Anetic Aid Ltd
AT4 Tourniquet System Final Test Sheet
Final Test Sheet
1 Page
Preview
Page 1
Final Test AT4E Model 40080
Serial Number 8 0 1 - 4 2 0 1 : Date
Item Serial Number Encoder PCB Display PCB Controller PCB Power Supply and Solenoid PCB Battery Software Version v... IFU Version and Language v... Voltage and National connector 230V UK Note this is a final functional and appearance check prior to release. Step
Action
Feature
1.
Check test results of base
2.
Check test results of tray and head
Test results to have been completed, passed and validated
3.
Press ON
4.
Confirm SW Version
5.
Calibrate
Follow specified sequence
6.
Real Time Clock
Set to
7.
Press Timer
Timer Reminder Options
8.
Press IVRA
IVRA Yellow LED
9.
Inflate Blue to 350
Inflates to >335 in 5 seconds Within 345-355 within 10 seconds
10.
Inflate Red to 200
Inflates to >180 in 5 seconds Within 195-205 within 10 seconds
11.
Deflate Blue fast
Must go to “0”
12.
Deflate Red fast
Must go to “0”
13.
Inflate Red to 600
Must be > than 500 in <7 Seconds
14.
Inflate Blue to 600
Must be > than 500 in <7 Seconds
15.
Deflate both
Must got to zero in <5 Seconds
14.
Earth bond
<0.2Ω
15.
Insulation @500V
>20MΩ
16.
Earth Leakage
<0.5mA
17.
Enclosure Leakage
<0.5mA
18.
Connect mains
Alternate Green & Yellow indicator
19.
Check labels on unit
Total 10 including front panel, 230V, CE
20.
Accessories
Hoses, Cable, IFU - Remove Battery Fuse
21
Packing Case
Label and banded
Function
Name
Pass
Fail
Check Displays
Sign
UK
Date
Inspect & Pack Validate form
AT4E-Test-01
Final Test
Page 1 of 1