SHOWER PANEL BCA 1224/1222/1221-04 Operating and Product Care Instructions
80 70
60
50
40
30
20
MIX
Reference Copy Printed 9:42 am, Feb 09, 2007
Check occasionally for new editions.
4.BC.01/3GB,US March 2003
Safety instructions
Reference Copy Printed 9:42 am, Feb 09, 2007
The equipment must be used in accordance with these safety instructions. Anyone using the equipment must also have read and understood the instructions in this booklet.If there is anything you are not sure about, ask your ARJO representative.
Always make sure that: ❏ the equipment is handled by trained staff. ❏ the shower temperature is checked by hand before showering.
Intended use This equipment is intended for showering hospital or care facility residents under the supervision of trained skilled nursing staff in accordance with the instructions outlined in the Operating and Product Care Instructions. All other uses must be avoided. For equipment fitted with a disinfection device, such equipment must be disinfected according to the instructions established in the Operating and Product Care Instructions. All other uses of the disinfection device must be avoided. The equipment must only be used for the purposes stated, and must be installed according to the Assembly and Installation Instructions, which can only be superseded by local code.
❏ the flowmeter is correctly set before use of the disinfection system and that the bottle for the disinfectant is full. ❏ your skin and eyes are protected against concentrated disinfectant. Carefully follow the instructions on the disinfectant bottle!
The useful life of this equipment, unless otherwise stated is ten (10) years, subject to required preventative maintenance as specified in Operating and Product Care Instructions, the Assembly and Installation Instructions and the Spare Part Catalouge.
© ARJO Hospital Equipment AB 2003 ARJO products are patented or patent pending. Patent information is available by contacting ARJO Hospital Equipment AB. Our policy is one of continous development, and we therefore reserve the right to make technical alterations without notice. The content of this publication may not be copied either whole or in part without the consent of ARJO Hospital Equipment AB. 2
Check occasionally for new editions.
Modifications made to the equipment without express acknowledgement from ARJO will invalidate supplier’s product liability.
ARJO Shower Panel BCA 1224/1222/1221-04 General The BCA 1221-04 is a wall mounted panel for use in combination with e.g. Concerto and Prelude. It is normal that a small water flow comes out in the tolilet while using the resident or disinfectant shower.*)
Resident shower • • •
•
•
Lift off the resident shower handle (9). Set the temperature control (7) at the desired temperature. Direct the shower handle away from the resident. Press the trigger handle and turn on the water by turning the resident shower control (6). Adjust the temperature (7) and water flow (6). Hand test the water before commencing with the shower of the resident. After the shower, turn off the resident shower control (6).
Reference Copy Printed 9:42 am, Feb 09, 2007
Part designations
Check occasionally for new editions.
5
6
4
1
Disinfection spray handle with trigger
2
Catch for disinfection spray handle
3
Disinfectant control
4
Lockable door for disinfectant
5
Flow meter (showing concentration of disinfectant and water)
6
Resident shower control
7
Temperature control
8
Thermometer
9
Resident shower handle with trigger
80 70
60
50
40
30
20
MIX
3 2
7
8 11 1
9 10
10 Screen rinsing valve (with cap to prevent unintentional use) 11 Flush valve for rim flushing
Cap of Flow meter (5) disinfectant bottle
Disinfection 80
•
• • • •
•
• •
•
• •
Open the panel door (4) and check that a bottle of ArjoClean has an ample supply and hose is connected. Detach the red disinfectant spray handle (1) from the catch. Set the temperature control (7) as low as possible. Direct the spray handle into the sink or inside the shower cabin. Press the trigger and open the disinfectant control valve (3) as far as it will go. Check that the flow meter reading (5) is correct. If not - use a screw driver on the adjustment screw (see figure). When ArjoClean is used the top of the ball in the flow meter should be opposite to figure 50. Spray disinfectant over the areas to be disinfected. Use disposable cloths/paper towels to scrub the surfaces. Clean the toilet with a special toilet brush. Allow two minutes for the disinfectant to work. (For complete disinfection, read the instructions on the container). Rinse all surfaces carefully with clean water from the resident shower. (9). Wipe with a clean cloth or towel to remove excess water.
70
60
50
50
40
30
This adjustment is normally carried out only before the first use.
!
20
*) To prevent remaining pressure at the trigger handle when the shower control is closed.
MIX
Warning! Always use protective gloves and protective eye wear when dealing with concentrated disinfectant. Avoid getting it on your skin or in your eyes. If that does happen, rinse copiously with water. If skin or eyes become irritated, consult a physician.
3
Care and maintenance Daily
•
•
Disinfection/cleaning.
•
Before using the disinfection system check that there is concentrated disinfectant in the container.
•
Reference Copy Printed
Move the temperature control on the thermostat mixer back and forth over the whole temperature range with the control valve turned off. This prevents lime deposits and should be done at least once a day, particularly in areas with hard water.
Rinse the screen on the incoming water as described below.
A rinsing valve with a filter is mounted on each of the two water supply hoses. The filter should be rinsed in both valves as follows: •
Close the valves. The handles (red for hot water and blue for cold) should be in a cross position.
WARNING: Make sure that the valve is closed before removing the cap - especially important on the hot water supply.
Weekly •
Clean panel with a mild deliming agent.
•
Examine the panel for cracks or other damage.
•
Check for leakage.
•
Visual check of hoses, pipes and connections.
•
Remove the cap (a) with a wrench.
•
Pull out the filter (b) and clean it carefully.
•
Open the valve for a while. Put a vessel to catch up the waste water.
•
Replace the cap
Monthly •
Remove and clean the screens in both shower heads. 1
Loosen the screw.
2
Remove the parts according to the illustration below.
3
Clean carefully.
4
Re-assemble. The parts (a) must be assembled before they will be set into the handle.
a
5 •
4
Tighten the screw carefully. Examine shower and panel hoses so that no leaks occur.
b a
Yearly The Shower Panel has to be serviced according to a routine by ARJO authorized personel. See next page.
9:42 am, Feb 09, 2007
Check occasionally for new editions.
Preventive Maintenance Schedule
Reference Copy Printed 9:42 am, Feb 09, 2007
The Daily Maintenance and Customer Obligations shall be carried out in accordance with the schedule below. Check occasionally for new editions.
Mandatory maintenance must be carried out by qualified personnel, fully trained in servicing using correct tools and knowledge of procedures. Failure to meet these requirements could result in personal injuries and/or unsafe equipment. The equipment is subject to wear and tear, and the following maintenance instructions must be performed when specified to ensure that the equipment remains within its original manufacturing specification.
ARJO authorized service (X)
Customer obligations X
Action/check
Every DAY
Liquid check
X
Check of the temperature control on the thermostatic mixer
X
Every WEEK
Cleaning
X
Examine for damage
X
Check for leakage
X
Visual check of hoses, pipes and connections
X
Every MONTH
Cleaning of the screens in the shower heads
X
Rinse the screen for incoming water
X
Examine shower and panel hoses
X
Every 12 MONTHS
Check/adjust thermostatic mixer incl.seal kit/complete thermostatic unit
(X)
Check/clean filters, incoming water
(X)
Check/change seal kit or complete shower handles
(X)
Check/replace flow restrictor for constant leak
(X)
Check/replace valve seals
(X)
Check hoses and connections for leaks and damage
(X)
Perform full feature functionality test
(X)
Check all vital parts for corrosion/damage
(X)
Check mechanical attachments
(X)
Check that the latest upgrades have been implemented
(X)
Replace all pressurized water hoses
Every 5 YEARS
(X)
5
FINLAND OY Vestek AB Martinkuja 4 FI-02270 ESPOO Tel: 9 8870120 Fax: 9 88701291
POLAND ARJO Poland Sp.z o.o. Ul. Lirowa 27 Reference Copy PL-02-387 WARSZAWA Printed Tel: 22 882 06 26 9:42 am, Feb 09, 2007 Fax: 22 882 06 29
AUSTRIA ARJO-Sic GmbH Föhrenweg 5 A-6065 THAUR Tel: 05223-49 33 50 Fax: 05223-49 33 50 - 75
FRANCE ARJO Equipements Hospitaliers S.A. 45, Avenue de l’Europe Eurocit BP 133 F-59436 RONCQ CEDEX Tel: 03 20 28 13 13 Fax: 03 20 28 13 14 E-mail: [email protected]
for new editions. PORTUGAL ARJO International AB Rua das Valas 82 PT 4510-154 JOVIM GONDOMAR Tel: 22 450 32 48 Fax: 22 450 03 27
BELGIUM ARJO Hospital Equipment nv-sa Ternesselei 248 B-2160 WOMMELGEM Tel: 03 353 91 00 Fax: 03 353 91 01 E-mail: [email protected]
GERMANY ARJO GmbH Christof - Ruthof - Weg 6 D-55252 MAINZ-KASTEL Tel: 06134-186-0 Fax: 06134 186 160 E-mail: [email protected]
CANADA ARJO Canada Inc. 1575 South Gateway Road Unit "C" MISSISSAUGA, ON, L4W 5JI Tel: 1-800-665-4831 Fax: 1-800-309-1116 E-mail: [email protected] Web page: www.arjo.com.
GREECE C. Psimitis Co Ltd Dimitriou Andr. 59 GR-16121 KAISARIANI ATTIKIS Tel: 21 0724 36 68 Fax: 21 0721 55 53
SWEDEN ARJO Scandinavia AB Verkstadsvägen 5 Box 61 SE-241 21 ESLÖV Tel: 0413-645 00 Fax: 0413-645 83 E-mail: [email protected]
ITALY ARJO Italia S.p.A. Via Tor Vergata 432 I-00133 ROMA Tel: 06-87426211 Fax: 06-87426222 E-mail: [email protected]
SWITZERLAND ARJO-Sic AG Florenzstr. 1d Postfach CH-4023 BASEL Tel: 061-337 97 77 Fax: 061-311 97 42
THE NETHERLANDS ARJO Nederland BV De Blomboogerd 8 4003 BX TIEL Postbus 6116 NL-4000 HC TIEL Tel: 0344-64 08 00 Fax: 0344-64 08 85 E-mail: [email protected]
UNITED KINGDOM ARJO Med AB Limited St Catherine Street GLOUCESTER GL1 2SL Tel: 08702 430 430 Fax: 01452-525 207
AUSTRALIA ARJO Hospital Equipment Pty Ltd 154 Lytton Road BULIMBA Brisbane QLD 4171 Australien Tel: 07-3395 6311 Fax: 07-3395 6712
CZECH REPUBLIC ARJO Hospital Equipment s.r.o. Hlinky 118 CZ- 603 00 BRNO Tel.: 549 254 252 Fax: 541 213 550 DENMARK ARJO Scandinavia Postboks 51 DK-4632 BJAEVERSKOV Tel: 45 93 27 37 Fax: 45 93 27 39 FAR EAST ARJO Far East Limited Unit 3A, 4/F., block B Hoi Luen Industrial Centre 55 Hoi Yuen Road, Kwun Tong, Kowloon HONG KONG Tel: 2508 9553 Fax: 2508 1416
NORWAY ARJO Scandinavia Ryenstubben 2 N-0679 OSLO Tel: 98 28 11 70 Fax: 22 57 06 52
...with people in mind. www.arjo.com
Check occasionally
SPAIN ARJO Spain, S.A C/ San Rafael, No. 6 E-28108 ALCOBENDAS Tel: 902 300 126 Fax: 902 300 191 E-mail: [email protected]
USA ARJO Inc. 50 North Gary Avenue ROSELLE, IL 60172 Tel: 1-800-323-1245 Fax: 1-888-594-2756
If your country is not listed here, please contact your local distributor or: ARJO INTERNATIONAL AB, Box 61, S-241 21 ESLÖV, SWEDEN Tel: +46 413 645 00; Fax: +46 413 645 63; www.arjo.com
MEMBER OF THE GETINGE GROUP