes
connector ECG patient lead
• Protection against electric shock (Defibrillation-proof applied parts): Not suitable • Protection against harmful ingress of water or particulate matter –
equipment: IPX0 (Ordinary equipment)
–
Probe applied part: IPX7 (Watertight equipment)
• Suitability for use in an oxygen rich environment: Not suitable • Method(s) of sterilization: Not suitable for sterilization/disinfection with medicinal solution, gas or radiation. • Mode of operation: Continuous operation
改 訂 出図 2012.3.14 MS 技 MN1-5206 rev.20 アロカ
5
MN1-5206 Rev.20
CONTENTS This book consists of two separate volumes. These two volumes have the same table of contents and the index respectively.
1. Method of Operating Switches 1-1.
Operation Panel... 1-1
1-2.
Graphic Display ... 1-2
1-3.
Operating the Various Switches... 1-3
改 訂1-4. 出図 2012.3.14 MS 技 アロカ
1-3-1.
ACOUSTIC POWER... 1-3
1-3-2.
B GAIN ... 1-3
1-3-3.
CANCEL... 1-3
1-3-4.
CURSOR/B.L.S ... 1-4
1-3-5.
DEPTH/RANGE ... 1-5
1-3-6.
ENTER... 1-5
1-3-7.
FLOW GAIN ... 1-5
1-3-8.
FOCUS... 1-6
1-3-9.
FREEZE ... 1-7
1-3-10.
M/D GAIN ... 1-7
1-3-11.
MEASUREMENT ... 1-8
1-3-12.
+ ... 1-8
1-3-13.
MENU ... 1-8
1-3-14.
MODE ... 1-9
1-3-15.
NEW PATIENT ... 1-10
1-3-16.
PRESET ... 1-11
1-3-17.
PRINT(Archive Group) ... 1-11
1-3-18.
PROBE... 1-13
1-3-19.
REC ... 1-14
1-3-20.
REVIEW ... 1-14
1-3-21.
SCAN AREA ... 1-14
1-3-22.
SEARCH ... 1-15
1-3-23.
SELECT ... 1-15
1-3-24.
SEND ... 1-16
1-3-25.
STC ... 1-16
1-3-26.
STORE ... 1-17
1-3-27.
VEL RANGE ... 1-17
1-3-28.
ZOOM ... 1-18
1-3-29.
Full keyboard ... 1-20
Before Starting an Examination... 1-21 1-4-1.
Explanation of ID entry screen ... 1-21
1-4-2.
How to register a patient ... 1-31
1-4-3.
Search function... 1-33
1-4-4.
Find ... 1-34 6
MN1-5206 Rev.20
1-5.
1-6.
1-7.
1-4-5.
Worklist ... 1-35
1-4-6.
Data Management... 1-36
Before Completing an Examination ... 1-39 1-5-1.
When the instrument is not connected to the Worklist or MPPS server ... 1-39
1-5-2.
When the instrument is connected to the Worklist and MPPS servers via a network... 1-40
1-5-3.
When connected via network to the DICOM SR Server... 1-44
User Authentication... 1-46 1-6-1.
User Levels ... 1-46
1-6-2.
Login... 1-47
1-6-3.
User Authentication ... 1-47
1-6-4.
Setting new passwords... 1-48
1-6-5.
Changing passwords ... 1-49
1-6-6.
Log Off ... 1-50
1-6-7.
User Management... 1-56
B Mode ... 1-62 1-7-1.
1-8.
1-8-1. 1-9.
1-14.
Basic operation procedure ... 1-74
Tissue Doppler Imaging Display... 1-77 1-12-1.
1-13.
Basic operation procedure ... 1-71
Flow Mode, Power Flow Mode, eFlow Mode... 1-74 1-11-1.
1-12.
Basic operation procedure ... 1-66
B/D and D Modes ... 1-71 1-10-1.
1-11.
Basic operation procedure ... 1-65
B/M and M Modes... 1-66 1-9-1.
1-10.
Basic Operation Procedure ... 1-62
Tissue Harmonic Echo... 1-65
Basic operation procedure ... 1-77
Comment and Annotation Function ... 1-78 1-13-1.
Text entry... 1-79
1-13-2.
↑and text entry... 1-80
1-13-3.
Movement of a character string ... 1-80
1-13-4.
Deleting a character string ... 1-81
1-13-5.
Deleting a character ... 1-81
1-13-6.
System dictionary function ... 1-82
1-13-7.
Registration and deletion of a key word ... 1-83
Treatment at the Time of the Discovery of Abnormality ... 1-85 1-14-1.
Safety security of patient ... 1-85
1-14-2.
Instrumental treatment ... 1-85
改 訂 2. Save of Image 2-1.
Search ... 2-1
出図
2-1-1.
Search and scroll function ... 2-2
2012.3.14
2-1-2.
Basic Operation Procedure ... 2-3
MS 技 アロカ
7
MN1-5206 Rev.20
2-2.
2-3.
2-1-3.
Cine scale and search mark ... 2-4
2-1-4.
Loop playback function... 2-5
Store ... 2-7 2-2-1.
How to Connect USB Memory ... 2-9
2-2-2.
Images and Types feasible to acquire ... 2-12
2-2-3.
Setting of acquiring methods ... 2-14
2-2-4.
Setting of Acquire data... 2-15
2-2-5.
Basic Operation Procedure... 2-16
Review ... 2-23 2-3-1.
Image Viewer ... 2-23
2-3-2.
Search for image data... 2-28
2-3-3.
Saving method of images ... 2-46
2-3-4.
Adjusting images... 2-59
2-3-5.
Printing an image ... 2-60
3. Video Recording/Playing Back with VCR/DVD 3-1.
Setting for VCR/DVD... 3-1
3-2.
Recording Images ... 3-3
3-3.
3-2-1.
Starting and Pausing a Recording ... 3-3
3-2-2.
Remaining DVD capacity display... 3-4
3-2-3.
Recording Counter Number ... 3-5
3-2-4.
Title Menu, Chapter Menu... 3-6
Playing Back ... 3-7 3-3-1.
HS-MD3000(2)(MITSUBISHI) or SVO-9500MD2(SONY) ... 3-7
3-3-2.
BD-X201ME(Victor) ... 3-8
3-3-3.
DVO-1000MD(SONY)... 3-11
3-3-4.
Playback (DV-800) ... 3-13
4. Physiological Signals 4-1.
Electrode Setting ... 4-1
4-2.
Display of Physiological Signals ... 4-1
4-3.
Electrocardiosynclonization Display ... 4-4
4-4.
Heart Rate Stability Display ... 4-5
5. Touch Panel 5-1.
Operation on Touch Panel... 5-3 5-1-1.
5-2.
改 訂 出図
Registration / Change of touch panel ... 5-4
User Switch Area ... 5-7 5-2-1.
Invert ... 5-7
5-2-2.
FAM ... 5-7
5-2-3.
Active(B1 - B4)... 5-8
2012.3.14 MS 技 アロカ
8
MN1-5206 Rev.20
5-3.
5-4.
改 訂
5-2-4.
Body Mark ... 5-8
5-2-5.
Archive Group1 - 3... 5-9
5-2-6.
EXT... 5-9
5-2-7.
Imaging Information... 5-9
5-2-8.
M1 - M4 (Measurement1 to 4) ... 5-9
5-2-9.
End Study... 5-9
5-2-10.
Map Editor ... 5-10
Group Area ...5-13 5-3-1.
Image Func B1 - B3... 5-13
5-3-2.
Puncture ... 5-13
5-3-3.
Focus(B) ... 5-13
5-3-4.
Angle Gain... 5-14
5-3-5.
IP Regist (B) ...5-14
5-3-6.
STC Memory ... 5-15
5-3-7.
Image Func M1 - M3... 5-16
5-3-8.
Focus(M)... 5-16
5-3-9.
IP Regist(M) ...5-17
5-3-10.
Image Func D1 - D3 ... 5-17
5-3-11.
Focus(D) ... 5-17
5-3-12.
IP Regist(D) ... 5-18
5-3-13.
Image Func F1 - F3... 5-18
5-3-14.
Image Func PF1 - PF3 ... 5-18
5-3-15.
Image Func eF1 - eF3 ... 5-18
5-3-16.
Color Map(Flow) ... 5-18
5-3-17.
Display Priority(Flow)... 5-19
5-3-18.
IP Regist(Flow)... 5-19
5-3-19.
Image Function Other1 - 3... 5-20
5-3-20.
Post Processing ... 5-20
5-3-21.
Graphic Editor ... 5-21
5-3-22.
Physio1, 2 ... 5-21
5-3-23.
ECG Sync ... 5-22
5-3-24.
VCR Control...5-22
5-3-25.
Color Map(B/M/D)... 5-22
5-3-26.
Store Set Up... 5-22
Menu Function Area(B Mode) ... 5-23 5-4-1.
AGC(B)... 5-23
5-4-2.
Beam Processing... 5-23
5-4-3.
Beam Steer(B) ... 5-23
5-4-4.
Contrast(B) ... 5-23
5-4-5.
Edge Optimizer... 5-23
出図
5-4-6.
Edge Optimizer Level... 5-23
2012.3.14
5-4-7.
ExPHD(T.H.E.) ... 5-24
MS 技 アロカ
9
MN1-5206 Rev.20
5-5.
改 訂 出図 2012.3.14 MS 技 アロカ
5-6.
5-4-8.
Extend F-View(Extended Field of View) ... 5-24
5-4-9.
FTC(B) ... 5-24
5-4-10.
Frame CORRE Type ... 5-24
5-4-11.
Frame CORRE(B)... 5-24
5-4-12.
Frame Rate(B)... 5-24
5-4-13.
Image Direction(LR)... 5-25
5-4-14.
Image Direction(UL)... 5-25
5-4-15.
Image Enhance ... 5-25
5-4-16.
Image Freq(B/M) ... 5-25
5-4-17.
Image Rotation(B)... 5-25
5-4-18.
IP Select(B) ... 5-25
5-4-19.
OMNI Angle ... 5-26
5-4-20.
PRF Limit... 5-26
5-4-21.
Puncture G.Line ... 5-26
5-4-22.
Puncture Measure... 5-26
5-4-23.
Biopsy Select... 5-27
5-4-24.
Relief(B)... 5-27
5-4-25.
Angle Sel(Punc) ... 5-27
5-4-26.
Smoothing(B)... 5-27
5-4-27.
T.H.E.(Tissue Harmonic Echo)... 5-27
5-4-28.
Spatial Compound... 5-28
5-4-29.
Compound Angle ... 5-28
5-4-30.
Trapezoidal Scan... 5-28
5-4-31.
AIP(Adaptive Image Processing)... 5-29
5-4-32.
AIP Effect... 5-29
5-4-33.
AIP Edge Sens... 5-29
5-4-34.
AIP Edge Enhance ... 5-29
5-4-35.
AIP SR Level ... 5-29
5-4-36.
AIP Resolution ... 5-29
5-4-37.
AIP Level ... 5-30
5-4-38.
Image Optimizer... 5-30
5-4-39.
Brightness Level... 5-30
5-4-40.
Brightness Reset... 5-30
Menu Function Area(M mode) ... 5-31 5-5-1.
AGC(M) ... 5-31
5-5-2.
Echo Erase... 5-31
5-5-3.
Contrast(M) ... 5-31
5-5-4.
FTC(M) ... 5-31
5-5-5.
IP Select(M) ... 5-31
5-5-6.
Relief(M)... 5-32
5-5-7.
Sweep Speed(M) ... 5-32
Menu Function Area(D mode/TDI-D menu) ... 5-33 10
MN1-5206 Rev.20
5-7.
5-6-1.
Angle Correct... 5-33
5-6-2.
Auto Angle Correct... 5-33
5-6-3.
Base Line Shift(D)... 5-33
5-6-4.
Beam Steer(D) ... 5-33
5-6-5.
Comp Curve(D) ... 5-33
5-6-6.
Contrast(D) ... 5-33
5-6-7.
Filter Control(D)... 5-34
5-6-8.
Filter(D) ... 5-34
5-6-9.
Freeze Trigger... 5-34
5-6-10.
Image Polarity(D) ... 5-34
5-6-11.
Image Select(D) ... 5-34
5-6-12.
IP Select(D) ... 5-35
5-6-13.
M/PW(M/PW Mode) ... 5-35
5-6-14.
Resolution Select(D)... 5-35
5-6-15.
Sample Volume ... 5-35
5-6-16.
Image Freq(D) ... 5-35
5-6-17.
Spectrum Invert ... 5-35
5-6-18.
Sweep Speed(D) ... 5-36
5-6-19.
Transfer List... 5-36
5-6-20.
D.Trace Direction ... 5-36
5-6-21.
D.Trace Smooth... 5-36
5-6-22.
D.Trace ... 5-36
5-6-23.
D.Trace Locate ... 5-36
5-6-24.
D.Trace Level ... 5-36
Menu Function Area(Flow/Power Flow/TDI Flow/TDI Power Flow) ... 5-37 5-7-1.
Average(Flow)... 5-37
5-7-2.
Baseline Shift(Flow)... 5-37
5-7-3.
Beam Steer(Flow)... 5-37
5-7-4.
Capture Mode(Flow) ... 5-37
5-7-5.
Capture Time(Flow) ... 5-37
5-7-6.
Color Line Correlation... 5-38
5-7-7.
Color Polarity... 5-38
5-7-8.
Frame Rate Accelerator ...5-38
5-7-9.
DDD... 5-38
5-7-10.
Directional(Flow) ... 5-39
5-7-11.
Edge Smooth...5-39
5-7-12.
Flow Edge... 5-39
5-7-13.
Filter(Flow)... 5-39
5-7-14.
Frame CORRE(Flow)... 5-39
5-7-15.
Frame Rate(Flow)... 5-40
出図
5-7-16.
Image Freq(Flow) ... 5-40
2012.3.14
5-7-17.
Image Select(Flow)... 5-40
改 訂
MS 技 アロカ
11
MN1-5206 Rev.20
5-8.
5-7-18.
IP Select(Flow)... 5-40
5-7-19.
PFD Range ... 5-41
5-7-20.
Priority Level ... 5-41
5-7-21.
Rejection(Flow)... 5-41
5-7-22.
Smoothing(Flow) ... 5-41
5-7-23.
Wall Motion Reduction... 5-41
5-7-24.
AWR(Advanced Wall Motion Reduction)... 5-42
5-7-25.
TDI PF PreSTC ... 5-42
Menu Function Area(Other)... 5-43 5-8-1.
Area Lock... 5-43
5-8-2.
B Refresh... 5-43
5-8-3.
B/* Format ... 5-43
5-8-4.
Beam Steer(B/F/D)... 5-43
5-8-5.
Beam Steer(F/D) ... 5-43
5-8-6.
Element Select... 5-44
5-8-7.
Graphic Color... 5-44
5-8-8.
Imaging Information ... 5-44
5-8-9.
Invert Link... 5-44
5-8-10.
Steering Link... 5-44
5-8-11.
Thermal Index ... 5-44
5-8-12.
Print Queue... 5-45
5-8-13.
View Gamma ... 5-45
5-8-14.
Zoom Lock ... 5-45
5-8-15.
ECG Display ... 5-45
5-8-16.
ECG Posi ... 5-45
5-8-17.
ECG Sens ... 5-45
5-8-18.
ECG Invert ... 5-46
5-8-19.
Drift Filter ... 5-46
5-8-20.
R Wave Beep... 5-46
5-8-21.
R-Delay Time... 5-46
5-8-22.
PCG Display... 5-46
5-8-23.
PCG Filter ... 5-46
5-8-24.
PCG Posi ... 5-46
5-8-25.
PCG Sens ... 5-47
5-8-26.
Pulse Display ... 5-47
5-8-27.
Pulse Posi ... 5-47
5-8-28.
Pulse Sens... 5-47
5-8-29.
Resp Display ... 5-47
5-8-30.
Resp Disp Method... 5-47
5-8-31.
Resp Posi ... 5-48
出図
5-8-32.
Resp Sens ... 5-48
2012.3.14
5-8-33.
ECG Sync... 5-48
改 訂
MS 技 アロカ
12
MN1-5206 Rev.20
5-8-34.
B/Sync Mode ... 5-48
5-8-35.
VCR Counter Display... 5-48
5-8-36.
DVD Remain ... 5-48
5-8-37.
Counter ... 5-49
5-8-38.
Acquire Method ... 5-49
5-8-39.
Acquire Mode ... 5-49
5-8-40.
Video Clip Auto Stop ... 5-49
5-8-41.
ECG Cycle ... 5-49
5-8-42.
Time Cycle... 5-50
5-8-43.
Store Media... 5-50
5-8-44.
Loop Mode... 5-50
5-8-45.
Thumbnail Display ... 5-50
5-8-46.
Thumbnail Page ... 5-50
5-8-47.
Print Area... 5-51
5-8-48.
Auto Clipping ... 5-51
5-8-49.
Log Off ... 5-51
5-8-50.
Fixed MI ... 5-51
5-8-51.
HR Stability Display... 5-51
5-8-52.
HR Stability Display(Avg.) ... 5-52
5-8-53.
HR Stability Display(Conti) ... 5-52
5-8-54.
Power Limit Override ... 5-52
6. Preset 6-1.
PRESET Function... 6-1
6-2.
Procedure for Changing the Set Contents... 6-2
6-3.
Common Preset... 6-5
6-4.
6-3-1.
Common Preset1 - 2 ... 6-5
6-3-2.
Probe Select ... 6-9
6-3-3.
Print (Freeze) ... 6-11
6-3-4.
Print(Realtime) ... 6-12
6-3-5.
Print Select... 6-13
6-3-6.
DICOM Store/Send, Address, Printer ... 6-14
6-3-7.
DICOM SR ... 6-18
6-3-8.
IHE/Auto Delete ... 6-19
6-3-9.
Stress Echo... 6-22
Preset Control ... 6-23 6-4-1.
Method of copying preset No. data to media... 6-24
6-4-2.
When copying Preset No. data inside the instrument ... 6-25
6-4-3.
When copying Preset No. data from media ... 6-25
6-4-4.
When rearranging Preset No. data ... 6-26
出図
6-4-5.
When copying data in Annotation Dictionary to media ... 6-26
2012.3.14
6-4-6.
When copying data in Annotation Dictionary from media ... 6-27
改 訂
MS 技 アロカ
13
MN1-5206 Rev.20
6-4-7.
When copying Common Preset data to media ... 6-27
6-4-8.
When copying Common Preset data from media ... 6-28
6-4-9.
When copying Color Map to media ... 6-28
6-4-10.
When copying Color Map from media ... 6-29
6-4-11.
When copying Measurement to media... 6-29
6-4-12.
When copying Measurement from media ... 6-30
6-5.
Preset Set-Up Menu ... 6-31
6-6.
Display1 to 2 DISP-B,M -D,Flow ... 6-32
6-7.
ID Comment... 6-38
6-8.
Graphics ... 6-39
6-9.
Store, Cine ... 6-41
6-10.
Body Mark ... 6-43
6-11.
Menu-User SW Assign, -Group Assign, -Function Assign, Menu Analysis... 6-45
6-12.
Custom SW, -Foot SW, -Keyboard ... 6-49
6-13.
Physio... 6-51
6-14.
Image-B,M1,M2 IP Select(B),(THE),(ExPHD),(Spatial Compound), (Spatial Compound,THE/ExPHD) AIP Level (B),(THE),(ExPHD) IP Select(M) ... 6-54
6-15.
Focus ... 6-62
6-16.
Post Processing ... 6-64
6-17.
Doppler1, 2... 6-65
6-18.
Flow ... 6-69
6-19.
Power Flow ... 6-72
6-20.
eFlow... 6-75
6-21.
Tissue Doppler ... 6-78
6-22.
Tissue Flow ... 6-81
6-23.
Tissue Power Flow... 6-84
6-24.
Stress Echo1, 2... 6-87
6-25.
FAM ... 6-90
6-26.
Contrast Echo1, 2... 6-91
6-27.
Echo Tracking ... 6-96
6-28.
Wave Intensity ... 6-97
6-29.
FMD ... 6-98
6-30.
RT3D... 6-99
6-31.
3D Scan ... 6-105
6-32.
Flow 3D... 6-106
6-33.
KI1, 2 ... 6-108
7. Optional Function
改 訂7-1. 出図 2012.3.14 MS 技 アロカ
7-2.
EFV Mode (Option: SOP-ALPHA10-1 and EU-9102 are necessary)... 7-1 7-1-1.
Switches to be used ... 7-2
7-1-2.
Basic operation procedure... 7-3
Real Time Doppler Auto Trace (Option:SOP-ALPHA10-3 is necessary) ... 7-5
14
MN1-5206 Rev.20
7-3.
7-4.
7-5.
7-6.
7-7.
改 訂 出図 2012.3.14 MS 技 アロカ
7-2-1.
Real time Doppler Auto Trace display screen... 7-6
7-2-2.
Real Time Doppler Auto Trace Touch panel menu... 7-7
7-2-3.
Basic operation procedure ... 7-8
RT3D (Option: SOP-ALPHA10-4 and EU-9102 are necessary) ... 7-10 7-3-1.
Rotation / Movement and Image display of Volume data... 7-11
7-3-2.
Meaning of terms... 7-14
7-3-3.
Real Time 3D display screen... 7-15
7-3-5.
Switches to be used and menu ... 7-17
7-3-6.
Basic operation procedure ... 7-30
7-3-7.
Applied functional operation procedure ... 7-37
7-3-8.
Basic operation method for 3D automatic volume measurement (Option: SOP-ALPHA10-20 is necessary) ... 7-40
7-3-9.
Transferring automatic volume measurement value... 7-46
7-3-10.
Basic operation method for 3D Scan mode ... 7-51
7-3-11.
Cutting a 3D image on any sections ... 7-53
7-3-12.
Displaying the multiple sliced images from volume data simultaneously ... 7-55
Contrast Harmonic Echo (Option: CHM- ALPHA10 is necessary)... 7-57 7-4-1.
Contrast Harmonic Echo Touch panel menu... 7-58
7-4-2.
Basic operation procedure ... 7-64
CHE Analysis (Option: SOP-ALPHA10-14 is necessary)... 7-69 7-5-1.
Outline of Contrast Harmonic Echo analysis ... 7-69
7-5-2.
CHE Analysis Screen ... 7-71
7-5-3.
CHE Analysis Base Menu ... 7-73
7-5-4.
Basic operation procedure ... 7-81
7-5-5.
Time Intensity Curve ... 7-84
7-5-6.
Subtraction... 7-86
7-5-7.
CHE Analysis touch panel menu...7-88
eTRACKING (Option: SOP- ALPHA10-11 is necessary) ... 7-91 7-6-1.
Stiffness parameter(b) data acquisition ... 7-91
7-6-2.
Stiffness parameter(b) analysis... 7-97
7-6-3.
Report ... 7-105
7-6-4.
Reference ... 7-107
FMD (Flow Mediated Dilation) analysis (Option: SOP-ALPHA10-16 is necessary) ... 7-108 7-7-1.
Summary of FMD examination ... 7-108
7-7-2.
Before starting an examination... 7-109
7-7-3.
FMD operation screen ... 7-110
7-7-4.
FMD touch panel menu ...7-111
7-7-5.
FMD measurement procedure ... 7-114
7-7-6.
FMD analysis... 7-117
7-7-7.
FMD analysis result screen... 7-118
7-7-8.
FMD analysis operation procedure... 7-122
7-7-9.
References... 7-133
15
MN1-5206 Rev.20
7-8.
7-9.
7-10.
7-11.
7-12.
改 訂
Wave Intensity (Option: SOP-ALPHA10-25 is necessary)... 7-134 7-8-1.
Summary of Wave Intensity... 7-134
7-8-2.
Before starting an examination ... 7-135
7-8-3.
Wave Intensity data collection ... 7-135
7-8-4.
Wave Intensity procedure ... 7-139
7-8-5.
Wave Intensity analysis... 7-141
7-8-6.
Report... 7-153
7-8-7.
Reference... 7-155
TDI Analysis (Option: SOP-ALPHA10-13B is necessary)... 7-156 7-9-1.
TDI Analysis Screen ... 7-157
7-9-2.
Operation Panel... 7-159
7-9-3.
Base Menu Function ... 7-160
7-9-4.
Basic operation procedure... 7-177
7-9-5.
Temporal VP(B mode)... 7-179
7-9-6.
Regional VP(B Mode)... 7-181
7-9-7.
Strain Rate(B Mode) ... 7-183
7-9-8.
Strain(B Mode)... 7-184
7-9-9.
Myocardial Thickness(B Mode) ... 7-185
7-9-10.
Velocity Trace(M mode)... 7-186
7-9-11.
Myocardial Thickness(M mode) ... 7-188
7-9-12.
Velocity Profile(M mode) ... 7-190
7-9-13.
TDI Analysis touch panel menu... 7-192
Stress Echo (Option: SOP-ALPHA10-15B is necessary)... 7-195 7-10-1.
Stress Echo Protocol ... 7-196
7-10-2.
Image acquisition method ... 7-201
7-10-3.
Convenient functions for image acquisition ... 7-212
7-10-4.
Comparative display(Shuffle) ... 7-217
7-10-5.
Scoring ... 7-221
7-10-6.
Report... 7-223
7-10-7.
Pause Protocol... 7-227
7-10-8.
Stress Echo touch panel menu... 7-228
Stress Echo Analysis (Option: SOP-ALPHA10-15B is necessary)... 7-232 7-11-1.
Starting analysis using the Review switch ... 7-232
7-11-2.
Object Image Registration... 7-233
7-11-3.
Scoring ... 7-240
7-11-4.
Report... 7-247
7-11-5.
Stress Echo Analysis touch panel menu... 7-251
KI/A-SMA/CQ/WT Analysis (Option: EU-9100 is necessary)... 7-253 7-12-1.
KI ... 7-253
7-12-2.
A- SMA... 7-255
出図
7-12-3.
CQ(Cardiac Quantification) ... 7-261
2012.3.14
7-12-4.
WT(Wall Thickness)... 7-264
MS 技 アロカ
16
MN1-5206 Rev.20
7-13.
7-14.
7-12-5.
KI, A-SMA Touch Panel Menu... 7-266
7-12-6.
CQ Touch Panel Menu ... 7-272
7-12-7.
WT touch panel menu ... 7-275
7-12-8.
KI, A-SMA Operation Procedure ... 7-278
7-12-9.
Notice for the use of A-SMA, CQ, and WT ...7-288
Brachytherapy Guide Lines ... 7-290 7-13-1.
Basic operation procedure ... 7-291
7-13-2.
Brachytherapy Touch Panel Menu ... 7-295
Ultrasonic Gastrovideo Scope (Option: SOP-ALPHA10-18 is necessary)...7-296 7-14-1.
Degree of display viewing angles... 7-297
7-14-2.
Adjusting Display position ... 7-298
改 訂 出図 2012.3.14 MS 技 アロカ
17
MN1-5206 Rev.20 1-1.Operation Panel
1. METHOD OF OPERATING SWITCHES 1-1.
Operation Panel The methods of operating the various switches are described below. Touch panel Rotary encoder 1
Rotary encoder 2
Rotary encoder 3
Rotary encoder 4
NEW PATIENT
Fig. Operation panel
改 訂 出図 2012.3.14 MS 技 アロカ
Rotary encoder 1
It is used for the menu displayed at the upper section of the touch panel.
Rotary encoder 2
It is used for the menu displayed at the middle section of the touch panel.
Rotary encoder 3
It is used for the menu displayed at the lower section of the touch panel.
Rotary encoder 4
When functions of measurement or zoom are set, it is used.
This section consists of 86 pages. 1-1
MN1-5206 Rev.20 1-2.Graphic Display
1-2.
Graphic Display (3) '00/02/28 : Y :M 11:23:36 M1 =0.73 TIS< 0.4 17%
:No ID :
ALOKA
(1)
3.75M 22Hz 1 R17.0 G60 C90
(4)
(2)
1:Abdomen Probe:52101
AIP
BbH
SCI
(5) Fig. B mode image (1)
Gray scale bar
(2)
Scale mark
(3)
Active mark A[ ] mark indicates an active state (An active image is the image on which operations can be performed when two or more images are displayed.) A[ ] mark indicates a non − active state.
(4)
Focus mark
(5)
AIP BbH SCI
: Indicates that AIP is operating. For details, refer to Section 5-4-31. "AIP(ADAPTIVE IMAGE PROCESSING)". : Indicates that broadband transmission is being performed in ExPHD (T.H.E.). : Indicates that Spatial Compound is operating. For details, refer to Section 5-4-28. "SPATIAL COMPOUND".
[Remark] For details, refer to Section 3-6 "SCREEN DISPLAY" of “Safety Instruction”.
改 訂 出図 2012.3.14 MS 技 アロカ
1-2
MN1-5206 Rev.20 1-3.Operating the Various Switches
1-3.
Operating the Various Switches Operating instructions of each switch are explained below.
1-3-1. ACOUSTIC POWER The ACOUSTIC POWER function enables the acoustic power to be manually adjusted from the automatic setting.
<Operation method> (1)
Press the ACOUSTIC POWER switch.
(2)
Using rotary encoder 4, adjust the output to the optimum value. →
(3)
Turning rotary encoder 4 clockwise causes the output to increase, and vice versa. At the same time, the output value on the image changes.
Press the ACOUSTIC POWER switch to finish this function.
1-3-2. B GAIN The B GAIN knob adjusts the sensitivity of the B mode image. The sensitivity (gain) can be varied between 10 and 90 dB, and is displayed on the screen. [Remark] You can set the gain at startup by means of a preset, however depending upon the conditions of the probe, and so on, the gain may not always match the dB value displayed on the screen.
<Operation method> (1)
Turn the knob counterclockwise or clockwise. →
(2)
Turning the knob clockwise increases the image sensitivity, and vice versa.
While observing an image, adjust the sensitivity to a suitable level.
1-3-3. CANCEL Used to interrupt a selected operation or return to the previous step during a measurement.
改 訂 出図 2012.3.14 MS 技 アロカ
1-3