Model AR- 6400 series Continuous Wave & DualWave Safety Test Quick Guide Rev C
Safety Test Quick Guide
2 Pages
Preview
Page 1
Safety test Quick guide
This is the “clamp test” for the main pump tubing set-up on the Arthrex Continuous Wave & DualWave Arthroscopic Pump (AR-6450, AR-6475, AR-6480) I
II
III
IV
1. Set up the pump tubing. a) First remove the protection cap ( I – IV ). b) Insert the connector fitting of the Pump Tubing into the Tubing Sensor Coupler. This step must be completed first to ensure accurate pressure measurements. c) Release the locking lever of the roller housing door and open the door completely. Allow the door to rest against the stop. The roller mechanism is now exposed.
green dot
d) Place the green-collared section of the Pump Tubing into the Tubing IN Guide indicated by the green dot. e) Guide the tubing boot over the rollers and insert the output side of the tubing boot into the Tubing OUT Guide.
locking lever
AR-6475
NOTE: The Pump Tubing is connected properly when the green connector on the Main Pump Tubing is aligned with the green dot on the front panel of the console. f ) Close and lock the roller housing door. g) Connect the tubing spikes to an bag.
2. After adjusting the required pressure using the Pressure Set buttons, remove the cap from the patient end of the tubing. AR-6480
green dot
www.arthrex.com
3. Assure that all appropriate tubing clamps are open. 4. Activate the pump motor by pressing the ON/OFF button. AR-6450
AR-6475
AR-6480
5. At the beginning of the pumping procedure the mem brane must be filled with air. Fill the entire length of the tubing with fluid to remove any air bubbles. Membrane Sensor Chamber
NOTE: It is not necessary to remove the air within the Sensor Chamber on the Pump Tubing Set.
6. After the air has been purged from the tubing, close the clamp at the patient end of the tubing. The Sensor Chamber must not be filled completely. Generally the top water level is at ⅓ Chamber height.
The rollers must stop turning, and interrupt the pumping procedure immediately. This is a safety check to ensure that the sensor system is working properly.
■■If the rollers do not stop, ensure clamp is firmly closed. ■■ If the rollers turn continuously, the connector fitting may not be functioning properly. → Replace the Pump Tubing. In case of continuing failure, please contact your sales representative and return the pump for investigation to Arthrex. To achieve the maximum patient safety, we would like to ask you to run this test prior to every use.