Product description
Setting up the device/assembly
Smartbox Invest is a measuring device and a temperature control unit for all investment
After transporting the machine at low temperatu-
materials and liquids. Its integrated cooling system,
res, you must allow it to acclimatise itself to room
excellent storage conditions and precisely
temperature before switching on.
functioning measuring technology means that accurate and reproduceable mixing results can be obtained.
Switching the voltage It is possible to switch from 230 V to 100 V and vice versa. To do so, pull out the fuse holder and position
Smartbox Invest functions
the insert in such a way that the voltage is clearly
_ Powder amounts can be selected in 5 g
visible in the small window. After that, insert the
increments from 50-1000 g
fuse holder along with the fuse back in again.
_ Liquid concentration 0 – 100 % in 1 % increments/1 ml increments _ Ratio of powder/liquid:
Electrical connection
50 – 250 g powder reference quantity
Position the machine and insert the accompanying
Total amount of liquid (based on powder
power cable into the back of the device (check the
reference quantity) 1 – 99 ml in 1 ml increments
right voltage setting under the power cable connec-
_ Dense liquid from 1.0 g/cm3 – 2.0 g/cm3 in 0.01 g
tion, see also section on ‚Switching the voltage‘).
increments
Now connect the plug to the mains supply.
_ Cooling temperature: 16 – 25 °C/60 – 77 °F (no heating function) _ Operation via multi-functional rotary knob or
Connecting the extraction unit
press button (turning = selection, pressing =
When using the extraction unit, first plug the
confirmation) and 2 further keys.
extraction hose into the back of the device. The Smartbox Invest control unit is programmed in such a way that it can be switched on only during
Unpacking and inspection
the measuring of the powder. In order for this function to operate, the power cable of the extrac-
After unpacking, the machine must be immedi-
tion unit has to be connected to Smartbox Invest
ately inspected for any possible damage during
using an appropriate connection lead (114956).
transit. Should any damage have occurred during transit, please report it to the supplier immediately. The machine is delivered ready for use.
Installation location In order to avoid extreme thermal and mechanical
Please keep all packaging!
influences and any influences from moisture, the following details must be observed: _ Do not install the machine in direct sunlight.
Check mains voltage Check the voltage specified on the type plate on
_ Do not install next to vibrating units or similar devices.
the back of the casing corresponds to the mains
_ Do not place near steam cleaners.
voltage available.
If the device is placed under a shelf or similar, the lid opening can be altered.
16
GB
Lid opening options
Starting up
The lid can be used either hinged or detached. By default, the lid is assembled so that it can be
Button functions / operation
opened on hinges. This can be changed as follows:
Multifunctional knob/press button (1)
First lift the lid (9). The lid‘s opening mechanism
(hereafter known as MFK)
is fixed with two Allen screws that are located on
TURN = select
each side in a metal locking device. These locking
PRESS = confirm
devices have an inner and an outer borehole. Return key (2) Position 1: Metal locking device pushed inwards and screws in the outer boreholes = lid can be opened on its hinge. Position 2: Metal locking devices pushed out-
Back to previous menu level Start button (3) Start measuring process/taring the scales/ activating the cleaning function
wards and screws in inner boreholes = lid can be
An arrow ( > ) marks the parameter in the menu
taken off (to do so, pull the lid forwards at an
that can be changed or selected. The arrow can
angle of about 20°).
be moved by turning the MFK.
Setting the language There is a choice of five languages for Smartbox Invest‘s controls (German, English, Italian, French and Spanish). If the preset language is not the language you want, then switch the machine off. Then press the return button (2) and the start button (3) at the same time and switch the machine on when the buttons are still being pressed. Keep pressing the buttons (2 and 3) until the language selection menu appears. Using the MFK (1), select the desired language (by turning) and confirm it (by pressing). After the language has been set, menu level 1 will appear automatically.
Charging Smartbox Invest Before charging, lay out all components so that charging can be done quickly. After about 3-5 hours, all the components will have achieved the required set temperature. Pre-heated components minimise the time needed to reach the required set temperature! The cooling time also always depends on the ambient or room temperature! Example: cool down components that are 25 °C to the 18 °C set temperature in the device = cooling time approx. 3.5 hours. 17
GB
Charging the Smartbox Invest To charge it with powder and liquid, open the lid
- do not use excessive force. Caution: Do not remove or push the valve!
of Smartbox Invest (9) by applying pressure on the lid‘s locking mechanism (8).
Menu program Charging the powder silo
From level 1 to level 2 = selection by turning the
large container (e.g. 781685 Giroinvest Super) into
MFK (1) and confirming selection by pressing the
the silo. The maximum top-up quantity is 4 kg.
MFK (1)
Avoid making dust! Pour the powder in loosely, do
From level 2 to level 1 = with return button (2)
not compact it. We recommend using disposable? bags (holding device in the lid). Keeping the silo open for long periods of time could lead to moisture being absorbed and condensation on the walls of the silo. Avoid fouling the sealing surfaces and the lid seal - clean them with a dry cloth if necessary before shutting the silo (Caution: do not damage the seals).
Before insertion into the Smartbox Invest, take the desiccant bag out of the protective film. Place colour indicator visibly. When the colour of the indicator changes to pink, insert new desiccant bag!!!
Charging the fluid reservoir Pour the liquid and distilled water into the labeled containers provided (left water, right liquid). The maximum top-up quantity is 1000 ml. Use a funnel to pour in the liquids. Remove any excess liquid from the overflow tray. Screw the lid closure back 18
Menu program/control set-up
Open the inner lid of the silo. Carefully empty
GB
Menu level 1
Menu level 2
Menu level 3
19
GB
The status line at the bottom edge of the display
_ Quantity of liquid
(7) shows the function that can be carried out,
The total fluid amount (distilled water + liquid)
alternating with any possible tips (e. g. Fill up P
based on the powder reference quantity or the
(Powder)).
mix ratio (e.g. Giroinvest Super 150 g : 37 ml = liquid amount 37 ml). _ Liquid density The measuring of the liquids is done using the scales. The precise input of the liquid density is vital for measuring accuracy! Contact the manufacturer about the concentration of the liquid or
Warning hints will be displayed prefaced with the
calculate it using the following method.
word „Caution!“ (e.g. silo lid closed). Also, the
We recommend using this method with water a
status line will also display the required set tem-
few times before calculating the required liquid
perature until it is reached. If the silo lid and/or
density - this means that a weight of 100 g has
silo closure is left open too long (risk of moisture
to be determined for 100 ml of water -> divided
penetrating the equipment), a warning tone will
by 100 = density of water 1.00 g/ml.
sound. This can only be stopped by closing the lid
Unpack the accompanying syringe, place it onto
and/or closure.
scales that are as accurate as possible (e.g. an assay-balance) and calculate and note down the empty weight. Then fill the syringe with liquid up
Basic settings before using the machine
to the 100 ml mark (draw up) and weigh again. Before weighing the filled syringe, you must ensure that there is no more air in it - you may have to press any air that may be present out of
Before using Smartbox Invest with an investment
the syringe while it is pointing upwards. The
material, the following basic adjustments must be
syringe must be filled with liquid up to the
made. Switch the machine on at the main switch
100 ml mark without any air in it!
first. The main menu (menu level 1) will appear.
The weight of the filled syringe that has been
The best settings of the device for the investment
ascertained in this way (minus the previously
material in question (processing temperature,
determined empty weight) divided by 100 gives
mixing ratio, etc.) can be found in the processing
you the density of the liquid. Input and confirm
instructions for the investment material or con-
this value by turning the MFK (1).
tact the manufacturer. Use the MFK (1, turn = select, press = confirm) to
_ Cooling temperature
select the item ‚Additional functions“ on menu
Input it and confirm it by using the MFK (1) in
level 1 and sub-item ‚Basic setting‘ on menu level
line with the manufacturer‘s instructions. Maxi-
2. The values set there apply to all subsequent
mum cooling capacity see Technical Data.
dosages with this investment compound. The following parameters have to be set:
_ Beaker control When the beaker control is switched on (select:
_ Powder reference quantity
20
YES), no measuring will take place if there is no
Powder quantity that the mixing ratio of powder
beaker on the weighing tray. Similarly, this func-
and liquid refer to (e.g. Giroinvest Super 150 g :
tion uses the weight of the beaker set up on it
37 ml = powder reference quantity of 150 g).
to see if the beaker placed there is big enough
GB
for the amount of gypsum specified.
weight (in 5 g increments). Confirm the selection
If the beaker control is inactive (select: NO), the
by pressing the MFK. By turning the multi-functio-
measuring starts up without any further report.
nal knob, now select the second of the four record-
As standard, this is based on the permitted filling
able beaker weights, set it and confirm it by
quantity of all AmannGirrbach mixing beakers
pressing the MFK again. Record the two remaining
(Startmix vacuum mixer). For the storage of
beaker weights in the same way. If no other beaker
permitted filling quantities for other beakers,
weights are required, just skip them by pressing
see „Individual setting of beaker weights“.
the MFK. After entering the last beaker weight (description „1000 ml“), the query „Store yes/no“
_ Beaker weights
will appear. If you answer „yes“, the beaker control
The values set here are used to define the per-
will then react as desired to the individually
mitted filling quantities. As a standard, they are
recorded beaker weight.
based on the beaker weights of AmannGirrbach Smartmix mixing beakers. You can overwrite them with four individual beaker weights
Operation
(250 ml, 500 ml, 750 ml and 1000 ml corresponding to the beaker descriptions of small, medium,
Before using the machine, the above-mentioned
large, extra large).
basic settings have to be carried out. In order to operate Smartbox Invest, there are basically two
_ Service settings Only available to the Technical Service of AmmanGirrbach GmbH or authorised service companies.
modes of operation available: _ individual measuring: direct amendment of quantities and concentration without storage function _ stored measuring programs: retrieves up to 50
Individual adjustment of the
individually recorded programs (quantity,
beaker weights
concentration and program names can be
On menu level 1, select the item ‚Additional
stored)
functions‘ and then dial the sub-item ‚Scales‘. Pressing the start button (3) tares the scales.
You must place the mixing beaker under the
Then position the beaker, read the weight off and
outlet of Smartbox Invest before starting the
note it down. With the aid of MFK (1), select the
measuring process! Open and close the outlet
menu item ‚Basic settings‘ on menu level 1 and
carefully!
then select ‚Beaker weights‘ on menu level 2. The display now shows the beaker description „250 ml“ on the left-hand side and the pre-defined beaker
Individual measuring
weight in grammes on the right-hand side
Select item ‚Individual‘ on menu level 1
(AmannGirrbach Smartmix mixing beaker).
Input powder quantity: turn/confirm by pressing the MFK Input concentration: turn/confirm by pressing the MFK Manually open silo closure (4) (otherwise there will be an error message when pressing the start button). Start measuring process with the start
By turning the MFK, adjust the set beaker weight
button (3).
to the previously weighed, individual beaker 21
GB
Programmable measuring programs In order to store your own individual programs, you can use sub-items 2-6 on menu level 1. For every sub-point, 10 programs can be stored (preset descriptions: Alloy A to J). In order to measure the indicated parameters of
Every letter of the name can now be changed
the program, press the start button (3).
by turning the MFK (1). By pressing the MFK, the cursor jumps to the next letter. Once the desired
To change or program, press the MFK (1) for a
name has been input, press the MFK to get to the
long time (protective function against undesired
end of the line. After confirming the last position,
overwriting). The adjustable parameters will be
the display goes back a level and the altered
displayed:
name is automatically recorded.
Changing the designations of items 2-6 on menu level 1 Select the point to be changed. On menu level 2, place the cursor arrow in front of the designation _ Powder quantity: arrow to powder quantity,
by turning the MFK (e.g. 2 Individual crowns). In
select (press MFK), change (turn MFK) and
order to change the designation, proceed as with
confirm (press MFK)
„Programmable measuring programs“.
Scales This function serves to weigh any components - without using the measuring function of Smartbox Invest (e.g. to weigh the quantity of another _ Concentration: arrow to concentration, select,
investment material): menu level 1 additional
change and confirm. For checking purposes, the
functions, menu level 2 select scales - set up
proportion of the liquids in ml relating to the
beaker (6) and tare with start button (3), weigh
powder reference quantities are displayed (see
out the desired quantity.
basic settings!). Only liquid Only the liquids are measured - without measuring the powder (e.g. for the use of liquids with another investment material). Set the desired total liquid quantity and concentration of the liquid, similarly _ Program name: arrow to program name, select (pressMFK).
22
to individual measuring, and start the measuring using the start button (3).
GB
Cleaning functions On menu level 2, item „Additional functions“, you can find the sub-items Water extraction, liquid extraction and powder extraction. They serve to get the relevant component out of the machine (without measuring function) - e.g. before using another investment, material or liquid. The shaft return is part of the cleaning functions - it should be used when the screw conveyor has, for some unknown reason, not conveyed the powder on or the conveyor is blocked. All cleaning functions are selected by turning and pressing the MFK and carried out by pressing down the start button permanently - and will carry on until the start button is released again.
Maintenance/care As with the majority of AmannGirrbach machines, Smartbox Invest is designed to be maintenancefree and that is why they are provided with a maintenance-free conveying and dosing unit. The machine must be kept clean at all costs. In addition, please ensure that no fluid gets into the conveying screw. Never clean the Smartbox Invest with compressed air or a steam cleaner! Switch the machine off before carrying out any maintenance and cleaning work! Release the guard of the closure and lift the Cleaning the liquid outlet
closure completely upwards (the silo closure
Regularly check the outlet of the liquid hose and
thus supports the display that has been flipped
the seal of the silo cover and clean if necessary!
upwards).
To do so, flip the display up and open the silo closure.
The seal of the silo cover can now be removed, cleaned and re-fitted. Pull off the hose, loosen the
The seal of the silo cover can now be removed,
knurled thumb screw that fixes the small metal
cleaned and re-fitted. Pull off the hose, loosen the
outlet pipe, remove the outlet pipe and clean.
knurled thumb screw that fixes the small metal outlet pipe, remove the outlet pipe and clean.
23
GB
Cleaning the ventilator filter units Check the ventilator of the cooling unit and control unit (back of machine) at least every 4 weeks. In order to do so, remove the plastic covers, inspect the filter units and clean with water and/or compressed air if necessary. CAUTION: do not insert any damp filter units!
Cleaning the conveyor screw/silo Before cleaning (at least every six months, and whenever a change of investment material or an error message appears), the investment material in the silo has to be completely removed (powder extraction - see „Operation“). Remove any residues carefully with a vacuum cleaner. Make Cleaning the powder silo
sure you do not damage the surface of the silo‘s
Carefully remove any deposits lteft on the inner
inner walls!
walls of the silo occasionally using a soft brush
To remove the screw, proceed as follows: Flip the
or a dry cloth (approx. every third to fifth time
display upwards and open the silo closure. Release
before filling with fresh investment powder).
the guard of the closure and lift the closure
CAUTION: The inner walls of the powder silo
completely upwards (see also „Cleaning the liquid
have a special coating that can be damaged by
outlet“). Open the screw cap to the screw and
a scouring sponge or pointed object! Only soft
take out the conveyor screw.
brushes or cloths should be used for cleaning!
Clean the screw with a dry cloth and then re-assemble the machine in reverse order.
24
GB
Technical data: Dimensions:
525/300/620 mm
Weight:
25 kg
Electr. connected load:
230 V/50 Hz - 115 V/60 Hz
Ouput:
100 W
Electrical fuse:
T2A
max. dosing amount: 1000 g investment compound powder Content of fluid container:
1l
Content of powder silo:
approx. 4 kg
Cooling capacity:
max. 15 °C below the existing ambient temperature
Scope of delivery: Order No.: 117200
Smartbox Invest Basic device with powder silo and 2 storage tanks for fluids, power cable
Accessories: Order No.: 117201
Beaker height adaptor
117202
Desiccant bags, pack of 5
25
GB
What to do if…
Problem
Cause
Device does not acknowledge the lid is closed Liquid escapes
Solution Adjust lid and check microswitch, call service department*
Liquid outlet clogged
Liquid squirts out, even after cleaning
Clean clip, refer to cleaning instructions of liquid outlet Check pump settings, call service department*
Improper amount of dose used
Dirt underneath the scale
Clean all spaces, refer to instructions on how to clean the scale
Required amount of powder cannot be achieved
Hopper empty
Press “Start” button to abort dosing process, refill hopper
Powder is not dispensed
Press “Start” button to abort dosing process, tap on hopper wall to cause powder to drop down, clean hopper walls, refer to cleaning instructions of powder hopper
Worm gear blocked
Check and clean hopper outlet, refer to cleaning instructions of powder hopper; if you still require assistance, call the service department
Container empty
Press “Start” button to abort dosing process and check the respective container
Liquid outlet and hose clogged
Clean clip and hose, refer to cleaning instructions for liquid outlet, if hose still cogged, call service department*
Required amount of liquid not reached
* If the device must be returned for service, please ensure all containers are completely empty
26
GB
and safety ! Warning instructions
cristobalite. Do not breath the dust in since there is a risk of lung damage (silicosis, lung
Smartbox Invest was developed for use in the dental sector. Setting up and operation must be carried out in line with the operating instructions. The manufacturer assumes no liability for
cancer)! _ We strongly recommend connecting a suitable laboratory extraction unit to the Smartbox!
damage arising from other or improper use. Furthermore, before using it, the user is obliged, on his own responsibility, to check the machine‘s suitability and usability for the intended purposes insofar as these purposes are not listed in the instruction manual. All repairs and/or servicing work, apart from those described under „Maintenance/care“, may only be carried out by the manufacturer or authorised service companies. Before transporting, completely empty the powder silo and completely clean the conveyor screw by means of a vacuum cleaner. In addition, completely empty the liquid storage tank. Always transport the Smartbox in its original packaging. _ Only use original AmannGirrbach accessories or spare parts! _ Switch the machine off before carrying out any maintenance and cleaning work! _ Regularly check the outlet of the liquid hose and clean if necessary! _ The inner walls of the powder silo have a special coating that can be damaged by a scouring sponge or pointed object! Only soft brushes or cloths should be used for cleaning! _ Check the ventilator of the cooling unit and control (back of machine) at least every 4 weeks. In order to do so, remove the plastic covers, inspect the filter units and clean with water and/or compressed air if necessary. CAUTION: do not insert any damp filter units! _ Investment material contains quartz and
27
2
3
I
1 Livello di menu 1
1 2
2 3 2
44
3 1 3
Livello di menu 2
Livello di menu 3
EU Declaration of Conformity
We,
Amann Girrbach GmbH Herrschaftswiesen 1 6842 Koblach | Austria
hereby declare that the product described below, both, in its basic design and construction and in the version marketed by us is in conformity with the relevant safety and health related requirements of the appropriate EU Directive. This declaration shall cease to be valid if modifications are made to the product without our approval. Product: Number: Applicable EU Directives:
Smartbox Invest Dosing apparatus 117200 98/37/EU - Machinery guidelines EU Low-Voltage Equipment Directive (73/23/EU) EU Directive on Electromagnetic Compatibility (89/336/EU)
Applicable harmonized standards:
DIN EN ISO 12100-1 DIN EN ISO 12100-2 DIN EN 55014-1/2 DIN EN 61010-1
Applicable national standards:
-
It is ensured through internal measures that products will always meet the requirements of current EU Directives and relevant standards.
Pforzheim, 27.06.2007
by order Dr.-Ing. S. Cramer von Clausbruch
The signatory acts on behalf of company management and with full power of attorney.
67