Contoura Models 1000 and 1080 Instructions for Use Oct 2013
Instructions for Use
36 Pages
Preview
Page 1
Design Policy and Copyright ® and ™ are trademarks belonging to the ArjoHuntleigh group of companies. © ArjoHuntleigh 2000. As our policy is one of continuous improvement, we reserve the right to modify designs without prior notice. The content of this publication may not be copied either wholly or in part without the consent of ArjoHuntleigh.
Contents Preliminaries... ii 1.
Introduction ... 1
2.
Clinical Applications ... 3
3.
Installation ... 5
4.
Operation ... 7
5.
Patient Weighing System (Model C1080) ... 16
6.
Decontamination ... 21
7.
Maintenance ... 22
8.
Troubleshooting ... 24
9.
Accessories ... 25
10.
Warranty and Service... 26
11.
Dimensions and Data ... 27
12.
Transportation and Storage ... 29
i
746-173-CA-3
Preliminaries Scope This manual provides information and instructions for the use and maintenance of ArjoHuntleigh Contoura® C1000 series bariatric beds. Please read this manual carefully before using the equipment. Pay particular attention to safety instructions. Warnings, Cautions and Notes WARNINGS given in this manual identify possible hazards in procedures or conditions which, if not followed correctly, could result in death, injury or other serious adverse reactions. Cautions given in this manual identify possible hazards in procedures or conditions which, if not followed correctly, could result in equipment failure or damage. Notes given in this manual (indicated by ) are used to explain or amplify a procedure or condition.
General Warnings
WARNING Use a recommended and approved mattress and make sure that it is the correct size. Do not use accessories that have not been designed or approved for use with the bed. Electrical equipment can be hazardous if misused. Obey safety instructions. Do not use electrically powered beds in the presence of flammable gases such as anesthetic agents. Do not use electrically powered beds in operating theatres. The electrical power outlet on the foot end of model C1080 is only for use with ArjoHuntleigh approved pumps. Do not use it to supply power to any other equipment. The bed is very heavy and at least two people are needed to move it. Use extra caution when moving on an incline.
ii
746-173-CA-3
1. Introduction Product Description
ArjoHuntleigh Contoura C1000 series beds are designed to accommodate patients weighing up to 990 lbs. (450kg). They combine the highest standards of functionality and convenience with an aesthetically pleasing appearance. Model C1000 is the basic bed. Model C1080 includes a weighing system and a power outlet for an ArjoHuntleigh approved mattress pump. All beds incorporate the following features: • Electrically powered adjustable height. • Electrically powered adjustable two-way tilt. • Four-section profiling mattress platform with electrically powered adjustment of backrest and leg elevation plus auto-contour facility. • Adjustable kneebreak angle. • Separate attendant control panel with the facility to prevent operation by the patient. • Integral bed extension, including extending mattress platform. • Large diameter castors with braking and steering facility. • Integral folding siderails with width adjustment. • Full width mattress platform to minimize transfer gap. • Removable (lift off) head panel and foot panel, manufactured from reinforced, molded and contoured ABS. • Revolving buffers on all four corners. The electrical controls enable effortless positioning of the patient. The self-contained weighing system on the Model C1080 enables continuous monitoring of patient weight and weight change, with minimal effort and disturbance.
1
746-173-CA-3
The power driven functions of height, backrest angle and leg elevation are operated by simple push button controls. The push button handset may be used by the caregiver or patient. A second handset, accessible only to the caregiver, incorporates additional push buttons for two-way tilt control and CPR emergency settings. The bed also has an attendant control panel (ACP), by means of which the power driven functions can be individually disabled. Operation of the bed by the patient is thus at the discretion of the caregiver. The electrical actuators consist of a screw and nut assembly driven by a 24V d.c. motor and gearbox. Power for the actuators is derived from a mains operated d.c. power supply with emergency battery back up.
Contoura C1080 bed
2
746-173-CA-3
2. Clinical Applications Area of Use
General medical and surgical floors, care of the elderly, ICU, coronary care, oncology and emergency rooms.
Indications
The Contoura® 1000 and 1080 beds are designed for use exclusively by bariatric patients. A bariatric patient is defined as one who, regardless of their age, has limitations in health and social care due to their weight, physical size, shape, width, health, mobility or tissue viability, and fits any of the following criteria: • Body Mass Index (BMI) greater than 40 * • More than 88 lbs (40kg) above the ideal weight for their height * • Exceed the safe working load (SWL) and dimensions of standard support surfaces such as beds, chairs, wheelchairs, couches, trolleys and mattresses * Based on guidelines from NICE (UK National Institute for Health and Clinical Excellence), 2004
WARNING This bed must not be used by patients who do not fall into the bariatric category as defined above.
3
746-173-CA-3
In-Bed Weighing (Contoura C1080 only)
The in-bed weighing system is designed to be used for monitoring changes in a patient’s weight over a period of time deemed appropriate by clinical staff. This may be used, in association with other clinical findings, to monitor the overall progress of the patient. To gain a true representation of the change in a patient’s weight, urine bags and drainage bags should be emptied before the reading is taken.
Contraindications
Do not exceed the safe working load of 1100 lbs. (500kg) for the combined patient, mattress and accessories.
When using powered mattress replacements, e.g. BariBreeze, do not exceed the safe working load stated by the mattress manufacturer.
4
746-173-CA-3
3. Installation All Models
Unwind the power cord from the cable tidy at the head of the bed and connect it to a suitable outlet socket.
WARNING Make sure that the power cord is not stretched when connected. Make sure that the handset cable and power cord cannot become entangled with moving parts of the bed.
Caution When the bed is to be moved, disconnect the power cord from the socket outlet and store it on the cable tidy at the head end of the bed as shown. Always remove the power cord from the cable tidy when it is connected to the electrical supply. An equipotential terminal in accordance with the requirements of IEC 601-1 (EN 60601-1) is provided on the mattress platform frame at the head of the bed.
5
746-173-CA-3
Model 1080
When the bed is in the area where it is to be used, release the transport locking screws by turning them fully anti-clockwise. These are four, hand operated turnscrews which are accessible on the base.
Caution The transport locking screws must always be firmly locked when the bed is moved except for short distances over smooth surfaces. BariBreeze Mattress Replacement System Contoura C1000 and C1080 beds are suitable for use with the ArjoHuntleigh BariBreeze mattress replacement system, model number 647311.
WARNING Read the Instructions for Use supplied with the BariBreeze mattress before attempting to install or use it. Make sure that the mattress tubes and mains cable do not become entangled with moving parts of the bed. Before using the BariBreeze mattress “patient turn” feature: • Raise the siderails – they can be upright or inclined outwards as required • Make sure that the patient cannot roll off the mattress when the bed is profiled. The degree of turn and profiling that can be used safely will depend on the size and condition of the patient.
The Contoura C1080 bed can also be used with a standard mattress, model 450PM206/092. Details will be provided on request.
6
746-173-CA-3
4. Operation • Always use a mattress of the correct size and type. Suitable mattresses are available from ArjoHuntleigh.
Safety
• When the bed is operated, check that no obstacles such as bedside furniture are in its path. Also make sure that the patient’s and caregivers’ limbs are not trapped. • Always apply the brakes when the bed is stationary. • Lower the bed to minimum height and raise the siderails when the patient is left unattended. • Keep these instructions handy for reference.
WARNING The bed is very heavy and at least two people are needed to move it. Use extra caution when moving on an incline. Brakes and Steering
At the foot of the bed is a pair of linked brake pedals that operate on all four castors. When the pedals are in the middle position, the brakes are free and all four castors are able to rotate and swivel. When the pedals are pushed down, the brakes are applied on all four castors. When the pedals are raised, both head end castors are engaged so that they cannot swivel. Always engage both castors in line and trailing. The bed can then be steered by pushing from the foot end.
FREE
BRAKE
7
STEER
746-173-CA-3
Siderails
To lower the siderail: Press the release button (1) and lower the siderail toward the foot end (2). To raise the siderail: Lift the top rail up and toward the head end until the latch engages at maximum height.
The siderails can be inclined outward at the top to increase the in-bed width or moved to the upright condition for transport. When the siderails are upright, the bed will pass comfortably through a 48 ins. (120cm) opening. To increase in-bed width: Lift the release handle (3) and swing the top rail outwards (4) until the mechanism locks. To reduce the in-bed width: Lift the release handle (3) and swing the siderail up (4) until the mechanism locks in the upright position
8
746-173-CA-3
Bed Extension
Before extending or shortening the bed: Apply the brakes, ensure that the legs are lowered to the horizontal position and the foot end strut is raised (disengaged). The operation is easier if you first remove the foot panel. To extend the bed: Lift the locking handle (1) and pull out the foot end of the mattress platform (2). Return the handle to horizontal and re-engage the lock. Slide back, lift and hold the locking lever (3) and pull out the foot end of the calf section (4). Reengage the lever to lock in the extended position. Install a mattress extension squab.
To retract the bed: Remove the mattress extension squab. Slide back, lift and hold the locking lever (3) and push in the foot end of the calf section (4). Reengage the lever to lock. Lift the locking handle (1) and push in the foot end of the mattress platform (2). Return the handle to the horizontal and re-engage the lock.
9
746-173-CA-3
Kneebreak Angle
When the legs are raised, the kneebreak angle can be adjusted using the strut at the foot end of the calf section.
WARNING The calf section is very heavy. Do not try to raise or lower it by hand. Use the push button controls. Do not attempt to disengage the strut when the leg section is raised. Foot up position: Lower the leg sections to the horizontal position, push down the handle (1) and engage the strut in the slot (2). Raise the legs to the required height. Foot down position: Lower the leg sections to the horizontal position, lift the handle (1) and disengage the strut from the slot (2). Raise the legs to the required height.
The second slot (3) is only used when the bed is extended
10
746-173-CA-3
Backrest Adjustment
The angle of the backrest is adjusted by means of push buttons. Refer to Power Operation later in this section.
Leg Elevation
As the legs are raised, by increasing the angle of the thigh section, the kneebreak angle will increase correspondingly. The elevation of the leg section is adjusted by means of push buttons. Refer to Power Operation later in this section.
Height Adjustment
The height of the mattress platform is adjusted by means of push buttons. Refer to Power Operation later in this section.
Tilt Adjustment
The angle of tilt is adjusted by means of push buttons on the caregiver handset. Refer to Power Operation later in this section.
Power Operation General
The bed is provided with power driven adjustment of backrest angle, leg elevation, bed height and tilt angle. The functions are operated from a push button handset, which is connected to the bed by an extendable, flexible cable. When a push button is pressed, the appropriate bed function will operate until the push button is released or the limit of travel is reached. The Autocontour button operates both backrest and leg elevation simultaneously. The backrest, leg and height controls are duplicated on a second handset, accessible only to the caregiver, which has additional push buttons for the control of tilt angle and emergency CPR configuration. The caregiver handset is stored at the foot end of the bed.
11
746-173-CA-3
Backrest Leg elevation Height Auto-contour
Patient control handset By using the Attendant Control Panel (ACP), the Control Inhibit (using the ACP) caregiver has the discretion to disable the push button controls and thus limit or prevent operation of the bed by the patient. Pull out the sliding tray beneath the foot end of the mattress platform. This gives access to both the ACP and the user interface module for the weighing system if fitted. The caregiver handset is also stored on this tray. Use of the weighing system is described in Section 5. On the ACP is a rotary ‘lock-out’ switch for each bed function. To prevent the patient from making an adjustment, disable that particular bed function by turning the lock-out switch so that the pointer is horizontal. In this condition the selected push buttons are disabled on both the patient handset and the caregiver handset. Because the bed height and tilt angles are determined by the same two actuators, both these functions are disabled by the same lock-out switch.
12
746-173-CA-3
Pull out sliding tray
Leg Elevation
Backrest
CPR
Tilt Height
Leg elevation Battery charge indicator
Backrest
Tilt and height
Caregiver handset and Attendant Control Panel (ACP)
13
746-173-CA-3
Tilt Angle Control
The tilt angle can only be controlled from the caregiver handset When reverting from a tilt condition, the bed will automatically stop in the horizontal position for two or three seconds.
CPR Emergency
There is a CPR button on the caregiver handset. Press and hold this button to automatically return the bed to a flat, horizontal, low height condition.
WARNING Before using the CPR button, ensure that all lock out switches on the ACP are in the vertical, unlocked position. Calibration Errors
If a bleeping sound is heard when operating the push buttons and the function is inoperative, this indicates an actuator calibration error. Contact the ArjoHuntleigh service department.
Batteries
The emergency back-up battery is automatically switched on if the bed is disconnected from the electrical supply. Battery life is limited and is intended for emergency use only. To ensure that batteries are kept in a fully charged condition, the bed must be connected to the electrical supply at all times during normal use. If, when the bed is operated using the batteries, a bleeping sound is heard, this indicates a low charge condition. The bed should be reconnected to the electrical supply, for a minimum of 24 hours in order to recharge the batteries. The battery charge indicator on the ACP illuminates when the batteries are being charged.
14
746-173-CA-3
Emergency CPR Procedure
IN THE EVENT OF CARDIAC ARREST: 1 - Operate the BariBreeze pump CPR release (if in use) – refer to the instructions for use supplied with the mattress 2 - Operate the Contoura bed CPR release 3 - Lower the siderails
15
746-173-CA-3
5. Patient Weighing System (Model C1080) General
Four load cells are installed in the bed frame. The electrical output signals from the load cells are integrated in a load converter, which provides a single output to the user interface. Separate displays are provided for patient weight and patient weight change.
Accuracy
The weighing system is very sensitive and will be affected by a number of outside influences. It is therefore important that the following criteria are observed. a) The bed must be installed on a flat and level floor. b) The transport screws must be fully disengaged prior to use. c) The bed should be positioned clear of obstructions such as walls, furniture, trailing leads or curtains. d) The weighing system should be used within the normal operating ambient temperature range of 50°F to 104°F (10°C to 40°C). e) For best results, the patient should be positioned approximately centrally on the bed. f) The weighing system is capable of detecting relatively small movements of the patient. It is therefore necessary for patient movement to be minimized during the few seconds required to obtain a stabilized reading and that the operators / caregivers stand clear of the bed during this time. g) When weighing, the handsets must be replaced in position on the platform.
16
746-173-CA-3
User Interface Module (UIM)
Lift and pull out the sliding tray beneath the foot end of the mattress platform. Open the lid of the UIM (right hand side).
Pull out sliding tray Operation
Operation of the system is described in the following paragraphs. The way the features are used varies in accordance with the situation viz.: Patient already on the bed. Enter the previously known patient weight on the top display. Enter the previously known patient weight change on the bottom display. (If not known, zero the bottom display). The indicator will now display patient weight and weight change as patient weight varies, starting at the values entered Empty bed. Zero the top display. Place patient on the bed. Top display reads patient weight. Zero the bottom display. As patient weight varies, the change is shown on the bottom display
System Stable Indicator
After pressing buttons, a small delay occurs before the display responds. The System Stable indicator (F) will illuminate to show that display readings are valid.
17