ARJO
smartsigns MiniPlus Instructions for Use 2012
Instructions for Use
26 Pages
Preview
Page 1
Table of Contents
Contents 1. Quality, Reliability & Safety ... 3 2. Introduction ... 4 2.1 2.2 2.3
Unpacking ... 4 Preliminary Checks ... 4 Recommended Clinical Applications ... 5 2.3.1 Intended Use ... 5 2.3.2 Indications for Use ... 5 2.3.3 Contra Indications... 6
2.4 2.5
Recommended Sensors... 6 Accessories ... 6
3. Controls, Indicators, Symbols & Displays ... 7 3.1 3.2 3.3 3.4 3.5
Front panel ... 7 Keypad ... 7 Displays and Audible Alarms... 8 Rear panel... 9 Desk Stand / Charger (MP1R Only) ... 10 3.5.1 Connecting to Power Supply ... 10 3.5.2 Battery charging LED status ... 10
4. Set-Up ... 11 4.1 4.2 4.3
Battery Installation - MP1 ... 11 Battery Installation - MP1R ... 12 Probe Installation and Use ... 13
5. Operation ... 14 5.1 5.2 5.3 5.4 5.5 5.6
Switching the Unit ON ... 14 System Operation ... 14 Setting the Alarms ... 16 Setting the pulse tone ... 17 Pulse Strength... 17 Switching the Unit OFF ... 17
6. Care of Your Equipment ... 18 6.1 6.2
Storage... 18 Cleaning and Disinfecting ... 18 6.2.1 Care of the Sensor ... 18 6.2.2 Care of the Unit ... 18
7. Specifications ... 19 8. Trouble Shooting ... 21 9. Maintenance ... 22 10.Warranty ... 22 11.Service ... 22 12.End of Life Disposal ... 23 2 13.I.V. Pole Fixing ... 24
Quality, Reliability & Safety This section contains important safety information related to the use of the Smartsigns® MiniPulse pulse oximeter. Other important safety information appears throughout the manual Before using this equipment, please study this manual carefully and familiarise yourself with the controls, displays and operating techniques. All modifications and repairs must be carried out by a qualified service engineer, agent or hospital technician authorised by Huntleigh Healthcare.
Atttention, consult accompanying documents / Instructions for Use Warning
The Smartsigns® MiniPulse pulse oximeter should only be operated by qualified personnel.
Warning
Do not use the Smartsigns® MiniPulse pulse oximeter in the presence of flammable anaesthetics.
Warning
Apart from the batteries, there are no user-serviceable parts inside the instrument.
Warning
To ensure accurate performance and prevent device failure, do not expose the Smartsigns® MiniPulse pulse oximeter to extreme moisture.
Caution
The Smartsigns® MiniPulse pulse oximeter is intended only as an adjunct to patient assessment. It must be used in conjunction with all clinical signs and symptoms
Warning
Pulse oximetry readings and pulse signals can be affected by certain ambient, environmental and certain patient conditions.
Warning
Do not use the Smartsigns® MiniPulse during magnetic resonance imaging (MRI) scanning
Caution
Do not autoclave, or use ethylene oxide sterilization process on any part or accessory of the system. Do not immerse the system in any liquid.
Caution
Electromagnetic Compatibility (EMC). This product complies with the requirements of applicable EMC Standards. The use of accessories not specified by the manufacturer may result in increased emissions by, or decreased immunity of, the equipment, affecting its performance.
Quality Reliability & Safety
1.
3
Introduction
2.
Introduction
2.1
Unpacking Carefully remove the Smartsigns® MiniPulse and its accessories from the shipping carton. Save the packing materials in case the device must be shipped or stored.
2.2
Preliminary Checks Huntleigh Healthcare takes every precaution to ensure that their products reach you in perfect condition. We recommend that a thorough visual inspection is made prior to installation. Should any damage be evident or any part missing, please ensure that Huntleigh Healthcare, Diagnostic Products Division are informed at once. Each system is supplied with the following accessories: Item 1 x Smartsigns MiniPulse 1 pulse oximeter ®
MP1 •
1 x Smartsigns MiniPulse 1R pulse oximeter
•
®
1 x SpO2 Sensor
•
4 x AA high capacity alkaline batteries
•
•
4 x Rechargeable AA NiMH batteries
•
1 x Desk stand with integrated charger
•
1 x Charger wall cube
•
1 x Instructions for Use
4
MP1R
•
•
Recommended Clinical Applications The Smartsigns® MiniPulse handheld pulse oximeter is intended for noninvasive continuous or spot check monitoring of functional arterial oxygen saturation (SpO2%), and pulse rate.
2.3.1
Intended Use
Introduction
2.3
The Smartsigns® MiniPulse pulse oximeter is intended for use by healthcare professionals for monitoring adult, paediatric and neonatal patients in hospitals and hospital-type facilities. A range of sensors are available for use with the MP1/MP1R. These are listed in section 2.4 along with details of each application site. Under low perfusion or patient motion conditions, the displayed pulse rate and SpO2 may become erratic or intermittent. In these circumstances, reposition the sensor and ensure the patient remains still.
2.3.2
Indications for Use
The specific medical indications for the use of this device are: • • • •
This device is a prescription device. This device is reusable and is intended for spot or continuous measurement. This device is intended to display oxygen saturation, heart rate and pulse strength or pulsatile signal. This device is intended to indicate an alarm if the measured saturation or heart rate are outside user-set alarm limits.
Warning
The Smartsigns® MiniPulse pulse oximeter is designed for attended monitoring only and must be used under the direct observation of a qualified health care provider.
Warning
The MP1R must not be charged when connected to a patient. The MP1R cannot be used while being charged.
Warning
Any doubt concerning the accuracy of any measurement, an alternative method should be used.
Warning
In extreme situations, tissue damage can be caused by incorrect application or duration of use of the SpO2 probe. Regularly inspect the probe site.
5
Introduction
Caution
Ambient light, movement, low patient perfusion, electromagnetic interference, artefacts, dysfunctional haemoglobin, and certain dyes, may interfere with the pulse oximeter’s function.
Caution
Do not autoclave, or use ethylene oxide sterilization process on any part or accessory of the system. Do not immerse the system in any liquid.
2.3.3
Contra Indications
Inaccurate measurements may be caused by: • • • • •
2.4
2.5
6
Excessive patient movement Venous pulsations Intravascular dyes, such as indocyanine green or methylene blue Significant levels of dysfunctional haemoglobin Defibrillation
Recommended Sensors Description
Application Site
Part No.
Adult Sensor
Finger
ACC-VSM-171M
3178 Paediatric Sensor
Finger
ACC-VSM-170
3043 Universal Y Sensor
Finger/Toe - Adult/ Paediatric Hand/Foot - Infant
ACC-VSM-169
3025 Infant Wrap Sensor
Toe
ACC-VSM-167
3026 Neonatal Wrap Sensor
Foot
ACC-VSM-168
Accessories Item
Part No.
Pole Clamp
ACC47
Carry Case
ACC-VSM-172
Desk Stand (MP1)
ACC-VSM-173
Desk Stand with Integrated charger (MP1R)
ACC-VSM-174
Controls, Indicators, Symbols & Displays
3.1
Front panel Item
Description
1
Sensor connection
2
Loudspeaker
3
Protective grip
4
Display
5
Front Panel
6
Keypad
1
2
3 4
5 6
3.2
Controls, Indicators, Symbols and Displays
3.
Keypad Button
Description
Button
Description
Power - ON / OFF Key
Up Key
MENU – Alarm limits and Pulse tone setting
Down Key
Alarm suspend - turns the alarms ON or OFF. Press and hold for 2 seconds to permanently disable the alarms; the system will sound a confirmation tone.
7
Controls, Indicators, Symbols & Displays 8
3.3
Displays and Audible Alarms
4
1 3 2 Item
Description
1
%SpO2
• Patient’s saturation • ‘Flashing’ alarm threshold exceeded
2
♥/min
• Patient’s pulse rate • ‘Flashing’ alarm threshold exceeded
3
Pulse Strength
• 8-segment bar graph pulse indicator (PI) and pulse strength
4
Alarm suspend
• Illuminated - Alarm OFF
Rear panel Item
Description
1
Probe socket
2
Probe storage
3
Belt Clip
4
Rear panel
5
Battery cover
1
2 3 4
Symbol
Description Attention, consult accompanying documents / Instructions for Use. Type BF applied part
5
Controls, Indicators, Symbols & Displays
3.4
9
Controls, Indicators, Symbols & Displays
3.5
Desk Stand / Charger (MP1R Only)
Symbol
Only use the PSU provided with the product. Class II Electrical Equipment
MODEL No. SERIAL No.
DSR
XXXXXXXXXXXX Huntleigh Healthcare Diagnostic Products Division 35 Portmanmoor Rd, Cardiff CF24 5HN, UK Tel +44 (0) 29 20485885 Fax +44 (0) 29 20492520 Email: [email protected]
INPUT RATING:
9V
Made in the U.K. 750316-2
3.5.1
1A
Connecting to Power Supply
Connect the power supply to the desk stand and connect to mains supply.
3.5.2
10
Description
• • •
Battery charging LED status Red Amber Green
Battery conditioning Charging Charged
Set-Up
4.1
Battery Installation - MP1 1.
2.
Set-Up
4.
Slide the battery compartment latch across to remove the battery cover.
Install four “AA” size batteries, ensuring correct polarity.
3.
Replace the Battery cover before use.
11
Set-Up
4.2
Battery Installation - MP1R
1.
Remove the securing screw and the battery cover.
2.
Install four “AA” size rechargeable batteries, ensuring correct polarity.
3.
Replace the Battery cover before use.
Caution: Use only the prescribed batteries with the system. Note:
12
Do not mix alkaline “AA” batteries with rechargeable batteries. When replacing batteries, replace with four fresh batteries. Do not mix new and used batteries.
Probe Installation and Use Warning
Before use, carefully read the probe manufacturers directions for use, including all warnings, cautions, and instructions.
Warning
Do not use a damaged probe. Do not use a probe with exposed optical components.
Warning
Use only BCI sensors for SpO2 measurements. Other sensors may cause damage.
Warning
Incorrectly applied sensors may give inaccurate readings.
1.
Connect the probe to the pulse oximeter.
2.
Insert the patient’s finger into the tip of the oximeter probe.
3.
Ensure the cable extends along the top of the patient’s hand.
Set-Up
4.3
13
Operation
5.
Operation
5.1
Switching the Unit ON Warning
Do not lift the unit by the probe cable.
Press the ON/OFF button
.
During power up, the system will perform a self test, illuminate the displays for 2 seconds and enter the measurement mode.
5.2
System Operation With the system connected to a patient, the oximeter will search for a pulse. After a short period the system will display the patient’s saturation level and pulse rate. The pulse indicator will be synchronised with the audible pulse tone (If enabled).
The frequency of the pulse tone is closely related with saturation level; lower tones indicate lower saturation levels. After 2 mins, the display will reduce to a level which is equivalent to 50% output intensity; press any key to re-establish the normal display intensity setting. The display will automatically brighten in the event of an alarm condition.
14
Operation
If there is no finger connected to the sensor, the %SpO2 and Pulse Rate (♥/min) will display:
If no probe is connected to the system, the unit will turn off automatically after 3 minutes.
Warning
Reposition the probe at least once every 4 hours to allow the patient’s skin to respire.
15
Operation
5.3
Setting the Alarms The default setting for the alarms are:• • • •
SpO2 Low SpO2 High HR Low HR High
Press the MENU
92% 99% 50BPM 150BPM button to enter the Alarm Setup mode.
The display will change to :
Press the UP
or DOWN
button to adjust the SpO2 low setting.
Press the MENU button to select the SpO2 high setting. Press the UP or DOWN button to adjust the SpO2 High setting. Press the MENU button to select the HR Low setting. Press UP or DOWN button to adjust the setting. Press the MENU button to select the HR High setting. Press the UP or DOWN button to adjust the setting. Press the MENU button to save the settings.
16
Setting the pulse tone Press the MENU button five times to display the Pulse Tone set-up option. Press the UP or DOWN key to set the pulse tone Setting
Display
• OFF • ON
PB OFF PB ON
Operation
5.4
Press the MENU button to save the settings.
5.5
Pulse Strength Press and hold the UP button to display the Pulse Strength. The bar graph will illuminate according to the strength of the pulse signal.
Good Quality
5.6
Poor Quality
Switching the Unit OFF Press the ON/OFF
button to switch the unit OFF.
17
Care of your Equipment
6.
Care of Your Equipment It is recommended that the unit and accessories are inspected and tested on a regular basis. If any part of the system appears to be damaged, the system should be returned to your service centre for repair.
6.1
Storage Remove the batteries from the unit before long-term storage.
6.2
Cleaning and Disinfecting Warning
6.2.1
Turn the Pulse Oximeter OFF before cleaning.
Care of the Sensor
Clean the probe after each use. Clean the surface of the sensor and cable with a soft gauze pad by saturating it with a solution of 70% isopropyl alcohol. If low level disinfection is required, use a 1:10 bleach solution.
6.2.2
Care of the Unit
If cleaning is required, wipe the Pulse Oximeter’s surfaces with a soft cloth moistened with a non-abrasive cleaner. Do not allow any liquid to enter any of the system.
18
Specifications Electrical Power requirements MP1
4 x AA High Capacity Alkaline cells
MP1R
4 x AA rechargeable NiMH ≥ 2500mAH cells
Battery life MP1
60 hours typical
Battery life MP1R
60 hours typical before recharge is required
Specifications
7.
Charger MP1R only Type
Desk top stand with wall mount power supply
Charge time
6.5 Hours to full charge Built in protection for under / over charge
Wall cube
Input voltage 100 – 240 VAC 60Hz @ 6VA Output 9V
General Dimensions
75 x 140 x 25mm (excluding desk stand, sensor and wall cube)
Weight
300gm (excluding desk stand, sensor and wall cube)
Display indicators
• • • • •
Alarms
Audible and visual alarms for high and low saturation and pulse rate. Synchronized saturation tone
SpO2 Pulse rate Pulse Strength Alarm status Charge status (desk stand only)
19
Specifications
Environmental Operating temperature
5°C to 40°C
Storage temperature
-40°C to +70°C
Humidity
5% to 95% non condensing
Pressure
600mbar – 1080mbar
SpO2 /Pulse Rate %Saturation Range
0 – 99% Functional SpO2 (1% increments)
Accuracy
Adults +/-2% @ 70 – 99% SpO2 < 70% undefined
Averaging
8 beats
Pulse Rate Range
30 – 254 BPM (1 BPM increments)
Accuracy
Better than +/-2 BPM or +/-2%
Averaging
8 seconds
Bargraph
0 – 8 segments providing pulsatile signal or pulse strength
Calibration
Factory calibrated over range 70% to 100%
Compliance Medical Electrical Equipment 25EA
Equipment classification
20
IEC60601-1, WITH RESPECT TO ELECTRIC SHOCK, FIRE AND MECHANICAL HAZARDS ONLY, IN ACCORDANCE WITH UL60601-1, CAN/CSA C22.2 No 601.1 and ISO 9919.
EMC compliance
EN55011 Group 1 Class B, EN60601-1-2, EN61000-3-2, EN61000-3-3
Degree of protection
Type BF
Degree of protection against ingress of water
IPX1
Trouble Shooting
Problem
Possible Cause
Corrective Action
Unit will not switch on
Depleted batteries
Replace batteries (MP1) Recharge batteries (MP1R)
Unit switches on but switches off after a short time when not connected to a patient.
Auto off set
Contact Service Dept. (See section 11).
Display goes DIM after a few minutes
Ad (Auto dim) set
Contact Service Dept. (See section 11).
No Fin displayed when connected to a patient.
Faulty sensor
Replace sensor
No Prb displayed when sensor connected
Faulty sensor
Replace sensor
Pulse rate erratic, intermittent or incorrect.
Sensor incorrectly positioned. Poor patient perfusion. Patient motion.
Reposition sensor on patient. Patient must be still for unit to function properly.
Charge lamp does not illuminate when unit placed on charger
Charger wall cube not connected to mains power or to the charger cradle.
Check power and connection.
Open circuit/ faulty batteries.
Check/ replace batteries.
Trouble Shooting
8.
Dirty charger/ MP1R contacts. Clean contacts. Faulty charger
Test by shorting charger contacts with metal object charge lamp should go RED.
Charge lamp illuminates and stays RED continuously.
Faulty batteries
Replace batteries
Batteries charge normally but unit runs only for a short time
Faulty batteries
Replace batteries
No bleep from MP1R when placed in charger cradle.
Batteries have insufficient charge. No power to charger Dirty charger contacts
Recharge batteries
MP1 / MP1R displays “Err”
Internal system error
Return to supplier for service
Connect power to charger. Clean contacts.
21