Instructions for Use
36 Pages
Preview
Page 1
FLOWTRON Instructions For Use
Contents General Safety... iii Introduction... 1 About this Manual... 1 About Flowtron... 1 Pump: Front View... 2 Pump: Rear View... 2 Clinical Applications... 3 Indications... 3 Contraindications... 3 Cautions... 3 Guidelines and Recommendations... 4 Controls, Alarms and Indicators... 5 Control Panel With Typical LCD Screen View in Run Mode... 5 LED Indicators on the Front Case... 7 System Operation... 8 General... 8 Start Up... 8 Standby Screens... 9 Starting Therapy... 10 Stopping Therapy... 13 Switching Off the Pump... 13 Alarms... 13 Changing Options... 14 Settings Adjustment... 17 Decontamination... 18 Maintenance and Service... Flowtron System... Flowtron Pump... Serial Labels...
19 19 19 19
Troubleshooting... 20 Garment Information... 26 Technical Description... 27
(i)
(ii)
GENERAL SAFETY Before you connect the system pump to a mains socket, read carefully all the installation instructions contained within this manual. The system has been designed to comply with regulatory safety standards including: • EN60601-1:1990/A13:1996 and IEC 60601-1:1988/A2:1995. • UL60601-1 and CAN/CSA C22.2 No. 601.1-M90.
Safety Warnings • It is the responsibility of the care giver to ensure that the user can use this product safely. • Make sure that the mains power cable and tubeset or air hoses are positioned to avoid causing a trip or other hazard, and are clear of moving bed mechanisms or other possible entrapment areas. • Electrical equipment may be hazardous if misused. There are no user-serviceable parts inside the pump. The pump's case must only be removed by authorised technical personnel. No modification of this equipment is allowed. • The mains power socket/plug must be accessible at all times. To disconnect the pump completely from the electricity supply, remove the plug from the mains power socket. • Disconnect the pump from the mains power socket before cleaning and inspecting. • Keep the pump away from sources of liquids and do not immerse in water. • Do not use the pump in the presence of uncontained flammable liquids or gasses. • Only the pump and garment/insert combination as indicated by ArjoHuntleigh should be used. The correct function of the product cannot be guaranteed if incorrect pump and garment combinations are used. • The Flowtron system is NOT intended for use in the Home Healthcare Environment (e.g. private dwellings or nursing homes). Caution • USA Federal law restricts this device to sale by or on the order of a physician. Precautions For your own safety and the safety of the equipment, always take the following precautions: • Do not expose the system to open flames, such as cigarettes, etc. • Do not store the system in direct sunlight. • Do not use phenol-based solutions to clean the system. • Make sure the system is clean and dry prior to use or storage. Electromagnetic Compatibility (EMC) This product complies with the requirements of applicable EMC Standards. Medical electrical equipment needs special precautions regarding EMC and needs to be installed in accordance with the following instructions: • The use of accessories not specified by the manufacturer may result in increased emissions by, or decreased immunity of, the equipment, affecting its performance. • Portable and mobile radio frequency (RF) communications equipment (e.g. mobile/cell phones) can affect medical electrical equipment. • If this equipment needs to be used adjacent to other electrical equipment, normal operation must be checked before use. • For detailed EMC information contact ArjoHuntleigh service personnel.
(iii)
Environmental Protection Incorrect disposal of this equipment and its component parts, particularly batteries or other electrical components, may produce substances that are hazardous to the environment. To minimise these hazards, contact ArjoHuntleigh for information on correct disposal. Design Policy and Copyright ® and ™ are trademarks belonging to the ArjoHuntleigh group of companies. As our policy is one of continuous improvement, we reserve the right to modify designs without prior notice. © ArjoHuntleigh 2010.
(iv)
1. Introduction About this Manual
This manual is your introduction to the Flowtron system. Use it to initially set up the system and keep it as a reference for day-to-day routines and as a guide to maintenance.
About Flowtron
The Flowtron system is a non-invasive external intermittent pneumatic compression system for reducing the incidence of Deep Vein Thrombosis (DVT). The application of external pneumatic compression has two effects: • Augments venous blood flow velocity, thereby reducing stasis. • Enhances fibrinolytic activity to reduce the risk of early clot formation. The system comprises a pump that can be used in conjunction with an extensive range of ArjoHuntleigh inflatable garments, including foot, calf and calf & thigh versions. The pump automatically adjusts to the correct therapy profile depending upon which garment type (foot, calf or calf & thigh) is connected. The tubeset is integral to the system and cannot be disconnected from the pump. The mains power supply is the primary power source for the pump. The pump incorporates an internal battery pack, which is a secondary power source to back up the pump in the event of failure or disconnection (accidentally or deliberately) from the mains power supply. The Flowtron system is intended for use ONLY in Professional Healthcare Facilities (e.g. hospitals or physicians’ offices).
1
Pump: Front View LCD Screen and Controls
Integral Carry Handle Integral Tubesets
LED Indicator Tubeset 1 (Blue Button)
Tubeset 2 (Orange Button)
Garment Connectors
Pump: Rear View
Swing-out Bed-Hooks
Tubeset 1
Integral Tubesets Mains Power Cord Tubeset 2
2
2. Clinical Applications Indications
The primary application of the Flowtron system is for the prevention of Deep Vein Thrombosis (DVT). Depending on the garment type used, other clinical applications are also appropriate. The FG foot garment, in particular, has a wide range of clinical applications. Full details for clinical applications are included in the packaging of every garment. The type of garment used on an individual patient must be specified by a physician.
Contraindications
The Flowtron system should NOT be used in the following conditions: 1. Where an increase of fluid to the heart may be detrimental. 2. Severe congestive cardiac failure. 3. Known or suspected acute deep vein thrombosis, thrombophlebitis or pulmonary embolism. 4. On untreated, infected legs. If you are unsure of any contraindications, refer to the patient’s physician before using the Flowtron system.
Cautions
1. Proper garment application and connection to the pump is essential. 2. Garments should be positioned in such a way that they do not create any potential for constant pressure points on the patient’s limb. Additional care should be taken when placing the garments on any deformed leg or foot, or on legs with significant oedema. 3. Garments should be removed immediately if the patient experiences tingling, numbness, or pain, and the physician notified. 4. When used for DVT prevention, continuous external pneumatic compression is recommended until the patient is fully ambulatory. Uninterrupted use of the Flowtron system is encouraged.
3
5. The Flowtron system should be USED WITH CAUTION on patients with: • Insensitive extremities. • Diabetes. • Impaired circulation. • Fragile or impaired skin. These are guidelines only and should not replace clinical judgement and experience. Guidelines and Recommendations General • In line with good clinical practice, garments should be removed regularly to inspect the skin. Recommendations • Where appropriate patients should be instructed in the proper use of the system, the purpose of therapy and that any problems should be reported to the nursing staff. DVT Prophylaxis • The Flowtron system should be applied to the patient pre-operatively, prior to the induction of anaesthesia. • The system should be used continuously for no less than 72 hours post-operatively or until the patient becomes fully ambulatory. • If the garment cannot be applied to the operative limb during surgery, it may be applied to the limb once the patient reaches the recovery unit. • In the non-surgical patient, the system should be initiated immediately the risk of DVT formation is identified.
4
3. Controls, Alarms and Indicators Control Panel With Typical LCD Screen View in Run Mode Foot Therapy Indicator
Calf (or Calf & Thigh) Therapy Indicator
Mains/Power Indicator
Battery Indicator LED Indicator
Pressure Indicator
On/Off Button Stop
2h
LCD Screen
Patient Hours Meter Start/Stop Therapy
“Down” Select Button
On/Off Button and LED Indicator
Start/Stop Button
“Up” Select Button
Press the On/Off button to power up the pump from its unpowered state and put it into Standby (refer to “Standby Screens” on page 9). Connecting the pump to the mains power supply will automatically power up the pump from its unpowered state and put it into Standby, without having to press the On/Off button. The adjacent LED indicator shows the pump status:
LED Colour
Mains Power
Pump Status
Extinguished
Disconnected
Off
Amber
Connected
Off
Amber
Connected or disconnected
Standby
Green
Connected or disconnected
Run
To switch off the pump, press and hold the On/Off button for approximately 2 seconds until the LCD screen goes blank, and then release. If you switch off the pump and then press the On/Off button to power it back up: • If the mains power is disconnected, the pump runs the diagnostic tests and then goes into Standby. • If the mains power remains connected, the pump goes into Standby and does not run diagnostic tests. 5
Start/Stop Button
“Up” and “Down” Select Buttons
LCD Screen
This has the following functions depending on the text or icon displayed above the button: • To start or stop therapy. • To confirm the option selected in the Options menu. When the pump is in Standby, press the Start/Stop button to put the pump into the Run mode and start therapy; the LED indicator adjacent to the On/Off button will change to green and the LEDs on the front case will be illuminated green. To stop the therapy and put the pump into Standby, press and hold the Start/Stop button for approximately 2 seconds until the Standby screen is displayed and then release the button. The LED indicator adjacent to the On/Off button changes to amber and the LEDs on the front case are extinguished. These have the following functions depending on the text or icon displayed above the buttons: • To select the Options menu. • To move up and down the Options menu. • To mute certain audible alarm conditions. This displays the pump operating mode and status: 1. Foot or Calf (or Calf & Thigh) therapy indications. These show what garment type is connected to each tubeset and when each garment is being inflated: Calf or Calf & Thigh Garment
Foot Garment
• All therapy indications are initially shown as grey to indicate they are deflated. • A therapy indication changes to black to show it is inflating. • The therapy indication reverts to grey when it starts to deflate. • The garment connector on the end of each tubeset has a push button which is colour-coded and has a number marked on it: blue “1” or orange “2”. The numbers corresponds with the “1” and “2” on the right side of the LCD screen. Where the tubesets enter the side of the pump, the top tubeset is marked “1”. 2. Pressure indication: this is the inflation pressure applied to the garment, as follows: 6
• Foot Garment: 130 mmHg. • Calf and Calf & Thigh Garments: 40 mmHg. 3. Mains power indications: The pump is connected to the mains power supply. The pump is not connected to the mains power supply, but is powered by the internal battery only.
4. Battery indication: • When the pump is connected to the mains power supply, then: •• If the battery is fully charged, the indicator will be static and show “full”. •• If the battery is not fully charged, the indicator will “scroll” from “empty” to “full” to indicate that the pump is being charged. • When the pump is NOT connected to the mains power supply, then the indicator is static and shows the charge remaining in the battery: Battery Empty
Battery 1/3 Full
Battery 2/3 Full
Battery Full
5. Patient Hours Meter: if this is selected, the total therapy time is shown (in hours). Refer to “Changing Options” on page 14 to select and/or clear the Patient Hours Meter. • This is the pump run time since the Patient Hours Meter was last cleared. • The factory default is to NOT show the Patient Hours Meter on the LCD screen. 6. Warning and alarm indications (refer to “Troubleshooting” on page 20). LED Indicators on the Front Case
There are two sets of LED indicators on the front case of the pump. Their status is as follows:
LED Colours
Pump Status
Warnings/Alarms
Extinguished
Off or Standby
--
Green (Static)
Run
• No fault detected. • Warning only detected.
Red (Flashing)
Run
Fault detected with full audible and visual alarm
7
4. System Operation General
Start Up
These instructions cover the day-to-day operation of the system. Other operations, such as maintenance and repair, should only be carried out by suitably qualified personnel. Refer to “Controls, Alarms and Indicators” on page 5 for a description of the controls, indicators and LCD screen. To switch on the pump, do one of the following: • Connect the pump to the mains power supply. The pump will power up automatically and start running a diagnostic test, and display the following screen:
Progress Bar
• If operating from just the battery (disconnected from the mains power supply), press the On/Off button. The pump will power up and start running the diagnostic test, and display the following screen. Note the “Checking Battery” text and the “Mains power not connected” icon at the top of the screen: Checking Battery
“Mains power not connected” icon
After a few seconds, the following screen is displayed, showing “Battery OK”: Battery OK
If the text at the top of the screen says “Low
Battery” then you must connect mains power before you can start therapy. 8
• At the end of the diagnostic test, the following screen is displayed, showing the “Hours Run” at the top and the software version of the pump at the bottom: Hours Run: 145
XX.XXXX-XXX
The “Hours Run” value displayed on this screen is
for service use only, and is different to the “Patient Hours Meter” displayed during therapy. Refer to “Controls, Alarms and Indicators” on page 5 for the “Patient Hours Meter”.
Standby Screens “Mains power connected” icon
The “No Garments” Standby screen is then displayed: Attach garments
Battery Indication
Options
There are two “leg” icons with no garments attached and the text “Attach garments” at the top of the screen; this indicates that no garments are currently connected to either tubeset. There is no text above the Start/Stop button until at least one garment is connected to a tubeset. Apply the prescribed ArjoHuntleigh garment(s) to the patient by carefully following the instructions included in the garment packaging. Connect the garment(s) to the pump by pushing the garment connector firmly into the pump tubeset connector until it “clicks”. The pump LCD screen will show which garments are connected to each tubeset connector: the garment connector with the blue push button is “1” and the one with the orange push button is “2”. If any pump options (Audio Alarm Volume, Patient Hours Meter or Language) need to be changed, they must be changed before the patient therapy is started. Refer to “Changing Options” on page 14.
9
The following four screens show typical Standby screens with different garment configurations:
Start
Start
Options
Foot Garment Connected to Tubeset 1
Start
Calf (or Calf & Thigh) Garment Connected to Tubeset 1
Options
Start
Two Calf (or Calf & Thigh) Garments Connected
Starting Therapy
Options
Options
Foot and Calf (or Calf & Thigh) Garments Connected
Make sure the garment(s) are fitted correctly to the patient and the pump. Make sure any pump options have been changed if necessary. If the pump is to be positioned under a bed, make sure it is placed horizontally (with the bed-hooks underneath) to prevent damage to the pump. While the pump is in Standby, the LEDs on the front case remain extinguished. Press the button below “Start” to start the therapy. The LEDs on the front case change to static green. It is recommended that the following checks are carried out at the start of, and throughout, the therapy: • Check the LCD screen icons to confirm that the correct type of garment(s) have been connected. • During garment inflation, check the LCD screen to confirm that the correct pressure is being applied and that there are no fault messages displayed. The inflation pressures for the different garments are: • Foot Garment: 130 mmHg. • Calf and Calf & Thigh Garments: 40 mmHg. Initial inflation cycles (up to 10) may not show the correct inflation pressure. This is to be expected. 10
• Check that there are no kinks in the pump tubesets. • Check that the pump tubesets and connectors do not cause the patient any discomfort. • Regularly check that the garments remain correctly fitted to the patient. The pump will inflate and deflate each garment in turn, as follows, starting with garment 1: The following example shows two foot garments fitted to the pump. 1. Both garments are initially deflated. The garment pressure is displayed on the left side of the screen. If the Patient Hours Meter is selected in the “Options” menu (refer to “Changing Options” on page 14), the total elapsed time (in hours) is shown in the bottom right of the screen:
2h
Patient Hours Meter
2. Foot garment 1 is inflated to 130 mmHg, with a ramp up and hold time of 3 seconds. The foot garment indicator is black while it is inflated:
Stop
2h
3. The foot garment is deflated to zero. Both garment indicators are grey:
2h
11
4. Foot garment 2 is then inflated to 130 mmHg, with a ramp up and hold time of 3 seconds. The foot garment indicator is black while it is inflated:
2h
5. The foot garment is deflated to zero. Both garment indicators are grey:
2h
6. This cycle of alternate garment inflations repeats from step 2 (above) until the therapy is stopped. If one or two calf (or calf & thigh) garments are attached to the pump, then each garment is inflated to 40 mmHg, with a ramp up and hold time of 12 seconds. If a foot garment and a calf (or calf & thigh) garment are attached to the pump, then since the inflate and hold times for a foot garment are shorter than for a calf (or calf & thigh) garment, there is a modified inflation sequence: the foot garment is always inflated twice in succession and then the calf (or calf and thigh) garment is inflated, as follows. The first garment to be inflated is always garment 1. 1. Inflate the foot garment to 130 mmHg. 2. Deflate the foot garment. 3. Repeat the inflation of the foot garment to 130 mmHg. 4. Deflate the foot garment. 5. Inflate the calf (or calf & thigh) garment to 40 mmHg. 6. Deflate the calf (or calf & thigh) garment. 7. This cycle of garment inflations repeats from step 1 (above) until the therapy is stopped.
12
Stopping Therapy
To stop the therapy and put the pump into Standby, press and hold the Start/Stop button for approximately 2 seconds until the Standby screen is displayed (refer to “Standby Screens” on page 9) and then release it. The LED indicator by the On/Off button changes to amber and the LEDs on the front case are extinguished. If the pump stays in Standby, then: 1. After 5 minutes the LCD display backlight is dimmed. 2. After a further 15 minutes (20 minutes total) the pump automatically switches off. The LED indicator will remain amber if the pump is connected to the mains power and extinguished if the mains power is disconnected.
Switching Off the Pump
1. Make sure the therapy is stopped and the pump is in Standby (refer to “Stopping Therapy” on page 13). 2. Press and hold the On/Off button for approximately 2 seconds until the LCD screen goes blank and then release it. The LED indicator will remain amber if the pump is connected to the mains power and extinguished if the mains power is disconnected.
Alarms
On detection of a fault condition, the Flowtron pump provides a visual-only warning followed by an audible and visual alarm if the fault is not cleared. 1. The visual-only warning is a fault message on the LCD screen. The LED indications remain unchanged and there is no audible alarm. 2. If the fault is not cleared then the warning is replaced by an audible and visual alarm, which consists of: • A fault message on the LCD screen. • The LED indicators on the front of the pump change to red and start flashing. • An audible alarm will sound which increases in pitch and intensity if the alarm is ignored. The warning and alarm can be cleared by either: • Checking and rectifying the fault on the system, or • Pressing the button below “Stop” to put the pump into Standby. Refer to Section 7, Page 20 “Troubleshooting” for the alarms, their possible causes and their remedies.
13
Changing Options
To change the pump operating options, the pump must be in Standby. If the pump is in Run mode, press and hold the button below “Stop” for 2 seconds to put the pump into Standby. On the Standby screen, press the button below “Options” and the “Options” menu is displayed:
• The “border” around an item indicates the currently selected item. • To select a new menu item, press the buttons below the “up” and “down” arrows. • To confirm the new menu item, press the button below the “tick”. • To exit and return to the Standby screen, select and confirm “Exit” on the menu. 1. To change the Audio Alarm Volume setting: • On the “Options” menu, select and confirm “Audio Alarm Volume” and the “Audio Alarm Volume” screen is displayed: Set Volume High Set Volume Low Back Exit
• The “tick” on the left side of the screen shows that the alarm is currently set to High Volume. • Use the “up” and “down” buttons to select the new volume setting. Press the “tick” button to confirm the new volume. The “tick” now shows the new setting. • To exit and return to the Standby screen, select and confirm “Exit”. • To change another option, select and confirm “Back” and the “Options” menu is displayed.
14
2. The Patient Hours Meter settings. When selected, the Patient Hours Meter shows the total therapy time at the bottom right of the LCD screen during therapy. This is the elapsed time since the Patient Hours Meter was last cleared using these commands. To change the settings: • On the “Options” menu, select and confirm “Patient Hours Meter” and the “Patient Hours Meter” screen is displayed: Show Patient Hours Clear Patient Hours Back Exit Hours Run 00001
• A “tick” to the left of “Show Patient Hours” indicates that the total therapy time is shown in the bottom right of the LCD screen during therapy. • To change this setting, use the “up” and “down” buttons to make sure the “Show Patient Hours” option is selected, and then press the “tick” button to toggle the option on or off. • The text “Hours Run 00001” indicates the current value of the Patient Hours Meter. • To clear the elapsed time showing in the Patient Hours Meter (for example, when therapy is started on a new patient), use the “up” and “down” buttons to select the “Clear Patient Hours” option. • Press the “tick” button TWICE to clear the “Hours Run” to “00000”. The Patient Hours Meter on the LCD screen will show “0h” during therapy, if selected. You must press the “tick” button twice. • To exit and return to the Standby screen, select and confirm “Exit”. • To change another option, select and confirm “Back” and the “Options” menu is displayed.
15