ARJO Huntleigh Healthcare
Flowtron Excel DVT System Instructions For Use Rev 24 April 2019
Instructions For Use
28 Pages
Preview
Page 1
WARNING To avoid injury, always read this Instructions for Use and accompanied documents before using the product. Mandatory to read the Instructions for Use
Design Policy and Copyright ® and ™ are trademarks belonging to the Arjo group of companies. © Arjo 2019. As our policy is one of continuous improvement, we reserve the right to modify designs without prior notice. The content of this publication may not be copied either whole or in part without the consent of Arjo.
Contents General Safety... iii Introduction...1 About this Manual... 1 Intended Use... 1 About Flowtron Excel... 1 Clinical Applications...3 Indications... 3 Contraindications... 3 Cautions... 3 Guidelines and Recommendations... 4 System Set Up...5 Installing the Pump... 5 Garment Application... 5 Garment Removal... 5 To Use Only One Garment... 5 Pre-Use Check...6 Operation...7 Start Up... 7 Shut Down... 7 Pressure Adjustment...8 Pressure Output Check... 8 Adjusting the Output Pressure... 8 System Calibration Check... 8 Decontamination...9 Routine Maintenance...10 Flowtron Excel System... 10 Flowtron Excel Pump... 10 Serial Labels... 10 Troubleshooting - Alarm Condition...11 Alarm Condition...11 Alarm Cancel... 11 Troubleshooting...12 Accessories...13 Garments and Tubing... 13 Technical Specification...14 Electromagnetic Compatibility (EMC)...16
(i)
Intentionally left blank
(ii)
General Safety Before you connect the system pump to a mains socket, read carefully all the installation instructions contained within this manual. The system has been designed to comply with regulatory safety standards including: • EN60601-1:2006/A1:2013 • IEC60601-1:2005/A1:2012 and IEC60601-1-8:2012 • ANSI/AAMI ES60601-1(2005)+AMD(2012) and CAN/CSA-C22.2 No.60601-1(2008)+(2014)
Safety Warnings • It is the responsibility of the care giver to ensure that the user can use this product safely. • Make sure that the mains power cable and tubeset or air hoses are positioned to avoid causing a trip or other hazard, and are clear of moving bed mechanisms or other possible entrapment areas. • Electrical equipment may be hazardous if misused. There are no user-serviceable parts inside the pump. The pump's case must only be removed by authorised technical personnel. No modification of this equipment is allowed. • The mains power socket/plug must be accessible at all times. To disconnect the pump completely from the electricity supply, remove the plug from the mains power socket. • Disconnect the pump from the mains power socket before cleaning and inspecting. • Keep the pump away from sources of liquids and do not immerse in water. • Do not use the pump in the presence of uncontained flammable liquids or gases. • Only the pump and garment/insert combination as indicated by Arjo should be used. The correct function of the product cannot be guaranteed if incorrect pump and garment combinations are used. • The Flowtron® Excel system is NOT intended for use in the Home Healthcare Environment (e.g. private dwellings or nursing homes).
WARNING - Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30cm (12 in) to any part of the Flowtron Excel (ACS550), including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
Caution (applicable to the USA market only) • US Federal law restricts this device to sale by or on the order of a physician.
(iii)
Precautions For your own safety and the safety of the equipment, always take the following precautions: • Do not expose the system to naked flames, such as cigarettes, etc. • Do not store the system in direct sunlight. • Do not use phenol-based solutions to clean the system. • Make sure the system is clean and dry prior to use or storage. • When the pump is in use the operator should remain in area in case the system alarms. • If a serious incident occurs in relation to this medical device, affecting the user, or the patient then the user or patient should report the serious incident to the medical device manufacturer or the distributor. In the European Union, the user should also report the serious incident to the Competent Authority in the member state where they are located. Electromagnetic Compatibility (EMC) This product complies with the requirements of applicable EMC Standards. Medical electrical equipment needs special precautions regarding EMC and needs to be installed in accordance with the following instructions: • The use of accessories not specified by the manufacturer may result in increased emissions by, or decreased immunity of, the equipment, affecting its performance. • Portable and mobile radio frequency (RF) communications equipment (e.g. mobile/cell phones) can affect medical electrical equipment. • If this equipment needs to be used adjacent to other electrical equipment, normal operation must be checked before use. • For detailed EMC information contact Arjo service personnel. Expected Service Life The Flowtron Excel pump has an expected service life of seven years. To maintain the condition of the pump have the pump serviced regularly according to the schedule recommended by Arjo. Do NOT use unapproved accessories or attempt to modify, disassemble or otherwise misuse the Flowtron Excel system. Failure to observe this caution could result in injury, or in extreme cases, death. End of Life Disposal • Garment material or any other textiles, polymers or plastic materials etc. should be sorted as combustible waste. • Pump units have electrical and electronic components should be disassembled and recycled per Waste of Electrical and Electronic Equipment (WEEE) or in accordance with local or national regulation. Environmental Protection Incorrect disposal of this equipment and its component parts, particularly batteries or other electrical components, may produce substances that are hazardous to the environment. To minimise these hazards, contact Arjo for information on correct disposal.
(iv)
1. Introduction About this Manual
7KLVPDQXDOLVRXULQWURGXFWLRQWRWKH)ORZWURQ®([FHO VVWHP<RXPXVWUHDGDQGIXOOXQGHUVWDQGWKLVPDQXDO EHIRUHXVLQJWKHVVWHP 8VHWKLVPDQXDOWRLQLWLDOOVHWXSWKHVVWHPDQGNHHS LWDVDUHIHUHQFHIRUGDWRGDURXWLQHVDQGDVDJXLGHWR PDLQWHQDQFH ,IRXKDYHDQGLIILFXOWLHVLQVHWWLQJXSRUXVLQJWKH Flowtron([FHOVVWHPFRQWDFWRXUORFDO$UMRVDOHV RIILFHOLVWHGDWWKHHQGRIWKLVPDQXDO
Intended Use
7KHLQWHQGHGXVHRIWKLVSURGXFWLVWRSUHYHQW'HHS9HLQ 7KURPERVLV '97 7KHJDUPHQWVDUHVLQJOHSDWLHQWXVH RQO,WLVQRWIRUXVHLQWKHKRPHKHDOWKFDUH HQYLURQPHQW 7KHFlowtron([FHOVVWHPVKRXOGEHXVHGDVSDUWRID SUHVFULEHGSODQRIFDUH UHIHUWR³,QGLFDWLRQV´RQ SDJH
About Flowtron Excel
7KHFlowtron([FHOSXPSRSHUDWHVRQDVHFRQG DXWRPDWLFDOOWLPHGFFOHFRQVLVWLQJRIDSSUR[LPDWHO VHFRQGVRILQIODWLRQIROORZHGEDSSUR[LPDWHO VHFRQGVRIGHIODWLRQ 7KHFlowtron([FHOVVWHPPDEHXVHGRQSDWLHQWVDW ULVNRIGHYHORSLQJGHHSYHLQWKURPERVLVDQGLQ FRQMXQFWLRQZLWKVVWHPLFLQWHUYHQWLRQV HJ DQWLFRDJXODWLRQGUXJV IRUWKHKLJKULVNSDWLHQW 7KHFlowtron([FHOLVLQWHQGHGIRUXVH21/<LQ 3URIHVVLRQDO+HDOWKFDUH)DFLOLWLHV HJKRVSLWDOVRU SKVLFLDQV¶RIILFHV $IXOOWHFKQLFDOGHVFULSWLRQRIWKHFlowtron([FHO VVWHPFDQEHIRXQGLQWKH6HUYLFH0DQXDOSDUW1R 6(5DYDLODEOHIURPRXUORFDO$UMRVDOHVRIILFH
1
Carry Handle
Pressure Display
Single Leg Mode Button and Indicator Garment Connectors
Mains Power Switch
Pressure Control
Flowtron Excel Pump - Front View
2
2. Clinical Applications Indications
7KHDSSOLFDWLRQRIWKHFlowtron([FHOVVWHPLVIRUWKH SUHYHQWLRQRI'HHS9HLQ7KURPERVLV '97 ZKHQ FRPELQHGZLWKDQLQGLYLGXDOLVHGPRQLWRULQJ SURJUDPPH 7KLVVVWHPUHSUHVHQWRQHDVSHFWRID'97VWUDWHJLI WKHSDWLHQW VFRQGLWLRQFKDQJHVWKHRYHUDOOWKHUDS UHJLPHQVKRXOGEHUHYLHZHGEWKHSUHVFULELQJFOLQLFLDQ 7KHDERYHDUHJXLGHOLQHVRQODQGVKRXOGQRWUHSODFH FOLQLFDOMXGJHPHQW
Contraindications
7KH Flowtron([FHOVVWHPVKRXOGQRWEHXVHGLQWKH IROORZLQJFRQGLWLRQV 6HYHUHDUWHULRVFOHURVLVRURWKHULVFKDHPLFYDVFXODU GLVHDVHV .QRZQRUVXVSHFWHGDFXWH'HHS9HLQ7KURPERVLV '97 RUSKOHELWLV 6HYHUHFRQJHVWLYHFDUGLDFIDLOXUHRUDQFRQGLWLRQ ZKHUHDQLQFUHDVHRIIOXLGWRWKHKHDUWPDEH GHWULPHQWDO 3XOPRQDUHPEROLVP $QORFDOFRQGLWLRQLQZKLFKWKHJDUPHQWVZRXOG LQWHUIHUHLQFOXGLQJJDQJUHQHUHFHQWVNLQJUDIW GHUPDWLWLVRUXQWUHDWHGLQIHFWHGOHJZRXQGV ,IRXDUHXQVXUHRIDQFRQWUDLQGLFDWLRQVUHIHUWRWKH SDWLHQW¶VSKVLFLDQEHIRUHXVLQJWKHGHYLFH
Cautions
3URSHUJDUPHQWDSSOLFDWLRQDQGFRQQHFWLRQWRWKH SXPSLVHVVHQWLDO *DUPHQWVVKRXOGEHSRVLWLRQHGLQVXFKDZDWKDW WKHGRQRWFUHDWHDQSRWHQWLDOIRUFRQVWDQW SUHVVXUHSRLQWVRQWKHSDWLHQW¶VOLPE$GGLWLRQDO FDUHVKRXOGEHWDNHQZKHQSODFLQJWKHJDUPHQWVRQ DQGHIRUPHGOHJRURQOHJVZLWKVLJQLILFDQW RHGHPD :KHQXVHGIRU'97SUHYHQWLRQFRQWLQXRXV H[WHUQDOSQHXPDWLFFRPSUHVVLRQLVUHFRPPHQGHG XQWLOWKHSDWLHQWLVIXOODPEXODWRU8QLQWHUUXSWHG XVHRIWKHFlowtron([FHOVVWHPLVHQFRXUDJHG *DUPHQWVVKRXOGEHUHPRYHGLPPHGLDWHOLIWKH SDWLHQWH[SHULHQFHVWLQJOLQJQXPEQHVVRUSDLQDQG WKHSKVLFLDQQRWLILHG 3
7KHFlowtron([FHOVVWHPVKRXOGEH86(' :,7+&$87,21RQSDWLHQWVZLWK ,QVHQVLWLYHH[WUHPLWLHV 'LDEHWHV ,PSDLUHGFLUFXODWLRQ )UDJLOHRULPSDLUHGVNLQ NOTE These are guidelines only and should not replace clinical judgement and experience.
Guidelines and Recommendations General Recommendations
DVT Prophylaxis
:KLOHXVLQJWKHVVWHPWKHSDWLHQW¶VOLPEVVKRXOGEH FKHFNHGGXULQJHYHUVKLIWDQGPRUHRIWHQLIWKH SDWLHQWKDVNQRZQFLUFXODWRURUVNLQSUREOHPVRULV GLDEHWLF &OLQLFDOMXGJPHQWVKRXOGEHXVHGWRGHWHUPLQHLIWKH SDWLHQW¶VVNLQFRQGLWLRQUHTXLUHVDGGLWLRQDOPHDVXUHV RULIWKHWUHDWPHQWVKRXOGEHGLVFRQWLQXHGDQG DOWHUQDWLYHPRGDOLWLHVXVHG $UMRGRHVQRWUHFRPPHQGWKHXVHRIFRPSUHVVLRQ VWRFNLQJVZLWKLWVVVWHP,IWKHVHDUHRUGHUHGEWKH SKVLFLDQWKHFOLQLFLDQVKRXOGHQVXUHWKDWWKH FRPSUHVVLRQVWRFNLQJVDUHSURSHUOPHDVXUHG DSSOLHGDQGZRUQEWKHSDWLHQW$QFRPSUHVVLRQ VWRFNLQJXVHGVKRXOGEHURXWLQHOFKHFNHGWRHQVXUH FRQWLQXHGSURSHUILWDQGDSSOLFDWLRQLQDGGLWLRQWR DVVHVVLQJWKHFRQGLWLRQRIWKHVNLQ :KHUHDSSURSULDWHSDWLHQWVVKRXOGEHLQVWUXFWHGLQ WKHSURSHUXVHRIWKHVVWHPWKHSXUSRVHRIWKHUDS DQGWKDWDQSUREOHPVVKRXOGEHUHSRUWHGWRWKH QXUVLQJVWDII 7KHFlowtron([FHOVVWHPVKRXOGEHDSSOLHGWRWKH SDWLHQWSUHRSHUDWLYHOSULRUWRWKHLQGXFWLRQRI DQDHVWKHVLD 7KHVVWHPVKRXOGEHXVHGFRQWLQXRXVOIRUQROHVV WKDQKRXUVSRVWRSHUDWLYHORUXQWLOWKHSDWLHQW EHFRPHVIXOODPEXODWRU ,IWKHJDUPHQWFDQQRWEHDSSOLHGWRWKHRSHUDWLYH OLPEGXULQJVXUJHULWPDEHDSSOLHGWRWKHOLPE RQFHWKHSDWLHQWUHDFKHVWKHUHFRYHUXQLW ,QWKHQRQVXUJLFDOSDWLHQWWKHVVWHPVKRXOGEH LQLWLDWHGLPPHGLDWHOZKHQWKHULVNRI'97IRUPDWLRQLV LGHQWLILHG
4
3. System Set Up Installing the Pump
$WWDFKWKHSXPSWRWKHEHGIUDPHXVLQJWKHEHGEUDFNHW RUSODFHWKHSXPSRQWKHIORRUXQGHUWKHEHG
Garment Application
&KHFNWKDWWKHPDLQVSRZHUVZLWFKRQWKHSXPSLV LQWKHRII O SRVLWLRQ 5HPRYHWKHJDUPHQWVIURPWKHSDFNDJLQJDQG XQIROG NOTE Garments are for single patient use only. Do not use the garments on a different patient after treatment.
3ODFHWKHEDFNRIWKHSDWLHQW¶VOHJLQWKHFHQWUH VHFWLRQRIWKHJDUPHQWZLWKWKHFRQQHFWRUWXELQJ SRLQWLQJGRZQZDUGVWRZDUGVWKHIRRW 6WDUWLQJZLWKWKHVLGHRIWKHJDUPHQWWKDWGRHVQRW KDYHWKHKRRNDQGORRSVWUDSWDEVZUDSVHFXUHO DJDLQVWWKHOHJ:KLOHKROGLQJWKHJDUPHQWDJDLQVW WKHOHJZUDSWKHWDEVRYHUWKHWRS(QVXUHWKDWWKH JDUPHQWLVFORVHILWWLQJDQGKDVQRFUHDVHVRUIROGV 7KHFRQQHFWRUWXELQJVKRXOGEHSRLQWLQJWRZDUGV WKHSDWLHQW¶VKHHO 0DNHVXUHWKHWXELQJDVVHPEOLVFRQQHFWHGWRWKH JDUPHQWFRQQHFWRURQWKHSXPS &RQQHFWWKHJDUPHQWFRQQHFWRUWRWKHWXELQJ DVVHPEO(QVXUHWKDWDVKDUSµFOLFN¶LVKHDUG3XOO OLJKWOWRFRQILUPSURSHUFRQQHFWLRQ 5HSHDWVWHSVWRIRUWKHVHFRQGJDUPHQWLIXVHG Garment Removal
7RGLVFRQQHFWDJDUPHQWSUHVVRQWKHWXELQJ DVVHPEOVQDSORFNFRQQHFWRUDQGSXOOWKHJDUPHQW FRQQHFWRUDZDIURPWKHWXELQJDVVHPEO
To Use Only One Garment
7RXVHRQORQHJDUPHQWFRQQHFWDVLQJOHJDUPHQWWR HLWKHUFRQQHFWRU NOTE The snap-lock connectors on the pump are self-sealing and do not require unused garments to be attached.
5
4. Pre-Use Check %HIRUHSRZHULQJRQWKHFlowtron([FHOVVWHPHQVXUH WKDW 7KHSUHVVXUHFRQWURONQREKDVEHHQVHWWRWKHPLG SRVLWLRQPDUNHGPP+J
*DUPHQWVKDYHEHHQDSSOLHGWRWKHSDWLHQW¶VOHJV FRUUHFWOFORVHILWWLQJDQGZLWKRXWFUHDVHVRU IROGV 7KHUHDUHQRNLQNVLQWKHWXELQJ 7KHSXPSLVFRQQHFWHGWRWKHPDLQVSRZHUVXSSO EXWQRWVZLWFKHGRQ $OOWXELQJFRQQHFWLRQVDUHVHFXUH 7KHVVWHPKDVEHHQDUUDQJHGVRWKDWWKHSRZHU FDEOHDQGJDUPHQWWXELQJGRQRWSRVHDWULSRU VWUDQJXODWLRQKD]DUG
6
5. Operation Start Up
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ingle Leg EXWWRQ WRSUHYHQWWKHDODUPVVWHPIURPLQGLFDWLQJD IDXOW7KHVVWHPUHVSRQGVZLWKDµEHHS¶DQGWKHHOORZ /('RQWKH)URQW3DQHOLOOXPLQDWHV NOTE If the Single Leg button is pressed while two garments are connected, the system will automatically reset to two-garment operation after two single-garment inflation cycles.
9HULIWKDWWKHSUHVVXUHGLVSODLVLQGLFDWLQJWKHGHVLUHG RXWSXWSUHVVXUHSUHVFULEHGEWKHSKVLFLDQ5HIHUWR VHFWLRQ³3UHVVXUH$GMXVWPHQW´RQSDJH IRUVSHFLILF SUHVVXUHVHWWLQJLQVWUXFWLRQV NOTE If the operation or performance of the pump changes during use, refer to “Troubleshooting - Alarm Condition” on page 11 of this IFU before calling a service engineer or contacting your local Arjo sales office. NOTE Loss of mains power will halt therapy.
Shut Down
7XUQWKHSRZHUVZLWFKWRWKHRII O SRVLWLRQ7XUQLQJWKH SRZHURIIZLOOVWRSWKHSDWLHQWWKHUDS NOTE If it is required to completely isolate the pump from the mains power, remove the plug from the mains power socket.
'LVFRQQHFWDQGUHPRYHWKHJDUPHQW V DVUHTXLUHG NOTE Garments are for single patient use only. Do not use the garments on a different patient after treatment.
7
6. Pressure Adjustment 7KHSUHVVXUHFRQWUROPHFKDQLVPLVORFDWHGRQWKHIURQW RIWKHSXPSDQGUDQJHVIURPPP+J7KHSUHVVXUH H[HUWHGEWKHJDUPHQWVRQWKHOHJFDQEHDGMXVWHGE WXUQLQJWKLVNQRE7XUQLQJWKHNQREFORFNZLVHLQFUHDVHV WKHSUHVVXUHFRXQWHUFORFNZLVHGHFUHDVHVWKHSUHVVXUH NOTE The recommended pressure setting is 40 mmHg. Alternatively, use the pressure prescribed by the treating physician.
7KHFlowtron([FHOSXPSSUHVVXUHPRQLWRULQJVVWHP LVLQGHSHQGHQWRIWKHSUHVVXUHFRQWURODQGGHOLYHU VVWHPSURYLGLQJDGGHGUHOLDELOLWDQGVDIHW7KH GLJLWDOGLVSODLQGLFDWHVWKHDFWXDOSUHVVXUHWKDWLV GHOLYHUHGWRWKHJDUPHQWVDQGSURYLGHVLPPHGLDWHDQG FRQWLQXRXVIHHGEDFNUHJDUGLQJSXPSSHUIRUPDQFH 7KHSUHVVXUHGLVSODLVXVHGIRUWKHIROORZLQJIXQFWLRQV Pressure Output Check
$IWHUWXUQLQJWKHSXPSRQFKHFNWKDWWKHSUHVVXUH GLVSODLVVKRZLQJWKHGHVLUHGRXWSXWSUHVVXUHZKHQWKH JDUPHQWVDUHLQIODWHG9LVXDOOUHFKHFNWKHGLVSODDW UHJXODULQWHUYDOV
Adjusting the Output Pressure
,IQHFHVVDUWKHSUHVVXUHFDQEHDGMXVWHGGXULQJWKH DFWLYHLQIODWLRQSHULRGEURWDWLQJWKHSUHVVXUHFRQWURO NQREXQWLOWKHGHVLUHGSUHVVXUHLVGLVSODHG
System Calibration Check
7RFRQILUPWKHFDOLEUDWLRQDFFXUDFRIWKHSUHVVXUH FRQWURODQGGLVSODSHUIRUPWKHIROORZLQJFKHFNHDFK WLPHWKHSXPSLVWXUQHGRQ 'XULQJQRUPDORSHUDWLRQZKHQWKHSUHVVXUHGLVSOD UHDGV WKHSRLQWHURQWKHSUHVVXUHFRQWURONQRE VKRXOGEHORFDWHGZLWKLQWKHPP+JDUFRQWKHIURQW SDQHO ,IWKHGLVSODUHDGVEXWWKHFRQWURONQRELVQRWZLWKLQ WKHPP+JDUFWKHSXPSVKRXOGQRWEHXVHGDQG UHIHUUHGWRVHUYLFHIRUUHFDOLEUDWLRQ
8
7. Decontamination 7KHIROORZLQJSURFHVVHVDUHUHFRPPHQGHGEXWVKRXOGEHDGDSWHGWRFRPSOZLWK WKHORFDORUQDWLRQDOJXLGHOLQHV 'HFRQWDPLQDWLRQRI0HGLFDO'HYLFHV ZKLFKPD DSSOZLWKLQWKH+HDOWKFDUH)DFLOLWRUWKHFRXQWURIXVH,IRXDUHXQFHUWDLQRX VKRXOGVHHNDGYLFHIURPRXUORFDO,QIHFWLRQ&RQWURO6SHFLDOLVW 7KHFlowtron([FHOSXPSVKRXOGEHURXWLQHOGHFRQWDPLQDWHGEHWZHHQSDWLHQWVDQG DWUHJXODULQWHUYDOVZKLOHLQXVHDVLVJRRGSUDFWLFHIRUDOOUHXVDEOHPHGLFDOGHYLFHV WARNING Remove the electrical supply to the pump by disconnecting the mains power cable from the mains power supply before cleaning. Protective clothing should always be worn when carrying out decontamination procedures.
Caution Do not use Phenol-based solutions or abrasive compounds or pads during the decontamination process as these will damage the surface coating. Avoid immersing electrical parts in water during the cleaning process. Do not spray cleaning solutions directly onto the pump. Do not immerse the tubeset in water. To clean
&OHDQDOOH[SRVHGVXUIDFHVDQGUHPRYHDQRUJDQLF GHEULVEZLSLQJZLWKDFORWKPRLVWHQHGZLWKDVLPSOH QHXWUDO GHWHUJHQWDQGZDWHU 'RQRWDOORZZDWHURUFOHDQLQJVROXWLRQVWRFROOHFWRQWKH VXUIDFHRIWKHSXPS
Chemical Disinfection
:HUHFRPPHQGDFKORULQHUHOHDVLQJDJHQWVXFKDV VRGLXPKSRFKORULWHDWDVWUHQJWKRISSP DYDLODEOHFKORULQH WKLVPDYDUIURPSSPWR SSPGHSHQGLQJRQORFDOSROLFDQG FRQWDPLQDWLRQVWDWXV :LSHDOOFOHDQHGVXUIDFHVZLWKWKHVROXWLRQWKHQZLSH XVLQJDFORWKPRLVWHQHGZLWKZDWHUDQGGUWKRURXJKO $OFRKROEDVHGGLVLQIHFWDQWV VWUHQJWK PDEHXVHG DVDQDOWHUQDWLYH (QVXUHWKHSURGXFWLVGUEHIRUHVWRUDJH ,IDQDOWHUQDWLYHGLVLQIHFWDQWLVVHOHFWHGIURPWKHZLGH YDULHWDYDLODEOHZHUHFRPPHQGWKDWVXLWDELOLWIRUXVH LVFRQILUPHGZLWKWKHFKHPLFDOVXSSOLHUSULRUWRXVH Caution
Garments are single patient use and hence cannot be cleaned or reused.
9
8. Routine Maintenance Flowtron Excel System Maintenance Servicing
Service Period
Flowtron Excel Pump General Care, Maintenance and Inspection
Serial Labels
7KHHTXLSPHQWKDVEHHQGHVLJQHGWREHPDLQWHQDQFH IUHHEHWZHHQVHUYLFHSHULRGV $UMRZLOOPDNHDYDLODEOHRQUHTXHVWVHUYLFHPDQXDOV FRPSRQHQWSDUWVOLVWVDQGRWKHULQIRUPDWLRQQHFHVVDU IRU$UMRWUDLQHGSHUVRQQHOWRUHSDLUWKHVVWHP $UMRUHFRPPHQGVWKDWWKHFlowtron([FHO SXPSLV VHUYLFHGHYHUPRQWKVEDQ$UMRDXWKRULVHGVHUYLFH DJHQW &KHFNDOOHOHFWULFDOFRQQHFWLRQVDQGSRZHUFDEOHIRU VLJQVRIH[FHVVLYHZHDU &KHFNWKHWXEHVHWDQGFRQQHFWRUVIRUDQGDPDJH ,QWKHHYHQWRIWKHSXPSEHLQJVXEMHFWHGWRDEQRUPDO WUHDWPHQWHJLPPHUVHGLQZDWHURUGURSSHGWKHXQLW PXVWEHUHWXUQHGWRDQDXWKRULVHGVHUYLFHFHQWUH 7KHVHULDOQXPEHUIRUWKHSXPSLVRQWKHODEHORQWKH EDFNRIWKHSXPSFDVH4XRWHWKLVVHULDOQXPEHUZKHQ UHTXHVWLQJVHUYLFH
10
9. Troubleshooting - Alarm Condition 7KHFlowtron([FHOVVWHPIHDWXUHVDQDXGLEOHDQGYLVXDODODUP,IDSUREOHP RFFXUVWKHVVWHPZLOOVHQVHWKHIDXOWDQGEULHIOIODVKDPHVVDJHRQWKHIURQWSDQHO SUHVVXUHGLVSOD ,IWKHVDPHIDXOWFRQWLQXHVIRUVXFFHVVLYHLQIODWLRQVWKHDXGLEOHDODUPZLOOVRXQG DQGDIODVKLQJPHVVDJHZLOOUHPDLQRQWKHSUHVVXUHGLVSODXQWLOFRUUHFWLYHDFWLRQLV FRPSOHWHG$OOWKHDODUPFRQGLWLRQVDUHORZSULRULWLQFRPSOLDQFHZLWK ,(& 7KHH[FHSWLRQWRWKLVLVDQ IDXOWZKLFKZLOODODUPLPPHGLDWHO Alarm Condition Display
Problem
Corrective Action
1. Hose disconnected at garment. 2. Garment leak. 3. Low pressure. 1. Hose kinked causing a blocked tube. 2. Hose disconnected at pump. 3. Single garment attached without pressing ‘single leg’ button. Pump failure.
Alarm Cancel
1. Check the hose connection at garment end. 2. Check garment and replace if faulty. 3. Call the service engineer. 1. Check hoses for kinks or obstructions. 2. Check the hose connection at pump outlet. 3. Press ‘single leg’ button, if only one garment to be used. DO NOT USE PUMP. Call the service engineer.
$IWHUDIDXOWKDVEHHQFRUUHFWHGWKHDODUPFDQEH FDQFHOOHGEWZRPHWKRGV 6ZLWFKWKHSXPSRIIWKHQRQDJDLQ $OORZWKHSXPSWRUXQXQWLOLWVHQVHVDQRUPDO LQIODWLRQLWZLOOWKHQUHVHWLWVHOI &RQWLQXHWRZDWFKWKHGLVSODIRUDSSUR[LPDWHORQH PLQXWHDIWHUUHVHW,IQRIODVKLQJPHVVDJHVUHDSSHDUWKH IDXOWKDVEHHQFOHDUHG
11
Troubleshooting Display
Problem
No displays, no indications, no operation
1. Power failure.
Yellow LED on the top panel flashes (approximately 4 times per second)
Internal electronic fault
Corrective Action 1. Check mains power supply. Check power cable. 2. Call the service engineer.
2. Fuse blown.
DO NOT USE PUMP. Call the service engineer.
If the troubleshooting procedures do not return the system to normal performance, stop using the system immediately and call the service engineer.
12
10. Accessories GARMENTS AND TUBING Description
Garment Part No.
Size
Standard Calf Garment
DVT10
Circumference
up to 43cm (17”)
Calf Garment (Sterile)
DVT10S
Circumference
up to 43cm (17”)
Large Calf Garment
DVT20
Circumference
up to 58cm (23”)
Standard Thigh Garment
DVT30
Circumference
up to 71cm (28”)
Thigh Garment (Sterile)
DVT30S
Circumference
up to 71cm (28”)
Large Thigh Garment
DVT40
Circumference
up to 89cm (35”)
Extra Large Calf Garment
DVT60L
Circumference
up to 81cm (32”)
Calf Garment
L501-M
Circumference
up to 43cm (17”)
Thigh Garment
L503-M
Circumference
up to 71cm (28”)
Connector Tubing
L550 (SP057)
Length
150 cm (60”)
Connector Tubing
L552 (SP061)
Length
300 cm (118”)
Small Calf Garment
DVT5
Circumference
up to 36 cm (14”)
13
11. Technical Specification 7DEOH Pump Supply Voltage and Frequency:
230V / 50 Hz, 230V / 60 Hz, 220V / 60 Hz, 120V / 60 Hz, 100V / 50 Hz 100V / 60 Hz
Power Input:
35 VA MAX
Size:
133 x 152 x 275 mm
Weight:
2.7 kg
Case Material:
Fire Retardant ABS Plastic
Plug Fuse Rating:
5A BS1362 (UK ONLY)
Pump Fuse Rating:
T1AL 250 V
Degree of protection against electric shock:
Class II, Double Insulated with Functional Earth Type BF
Degree of protection against liquid ingress:
IPX0 - No protection
Mode of operation:
Continuous
Cycle or Therapy Modes:
60 seconds total 12 seconds inflation 48 seconds deflation
Pressure Range:
30 - 60 mmHg (±10 mmHg)
ENVIRONMENTAL INFORMATION Condition
Temperature Range
Relative Humidity
Atmospheric Pressure
Operating
+10 °C to +40 °C (+50 °F to +104 °F)
30% to 75% (non-condensing)
700 hPa to 1060 hPa
Storage and Transport (Long Term)
+10 °C to +40 °C (+50 °F to +104 °F)
20% to 95% (non-condensing)
700 hPa to 1060 hPa
Storage and Transport (Short Term)
-20 °C to +50 °C (-4 °F to +122 °F)
20% to 95% (non-condensing)
500 hPa to 1060 hPa
Products should be well packaged and stored in a well-ventilated, dry and non-corrosive environment.
NOTE If this pump unit is subjected to extreme cold during transit or storage, initial performance may be impaired. Connect to power supply and leave running for approx. 30 minutes before use.
14
Symbols The operator must read this document (Instructions for Use) before use. Note: This symbol is blue on the product label.
E348583
CAN/CSA-C22.2 No 60601-1 (2008)+(2014) and ANSI/AAMI ES 60601-1(2005) +AMD(2012)
i
With respect to electric shock, fire and mechanical hazards only in accordance with CAN/CSA-C22.2 No. 60601-1 (2008) + (2014) and ANSI/AAMI ES 60601-1 (2005) + AMD (2012) MEDICAL EQUIPMENT
O (Off)
I
Power: Disconnects from the mains supply
(On)
Power: Connects to the mains supply
Refer to this document (Instructions for Use) for a description of the product classification (3rd Edition).
SN
Serial Number
Refer to this document (Instructions for Use) for a description of the product classification (2nd Edition).
REF
Model number
Manufacturer: This symbol is accompanied by the name and the address of the manufacturer.
Type BF
CE marking indicating conformity with European Community harmonised legislation Figures indicate Notified Body supervision.
Fuse
Indicates the product is a Medical Device according to EU Medical Device Regulation 2017/745
Double Insulated
Do not dispose of in domestic refuse
15