Quick Reference Guide
2 Pages
Preview
Page 1
QUICK REFERENCE GUIDE - CONTROLS
Nimbus 4 & Nimbus Professional
Nimbus
Run/Standby button Press for Run mode; the indicator changes to green. Press and hold 3 seconds for Standby; the indicator turns off. Static button Press and hold 3 seconds to select Static mode. There is a tone and the indicator illuminates - the pump is in Static mode. Press and hold the button 3 seconds to return to Dynamic mode (alternating pressure). Mute button Press and hold 3 seconds to mute the audible alarm. The indicator is illuminated when the audible alarm is muted. Comfort Control Firmer setting = turn clockwise Softer setting = turn counterclockwise
Alarm symbol Illuminated, with audible alarm, during Low Pressure, High Pressure, Pump Fault or Power Fail. Wait indicator Illuminated until the mattress is fully inflated (15 minutes).
High Pressure indicator Illuminated, with audible alarm, when the pump detects high pressure in the mattress. Make sure the tubeset does not have kinks. Low Pressure indicator Illuminated, with audible alarm, when the pump detects low pressure in the mattress. Make sure the CPR control is closed, the tubeset is connected properly, and the mattress is set to Normal. Power (Fail) alarm indicator Illuminated, with audible alarm, when a mains power failure is detected. The alarm continues until the mains power is resumed or the Run/ Standby button is pressed to put the pump in Standby. If the power failure is prolonged, switch to Transport mode and disconnect the tubing. Pump Fault indicator Illuminated when an internal pump malfunction is detected. Call the Service Engineer. Service indicator Illuminated after a set number of running hours, indicating the pump is ready for service. The pump continues to function normally while the Service indicator is on.
649934EN_06 • 11/2022 • Page 1/2
Mattress CPR and transport controls CPR Control To be used in the event of a patient suffering cardiac arrest and CPR needs to be administered. Rapidly exhausts air from the mattress.
FAST DEFLATE
NORMAL
CP
R 1
TRANSPORT
2 3
Transport Control Turn clockwise for Transport mode. The mattress is sealed and the support surface is equally pressurised; the pump and/ or tubeset can be removed. Supports the patient up to 12 hours.
Vent valves control • • •
For permanent pressure relief: deflate no more than one cell directly under the area needing relief. For temporary nursing procedures only: deflate one or more adjacent cells, reinflate once nursing procedure is done. Deflating more than one adjacent cell may affect the support of the patient in normal alternating cycle. For complex patient needs, off-load more than one area for longer periods: • Deflate only one cell in the torso section. • Deflate only one cell in the calf or heel section. • Deflate only one cell in the head section in the supine position (face up), or all three cells in the head section in the prone position (face down). Deflating more cells in each area may affect the support of the patient in normal alternating cycle
Nimbus 4
Vent valves at Heelguard cells Nimbus Professional
Head cells Torso cells Shoulder support cell cannot be deflated Closed
Open
Thigh cells
Heelguard cells
To deflate a cell; turn the Vent Valve clockwise, to point downwards.
ArjoHuntleigh AB Hans Michelsensgatan 10 211 20 Malmö, Sweden www.arjo.com
649934EN_06 • 11/2022 • Page 2/2