Arthrex
AR- 6485 Continuous Wave 4 Arthroscopy Pump Users Guide Rev 0 July 2020
Users Guide
56 Pages
Preview
Page 1
Continuous Wave™ 4 Arthroscopy Pump User’s Guide The Arthrex Continuous Wave™ 4 (CW4) Arthroscopy Pump User’s Guide provides safety operation information for all components of the Arthrex CW4 Arthroscopy Pump (model AR-6485), including accessories. All operating personnel must read this User’s Guide for version 1.9 or higher thoroughly prior to using this system and follow all safety warnings, cautions, and precautions.
Arthrex, Inc.
Arthrex GmbH
1370 Creekside Blvd. Naples, FL 34108-1945 USA Toll Free: 1-(800) 934-4404
Erwin-Hielscher-Strasse 9 81249 München, Germany Tel: +49 89 909005-0
www.arthrex.com
www.arthrex.de
Arthrex Brazil Equipment Import and Trading LTDA ME Est da Lagoinha 489 Building 1 Lagoa Vargem Grande Paulista-SP 06730-000 BRAZIL Tel: +55 11 4310-8950 DFU-0249-5 Revision 0
07/2020
© 2020 Arthrex, Inc. All rights reserved
This is not a warranty document. For all warranty information, including disclaimers, exclusions, terms, conditions and related provisions, refer to the "Arthrex U.S. Product Warranty" section of the Arthrex, Inc. website, found at www.arthrex.com whose provisions are incorporated herein by reference.
Table of contents 1.0 General Warnings and Safety Notices - Read This First ... 1 1.1
Important Safety Conventions ... 1
1.2
Symbols Definition ... 5
1.3
Shipping, Unpacking, and Warranty Information ... 7
2.0 Product Description ... 8 2.1
Product Description and Intended Use... 8
2.2
Product Features... 10
2.2.1
AR-6485 Console: Front View ...10
2.2.2
AR-6485 Console: Rear View ...11
2.2.3
AR-6485 Operator Display Messages and Iconography ...11
2.3
Foot Pedal Unit (AR-6483) ... 14
2.4
Remote Control Unit (AR-6482) ... 15
2.5
Tubing ... 16
2.5.1
Tubing Configurations ...16
2.5.2
Main Pump Tubing Set (AR-6410: Region A) ...18
2.5.3
ReDeuce™ Pump Tubing Set (AR-6411: Region A) ...18
2.5.4
ReDeuce Patient Tubing Set (AR-6421: Region A) ...18
2.5.5
Extension Tubing System (AR-6220: Region A)...18
2.5.6
One-Piece Tubing Set System (AR-6415/AR-6415CL: Region B) ...18
2.5.7
Main Pump Tubing Set (AR-6420/AR-6420CL: Region B) ...19
2.5.8
Patient Extension Tubing System (AR-6425: Region B) ...19
2.5.9
Y-Tubing Adapter (AR-6215) ...19
3.0 Technical Specifications ... 21 3.1
Console ... 21
3.2
Foot Pedal ... 22
3.3
Remote Control ... 22
3.4
Safety, EMC, and Regulatory Requirements ... 22
4.0 Setup ... 23 4.1
How to Set Up the Console ... 23
4.2
AC Power Safety Considerations ... 23
4.3
Replacing the Fuses ... 24
4.4
Electromagnetic Compatibility (EMC)... 24
4.5
Basic Setup Procedure for the AR-6485 ... 25
DFU-0249-5r0_fmt_en-US
Page i
Table of contents 4.6
Shaver Detect ... 27
4.6.1
Shaver Detect, Set-up...27
4.6.2
Shaver Detect, Activation ...27
4.7
How to Set Up the Synergy Heads-Up Display ... 28
4.8
How to Set Up Pump Tubing ... 28
4.9
How to Set Up the Two-Piece Tubing System ... 28
4.10
How to Change the Language Setting ... 29
4.11
How to Test the Power Supply Voltages ... 29
4.12
Safe Setup and Performance ... 29
4.12.1
Abnormal Operation ...29
4.12.2
Overpressure Sensing ...30
4.12.3
Inflow Roller Housing [4] ...30
4.12.4
Tubing Sensor Coupler [14] ...30
4.13
Shutdown Procedure ... 30
5.0 Operation and Frequently Used Functions ... 31 5.1
Initial Pressure Settings ... 31
5.2
How to Operate the AR-6485 ... 32
5.3
How to Operate the AR-6485 in LAVAGE Mode ... 32
5.4
How to Operate the Flow Control on the AR-6485 ... 33
6.0 Cleaning and Disinfection ... 34 6.1
Console (AR-6485) and Foot Pedal Control Unit (AR-6483) ... 34
6.2
Remote Control Unit (AR-6482) ... 34
6.3
Tubing ... 35
7.0 Sterilization ... 36 7.1
Transmissible Spongiform Encephalopathy Agents ... 36
8.0 Maintenance ... 37 8.1
Periodic Maintenance... 37
9.0 Technical Support... 38 9.1
How to Display the Software Version ... 38
9.2
Additional Technical Information ... 38
10.0
Troubleshooting ... 39
10.1
Troubleshooting Interference with Other Devices ... 40
DFU-0249-5r0_fmt_en-US
Page ii
Table of contents 11.0
Repair Policy ... 41
12.0
End of Life, Environmental Directives ... 42
13.0
Electromagnetic Emissions ... 43
List of Figures Figure 1
Front Panel of Console ... 10
Figure 2
Rear Panel of Console ... 11
Figure 3
Foot Pedal Unit (AR-6483) ... 14
Figure 4
Remote Control Unit (AR-6482) ... 15
Figure 5
One-Piece Tubing Configuration ... 16
Figure 6
Two-Piece Tubing Configuration ... 17
DFU-0249-5r0_fmt_en-US
Page iii
Table of contents List of Tables Table 1
Front Panel Elements ... 10
Table 2
Rear Panel Elements ... 11
Table 3
AR-6485 Operator Display Messages and Iconography... 12
Table 4
Foot Pedal Elements (AR-6483) ... 14
Table 5
Remote Control Unit Elements (AR-6482) ... 15
Table 6
Elements of the One-Piece Tubing Configuration ... 16
Table 7
Elements of the Two-Piece Tubing Configuration ... 17
Table 8
Tubing Set Correlation and Comparisons ... 20
Table 9
Control Unit (AR-6485) Specifications ... 21
Table 10
Ambient conditions for operation ... 21
Table 11
Ambient conditions for storage (in shipping packaging)... 21
Table 12
Foot Pedal Unit (AR-6483) Specifications ... 22
Table 13
Remote Control Unit (AR-6482) Specifications ... 22
Table 14
Initial Pressure Settings ... 31
Table 15
Troubleshooting: Faults, their Causes, and Solutions... 39
Table 16
Guidance and Manufacturer's Declaration - Electromagnetic Emissions ... 43
Table 17
System Cables ... 43
Table 18
Guidance and Manufacturer's Statement - Electromagnetic Immunity... 44
Table 19
Guidance and Manufacturer's Statement - Electromagnetic Immunity (cont’d) ... 45
Table 20 Guidance and Manufacturer's Statement – Recommend Separation Distance between portable and mobile RF communications equipment and the Model AR-6485 ... 46
DFU-0249-5r0_fmt_en-US
Page iv
General Warnings and Safety Notices - Read This First CW4 Arthroscopy Pump User’s Guide
1.0 General Warnings and Safety Notices - Read This First It is imperative that the symbols and conventions listed below be clearly understood. The CW4 Arthroscopy Pump User’s Guide identifies critical, important, and useful information using these symbols and conventions.
1.1
Important Safety Conventions Users of this device are encouraged to contact their Arthrex representative if they require a more comprehensive surgical technique.
W A R N I N G ! WARNING! is the most important safety symbol. It identifies critical information that must be followed precisely to avoid injury or death. 1. All fluid inflow devices, including gravity assist, may cause fluid extravasations into the surrounding tissues. This extravasation may be mild, moderate or severe. In severe cases, the resulting edema may result in a serious adverse patient event which may include compartment syndrome, nerve compromise, or death. Undiagnosed capsular defects will exacerbate fluid extravasation conditions. 2. When utilizing any fluid management device, the patient (extremity and surrounding area) must be monitored closely by the surgical team for signs of excess fluid buildup. Fluid usage volumes should be monitored and compared to similar surgical procedures. With all arthroscopy pumps, correct setup and proper user operation is required. Always select the lowest possible pressures in order to achieve the required intra-articular distention. All alarms or alerts must be acknowledged and the appropriate troubleshooting procedure followed. 3. Failure to follow the setup instructions and/or continuing to use the pump without resolving an alarm condition could result in a serious patient adverse event. 4. Failure to adhere to the setup instructions and use of Arthrex certified tubing may result in inaccurate pressure sensing and monitoring by the device. It is imperative that the user is aware that patient safety may be compromised when an alarm on the pump is ignored or silenced incorrectly. NEVER ignore or silence alarms. Follow appropriate troubleshooting procedures and carefully monitor the patient. Only Arthrex certified tubing must be used 5. This device is only for use in normal arthroscopic procedures as described in the User’s Guide, under the supervision of a trained and licensed physician. This device should not be used by untrained personnel or used for indications other than those described in this User’s Guide. 6. No modification of the console (AR-6485) or accessories is allowed. DFU-0249-5r0_fmt_en-US
Page 1 of 50
General Warnings and Safety Notices - Read This First CW4 Arthroscopy Pump User’s Guide 7. DO NOT open or attempt to service this system, as this may void your warranty. There are no user-serviceable parts inside. Removing the cover may introduce an electric shock hazard by exposing you to dangerously high voltages or other risks. If the system malfunctions, return it for servicing immediately. 8. To avoid the RISK of electric shock, this equipment must only be connected to a MAINS POWER SUPPLY with a protective earth terminal. 9. DO NOT have the device in direct contact with the patient if high-frequency devices are in use, or if the patient requires defibrillation. 10. To ensure that correct pressure monitoring occurs, the pump and operative site MUST be in the same horizontal plane. 11. DO NOT stack or place equipment adjacent to the AR-6485 console, if possible. If such a configuration is necessary, carefully observe the configuration in question to ensure that electromagnetic interference does not degrade performance. 12. USE ONLY Arthrex approved accessories. Other accessories may result in increased emissions or decreased immunity of the system. Contact your Arthrex representative for a complete list of accessories. DO NOT modify any accessory. Failure to comply may result in injury to the patient and/or operating room staff. 13. DO NOT use in the presence of flammable anesthetics or oxidizing gases such as nitrous oxide, oxygen, or endogenous gases. All oxygen connections must be leak free for the duration of the surgical procedure. 14. Use ONLY Arthrex approved tubing accessories. Other accessories may result in decreased pressure accuracy. Contact your Arthrex representative for a complete list of accessories. DO NOT modify any accessory. Failure to comply may result in patient and/or operating room staff injury. 15. The extension and/or patient tubing must be replaced before each new patient and/or procedure. 16. The sterile connector cap must be used to cover the pump tubing set connector after each surgical procedure. This maintains sterility of the pump tubing and ensures its safe operation throughout the entire surgical day. 17. If the tubing is disconnected from the pump, it MUST be replaced. Do not attempt to reconnect the tubing to the pump as it could lead to unreliable pressure. 18. The safety and effectiveness of the AR-6485 is verified and documented; however, the AR-6485 must be used with an awareness of the risk of extra-
DFU-0249-5r0_fmt_en-US
Page 2 of 50
General Warnings and Safety Notices - Read This First CW4 Arthroscopy Pump User’s Guide articular edemas for patients with pathologically-changed articular capsules and for procedures involving an opening of the capsule (e.g. lateral release). 19. Slight swellings have been observed and described in the literature in cases where roller pumps are used in arthroscopy. This build-up of fluid can lead to postoperative swellings and pathological changes in patients. It is of the utmost importance that the surgeon monitors both the system and the patient closely while the roller pump is in operation. 20. Always start with the lowest possible pressure to achieve the desired joint distention. Continue to increase distention pressure until a clear liquid medium is obtained. 21. The initial pressure settings are recommendations. It is always appropriate and prudent to use the lowest possible pressure setting to minimize extravasation and any other pressure-related injury to the patient. 22. User-programmed “Pressure Set” values are increased by as much as 50%, but not to exceed a pump pressure of 120 during the LAVAGE function. Exercise caution to avoid injury to the patient. 23. After autoclaving, the accessory devices are VERY HOT. Handle with care to avoid burns. 24. Caution: Federal law restricts this device to sale by or on the order of a physician. 25. This device is intended to be used by a trained medical professional. 26. Detailed instructions on the use and limitations of this device should be given to the patient. 27. Biohazard waste, such as explanted devices, needles and contaminated surgical equipment, should be safely disposed of in accordance with the institutions policy. 28. Serious incidents should be reported to Arthrex Inc., or an in-country representative, and to the health authority where the incident occurred.
The PRECAUTION! symbol identifies methods and procedures that must be followed to avoid damaging the device or causing it to malfunction. 1. Do not disconnect the plug of the remote control or foot pedal unit by pulling on the cable. Remove it by grasping and pulling on the body of the connector. 2. Only use replacement power cords that comply with medical grade standards, IEC 60320-1 Subclause 3.21, Detachable Power Supply Cords or electrical standards for the designated country where the AR-6485 is being used. Contact your Arthrex representative for further information. 3. Avoid positioning the console so that it is difficult to disconnect the coupler or plug from the mains power supply.
DFU-0249-5r0_fmt_en-US
Page 3 of 50
General Warnings and Safety Notices - Read This First CW4 Arthroscopy Pump User’s Guide 4. To prevent electrical shock do not use extension cords or two-prong/threeprong adaptors. 5. Always use fuses with the correct values to avoid allowing overcurrent to enter the system. 6. An incorrect fuse may increase the risk of electrical shock or fire hazard. 7. This device has passed testing for EMI / RFI radiation and susceptibility, and EMC compatibility. This device may cause interference with other devices in the near vicinity if not set up and used as instructed by Arthrex. 8. Do not attach the remote control or the foot pedal during the Self Test or Programming Modes. 9. NEVER use liquid to clean the accessory device connector contacts. Remove dust regularly using dry compressed air. 10. Liquid on the cable connector of the accessory device can damage the device. Before connecting the cable, ensure the receptacles are clean and dry. 11. Always comply with the instructions issued by the manufacturer of the cleaning disinfectant regarding concentration, exposure times, temperature, and material compatibility. 12. Never allow the console receptacles to come into contact with liquids. If there is dust or moisture on the receptacles, remove using dry compressed air. ONLY dry connectors should be plugged into the console. 13. Do NOT clean the device with abrasive cleaning or disinfectant compounds, solvents, or other materials that could scratch or damage the device. 14. Refer to the Instructions for Use package insert (DFU-0144-XX) for detailed remote control cleaning and sterilization instructions included with each remote control. Additional copies of this insert can be obtained from the Arthrex website at www.arthrex.com, or by contacting your local Arthrex representative. 15. Refer to the Instructions for Use package insert (DFU-0140-XX) for detailed tubing cleaning and sterilization instructions included with each tube set. Additional copies of this insert can be obtained from the Arthrex website at www.arthrex.com, or by contacting your local Arthrex representative. 16. The foot pedal is NOT suitable to be cleaned and disinfected in a thermo washer disinfector. 17. After sterilization in the autoclave, let the accessory device cool down slowly. NEVER use cold water to cool the remote control. This will damage the electronic components and seals. 18. Surgeons are advised to review the product-specific surgical technique prior to performing any surgery. Arthrex provides detailed surgical techniques in print, video, and electronic formats. The Arthrex website also provides detailed surgical technique information and demonstrations. Or, contact your Arthrex representative for an onsite demonstration. •
In CE accepting Countries: Procedures carried out using these devices may be used on the general population.
DFU-0249-5r0_fmt_en-US
Page 4 of 50
General Warnings and Safety Notices - Read This First CW4 Arthroscopy Pump User’s Guide •
In CE accepting Countries: The clinical benefits associated with the use of these devices outweigh the known clinical risks.
•
In CE accepting Countries: There are no unacceptable residual risks or uncertainties associated with the clinical use of these devices.
1.2
Symbols Definition
All of the symbols used on the labeling along with the title, description and standard designation number may be found on our website at www.arthrex.com/symbolsglossary.
Safety Sign Follow operating instructions
DFU-0249-5r0_fmt_en-US
Caution: Federal law restricts this device to sale by or on the order of a physician.
Stand-by power switch (push-push)
Type BF Equipment
Caution
Fragile, handle with care
Keep dry
This side up
Electrical hazard, dangerous voltages are present. Never attempt to repair the equipment. Only trained service personnel may remove the cover, or obtain access to system components.
Temperature limits for storage and transport
Page 5 of 50
General Warnings and Safety Notices - Read This First CW4 Arthroscopy Pump User’s Guide
DFU-0249-5r0_fmt_en-US
Alternating current
Atmospheric pressure limitation
Fuse
Humidity limits for storage and transport
Equipotential [equipment potential]
Protective earth terminal
Electrical waste
RF symbol. Non-ionizing electromagneti c radiation
Manufacturer
Date of manufacture; year and month.
Non sterile
The product meets the essential requirements of Medical Device Directive 93/42/EEC
Catalog number
Serial number
Quantity
Authorized representative in the European Community
Remote control connection
Foot pedal connection
Page 6 of 50
General Warnings and Safety Notices - Read This First CW4 Arthroscopy Pump User’s Guide Do not use if package is damaged Universal serial bus [for use ONLY with Arthrex approved thumb drive]
USB
[x]
1.3
IP22
International protection marking
Serial IO [Arthrex integration]
Square brackets that enclose a letter, number, or Roman numeral reference a callout on a line drawing. Section 2.2, Product Features, includes drawings of products associated with the AR-6485. Each line drawing has its own callout system to identify important elements of each product.
Shipping, Unpacking, and Warranty Information Carefully unpack and inspect all components for shipping damage. Any damage could compromise patient safety and should be reported immediately to Arthrex or any authorized Arthrex distributor. The warranty could be voided if shipping or first-installation damage is not reported within seven business days of receiving the device. Refer also to our General Terms of Business. All defective products will be repaired or replaced at the discretion of Arthrex at no charge. The warranty does not cover damage caused by unlawful use or improper handling of a product. The warranty is not valid if modifications are made to the product or repairs are carried out outside of Arthrex or an authorized Arthrex distributor. Arthrex will answer any questions referring to the quality, reliability, and/or shelf life of any product identified in this User’s Guide.
DFU-0249-5r0_fmt_en-US
Page 7 of 50
Product Description CW4 Arthroscopy Pump User’s Guide
2.0 Product Description 2.1
Product Description and Intended Use The Arthrex AR-6485 CW4 Arthroscopy Pump is a system that maintains constant, non-pulsed control of intra-articular rinsing and distention pressure throughout all phases of an arthroscopic surgical procedure. The AR-6485 is intended to provide continuous pulse-free flow that reacts immediately to changes in the intra-articular pressure so that joint distention can be sustained even under high shaver extraction volumes or secondary outflow.
W A R N I N G ! All fluid inflow devices, including gravity assist, may cause fluid extravasations into the surrounding tissues. This extravasation may be mild, moderate, or severe. In severe cases, the resulting edema may result in a serious adverse patient event which may include compartment syndrome, nerve compromise, or death. Undiagnosed capsular defects will exacerbate fluid extravasation conditions. When utilizing any fluid management device, the patient (extremity and surrounding area) must be monitored closely by the surgical team for signs of excess fluid buildup. Fluid usage volumes should be monitored and compared to similar surgical procedures. With all arthroscopy pumps, correct setup and proper user operation is required. Always select the lowest possible pressures in order to achieve the required intra-articular distention. All alarms or alerts must be acknowledged and the appropriate troubleshooting procedure followed.
W A R N I N G ! FAILURE TO FOLLOW THE SETUP INSTRUCTIONS AND/OR CONTINUING TO USE THE PUMP WITHOUT RESOLVING AN ALARM CONDITION COULD RESULT IN A SERIOUS PATIENT ADVERSE EVENT. Failure to adhere to the setup instructions and use of Arthrex certified tubing may result in inaccurate pressure sensing and monitoring by the device. It is imperative that the user is aware that patient safety may be compromised when an alarm on the pump is ignored or silenced incorrectly. NEVER ignore or silence alarms. Follow the appropriate troubleshooting procedures and carefully monitor the patient. Only Arthrex certified tubing must be used.
W A R N I N G ! This device is only for use in normal arthroscopic procedures as described in the User’s Guide, under the supervision of a trained and licensed physician. This device should not be used by untrained personnel or used for indications other than those described in this User’s Guide.
DFU-0249-5r0_fmt_en-US
Page 8 of 50
Product Description CW4 Arthroscopy Pump User’s Guide The AR-6485 includes: • • • •
A universal medical-grade switching power supply that allows the pump to function automatically at voltage ranges found worldwide. An automatic Shaver Detect feature that allows a pressure boost to be created to adjust for high flow shaver suction on demand. There is an operator display for user inputs. A Lavage function to provide elevated pressure to stop bleeding.
The user-defined settings for inflow pressure are adjustable through controls located on the operator display or on the remote control. There are three Applied Part pump tubing options for the AR-6485: 1.
One-piece tubing only. This tubing, when used alone, must be replaced after each patient.
2.
One-piece tubing and extension tubing combination. The AR-6410 can be reused for an entire surgical day, while the AR6220 must be replaced after each patient.
3.
Two-piece tubing combination. The pump tubing can be reused for an entire surgical day. The patient tubing must be replaced after each surgical procedure.
The optional Y-Tubing Adapter is intended to be used with Arthrex inflow tubing sets/system combinations to connect up to four irrigation bags. The AR-6485 is an inflow-only irrigation pump. Other optional accessories: • • •
Remote control Foot pedal Synergy integration cable
DFU-0249-5r0_fmt_en-US
Page 9 of 50
Product Description CW4 Arthroscopy Pump User’s Guide
2.2
Product Features 2.2.1
AR-6485 Console: Front View
Figure 1 uses a numeric callout system to identify the main elements of the console’s front panel, which are listed and labeled in Table 1. These callouts are referenced throughout this User’s Guide. Figure 1
Front Panel of Console
Table 1
Front Panel Elements
1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13.
14.
DFU-0249-5r0_fmt_en-US
Inflow tubing track Inflow tubing alignment marker Inflow roller assembly Inflow roller housing door Operator display Synergy Product Logo Power Switch IEC 60417-5009 symbol (Stand-by) Stand-by power switch Remote control or foot pedal connector Remote control or foot pedal symbols Type BF Symbols (Electric Shock Protection) Inflow door locking mechanism Tubing sensor coupler indicator LED. A steady green LED indicates that the tubing is connected properly. A flashing red LED indicates that the tubing is not connected, or that it is connected incorrectly Tubing sensor coupler
Page 10 of 50
Product Description CW4 Arthroscopy Pump User’s Guide
2.2.2
AR-6485 Console: Rear View
Figure 2 uses a numeric callout system to identify the main elements of the console’s rear panel, which are listed and labeled in Table 2. These callouts are referenced throughout this User’s Guide. Figure 2
Rear Panel of Console
Table 2
Rear Panel Elements
16. 17. 18. 19. 20. 21. 22. 23. 24. 25. 26. 27.
2.2.3
AC mains power input plug Main power input - Fuse holder Address Void seal sticker USB port (For use ONLY with an Arthrex approved thumb drive) Serial IO for Arthrex integration (refer to IFU 950-0052-00) Serial number label Software label Model number label Equipotential ground symbol Equipotential ground connector Main power input - Fuse label
AR-6485 Operator Display Messages and Iconography
The console’s operator display [5] provides information about the status of the AR6485 modes, pressure, and flow settings in real time. Table 3 describes each message or button, cause and explanation when the pump is in the ready state.
DFU-0249-5r0_fmt_en-US
Page 11 of 50
Product Description CW4 Arthroscopy Pump User’s Guide Table 3
AR-6485 Operator Display Messages and Iconography
Message
Cause
Explanation
Arthrex CW4
Message appears when the power stand-by switch is activated.
Power on message display.
** Tubing Out **
Message appears when tubing is not plugged into the tubing sensor coupler [14].
Check tubing installation.
** Door Open **
Message appears when the roller housing door [4] is open.
Roller housing door is not closed.
** Over Pressure **
Message appears when the sensed pressure exceeds over-pressure software limit of 300 mmHg.
Software overpressure condition.
Critical Failure
Message appears on the first line of the operator display if one of three conditions is met: Failure Condition 1: ** Power Failure ** Appears if the power supply self-test fails when the pump is turned on. Failure Condition 2: ** OVP Detect Fail ** Appears if the hardware overpressure diagnostic test fails when the pump is turned on. Failure Condition 3: ** Sensor Failure ** Appears if the pump detects a problem with the pressure sensors.
Critical failure, cannot continue operation.
** Power Failure **
Message appears if the power supply self-test fails when the pump is turned on.
Power supply test fails.
** OVP Detect Fail **
Message appears if the hardware overpressure diagnostic test fails when the pump is turned on.
Hardware overpressure diagnostic fails.
** Sensor Failure **
Message appears if the pump detects a problem with the pressure sensors.
Sensor failure.
** Pressure Fault **
Message appears when the pump is unable to reach a desired set pressure within a specific amount of time. This typically indicates improperly installed tubing or a split in the tube from continuous use.
Insufficient pressure.
Remote Control Icon
Icon appears when the remote is attached.
Remote connected.
Foot Pedal Icon
Icon appears when the foot pedal is attached.
Foot pedal connected.
The operator display shows the pressure reading until the PRESSURE (+) button is pressed. Once pressed, the displayed pressure reading will change to the pressure setting. Each subsequent press of the pressure button will increase the pressure setting in increments of 5.
Pressure set increase.
The operator display shows the pressure reading until the PRESSURE (-) button is pressed. Once pressed, the displayed pressure reading will change to the pressure setting. Each subsequent press of the pressure button will decrease the pressure setting in increments of 5.
Pressure set decrease.
+ Button
− Button
DFU-0249-5r0_fmt_en-US
Page 12 of 50
Product Description CW4 Arthroscopy Pump User’s Guide Message
Cause
Explanation
RUN Button
Button appears when the pump is stopped. Press this button to start the pump.
Motor on.
STOP Button
Button appears when the pump is running. Press this button to stop the pump.
Motor off.
LAVAGE Button
Button appears when the pump is in run mode. Press the button and the pump will increase pressure by a user-defined amount and length of time.
Pressure increased for a set time.
BOOST Button
Button appears when the pump is in inflow only mode. Press the button and the user may define the pressure increase when the shaver is activated.
Pressure increased.
MENU Button
Button appears when the pump is stopped. Press the button and the pump will enter the setup menu.
User setups displayed.
DFU-0249-5r0_fmt_en-US
Page 13 of 50
Product Description CW4 Arthroscopy Pump User’s Guide
2.3
Foot Pedal Unit (AR-6483) The AR-6485 CW4 Arthroscopy Pump can be remotely controlled with the optional foot pedal unit (AR-6483). It provides a Lavage function. See Figure 3 and Table 4. Do not disconnect the plug of the foot pedal unit by pulling on the cable. Remove the foot pedal unit plug by grasping and pulling on the body of the connector. Figure 3 uses a lowercase Roman numeral callout system to identify the main elements on the foot pedal unit, which are listed and labeled in Table 4. These callouts are referenced throughout this User’s Guide.
Figure 3
Foot Pedal Unit (AR-6483)
i Table 4
ii Foot Pedal Elements (AR-6483)
i ii
DFU-0249-5r0_fmt_en-US
Lavage increases the pressure by a percentage and time selected by the user. Rinse is for use only with an inflow outflow pump.
Page 14 of 50