ARJO Huntleigh Healthcare
Huntleigh Nimbus 3 and Nimbus 3 Professional User Manual
User Manual
40 Pages
Preview
Page 1
Nimbus® 3 and Nimbus® 3 Professional
User Manual
Pressure Area Care Products Division
Contents
General Safety... iii Introduction... 1 About this Manual... 1 About Nimbus 3 and Nimbus 3 Professional... 1 Nimbus 3 Pump... 2 Nimbus 3 Mattress... 3 Nimbus 3 Professional Mattress... 4 Clinical Applications... 6 Indications... 6 Contra-Indications... 6 Patient In Chair... 6 Installation... 7 Preparing the Nimbus 3 and Nimbus 3 Professional Systems for Use... 7 Installing the Nimbus 3 or Nimbus 3 Professional Mattress... 7 Installing the Pump... 9 Connecting the Tubeset... 10 Disconnecting the Tubeset... 10 System Operation... 10 Controls, Alarms and Indicators... Pump Controls... Pump Indicators... Mattress Controls... Additional Controls on the Nimbus 3 Professional Mattress...
11 11 12 13 14
Operation... 15 Installing the System... 15 Inflating the Mattress... 15 Testing the Power Fail Alarm... 15 Deflating the Mattress... 16 System Optimisation... 17 Selecting the Operating Mode... 17 Silencing Audible Alarms... 18 Comfort Control... 18 Transport Control... 18 CPR Control... 19 Patient Positioning Guidance for the Nimbus 3 Professional Mattress... 20 Decontamination... During Use... To Clean... To Disinfect... To Launder the Mattress Top Cover...
22 22 22 22 23
Routine Maintenance... 24
(i)
Nimbus 3 and Nimbus 3 Professional Systems... 24 Nimbus 3 Pump... 24 Nimbus 3 and Nimbus 3 Professional Mattresses... 24 Serial Number Labels... 25 Troubleshooting... 26 Fuse Replacement... 27 Warranty and Service... 28 Technical Description... 29 Cover Options and Features... 32
(ii)
General Safety Before you connect the system pump to a mains socket, read carefully all the installation instructions in Section 3, Page 7 “Installation”. The system has been designed to comply with regulatory safety standards including: EN60601-1:1990/A13:1996. Safety Warnings • The cover of this product is vapour permeable but not air permeable and may present a suffocation risk. It is the responsibility of the care giver to ensure that the user can use this product safely. • Electrical equipment may be hazardous if misused. The pump's rear case should only be removed by authorised technical personnel. • Do not use the pump in the presence of flammable gases such as anaesthetic agents. • Whilst the patient is unattended, safety sides should be used based on clinical assessment and in line with local hospital policy. • Alignment of the bed frame, safety sides and the system should leave no gap wide enough to entrap a patient's head or body. Care should be exercised to prevent occurrence of gaps by compression or movement of the mattress. Death or serious injury may occur. • Due to the inherently lower flame retardancy of the high performance eVENT® a fabric, it is NOT suitable for use in the homecare environment.
a. eVENT® is a registered trademark of BHA Technologies Inc.
(iii)
Precautions For your own safety and the safety of the equipment, always take the following precautions: • Placing extra layers between the patient and the mattress potentially reduces the benefits provided by the mattress and should be avoided or kept to a minimum. As part of sensible pressure area care, it is advisable to avoid wearing clothing which may cause areas of localised high pressure due to creases, seams, etc. Placing objects in pockets should be avoided for the same reason. • Keep the pump away from sources of liquids and do not immerse in water. • Do not expose the system, especially the mattress, to naked flames, such as cigarettes, etc. • Do not store the system in direct sunlight. • Switch off the mains power supply to the pump, by disconnecting the pump from the mains socket, before cleaning and inspecting. • Do not use phenol-based solutions to clean the system. • Make sure the system is clean and dry prior to use or storage. • Never use sharp objects or electrically heated under blankets on or under the system. • Store the pump and mattress in the protective bags supplied. • Only the pump and mattress combination as indicated by Huntleigh Healthcare should be used. The correct function of the product cannot be guaranteed if incorrect pump and mattress combinations are used.
(iv)
1. Introduction About this Manual
This manual is your introduction to the Nimbus® 3 and Nimbus 3 Professional systems. Use it to initially set up the system, and keep it as a reference for day-to-day routines and as a guide to maintenance.
About Nimbus 3 and Nimbus 3 Professional Nimbus 3 and Nimbus 3 Professional are Dynamic
Flotation Systems for the prevention, treatment and management of pressure ulcers. Nimbus 3 and Nimbus 3 Professional systems comprise
a pump and mattress replacement which can be used on standard hospital and normal domestic beds. Beds can be adjusted or profiled with the mattress in position. The Nimbus 3 Professional mattress has the following additional features to enable the patient to be proned, and to assist with pressure area and patient care management:
• A Head Section Deflate Control to allow the three head cells to be fully deflated. • Individual Vent Valves to allow 16 of the 20 cells to be independently deflated. The Nimbus 3 and Nimbus 3 Professional mattresses incorporate an advanced AutoMatt® sensor pad which makes sure that the patient is automatically supported at optimum pressures regardless of size, height, position or weight distribution. If cardiac arrest occurs, the Nimbus 3 and Nimbus 3 Professional mattresses can be deflated in less than 10 seconds to allow cardiac resuscitation procedures to be performed.
1
Nimbus 3 Pump
The same pump is used on the Nimbus 3 and Nimbus 3 Professional systems. The pump comprises a moulded case with non-slip feet on the base and rear, and an integral carry handle. Carry Handle
Mains Power Switch
Alarm Indicators
Tubeset
Mains Power Cord
Front Panel
The pump has two modes of operation: • Dynamic mode that cycles the support surface beneath the patient every 10 minutes providing periods of pressure relief for the whole body. • Static mode where the support surface remains constant (all cells equally inflated). The controls and indicators are located on the front panel, and a sophisticated alarm system differentiates between normal operation and genuine system faults. If an alarm situation is detected a flashing indicator will illuminate, together with an indication of the cause of alarm, and an audible warning will sound. The pump can be fixed to the foot end of a hospital bed by the separate bed bracket. The bed bracket fits in the pump handle and then clips onto most common bed frames. The pump can also be stood on the floor, either upright or on its rear cover.
2
Nimbus 3 Mattress
The Nimbus 3 mattress comprises the following components:
Detachable Cover 3 Head Cells 4 Thigh Cells Head End
5 “Heelguard” Cells
Drag Handle
8 Torso Cells
Securing Strap
Foot End
Carry Handle
CPR Control Pump Tubeset
Transport Control
Detachable Cover
The standard protective cover comprises a 2-way stretch cover zipped to a durable anti-slip base. The zips are protected by flaps to prevent ingress of contaminants, and allow easy removal of the cover for cleaning. Alternative covers with advanced properties, such as Advantex® and eVENT®, are also available (Refer to “Cover Options and Features” on page 32).
Cells
The Nimbus 3 mattress comprises 20 polyurethane (PU) cells providing support to the user in either Alternating or Static modes. The cells are grouped in four sections, each of which has a specific function: • The three Head cells remain at a constant pressure for pillow stability and patient comfort. • The eight Torso cells combine alternating and static pressure characteristics to support patients fully in both lying and sitting positions without the risk of ‘bottoming’. • The four Thigh cells cycle dynamically to maximise pressure relief. • The five Heelguard® cells are specially powered to maximise the pressure relief under the heels. 3
AutoMatt
The advanced AutoMatt sensor pad is under the cells, and makes sure that the patient is automatically supported at optimum pressures regardless of size, height, position or weight distribution.
CPR Control
The CPR (Cardio-Pulmonary Resuscitation) Control is at the foot end of the mattress, and allows the air to be evacuated in under 10 seconds.
Transport Control
The Transport Control is next to the CPR Control. When operated, it seals the mattress so that air is not exhausted when the tubeset is disconnected and also creates an even pressure in all the cells.
Tubeset
The tubeset incorporates a flexible, compact anti-kink tube that is resistant to crushing and any subsequent obstruction of air flow. Each end has a quick-lock system for easily connecting and disconnecting the air supply at the pump and mattress.
Nimbus 3 Professional Mattress The Nimbus 3 Professional mattress is of similar construction to the Nimbus 3 mattress, with the addition of a Head Section Deflate Control, individual Vent Valves on 16 of the 20 cells and a Shoulder Support Cell. Detachable Cover
5 “Heelguard” Cells
CPR/Transport Control
7 Torso Cells Shoulder Support Cell (with Central Cutout)
Foot End
Drag Handle
4 Thigh Cells 16 Vent Valves
Head End 3 Head Section Cells (“Power-Down” Cells)
Carry Handle Securing Strap
4
Head Section Deflate Control
Head Section Deflate Control
This is a two-position rotary-action control at the head end of the mattress: • Dynamic (Normal) Mode. The three cells in the Head Section are inflated at a constant pressure and the remaining 17 cells alternate. • TriCell Head Section Deflate. The three cells in the Head Section are fully deflated to assist with patient care management, and the Shoulder Support Cell (the fourth cell, next to the Head Section) is inflated to a constant pressure to support the patient’s shoulders. The remaining 16 cells alternate.
Cells
The Nimbus 3 Professional mattress has the same number of cells as the Nimbus 3 mattress (20 cells). The function of the first four cells at the head end of the mattress is different on the Nimbus 3 Professional: • The three cells in the Head Section are either fully inflated or fully deflated, depending on the position of the Head Deflate Control, to assist with patient care management. The cells are specially powered to enable them to be fully deflated. • The single Shoulder Support Cell (the fourth cell, next to the Head Section) has a shallow cutout in the mid-section of the cell. This is to allow access to the neck area for clinical procedures and to ensure the smooth, uniform extension of the neck during deflation. Its operation is controlled by the Head Section Deflate Control: the cell is either fully inflated to support the patient’s shoulders or alternates (together with the remaining 16 cells). • The remaining 16 cells (seven Torso cells, four Thigh cells and five Heelguard cells) have the same basic function as on the Nimbus 3 mattress.
Vent Valves
The seven Torso cells, four Thigh cells and five Heelguard cells have individual Vent Valves to allow each cell to be independently deflated, to assist with pressure area and patient care management.
5
2. Clinical Applications Indications
The Nimbus 3 and Nimbus 3 Professional systems are indicated for patients weighing up to 250 kg (39 stones) and are suitable for the prevention and management of all grades of pressure ulcers.
Contra-Indications
The Nimbus 3 and Nimbus 3 Professional systems, both in alternating and static modes, should not be used for patients with unstable spinal fractures. In the case of patients with other unstable fractures, where a moving surface can be harmful, advice should be obtained from the appropriate physician before using the Nimbus 3 or Nimbus 3 Professional systems in both alternating and static modes.
Patient In Chair
If the patient will be sitting in a chair for any period of time, it is strongly recommended that a pressure reducing or relieving seat cushion is used.
The Nimbus 3 and Nimbus 3 Professional systems are an aid to the prevention and management of pressure ulcers. If there is no improvement in the patient's condition, clinical advice should be sought.
The above are guidelines only and should not replace clinical judgement or experience.
6
3. Installation The Nimbus 3 and Nimbus 3 Professional systems are very simple to install using the following guidelines.
Refer to Section 4, Page 11 “Controls, Alarms and Indicators” for a comprehensive description of the controls and indicators on the pump and mattress.
Preparing the Nimbus 3 and Nimbus 3 Professional Systems for Use 1. Remove the system from the packaging. You should have the following items: • Nimbus 3 pump, with integral mains power cord. • Nimbus 3 mattress replacement or the Nimbus 3 Professional mattress replacement. • Bed bracket. • Tubeset. Installing the Nimbus 3 or Nimbus 3 Professional Mattress
1. Remove the conventional mattress from the bed frame and check that there are no protruding bed springs or sharp objects on the bed frame surface. Heavily ridged bed baseboards may require special considerations for correct system operation - consult your Huntleigh Healthcare representative. 2. Unroll the mattress onto the bed base and make sure that the CPR is at the foot end, and the CPR label is hanging freely. 3. Attach the mattress to the bed frame using the hook and loop securing straps.
If the bed can be profiled to any position (i.e. raised or lowered), attach the mattress to the movable parts of the bed only.
7
4. For Nimbus 3 mattresses only, check the AutoMatt sensor pad, as follows: • Unzip the cover on one side of the mattress only. • Pull the side of the mattress away from the cells. • The AutoMatt sensor pad is situated under the cells between the soft and hard foam sheets. • Make sure that the AutoMatt sensor pad is lying flat and is not “kinked”. • Zip the cover back onto the mattress, taking care not to trap any cell material in the zip. Cells
Cover
Head End
Foot End
Soft Foam Sheet AutoMatt Sensor Pad
Hard Foam Sheet Hard Foam Sheet
AutoMatt Sensor Pad
For Nimbus 3 Professional mattresses, the AutoMatt is encapsulated and does not need to be checked. 5. Leave the ends of the mattress cover free when profiling the bed. 6. Make sure the CPR control is closed and locked in position and the Transport control is set to NORMAL. NORMAL TRANSPORT
CPR Control
8
Transport Control
Additional Checks on the Nimbus 3 Professional Mattress
1. Make sure that all 16 Vent Valves are closed.
Closed
Open
2. Make sure that the Head Section Deflate Control is set to Dynamic (Normal) Mode. TriCell Head Section Deflate
Dynamic (Normal) Mode
Installing the Pump
1. If the pump is to be hung from the end of the bed, make sure that the bed bracket is securely attached to the pump, and then attach the pump and bed bracket to the bed frame.
2. Alternatively the pump can be placed underneath the bed, either upright or lying on its back. 3. Insert the connector on the end of the mains power cord into a suitable mains power outlet.
9
Connecting the Tubeset
To connect the tubeset to the mattress and pump: 1. Locate the bottom of the tubeset connector onto the bottom of the pump/mattress connector. 2. Pull the top of the tubeset connector up and over the top of the pump/mattress connector, until the tubeset connector “clicks” into position. 3. Make sure both connections are secure. 2
1
Disconnecting the Tubeset
To disconnect the tubeset from the mattress and pump: 1. Move the tubeset connector down by pulling the tubeset extrusion downwards, and then pull the bottom of the tubeset connector away from the bottom of the pump/mattress connector. 2. Lift the top of the tubeset connector off the top of the pump/mattress connector.
2 1
System Operation
The system is now ready for use. Refer to Section 4, Page 11 “Controls, Alarms and Indicators” and Section 5, Page 15 “Operation” for day-to-day operating instructions.
WARNING Make sure the mains power cord and tubeset are positioned to avoid causing a hazard. Caution Make sure the mains power cord and tubeset are clear of moving bed mechanisms or other possible entrapment areas. 10
4. Controls, Alarms and Indicators
Power Switch & Alarm Reset
Mute Control & Indicator
On/Reset Alarm Indicator Static Control & Indicator
Alarm Indicator
High Pressure Indicator
Low Pressure Indicator
Wait Indicator
Service Indicator
Comfort Control
Power Fail Indicator
Pump Fault Indicator
Pump Controls
The pump front panel has the following controls:
POWER Switch (and RESET ALARM)
Switches the mains power to the pump on and off. The green ON / RESET ALARM indicator is illuminated when the mains power is connected and the pump switched on. The switch is also used to reset the pump after an alarm condition has been detected.
STATIC Mode
Selects the operating mode, either Static or Dynamic. Static mode is confirmed when the yellow indicator on the button is illuminated. When Dynamic mode (default) is selected the yellow indicator will be extinguished.
Alarm MUTE
COMFORT CONTROL
An audible alarm mute is provided to cancel warning sounds during an alarm condition.
This is a rotary action control to set the relative firmness/ softness of the mattress for patient comfort.
11
Pump Indicators
The pump front panel has the following indicators:
ON / RESET ALARM
The green ON / RESET ALARM indicator below the POWER switch is illuminated when the mains power is connected and the pump switched on.
STATIC Mode
The indicator on the STATIC button is illuminated when Static mode has been selected for operation.
Alarm MUTE
The indicator on the MUTE button is illuminated when an audible alarm has been silenced.
The indicator will NOT be illuminated when a Power Fail alarm is muted. WAIT
The WAIT indicator is illuminated when the mattress is being inflated. The indicator will remain illuminated until the mattress has been fully inflated. This may take up to 15 minutes.
HIGH PRESSURE
The HIGH PRESSURE indicator is illuminated whenever the pump detects high pressure within the mattress. If this condition occurs, the air supply from the pump is switched off until normal pressure is detected. After 2 seconds of normal pressure being detected the indicator is switched off and the air supply restarted.
LOW PRESSURE
The LOW PRESSURE indicator is illuminated whenever the pump detects low pressure within the mattress. This may indicate that there is insufficient pressure to support a patient or that the Transport control is turned to the TRANSPORT position whilst the pump is on and connected to the mattress. The LOW PRESSURE indicator will be switched off once normal pressure is reached.
Alarm
The pump unit incorporates a sophisticated alarm detection system that differentiates between patient movement and genuine alarm conditions. Whenever an alarm condition is detected the red Alarm triangle starts flashing together with an indicator of the cause of the alarm. Additionally, an audible warning will sound, which can be cancelled by pressing the Alarm MUTE button (Refer to “Alarm MUTE” on page 11). The triangular Alarm symbol is displayed with one or more of the following indicators:
12
• LOW PRESSURE (Refer to “LOW PRESSURE” on
page 12).
• HIGH PRESSURE (Refer to “HIGH PRESSURE” on
page 12).
• PUMP FAULT (Refer to “PUMP FAULT” on
page 13). • POWER (Refer to “POWER Fail” on page 13). For all alarm conditions except Power Fail, once the alarm condition has been detected and displayed, it can only be cancelled by switching the pump unit off and then back on. Refer to Section 8, Page 26 “Troubleshooting” for possible causes of the above alarm conditions. PUMP FAULT
The PUMP FAULT indicator is illuminated when an internal pump malfunction is detected. The fault can only be rectified by carrying out a service on the pump.
POWER Fail
The POWER indicator will flash when a mains power failure has been detected. The alarm will continue until the mains power is resumed or the pump is switched off using the POWER switch on the pump control panel.
Service Indicator
The “telephone” symbol will be illuminated after a set number of running hours to indicate that the pump is ready for a service. This service period is set to 12 months.
The pump will continue to operate normally even when the illuminated.
Mattress Controls
symbol is
All Nimbus 3 and Nimbus 3 Professional mattresses have the following two controls, situated at the foot end of the mattress: NORMAL TRANSPORT
CPR Control
13
Transport Control
Transport Control
This sets the mattress into TRANSPORT mode where the support surface is equally pressurised and the pump and tubeset can be removed. In this mode the mattress will support the patient for up to 12 hours.
CPR Control
The CPR (Cardio-Pulmonary Resuscitation) Control provides a means of rapidly deflating the mattress to allow normal resuscitation procedures to be carried out.
The CPR control is used to deflate the mattress for packing and storage.
Additional Controls on the Nimbus 3 Professional Mattress The following two controls are on the opposite side of the mattress to the CPR/Transport Control: Head Section Deflate Control
This is a two-position rotary-action control at the head end of the mattress: • Dynamic (Normal) Mode. The three cells in the Head Section are inflated at a constant pressure and the remaining 17 cells alternate. • TriCell Head Section Deflate. The three cells in the Head Section are fully deflated to assist with patient care management, and the Shoulder Support Cell (next to the Head Section) is inflated to a constant pressure to support the patient’s shoulders. The remaining 16 cells alternate. TriCell Head Section Deflate
Dynamic (Normal) Mode 16 Vent Valves
The seven Torso cells, four Thigh cells and five Heelguard cells have individual Vent Valves to allow each cell to be independently deflated, to assist with pressure area and patient care management.
Closed
14
Open
5. Operation These instructions cover day-to-day operation of the system. Other operations, such as maintenance and repair, should only be carried out by suitably qualified personnel.
Refer to Section 4, Page 11 “Controls, Alarms and Indicators” for a comprehensive description of the controls and indicators on the pump and mattress.
Installing the System
Before using the Nimbus 3 or Nimbus 3 Professional system make sure: 1. The system has been installed correctly in accordance with Section 3, Page 7 “Installation”. 2. The CPR unit on the mattress is closed and locked in position. 3. The Transport control on the mattress is set to NORMAL.
4. If a Nimbus 3 Professional system is being installed, make sure that on the mattress: • All 16 Vent Valves are closed. • The Head Section Deflate Control is set to Dynamic (Normal) Mode.
Inflating the Mattress
1. Switch the pump POWER switch to ON. The ON / RESET ALARM indicator below the POWER switch should illuminate. 2. The pump will now run a self test for approximately 3 seconds when all the indicators on the front panel will be illuminated. 3. If the pump detects low pressure (e.g. a deflated mattress) it will enter an inflation sequence with the LOW PRESSURE and WAIT indicators illuminated. 4. Once normal operating pressure has been reached both the LOW PRESSURE and WAIT lights will extinguish. It may take up to 15 minutes to inflate the mattress.
The three Head Section cells and the five Heelguard cells will inflate more slowly than the rest of the mattress. Testing the Power Fail Alarm
The Power Fail Alarm is powered by a rechargeable battery. The duration of the alarm will depend on the level of charge in the battery. 15