ARJO

ARJO Pressure Management Systems

Auralis Quick Reference Guide

Quick Reference Guide

2 Pages

QUICK REFERENCE GUIDE - CLEANING AND DISINFECTION  Auralis Accessories needed: • Spray bottle with disinfectant solution (either 70 % alcohol solution or chlorine solution up to 10 000 ppm) • Neutral detergent (cleaning solution) • Spray bottle with water • Cloths Clean and disinfect between patients and at regular intervals while in use. WARNING To avoid electrical shock, always disconnect the pump from the power source before cleaning and inspecting. Always read the instructions for use and the safety data sheet of the disinfectant. To avoid eye or skin irritation, never disinfect in the presence of a patient.  Preparation  1. Disconnect the pump from the mattress or seat cushion. 2. Switch off the pump and disconnect from the power source.  Clean the pump  3. Use a clean cloth and cleaning solution and wipe to remove any deposits or visible dirt on: • main enclosure • membrane • labels • hanging brackets 4. Any areas where residual dirt has been left should be cleaned again with cleaning solution. 5. Use water and a new cloth to wipe off all traces of cleaning solution, followed by a dry cloth.  Disinfect the pump 6. Spray disinfectant solution onto a dry cloth and spread the solution onto all areas. 7. Allow time for disinfection. 8. Use water and a new cloth to wipe off all traces of disinfectant solution. 9. Wipe with a dry cloth.  08.AI.03EN_02 • 09/2022 • Page 1/2