INSTRUCTION MANUAL
5000 SERIES
From Serial Number #18019
PLEASE CAREFULLY READ INSTRUCTIONS BEFORE USING THE BED
PRE-OPERATION CHECK WITH BED IN POSITION CHECK THE FOLLOWING: • Electric beds should ONLY be used where earth leakage is fitted to the power source. • Check power cord is plugged into 240V supply and is active. • Check power cord is clear of castors and any moving part of bed. • Ensure power cord has enough slack so cord is not strained when the bed is raised and lowered. • Position bed at least 50mm from the floor in the lowest position. Lock castors. This position is critical so as not to damage the wall when raising the bed.
• Before using bed perform all functions to ensure they are working properly. If not please call your service technician or call Alrick Technician Team on +61 2 6898 1700. ** When fitting safety sides to this bed please beware of the possible crushing points. The clamps can create crushing points on the rails. **
MODELS AVAILABLE P5500 SERIES BED Electric Hi Low Electric Backrest Electric Forward & Reverse Trendelenburg 100mm Braking Castors
P5500W SERIES BED Electric Hi Low Electric Backrest Electric Forward & Reverse Trendelenburg 100mm Braking Castors
OPTIONS AVAILABLE C.EXT5000 Clip on Extension C.EXT5000W Clip on Extension ABB Battery Backup CPR Emergency Release
SPECIFICATIONS
ELECTRICAL DATA
HEIGHT RANGE:
295-740mm
U In:
BACKREST ANGLE:
60 Degrees
I In:
Max. 1.5A
KNEEBREAK ANGLE:
31 Degrees
U Out:
24V ... , max. 70W
TRENDELENBURG ANGLE:
10.5 Degrees
Op.:
10%, Max. 2min./18min.
SWL:
200KG (165KG Patient Weight)
OVERALL DIMENSION:
2165 x 995mm
WEIGHT OF BED:
P5500: 2200 x 960mm P5500W: 2200 x 1070mm
MATTRESS SIZE:
1960 x 860mm
TGA NUMBER:
230V ~, 50Hz
PROTECTION CLASS: 1Px4 INSULATION CLASS:
II Type B
NOISE:
48DBA
ARTG 192062
** LOAD CAPACITY HAS BEEN TESTED IN ACCORDANCE WITH AS/NZS 3200.2.38:2007 **
IMPORTANT • Check electric cords regularly and notify an electrician or Alrick Manufacturing on +61 2 6898 1700 immediately. If any damage occurs cease to use and remove mains power until checked and repaired. • No wiring is to be modified without authorisation from the manufacturer. • Ensure the small extension cable is correctly connected to the bed socket and to the 240 volt supply when the bed is in a stationary position. A small light on the control box will illuminate showing a power supply. • Low Low model beds should be left in the lowest position when a nurse is not in attendance. • Nurses, cleaners or anyone using Alrick Electric Beds must make sure that the main power cord is clear of any moving parts on the bed before turning the power on. • Electric Beds should ONLY be used where earth leakage is fitted to the power source. • When patient condition could lead to patient entrapment the mattress support platform should be left in the flat position when unattended.
For indoor use (House)
Type B equipment, as per EN60601-1
Product with a thermofuse
Equipment KI.2 (Double square)
Safety isolating transformer
Demko approval
RW-Tüv approval
OPERATOR CONTROLS: Electric Beds are controlled by a handset, which is plugged into the connection port. (Normally located below and at the edge of the mattress base towards the head end).
Foot Hi/low Frame
Head Hi/low Frame
Kneebreak Mattress base section
Backrest Mattress base section
Transport Frames
UPPER POSTURE, ELECTRIC BACKREST: The top pair of push buttons control the electric upper posture as indicated on the handset. Beware of potential pinch points when lowering backrest.
HI-LOW OPERATION: The third pair of buttons on the handset control the smooth operation of raising & lowering the bed. When raising and lowering the bed, ensure that no person of part of his or her body is under the bed. Failure to do so could cause serious physical injury.
ELECTRIC KNEEBREAK: The second pair of push buttons on the handset control the operation of the kneebreak. The calf section of the kneebreak is fitted with a prop, lift the calf section and fold down the prop engaging it in the channel provided at the foot end of the bed. Beware of potential pinch points when lowering bed.
TRENDELENBURG TILT: Tilt of mattress base is acheived by raising or lowering the head or footend of the bed using the forth pair of buttons on the hand control. Please ensure castors are all locked.
(Bed ready for Transport or storage.)
STEP 1:
Remove the 2 centre detent pins that hold the mattress base sections to the transport frames, on each side.
STEP 2: Remove the head & foot mattress base sections from between the hi/low frames.
STEP 2A: Position Hi/low frames in area for assembly, ensuring there is sufficient open area for when the bed is fully assembled.
STEP 3: Engage brakes on all castors Miminise risk of bed sliding or falling
= WARNING SYMBOL
STEP 4:
Remove Transport bars. To avoid foot & head end from unexpectedly sliding apart when removing the transport bars, it is advised to lean the headend against a wall or vertical surface & ensure the footend is securely held with brakes on.
(Bed ready to be assembled.)
STEP 5: Place both the mattress base sections in their corresponding Hi/low foot and End frames. Fit together best 120 degrees.
STEP 6: Place detent pins in all four corners of the mattress base.
STEP 7: Align head and foot mattress base sections for assembly.
STEP 8: Release all brakes, loosen black knobs and join both mattress base sections together.
STEP 9: Tighten black plastic knobs firmly.
STEP 10: Engage brakes on all castors
STEP 11: HB 4 3 2 1
Place plugs from actuators into control box. Port 1: Backrest actuator Port 2: Foot end Hi/low section Port 3: Kneebreak actuator Port 4: Head end hi/low section Port HB: Handset
STEP 11A: Then secure the plugs to the control box with the cable lock. Now your bed is ready for use. Please ensure the power cable is in a safe position and brakes are locked on at all times while the bed is in position.
SELF HELP POLE: (Optional)
Self Help Poles are supplied with a slotted end. If the pole is used on models with the pole support on either side, the pole MUST be fitted to the opposite side to which the patient gets out of bed. (110KG max.)
SAFETY BED RAILS: (Optional) 360HR HORIZONTAL BED RAILS
To raise, lift top bar and a positive click must be heard before leaving bed. To lower, pull pin outward & push top rail down. Please beware of potential crushing points when fitting rails. Always make sure cords are not caught in rails when raising and lowering.
SAFETY BED RAILS: (Pistol Grip Model) (Optional) 380VR VERTICAL BED RAILS
Squeeze trigger in top rail and push down. To raise, lift top rail and make sure a positive click is heard. Please beware of potential crushing points when fitting rails. Always make sure cords are not caught in rails when raising and lowering.
SAFETY BED RAILS: (Pistol Grip Model) (Optional) 400HR HORIZONTAL BED RAILS
To raise, lift top bar and a positive click must be heard before leaving bed. To lower, pull pin outward & push top rail down. Please beware of potential crushing points when fitting rails. Always make sure cords are not caught in rails when raising and lowering.
SUPPORT RAIL: (Optional)
Support Rails can be fitted to side rail of bed. Black knobs must be securely tightened. WARNING: Please check regularly. Please beware of potential crushing points when fitting rails.
SAFETY & WARNINGS • Electric Beds should ONLY be used where earth leakage is fitted to the room or building. • If any cords are cut or damaged the bed MUST not be used until checked, fixed and tagged. • No patient, service personnel or any person must lie underneath an Alrick bed while power is turned on, if not obeyed this may result in serious bodily injury or death if the bed controls are actuated. • Please check no person or object is under the bed before using. • Hand control to only be given to fully competent patients. • Never leave hand control unattended with bed switched on unless patient is fully competent. • Fit optional nurse lockout for safety of incompetent patients. • Only one person is to be on the bed at any time.
MAINTENANCE • Beds should be inspected monthly for abnormal wear. For example, bolts loose or missing, castors, rubbers or brakes damaged. • Rails must be inspected regularly for any wear or movement that may cause injury, do not use if unsure. • If one or more of the bed functions is found not working correctly, notify the manufacturer for advice. EVERY 6 MONTHS: Check there is no visible damage to electrical system or mechanical parts. Check wiring is secure and not interfering with bed movement. If actuator shafts are not clean - wipe chromed shaft of actuators to remove lint, grease & dirt Extended actuators fully. (Do not wash as this removes fine layer of oil). Correct operation of all bed functions including castor locking. Check all accessories including rails are still functioning properly. Check all pivot points on the bed for wear. Replace pivot pins or bolts and removable
bushes (where fitted) if a pivot has a significant play. Check all functions still run smoothly.
CLEANING • Cleaning can be done often - using domestic detergents. • Do not use acids, alkaline or any other chemical solvents • When cleaning actuators disconnect from mains power before wiping down with a moist cloth. • When cleaning, be careful of entrapment, as it could cause physical injury.
PARTS LIST
- FOR P5500 SERIES (including 5500W)
HI/LOW END FRAMES
(Please Note; Quantities are for one end)
Description 1. White Flanged Nylon Bushes 2. White End Cap Nylon Bushes 3. Black plastic End Cap 4. Locking Castors 100mm 5. Timber grain End Boards 6. Hi/Low Electric Actuator 7. Head end Coiled Cable 8. Foot end Coiled Cable 9. 3/8 x 50mm Pins and Clips 10. M8 x 50mm Set screw 11. M8 Spring washer 12. 8G–18 x 20 button HD s/ tapping
1. 2.
Description Aluminum Joining Bars M10 x 20mm Set Screws
QTY 2 2 2 2 2 1 1 1 2 2 2 8
CODE 55/WFNB 55/WECNB 55/BPEC 55/LC100 55/TGEB 270128-00 0277012-200 0277013-150-1250 B/PINL M8/50SS M8/SW 8G–18/ST
Cost/$ +GST $9.25ea $6.00ea $2.50ea $20.00ea $44.00ea $114.00ea $15.00ea $15.00ea $1.25ea $0.40ea $0.10ea $0.10ea
QTY 2 2
CODE 55/AJB M10/20SS
Cost/$ +GST $54.50ea $0.40ea
11. M8 Spring washer 12. 8G–18 x 20 button HD s/ tapping MATTRESS BASE SECTION
2 8
Description 1. Aluminum Joining Bars 2. M10 x 20mm Set Screws 3. M10 Spring washers 4. ¼ Detent Pins 5. Nylon Lanyard 6. Black Hand Wheels 7. Back Rest Electric Actuator 8. Knee Break Actuator 9. Back rest straight cable 10. Knee Break Coiled cable 11. 3/8 x 50mm Pins and Clips 12. Handset 13. Nylon Slide Strip for Knee break 14. Coloured Wafer Screws
QTY 2 2 2 4 4 2 1 1 1 1 4 1 2 8
M8/SW 8G–18/ST
$0.10ea $0.10ea
(Please Note; This the parts for both backrest and Kneebreak sections)
CODE Cost/$ +GST 55/AJB $54.50ea M10/20SS $0.40ea M10/SW $0.10ea 1/4D/Pin $2.10ea LAN $1.30ea 55/BHW $4.00ea 270127-00 $106.00ea 277100A211100040 $109.00ea 0277011-200 $15.00ea 0277012-400 $15.00ea B/PINL $1.25ea HB7460253-60600 $86.00ea P5022 $5.00ea AL1016 $0.30ea
TRANSPORT BRACKET
(Please Note; Quantities are for one transport bracket)
1. 2. 3. 4.
Description ¼ Detent Pins Nylon Lanyard 25 x 25 Black plastic Caps Coloured Wafer Screws
QTY 2 2 2 8
Page
CODE 1/4D/Pin LAN 25/SQBPC AL1016
Cost/$ +GST $2.10ea $1.30ea $1.00ea $0.30ea
RECORD OF PERFORMED SERVICE PROCEDURES AND REGULAR SERVICE MAINTENANCE Purpose and Description
Date
Performed By
I confirm that technical staff have undertaken the repairs as described above and the equipment is now functioning normally. ... / ... / ...
DATE
...
CUSTOMER SIGNATURE
NOTES:
________________________________________________________________ ________________________________________________________________ ________________________________________________________________ ________________________________________________________________ ________________________________________________________________ ________________________________________________________________ ________________________________________________________________ ________________________________________________________________
manufacturers of electric hi low hospital beds and couches
ALRICK SERIES ELECTRIC HI-LOW NURSING HOME & HOSPITAL BED
- WARRANTY CONDITIONS Your Alrick Electric Hi-Low Bed is guaranteed for 10 years on the frame from the date of purchase, 5 years on the electrics, 2 years on castors, 1 year on the battery. This warranty applies only on the parts manufactured or approved by Alrick Manufacturing.
EXCLUSIONS This warranty does not extend to those items that may need replacement due to normal wear and tear, or to damage caused by misuse or accident, for which Alrick Manufacturing or its distributors cannot be held responsible. If your bed should need attention due to failure as defined under the warranty above, please contact the distributor from whom you purchased your bed. Please fill in and retain for your own records: DATE PURCHASED: ______________________________________________________ MODEL:
______________________________________________________
SERIAL NUMBERS:
______________________________________________________
DISTRIBUTOR:
______________________________________________________
PHONE:
______________________________________________________
FAX:
______________________________________________________ ______________________________________________________
• E & O.E • Specifications subject to change without notification
104 Loughnan Street | PO Box 63 Lake Cargelligo NSW 2672 Australia P +61 2 6898 1700 F +612 6898 2144 E [email protected]
www.alrick.com.au
Tear along here
REPLY PAID 4 ALRICK MANUFACTURING 104 LOUGHNAN STREET LAKE CARGELLIGO NSW 2672 AUSTRALIA
No postage stamp required if posted in Australia