AlphaRelief ® Quick Reference Guide QUICK REFERENCE GUIDE (For more detailed instructions, refer to the User Manual)
ALPHARELIEF Pressure Control Knob Run/Standby Indicator Wait Indicator Low Pressure Alarm
alphaRELIEF
140
120
ON
GHT IN WEI kg
80
®
.
kg
Power Fail Alarm
WAIT LOW PRESSURE +40
POWER FAIL
Suspension Hook
Fuse Holder Run/Standby Switch Setting up instructions on rear
Run Standby Switch/ Switches the pump on and off. An internal battery will activate an Alarm Reset audible and visual alarm if the power supply is interrupted prior to switching off. Examples are mains lead removal and power failure. Once an alarm condition has been resolved, the pump can be reset by switching the pump to standby and back to run again using the Run/Standby switch on the side panel. ON
On
The green On indicator is illuminated on the front panel when the pump is running.
WAIT
Wait
The yellow Wait indicator is illuminated when the mattress is being inflated. The indicator will remain illuminated until the mattress has been fully inflated.
Low Pressure
The red Low Pressure indicator will flash whenever the pump detects low pressure within the mattress. An audible alarm will sound. The indicator will extinguish once normal pressure is reached. Check CPR stopper at the head end is securely inserted and the tubeset is connected properly.
Power Fail
The red Power Fail alarm indicator will start flashing when a mains power failure has been detected by the pump. An audible alarm will sound until power is resumed or the pump is switched off using the run/standby switch.
Manual Pressure Adjustment
The mattress cell pressure can be manually adjusted for individual patient weights using the dial on the pump face. Ensure that the patient is not bottoming out by unfastening one side of the AlphaRelief MR and sliding a hand underneath the cells. If you can freely put a flat hand under the patient the mattress is inflated correctly.
Transport Mode
The patient may be transported on the Alpha Relief MR by disconnecting the tubes from the pump and cross connecting them immediately. The pressure in the cells will equalise and the patient will be supported for up to 8 eight hours with the patient in the supine position.
LOW PRESSURE
POWER FAIL
140
0 12 80
H T IN WEIG kg .
kg
+40
For optimum use of the Transport facility set the mattress to Transport mode when all cells are inflated at cell cross over. Ensure that the patient remains in the supine position.
Installation
Quick Start:
Alpha Relief Mattress Replacement (MR)
When the Alpha Relief pump is switched on at the mains supply, the pump will run a self diagnostic check for approximately 2 seconds during which all indicators will be illuminated. The yellow Wait indicator will remain illuminated while the mattress inflates.
Place the MR directly onto the bed base, with the tubeset located near the foot end of the bed and the CPR stopper at the head end. The cells of the mattress must be uppermost.
Press the Run/Standby button on the control panel of the pump.
Use the straps provided to ensure that the MR is secured to the moving parts of the bed platform.
Allow approximately 25 minutes for the mattress to inflate fully. Once the mattress is fully inflated the yellow Wait indicator will extinguish. Place a bed sheet over the mattress and tuck in loosely. Once the patient is on the mattress, adjust the cell pressures using the manual pressure adjustment dial to match the patient's weight. If the patient is sat up in bed, add 40kg to the weight setting.
To Complete the Mattress Installation Zip the protective cover over the mattress. Ensure that the Huntleigh logo is uppermost and at the foot end of the overlay. Ensure that the CPR stopper at the head end is securely inserted. Ensure that the CPR stopper and power cord are clear of profiling (moving) bed mechanisms or other possible entrapment areas. The CPR must be accessible at all times.
CPR Control IN THE EVENT OF CARDIAC ARREST In the event of a patient suffering cardiac arrest and CPR needing to be administered:
Installing the Pump: Position the pump feet down on any convenient horizontal surface or alternatively suspend from the bed foot rail by means of swing out hooks.
To Activate CPR: Pull the CPR tag firmly at the head end of the mattress and disconnect the tube set connectors at the pump.
Connect the male and female air feed tubes into the appropriate outlets on the pump unit. Push connectors in until they click. Ensure that the MR air feed tubes are not “kinked” or twisted. Insert the mains power plug into a suitable mains power socket.
SAFETY SIDES Regarding the use of safety sides, always refer to your local trust policy AND the product user manual.
MRF-109
To Reset CPR: Replace the stopper securely and reconnect the air tubes to the pump