Synergy Resection
Quick Guide
Product Description AR-8305 Console – Front View On / Off power switch Handpiece connectors 3 Touch screen display 4 Footswitch connectors 1
2
3
2 1
4
2
4 AR-8305 Console – Rear View Main power input plug Equipotential bonding pin 3 Serial ports for Arthrex integration 4 USB port (for Arthrex service purposes) 1
2
1 2
3
4
AR-8330F, AR-8332H, AR-8332RH, AR-8330SJ Shaver Handpiece Suction connection Suction regulator 3 Cable plug with protection cab 4 Hand control buttons (AR-8332H, AR-8332RH) 5 Hub bore and locking pin 1
2
3 2
4
1
5
AR-8310W, AR-8310, AR-8315C, AR-8315 Footswitch Footswitch (SFS = Standard Footswitch, GFS = Gas Pedal) Definition of Direction
1
2
1 6 8
7
3 8
7
4 5
Handstück aktiviert
Handstück nicht aktiv
1.
Reverse / counterclockwise (REV)
2.
Forward / clockwise (FWD)
3.
Oscillation (OSC)
Main Screen Speed setting Speed Up button 3 Speed Down button 4 Oscillation mode (STD, EFF, AGG) 5 Settings button 6 Handpiece activated 7 Foot control handpiece connected 8 Standard footswitch connected 1
2
www.arthrex.com
Operation and Frequently Used Functions 1. Turn console power on.
2. Connect shaver handpiece(the console will recognize the handpiece type, and display a vailable options for the operator to adjust, as desired). ■■ Insert the connector from the handpiece into the handpiece receptacle of the console so that the red dots on the connector and receptacle align and engage easily.
■■ The shaver handpiece cable connects and locks to the console to prevent accidental separation during use. To avoid damage, only disconnect the shaver handpiece by pulling on the cable connector.
■■ Connect footswitch. 3. Adjust settings: a) Speed adjustment: Use Up and Down buttons. b) Oscillation modes: Choice of 3 different OSC Modes: ■■Standard (STD)
■■Efficient (EFF) ■■Aggressive (AGG)
■■The sterile suction hose is connected with the suction connection of the shaver handpiece.
■■Adjust the flow volume with the suction control. ■■To insert a shaver blade or burr, fully insert the hub into the hub bore of the handpiece. Watch the locking pin slide back and forth to check that it is locked in place. There will be an audible “click” sound when the handpiece is securely locked.
■■Use the RPM selection button on the front panel of the console, or the selection mechanism on the handpiece, or the footswitch to adjust the speed of the handpiece.
www.arthrex.com
Troubleshooting
Fault
Cause
Solution
Unit does not power on.
1. Not receiving power from the wall receptacle or the power strip.
1. Check the main power plug and wall receptacle.
2. Blown fuses.
2. Check the fuses.
Handpiece does not work.
Motor runs but shaver
3. Console has an internal power failure.
3. Send in for repairs.
1. Handpiece is not connected to the correct channel.
1. Check that the connector is plugged into the correct channel receptacle.
2. N ot fully or properly connected or connector is damaged.
2. Check that both connectors are plugged in properly.
3. Liquid in the receptacle or connector.
3. Dry thoroughly.
4. Connector is damaged.
4. Send in for repairs.
5. Footswitch is defective.
5. Operate the device using the console unit.
6. Footswitch is functional.
6. Replace the handpiece.
7. Handpiece is too hot after sterilization.
7. Allow it to cool down as indicated in the WARNING in the Synergy Resection User’s Guide.
8. Handpiece over-current is detected.
8. Allow 30 seconds for the console to reset the affected channel.
9. Defective handpiece.
9. Send in for repairs.
1. Shaver blade/burr incorrectly inserted.
1. Insert the shaver blade so that it engages. Fully close the locking mechanism.
2. Damaged hub.
2. Replace the shaver blade/burr.
3. Bent sleeve.
3. Replace the shaver blade/burr.
1. Clogged suction connection.
1. Rinse with a syringe.
2. C logged shaver blade or burr.
2. Clean with a brush.
3. Closed suction control.
3. Open suction control.
4. Damaged suction control.
4. Send in for repairs.
blades/burrs do not work.
No suction.