QUICK REFERENCE
Arthrex ® DualWave ™ QUICK REFERENCE
Read all safety and set-up instructions in the Arthrex DualWave Arthroscopy Pump User’s Guide and Directions For Use before using this this product. Please visit www.Arthrex.com for all product literature. Arthrex, Inc. 1370 Creekside Blvd. Naples, FL 34108-1945 USA Toll Free: 1 (800) 934-4404 www.arthrex.com
Arthrex GmbH Erwin-Hielscher-Strasse 9 81249 München, Germany Tel: +49 89 909005-0 www.arthrex.de
Arthrex Tech Support: +1-888-420-9393
+49 89 90 90 05 8800
1
ACCESSORIES
ARTHROSCOPY PUMP ACCESSORIES Autoclavable Remote
Lavage
Pressure Down
Shaver Suction Toggle
2
Foot Control
Rinse
Pressure Up
Run/Stop
Lavage: Timed increase of pressure for improved control of bleeding. Rinse: Timed increase of flow for improved control of debris removal. (Only available when using DualWave outflow) Shaver Suction Toggle: Allows switching between 4 preset shaver suction levels. (Only available when using DualWave outflow)
INFLOW
INFLOW TUBING SET-UP
1 Remove orange clip
2
3
4
Insert connector into Open inflow door pump pressure coupler
Insert green tubing hub
5 Wrap tubing around rollers
Inflow door Rollers
6
7
Insert clear tubing hub Close inflow door
Green tubing hub
Pump pressure coupler
8 Spike saline bags
Clear tubing hub
3
INFLOW
INFLOW ONLY USER INTERFACE Inflow Only
Stopped
Fluid Used
Set Pressure
Boost Control
Pressure Down
Pressure Up
Run/Stop
Menu
Running Pressure Reading Flow Rate
Run Time (Resets after power down) Run/Stop
Lavage 4
OUTFLOW
OUTFLOW TUBING SET-UP Shaver
Cut as needed
Outflow Boot
Waste
Luer Lock
1 Connect to shaver suction
3 Open outflow door
2 Connect to independent cannula
4
Pass off Cassette and Red Waste Tubing Cannula
5
*Install outflow cassette Close outflow door
6 Outflow display
7 Connect red waste tubing to waste system
*For easy set-up, move the rollers so that two wheels are flush. Extend the boot of the outflow cassette over the two wheels at the same time. With one hand, hold the boot in place; and, with the second hand, pull down on all three tubes, pushing the cassette in place. The cassette will be held in place by the edges of the outflow bezel.
5
INFLOW/OUTFLOW
INFLOW/OUTFLOW USER INTERFACE Inflow/Outflow
Stopped
Fluid Used
Cannula Suction Control
Shaver Suction Control
Set Pressure Pressure Down
Pressure Up
Run/Stop
Menu
Running Flow Rate
Pressure Reading
Run Time (Resets after power down) Run/Stop
Lavage 6
Rinse
SHAVER DETECTION
SYNERGY SHAVER DETECTION AND HEADS UP DISPLAY
1
2
Connect red cable that comes with DualWave to the rear of Synergy Resection Console
™
Power on Synergy Resection Console 3 2
3 Power on DualWave
4 *Connect to Synergy Imaging for heads up display cable
*For Synergy heads up display use AR-3200-1040 accessory cable kit to connect DualWave to camera console. (Only 1 cable is needed to connect the DualWave to the camera console.) Prior to using this system, see IFU 950-0052-00 for more detail.
7
SHAVER DETECTION
UNIVERSAL SHAVER DETECTION
1 Shaver detect cord
2 Plug into DualWave™
6 4 3 Plug into Shaver Console
5 Plug in handpiece 8
4 Power on Shaver Console
6 *Power on DualWave
*Shaver console must be fully powered up before DualWave is powered on
TROUBLESHOOTING
TROUBLESHOOTING PROBLEM
SOLUTION
** Critical Failure **
Return to Arthrex® for repair.
** Door Open **
1. Roller housing is not secured – ensure the locking lever is properly secured. 2. If the failure persists, return to Arthrex for repair.
** Overpressure **
1. Increase or open the outflow. 2. Manipulate the joint to a stress-free position. 3. If the failure persists, return to Arthrex for repair.
** Pressure Fault **
1. Ensure adequate fluid supply. 2. Decrease the outflow. 3. Check the tubing for damage or pinches, kinks, or blockages. 4. Check the tubing for the proper connections. 5. Replace the tubing. 6. If the failure persists, return to Arthrex for repair.
** Tubing Out **
1. If the tubing sensor indicator’s LED is red, the tubing is not properly connected. 2. Ensure the tubing sensor coupler is open, not seated. 3. Ensure that the tubing pressure line connector is seated completely. 4. Change the tubing. 5. If the failure persists, return to Arthrex for repair.
Console fails Self Diagnostic Test
1. Ensure no tubing is connected to the pump during the power on sequence. 2. If the failure persists, return to Arthrex for repair.
Console will not power up
1. Check the AC mains power cord. 2. Try an alternate power outlet. 3. Check the AC mains fuses. 4. If the failure persists, return to Arthrex for repair.
PROBLEM
SOLUTION
Does not pump when activated
1. Check for error messages. 2. Open all tubing clamps and shut-off valves. 3. Ensure the actual pressure is below the target pressure. 4. Check the tubing for pinches, kinks, or blockages. 5. If the failure persists, return to Arthrex for repair.
Inadequate distention, liquid bloody or cloudy
1. Increase the pressure. 2. Activate Lavage mode. 3. Reduce the outflow. 4. Use high-flow cannulas. 5. If the failure persists, return to Arthrex for repair.
No (or inadequate) flow
No Shaver Detect
1. Check for error messages. 2. Check that all tubing clamps are open. 3. Check the settings for flow and pressure. 4. Check the tubing for pinches, kinks, or blockages. 5. Check that the tubing seats correctly over the rollers. 6. Verify use of high-flow cannulas. 7. If the failure persists, return to Arthrex for repair. Synergy Shaver Detection 1. Verify DualWave™ software is at least 1.8 2. Verify Synergy Resection Console is at least 2.3. 3. Ensure the Synergy cable is connected. 4. Press stop and then run on DualWave while the Synergy Resection Console is on. 5. Replace cable. 6. If the failure persists, return to Arthrex for repair. Universal Shaver Detection 1. Ensure the universal shaver cable is connected. 2. Turn DualWave on and off while the shaver system is powered on. 3. Adjust the DELAY and/or THRESHOLD settings. 4. Isolate the power source. 5. If the failure persists, return to Arthrex for repair.
9
This description of technique is provided as an educational tool and clinical aid to assist properly licensed medical professionals in the usage of specific Arthrex products. As part of this professional usage, the medical professional must use their professional judgment in making any final determinations in product usage and technique. In doing so, the medical professional should rely on their own training and experience and should conduct a thorough review of pertinent medical literature and the product’s Directions For Use.
http://www.arthrex.com/imaging-resection/fluid-management
View U.S. patent information at www.arthrex.com/corporate/virtual-patent-marking 2019, Arthrex Inc. All rights reserved. DFU-0244-2 rev. 0 08/2019
©
DFU-0244-2 r0