AngioDynamics
Solero Ablation Measurement Guide Rev 01 June 2017
Measurement Guide
6 Pages
Preview
Page 1
A B L AT I O N M E A S U R E M E N T G U I D E
EX VIVO BOVINE LIVER (diameter Ø x length)
4 min
2 min 0.37 cm
60 W
1.20 cm
2.4 Ø x 2.8 cm
0.08 cm
100 W
1.50 cm
3.0 Ø x 3.7 cm
0.23 cm
140 W
1.60 cm
3.2 Ø x 4.0 cm
6 min 0.18 cm
0.03 cm
1.50 cm
1.75 cm
3.0 Ø x 3.6 cm
3.5 Ø x 3.9 cm
0.43 cm
0.68 cm
1.75 cm
2.10 cm
3.5 Ø x 4.4 cm
4.2 Ø x 4.9 cm
0.73 cm
0.93 cm
2.00 cm
2.20 cm
4.0 Ø x 5.0 cm
4.4 Ø x 5.4 cm
Note: Ablation volumes in perfused tissues may differ from static laboratory results
EX VIVO PORCINE KIDNEY (diameter Ø x length)
4 min
2 min 0.22 cm
60 W
1.20 cm
2.4 Ø x 3.1 cm
0.08 cm
100 W
1.30 cm
2.6 Ø x 3.7 cm
0.38 cm
140 W
1.45 cm
2.9 Ø x 4.3 cm
6 min 0.13 cm
0.23 cm
1.50 cm
1.60 cm
3.0 Ø x 3.8 cm
3.2 Ø x 4.0 cm
0.48 cm
0.73 cm
1.70 cm
1.75 cm
3.4 Ø x 4.5 cm
3.5 Ø x 5.0 cm
0.73 cm
1.03 cm
1.75 cm
1.95 cm
3.5 Ø x 5.0 cm
3.9 Ø x 5.6 cm
Note: Ablation volumes in perfused tissues may differ from static laboratory results
EX VIVO PORCINE LUNG (diameter Ø x length)
4 min
2 min 0.82 cm
60 W
0.65 cm
1.3 Ø x 1.9 cm
0.42 cm
100 W
0.80 cm
1.6 Ø x 2.7 cm
0.17 cm
140 W
0.85 cm
1.7 Ø x 3.2 cm
6 min 0.57 cm
0.72 cm
0.75 cm
0.80 cm
1.5 Ø x 2.1 cm
1.6 Ø x 2.4 cm
0.17 cm
0.17 cm
1.00 cm
1.15 cm
2.0 Ø x 3.2 cm
2.3 Ø x 3.2 cm
0.03 cm
0.03 cm
1.25 cm
1.30 cm
2.5 Ø x 3.6 cm
2.6 Ø x 3.6 cm
Note: Ablation volumes in perfused tissues may differ from static laboratory results
SOLERO* MICROWAVE TISSUE ABLATION SYSTEM SKU
Description
H78712740000US0
Solero MW Generator
H787700106001US0
Solero MW Applicator 14 cm
H787700106002US0
Solero MW Applicator 19 cm
H787700106003US0
Solero MW Applicator 29 cm
INDICATION FOR USE The Solero Microwave Tissue Ablation (MTA) System and Accessories are indicated for the ablation of soft tissue during open procedures. The Solero MTA System is not intended for cardiac use. Please refer to the Solero Generator Operator’s Manual and the Solero Applicator Directions For Use for complete instructions, warnings, and precautions. CAUTION: Federal Law (USA) restricts this device to sale by or use under the order of a physician.
USA > 14 Plaza Drive, Latham, NY 12110 > tel: 800-772-6446 or 518-798-1215 > fax: 518-798-1360 International > Haaksbergweg 75 (Margriettoren), 1101 BR, Amsterdam Z-O > The Netherlands > tel: +31 (0)20 753 2949 > fax: +31 (0)20 753 2939 www.angiodynamics.com *AngioDynamics, the AngioDynamics logo, Solero and the Solero logo are trademarks and/or registered trademarks of AngioDynamics, Inc., an affiliate or a subsidiary. © 2017 AngioDynamics, Inc. ANGM 498 US Rev 01 06/17