Instruction Manual
324 Pages
Preview
Page 1
ULTRASOUND DIAGNOSTIC INSTRUMENT
SSD-α10 Instruction Manual Measurement (volume 1/2)
Instruction manuals consist of this manual, Safety Instruction and How to Use. Before using this instrument, please read Safety Instruction.
改 訂 出図 2012.3.14 MS 技 アロカ
MN1-5207 rev.19
ProSound logo is registered mark of Hitachi Aloka Medical, Ltd. in Japan and other countries. Copyright©Hitachi Aloka Medical, Ltd. All rights reserved. Microsoft and Windows Media player is registered trademark of Microsoft Corporations in United States and/or other countries. All brand name and product name are trademarks or registered trademarks of their respective companies. In this manual, ® and ™ are omitted. VS-FlexGrid Pro copyright©1999-2000 Videosoft Corporation. Portions of this software are based in part on the work of the Independent JPEG Group.
改 訂 出図 2012.3.14 MS 技 アロカ
MN1-5207 rev.19
INTRODUCTION
Introduction This is an instruction manual for model SSD-α10, an ultrasound diagnostic instrument. Read the manual carefully before using the instrument. Take special note of the items in Chapter 1, "Safety Precautions." Keep this manual securely for future reference.
Symbols Used in this Document The following items are important in preventing harm or injury to equipment operator or patient. There are four levels of harm/damage that can be caused by ignoring instructions/displays or using the equipment incorrectly: "Danger," "Warning," "Caution," and "Note." These types are indicated by the following symbols. Indicates an imminently hazardous situation that will result in the death of or serious injury to the equipment operator. Indicates a hazardous situation that could result in death or serious injury. Indicates a hazardous situation that may result in slight or moderate injury, or property damage. Indicates a request concerning an item that must be observed in order to prevent damage or deterioration to the equipment and also to ensure effective use.
Cautions use the following graphics.
This mark indicates and alert, additional information.
This mark indicates that the action is not allowed.
This mark indicates that the action is required.
Conventions used in this manual. NOTE: Notes containing additional information. IMPORTANT: Information that is considered especially important.
改 訂
Input, output and screen-messages are presented in the following font: message. Menus and switches are written as Menu. Submenus are indicated by the use of angle brackets: Menu > sub-menu > sub-menu.
出図 2012.3.14 MS 技 MN1-5207 rev.19 アロカ
3
INTRODUCTION
About the model “SSD-α10” The SSD-α10 is intended to be used by doctors and other qualified personnel in fracture diagnostics and hemodynamic diagnostics. However, this equipment is not designed to be used in ophthalmic ultrasound diagnosis, as its sound intensity is not compliant with ophthalmic restrictions established by the FDA. Only physicians and other qualified personnel should operate this equipment for diagnostic purposes. Read section 1-1 of the Safety Instruction. 1)
PRECAUTIONS Concerning the Use/Management of the SSD-α10 •
Do not disassemble, repair or remodel this equipment or optional features without our consent. NOTE: Disassemble means to remove the parts or options from the equipment
using tools. NOTE: Remodel means to install or connect unauthorized parts or equipment
including the power cord. •
Assembly of the equipment or optional accessories shall be performed by a third party certified by us. Please contact one of our offices listed on back cover. NOTE: Assemble means to install or connect parts or optional accessories in/to the
device using tools.
2)
•
Transporting this equipment (via automobile/ship) shall be performed by a third party certified by the manufacturer. Please contact one of our offices listed on back cover.
•
Please conduct routine cleaning and inspection of the equipment. Refer to Chapter 5 of the Safety Instruction for details.
•
Ensure that the output level of the scan conforms to the required duration of diagnosis.
•
If any malfunction or abnormality is discovered during operation of the equipment, remove the probe from the patient immediately and discontinue use. If any abnormality is observed in the patient, provide proper care as quickly as possible. Refer to Chapter 4 of the Safety Instructions for more information on dealing with the equipment appropriately. If the malfunction is not listed in Chapter 4 of the Safety Instruction manual, contact one of our offices listed on back cover.
PRECAUTIONS for the SSD-α10 Installation This equipment is a electrical medical device that intended for use in hospitals and research facilities. The device should be installed in accordance with the following guidelines.
改 訂
•
Install in accordance with Chapter 3 of the Safety Instructions.
•
Install in an environment that conforms to the operating environments indicated in section 2-2-2 of the Safety Instruction manual.
•
Install in an environment that ensures electromagnetic compatibility, in accordance with section 1-2-6 of the Safety Instruction manual, "Precautions Concerning the Maintenance of Electromagnetic Compatibility," and Item 1-3, "Guidelines for Electromagnetic compatibility." NOTE: The electromagnetic compatibility (EMC) is the ability of device to function
satisfactorily in its electromagnetic environment without introducing intolerable electromagnetic disturbance to anything in that environment.
出図 2012.3.14 MS 技 4 アロカ
MN1-5207 rev.19
INTRODUCTION
Classification of SSD-α10 • Protection against electric shock (ME equipment): class I • ME equipment • Protection Against Electric Shock (Applied Parts): Type BF Applied Parts –
Probe/scanner applied parts and parts treated as applied parts: Refer to the following diagram (Probe/Scanner Pattern Diagram) and table.
Figure: Probe/Scanner Pattern Diagram Above illustrates a surface/intraoperative probe. Below shows a coelomic probe. B
C
A
connector
D
connector
C
A
Applicable part of body
Applied part
parts treated as applied parts
B - C length
surface of body
Ultrasonic irradiation area (D)
A to B
100 cm
Intraoperative
Ultrasonic irradiation area (D)
A to B
20 cm
Endocavity
A to C
A to C
N/A
–
Physiological signal applied part: ECG electrodes Part treated as applied part: 2m from the ECG electrode of the ECG patient cable (consult following diagram) 2 meters ECG electrodes
connector ECG patient lead
• Protection against electric shock (Defibrillation-proof applied parts): Not suitable • Protection against harmful ingress of water or particulate matter –
equipment: IPX0 (Ordinary equipment)
–
Probe applied part: IPX7 (Watertight equipment)
• Suitability for use in an oxygen rich environment: Not suitable • Method(s) of sterilization: Not suitable for sterilization/disinfection with medicinal solution, gas or radiation. • Mode of operation: Continuous operation
改 訂 出図 2012.3.14 MS 技 MN1-5207 rev.19 アロカ
5
MN1-5207 Rev.19
CONTENTS This book consists of two separate volumes. These two volumes have the same table of contents and the index respectively.
1. Measurement Functions 1-1. 1-2. 1-3. 1-4. 1-5.
1-6. 1-7.
1-8. 1-9.
1-10.
1-11. 1-12.
Preface... 1-1 Flow of Measurement Operations... 1-2 Switches Used for Measurement... 1-3 1-3-1. The function of the panel switch used by the measurement operation ... 1-3 Basic Operation Procedure for Measurement ... 1-4 1-4-1. Method of starting measurement... 1-4 Explanation of the Measurement Menus ... 1-9 1-5-1. When the measurement Study is changed... 1-9 1-5-2. When the measurement application is changed ... 1-10 Executing the Application Measurement Using the Transfer Function... 1-11 Measurement Mark and Measurement Method ... 1-13 1-7-1. Basic types of marks ... 1-13 1-7-2. Auxiliary line type marks... 1-13 1-7-3. Display mark ... 1-14 1-7-4. The basic operating method for each mark type ... 1-15 Basic Measurement Functional Outline... 1-22 1-8-1. Basic Measurement Functional List... 1-22 Measurement operation procedure... 1-25 1-9-1. B mode ... 1-25 1-9-2. M mode ... 1-44 1-9-3. D mode ... 1-49 1-9-4. B/D mode ... 1-63 1-9-5. B (Flow) mode ... 1-68 1-9-6. VCR Calibration... 1-72 Preset Function... 1-77 1-10-1. Composition of the preset function ... 1-77 1-10-2. Basic operations and function of each button ... 1-78 1-10-3. Initializing Preset ... 1-79 1-10-4. Create Measurement Tools ... 1-80 1-10-5. Application Measurement Tools ... 1-85 1-10-6. Study Assignment ... 1-94 1-10-7. SW Assignment... 1-98 Calculation Formula & Reference ... 1-102 1-11-1. Calculation ... 1-102 Abbreviation... 1-106
改 2.訂Cardiac Measurement 2-1.
出 図 2-2. 2012.3.14 MS 技 アロカ
Preface... 2-1 Cardiac Measurement Functional Outline ... 2-2 2-2-1. Cardiac Measurement Functional List ... 2-2 6
MN1-5207 Rev.19
2-3.
2-4.
2-5.
2-6.
2-7.
2-2-2. Items of Special Note... 2-7 2-2-3. Measurement Views for Measuring Cardiac Functions ... 2-8 Measurement operation procedure ... 2-10 2-3-1. B mode... 2-11 2-3-2. M mode... 2-32 2-3-3. D mode... 2-41 2-3-4. Flow mode ... 2-73 Report function ... 2-92 2-4-1. Basic Operation of a Report ... 2-92 2-4-2. Report Block...2-93 2-4-3. Description of Various Data Displayed in a Report ... 2-97 2-4-4. Function that Attaches an Ultrasound Image to a Report... 2-99 2-4-5. Printing Function ... 2-102 2-4-6. Output to a Personal Computer... 2-104 2-4-7. Output to a CSV file ... 2-105 Preset function ... 2-106 2-5-1. Preset Settings(Cardiac Measurement)... 2-106 2-5-2. PRESET list ... 2-107 Calculation Formula & Reference...2-115 2-6-1. Calculation... 2-115 2-6-2. Clinical References... 2-125 Abbreviation ... 2-133
3. Vascular Measurement 3-1. 3-2.
3-3.
3-4.
3-5.
改 訂 3-6.
出図 2012.3.14 MS 技 アロカ
Preface ... 3-1 Vascular Measurement Functional Outline ... 3-2 3-2-1. Vascular Measurement Functional List ... 3-2 3-2-2. Items of Special Note... 3-3 Measurement operation procedure ... 3-4 3-3-1. B mode... 3-6 3-3-2. D mode... 3-12 Report function ... 3-24 3-4-1. Basic Operation of a Report ... 3-24 3-4-2. Report Block...3-25 3-4-3. Description of Various Data Displayed in a Report ... 3-29 3-4-4. W. Trace Function ... 3-36 3-4-5. Function that Attaches an Ultrasound Image to a Report... 3-38 3-4-6. Printing Function ... 3-41 3-4-7. Output to a Personal Computer... 3-43 3-4-8. Output to a CSV file ... 3-44 Preset function ... 3-45 3-5-1. Preset Settings... 3-45 3-5-2. PRESET list ... 3-46 Calculation Formula & Reference...3-52 3-6-1. Calculation... 3-52 3-6-2. Anatomy Check List ... 3-53 7
MN1-5207 Rev.19
3-7.
3-6-3. Clinical References ... 3-55 Abbreviation... 3-57
4. Abdominal measurement 4-1. 4-2.
4-3.
4-4.
4-5.
4-6.
4-7.
Preface... 4-1 Abdominal Measurement Functional Outline... 4-2 4-2-1. Abdominal Measurement Functional List... 4-2 4-2-2. Items of Particular Note ... 4-4 Measurement operation procedure... 4-5 4-3-1. B mode ... 4-6 4-3-2. D mode ... 4-14 Report function ... 4-22 4-4-1. Basic Operation of a Report... 4-22 4-4-2. Report Block ... 4-23 4-4-3. Description of Various Data Displayed in a Report... 4-28 4-4-4. Graph... 4-31 4-4-5. W. Trace... 4-32 4-4-6. Function that Attaches an Ultrasound Image to a Report ... 4-34 4-4-7. Printing Function... 4-37 4-4-8. Output to a Personal Computer ... 4-39 4-4-9. Output to a CSV file... 4-40 Preset function... 4-41 4-5-1. Preset Settings ... 4-41 4-5-2. PRESET list ... 4-42 Calculation Formula & Reference ... 4-48 4-6-1. Calculation ... 4-48 4-6-2. Clinical References ... 4-48 Abbreviation... 4-49
5. Obstetrical Measurement 5-1. 5-2.
5-3.
改 訂5-4. 出図
Preface... 5-1 Obstetrical Measurement Functional Outline ... 5-2 5-2-1. Obstetrical Measurement Functional List ... 5-2 5-2-2. List of obstetrical measurement name built into system... 5-5 5-2-3. Items of Special Note ... 5-9 Measurement operation procedure... 5-10 5-3-1. B mode ... 5-11 5-3-2. M mode ... 5-25 5-3-3. D mode ... 5-27 5-3-4. Multiple pregnancies... 5-32 5-3-5. Interval Growth Rate... 5-34 Report function ... 5-35 5-4-1. Basic Operation of a Report... 5-35 5-4-2. Report Block ... 5-36 5-4-3. Description of Various Data Displayed in a Report... 5-40
2012.3.14 MS 技 アロカ
8
MN1-5207 Rev.19
5-5.
5-6.
5-7.
5-4-4. Function that Attaches an Ultrasound Image to a Report... 5-54 5-4-5. Printing Function ... 5-57 5-4-6. Output to a Personal Computer... 5-59 5-4-7. Output to a CSV file ... 5-60 Preset function ... 5-61 5-5-1. Preset Settings... 5-61 5-5-2. PRESET list ... 5-63 Calculation Formula & Reference & Table...5-72 5-6-1. Calculation... 5-72 5-6-2. Anatomy Check List ... 5-78 5-6-3. BPP Scoring... 5-80 5-6-4. References... 5-82 5-6-5. Data in the fetal growth table inside the system ... 5-96 Abbreviation ... 5-137
6. Gynecological Measurement 6-1. 6-2.
6-3.
6-4.
6-5.
6-6.
6-7.
Preface ... 6-1 Gynecological Measurement Functional Outline ... 6-2 6-2-1. Gynecological Measurement Functional List... 6-2 6-2-2. Items of Special Note... 6-3 Measurement operation procedure ... 6-4 6-3-1. B mode... 6-6 6-3-2. D mode... 6-13 Report function ... 6-15 6-4-1. Basic Operation of a Report ... 6-15 6-4-2. Report Block...6-16 6-4-3. Description of Various Data Displayed in a Report ... 6-20 6-4-4. Function that Attaches an Ultrasound Image to a Report... 6-27 6-4-5. Printing Function ... 6-30 6-4-6. Output to a Personal Computer... 6-32 6-4-7. Output to a CSV file ... 6-33 Preset function ... 6-34 6-5-1. Preset Settings... 6-34 6-5-2. PRESET list ... 6-35 Calculation Formula & Reference...6-40 6-6-1. Calculation... 6-40 6-6-2. Anatomy Check List ... 6-40 6-6-3. Clinical References... 6-41 Abbreviation ... 6-43
7. Urological Measurement
改 訂 出図
7-1. 7-2.
Preface ... 7-1 Urological Measurement Functional Outline ... 7-2 7-2-1. Urological Measurement Functional List ... 7-2 7-2-2. Items of Particular Note... 7-3
2012.3.14 MS 技 アロカ
9
MN1-5207 Rev.19
7-3.
7-4.
7-5.
7-6.
7-7.
Measurement operation procedure... 7-4 7-3-1. B mode ... 7-6 7-3-2. D mode ... 7-18 Report function ... 7-20 7-4-1. Basic Operation of a Report... 7-20 7-4-2. Report Block ... 7-21 7-4-3. Description of Various Data Displayed in a Report... 7-25 7-4-4. Graph function ... 7-30 7-4-5. Function that Attaches an Ultrasound Image to a Report ... 7-31 7-4-6. Printing Function... 7-34 7-4-7. Output to a Personal Computer ... 7-36 7-4-8. Output to a CSV file... 7-37 Preset function... 7-38 7-5-1. Preset Settings ... 7-38 7-5-2. PRESET list ... 7-39 Calculation Formula & Reference ... 7-43 7-6-1. Calculation ... 7-43 7-6-2. References ... 7-44 Abbreviation... 7-46
8. eTRACKING Measurement 8-1. 8-2.
8-3.
8-4.
Preface... 8-1 eTRACKING Measurement Functional Outline ... 8-1 8-2-1. Outline of Function ... 8-1 8-2-2. Measurement operation procedure... 8-1 Report function ... 8-2 8-3-1. Basic Operation of a Report... 8-2 8-3-2. Report Block ... 8-3 8-3-3. Description of Various Data Displayed in a Report... 8-5 8-3-4. Function that Attaches an Ultrasound Image to a Report ... 8-6 8-3-5. Printing Function... 8-9 8-3-6. Output to a Personal Computer ... 8-11 8-3-7. Output to a CSV file... 8-12 Preset function... 8-13 8-4-1. Preset Settings ... 8-13 8-4-2. PRESET list ... 8-14
改 訂 出図 2012.3.14 MS 技 アロカ
10
MN1-5207 Rev.19 1-1.Preface
1. MEASUREMENT FUNCTIONS 1-1.
Preface The measurement functions of the SSD-ALPHA10 are divided by application (clinical area) into Abdomen, OB, GYN, Cardio, Vascular, URO, eTRACKING, and Small Parts. The measurement values found in ultrasonic examination are recorded and managed for each patient. They are effective for evaluation of observation over time. In addition to selecting a measurement from the menu, you can also transfer the results of a basic measurement to an applied measurement and compute the results of applied measurement. Also, the available measurements depend on the application, so there are ultrasonic studies that bring together the measurement items for each application. (Examples of studies for each application)
Application
Ultrasonic Study
Remark
Cardio
Cardiac Func.
Coronary
TDI
Vascular
Carotid Artery
Lower Ext Artery Lower Ext Vein
Upper Ext Vein
TCD
Asynchrony
Refer to Section 2
Upper Ext Artery
Refer to Section 3
Abdom
Basic
Refer to Section 4
OB
Basic
Early
Twin Basic
GYN
GYN
Follicles
Bladder
Refer to Section 6
URO
Prostate&SV
Bladder&Testis
Kidney
Refer to Section 7
eTRACKING
eTRACKING
FMD
Wave Intensity
Refer to Section 8
Twin Early
Refer to Section 5
[Remark] These application measurement functions are based on documents that indicate the clinical effectiveness of the measurements concerned. Consequently, they use measurement conditions and special computation formulas. The abovementioned documents are listed at the end of the section for each application measurement function, so refer to the relevant documents. [Remark] The application measurement functions of the SSD-ALPHA10 include a function called EXAM. HISTORY for performing a progress observation (development evaluation, treatment evaluation, and so on.). This function records and manages measurement values and operation index values obtained during various ultrasonic examinations, and displays the results as numerical values and/or graphs. For this reason, in order to use this function correctly, be sure to enter the patient data using the ID screen before performing an ultrasonic examination. The various measurement functions of the SSD-ALPHA10 ultrasonic diagnostic instrument are described in the following sections.
改 訂 出図 2012.3.14 MS 技 アロカ
Section1 Section2 Section3 Section4
Basic Measurement Cardiac Measurement Vascular Measurement Abdominal Measurement
Section5 Section6 Section7 Section8
This section describes the basic measurement method. This section consists of 108 pages. 1-1
Obstetric Measurement Gynecological Measurement Urological Measurement eTRACKING Measurement
MN1-5207 Rev.19 1-2.Flow of Measurement Operations
1-2.
Flow of Measurement Operations Select the application.
(1)
Input the patient information.
(2)
Select the ultrasonic study.
(3)
Execute the measurement.
(4)
→
In order to obtain correct ultrasonic measurement results, the computation equations and patient management database are decided according to the application selected.
Input the patient information. →
Input the patient information with the New Patient (ID) screen in order to correctly operate the patient management functions effectively for evaluating observations over time. For details, refer to Section 1 How to Use.
Select the ultrasonic study. →
The measurement functions that can be executed are determined by the selected ultrasonic examination study. Refer to Section 1-5-1. "When the measurement Study is changed"
Execute the measurement. →
(5)
Report
Select the preset application (clinical area).
The measurement values are updated in real time until the measurement operation is finalized. Up to 10 caliper marks can be displayed in the measurement results display area. Section 1-4-1-1. "Starting from the + switch". Section 1-4-1-3. "Starting from the MEASUREMENT switch". Section 1-4-1-4. "Starting from a Hot key".
Report →
The measurement values displayed on the report screen can be set with the preset function to either the most recent values or an average. Also, up to six of these measurement values can be preset for each measurement item. (For an explanation of each item, see the pages for each application.)
[Remark] The SSD-ALPHA10 has the following three functions for recording the results of examinations.
改 訂
a.
When operation (4) above is performed and a report is displayed, it is recorded by the recording device (thermal printer) connected to the SSD-ALPHA10.
b.
The report can be printed on paper using to Printer in the Output function of the Report screen.
c.
The measurement results can be transmitted to the application software of an external personal computer using to PC in the Output function of the Report screen.
出図 2012.3.14 MS 技 アロカ
1-2
MN1-5207 Rev.19 1-3.Switches Used for Measurement
1-3.
Switches Used for Measurement
1-3-1. The function of the panel switch used by the measurement operation (1) - (2) are switches for starting measurement, and (3) - (9) are switches related to measurement operations. (7)
-
+
(8)
(9)
NEW PATIENT
(3)
(5) (9)
(1) (2)
(1)
(4)
(9)
(6)
: The measurement menu corresponding to the display mode is displayed. MEASUREMENT switch
(2)
+ switch
: Directly starts the basic measurement corresponding to each display mode registered in a preset. Also, displays a new mark if pressed during measurement.
(3) (4)
Trackball ENTER switch
(5)
CANCEL switch
: Used to move a measurement mark. : Press this switch to separate marks or to change over the mark to be moved by the trackball. : Returns the system to the immediately preceding (or initial) mark display condition.
(6)
改 訂
SEND switch
: Press to finalize the measurement operation presently taking place.
(7)
Touch panel
: Measurement menu is displayed.
(8) (9)
Rotary encoder 4 USER 1, 2, 3 switch
: The trace line is deleted little by little. : Directly starts the function or measurement registered using a preset.
出図 2012.3.14 MS 技 アロカ
1-3
MN1-5207 Rev.19 1-4.Basic Operation Procedure for Measurement
1-4.
Basic Operation Procedure for Measurement
1-4-1. Method of starting measurement You can use one of the following three methods to start a measurement. (1)
Display the measurement menu, and start the selected measurement.
(2)
Press the + switch to perform (basic) measurement.
[Remark] You can transfer the results obtained here to each parameter of the application measurement. Refer to Section 1-6. "Executing the Application Measurement Using the Transfer Function". (3)
Start measurement directly. →
1-4-1-1. (1)
Press the hot key (alphabet key) on the full keyboard to start the measurement.
Starting from the + switch Press the + switch. →
The preset measurement is started.
Control menu
Measurements set in Preset
[Remark] Up to eight frequently used basic measurement items can be preset to the + switch for each display mode.
改 訂
Control menu: Clear VCR Calib Trace Manual Locate Mark Display Report
出図
[Remark] Control menu is different with each an application.
: Measurement function is finished. : Calibration for VCR playback measurement can be performed. : It is switched to a manual trace directly from an auto trace of a doppler waveform. : Change the position of (move) the measurement results display area. : On/Off of displaying caliper mark is made. : Report of application measurement is displayed.
2012.3.14 MS 技 アロカ
1-4
MN1-5207 Rev.19 1-4.Basic Operation Procedure for Measurement
(2)
Select the basic measurement item which is carried out on the touch panel, and perform the measurement.
(3)
After measurement, to measure again, press the + switch. →
The last measurement is started.
[Remark] Pressing the CANCEL switch returns the system to the status before separation without finalizing the measured caliper mark. If the CANCEL switch is pressed before the caliper mark is separated, the mark is erased.
1-4-1-2.
Starting measurement with a compound mode image
When pressing the + switch on hybrid modes (B/M and B/D modes), the measurement corresponding to each ultrasound image (B, M, and D) is performed. For example, for B/M mode as in the figure below, when the + switch is pressed, the caliper mark is displayed on the M image with a priority order (D → M → B). The caliper mark is moved to the B mode image side with the trackball. The moment at which the caliper mark is moved into the B mode image it becomes Dist. The measurement items and control menu display at the bottom of the screen also change to the B mode image. Moving the + mark to the B (M) image with the trackball
M.VEL
DIST
改 訂 出図 2012.3.14 MS 技 アロカ
1-5
MN1-5207 Rev.19 1-4.Basic Operation Procedure for Measurement
1-4-1-3. (1)
Starting from the MEASUREMENT switch Press the →
switch.
The preset measurement menu and control menu are displayed.
Measurement menu
[Remark] Pressing the (2)
switch on the operation panel erases the measurement menu.
Select the measurement item displayed on the touch panel. →
The selected measurement starts.
Example: Uterus measurement
The proceeding order of the measurements can be selected voluntarily.
改 訂 出図 2012.3.14 MS 技 アロカ
1-6
MN1-5207 Rev.19 1-4.Basic Operation Procedure for Measurement
1-4-1-4.
Starting from a Hot key
Pressing a letter key starts the measurement item or control function allocated to that key directly without passing through the measurement menu. Here is an explanation using an example in which the GYN measurement Uterus (Volume) is assigned to the Q key. (1)
Press the Q key. →
The system starts Uterus measurement and measures each position.
[Remark] Measurement items can be assigned to letter keys on the keyboard with the preset function.
改 訂 出図 2012.3.14 MS 技 アロカ
1-7
MN1-5207 Rev.19 1-4.Basic Operation Procedure for Measurement
1-4-1-5.
Erasing individual marks
By performing the following operation, measurement values corresponding to the measurement marks which were fixed can be erased individually. (1)
Select the ch with the same number as the measurement results for erasing from the touch panel.
ch which corresponds to the number of measured times is displayed.
(2)
For example, press the ch1 switch once. →
(3)
The caliper mark becomes ready for re-measurement.
Once again, press the ch switch which was pressed in (2). →
The caliper mark with the number of 1 and the measurement results are erased.
[Remark] When the Clear switch on the control menu is pressed, the caliper mark and the measurement results are erased. To erase only the executed mark press the CANCEL switch on the operation panel twice. If the measurement has not yet been carried out (when there is one mark), pressing the CANCEL switch ends the measurement function.
1-4-1-6.
Ending a measurement function
The measurement marks and results on the screen are erased in the following cases. • When you press (select) the Clear switch. • When you press the MEASUREMENT switch. • When you cancel a freeze status [Remark] The above applies only when the Caliper auto off function in the Preset menu is ON.
改 訂 出図 2012.3.14 MS 技 アロカ
1-8
MN1-5207 Rev.19 1-5.Explanation of the Measurement Menus
1-5.
Explanation of the Measurement Menus (1)
Press the →
switch.
The measurement menu is displayed on the touch panel. The preset of the measurement is displayed.
Basic : The menu is displayed.
The measurement having multiple measurement methods is displayed on the lower row.
The application which is currently on
The basic measurement results are copied to the application measurement results.
1-5-1. When the measurement Study is changed (1)
Select the Study & Application on the touch panel. →
The measurement Study that is changeable is displayed
Measurement Study
(2)
Select the Study to change. →
The Study is changed.
改 訂 出図 2012.3.14 MS 技 アロカ
1-9
MN1-5207 Rev.19 1-5.Explanation of the Measurement Menus
1-5-2. When the measurement application is changed You can change the measurement menu from one to another clinical field during measurement, and perform an application measurement. (1)
Select the Study & Application on the touch panel. →
(2)
The Application which can be changed is displayed on the Change Application.
Select the Application and Study to change. →
The Application is changed. After selecting the Application, select the Study.
改 訂 出図 2012.3.14 MS 技 アロカ
1-10