Arthrex
AR-9800 SynergyRF Quick Start Guide Rev B
Quick Start Guide
2 Pages
Preview
Page 1
Power Cord
QUICK START GUIDE FOR SYNERGY AR-9800
Generator
RF
Probe Connector
UNPACKING: Carefully unpack and inspect all components for shipping damage. Save the packaging for later transport of the device. The footswitch is provided separately. The probe is provided separately and is in sterile packaging.
Probe
Users must read the SynergyRF User’s Manual and ApolloRF™ Probe DFU prior to using these devices. The User Manual and DFU provide the indications and contraindications for use, warnings, symbol definitions and the full list of product instructions. These documents can be found on Arthrex.com.
Setup
1 Place console on a flat, dry surface
2 Connect power cord to console
3
4
Connect power cord to outlet
Suction Tube Footswitch Cable Connector
5
Power on console
Footswitch
6
Allow console to fully initialize
Attach probe via connector to console
Main Screen Display APOLLO 90
APOLLO 90
Ablate Power Settings 7 For suction probes, attach suction tube
8 Attach footswitch if desired
Coag Power Settings
9 Change settings using touch panel display
Device Recognition Indicator
Footswitch Icon
Menu Button
Troubleshooting: Faults and Solutions FAULT
SOLUTION
Unit does not power on
1. Check the main power plug and wall receptacle. 2. Check the main fuses per the User Guide instructions. 3. Call Arthrex® Tech Support at 1-888-420-9393.
Accessory ApolloRF™ probe does not work
1. Disconnect and reconnect probe. 2. Turn off console, wait 30 seconds, repower console. 3. If applicable, disconnect and reconnect footswitch, replace if footswitch not functioning. 4. Open new ApolloRF probe.
Yellow fault message
Refer to User Guide for proper disposition.
Red fault message (Critical Failure)
1. Make note of message displayed. 2. Call Arthrex Tech Support at 1-888-420-9393.
To Adjust Settings APOLLO 90
1. Press menu setting button
2. Press desired setting
To return or repair device: Contact Arthrex Product Surveillance at 1-866-267-9138.
For Language:
For Footswitch Override:
For Volume:
Select desired language and then do one of the following:
Select desired option and then do one of the following:
Arthrex, Inc. 1370 Creekside Blvd. Naples, FL 34108-1945 USA Toll Free: 1 (800) 934-4404 www.arthrex.com
Press HOME to save setting and return to main screen
Press HOME to save setting and return to main screen
Select desired volume using + or – buttons and then do one of the following:
-OR-
-OR-
Press Setting Button to save setting and return to MENU
Press Setting Button to save setting and return to MENU
-OR-
-OR-
Press X to return to MENU without changing default
Press X to return to MENU without changing default
Arthrex Tech Support: +1-888-420-9393
Arthrex GmbH Erwin-Hielscher-Strasse 9 81249 München, Germany Tel: +49 89 909005-0 www.arthrex.de +49 89 90 90 05 8800
Button Activation
Press HOME to save setting and return to main screen -OR-
Press Setting Button to save setting and return to MENU -OR-
Press X to return to MENU without changing default
1
1. Ablation power setting 2. Ablate 3. Coag
2016, Arthrex Inc. All rights reserved. LI1-00027-EN_B
©
2 3