Arthrex
DrillSaw Mini 300 Battery-Powered System Instructions for Use
17 Pages
Preview
Page 1
DrillSaw Mini 300 Battery-Powered System ™
Equipment Setup and Overview
Inserting the battery
Assembling/removing the battery housing Switching forward/reverse modes
Assembling/removing the attachments
Sagittal saw attachment
Reciprocating saw attachment Chuck attachment
Quick connects Burring attachments
Pin/wire driver attachment Removing the battery
Sterilization and general tips
Inserting the battery
Press Sterile
1
2
3 Sterile
Press the locking device.
Open the battery housing and place the sterile plate on top.
This document is not intended to replace or supercede the DFU for the DrillSaw Mini 300™ Power System. It is only a tool to reference common questions related to these products (DFU-0023). For a full listing of instructions and warnings, please refer to the Instructions for Use: http://www.Arthrex.com.
Inserting the battery
Unsterile
4
5 Sterile
Note the arrow marks- (they serve as an aid)
6
Push the unsterile, fully charged battery as far as it will go into the battery housing. Push the loop on the battery down.
7
Unsterile
Remove the transfer plate.
8
9
Close the battery housing until it audibly snaps into place.
Assembling/removing the battery housing Press
Press
Assemble
1
Note the guide pins
Align the battery housing according to the guide pins and press the locking device. Press
2
Push the battery housing in a straight line to the end stop on the handpiece, while keeping the locking device pressed.
Removal
1
2
Press the locking device.
Pull the battery housing in a straight line off of the handpiece.
Switching forward and reverse modes
1
2
3
Safe/Off position (S or O)
Forward mode (F)
Reverse mode (R)
Forward mode
Safe/Off position Reverse mode
Assembling the attachments
1
2
3
Insert the position pins of the slightly twisted attachment into the recesses of the handpiece.
4
Turn the attachment until it audibly snaps into place.
Removing the attachments
1
Turn the handpiece chuck mechanism.
2
3
Pull the attachment off.
Sagittal saw attachment AR-300SAG
Press Press
1
2
3
Press the push button on the attachment.
Push the saw in as far as it will go.
Note that the saw audibly snaps into place and check that the connection is secure by pulling in an axial direction.
Reciprocating saw attachment AR-300SR
1
2
3
Flat shaft
Round shaft
Round shaft
Open the attachment chuck mechanism. Push the saw in and keep turning it until you can feel it snap into place.
Close the chucking system.
Check that the connection is secure by pulling in an axial direction.
Pin/wire driver attachments AR-300WD16 AR-300PD24
1
2
Pull the lever of the wire driver back completely and keep it pulled.
Insert the wire.
The AR-300CA Cannulated Adapter can be used for pins and wires longer than 6 in.
3
Release the lever to fix the wire in place. TIP: Lever pulled = wire not fixed Lever not pulled = wire fixed
Chuck attachment AR-300DK30
Hold ring while turning
AR-300DK45
1
2
Insert the drill bit into the open drilling chuck.
Fix the drill bit in the drilling chuck and verify the connection is secure by pulling in an axial direction.
Note: May use 4 in × 4 in (sterile cloth) to prevent glove damage.
Quick connects AR-300DAO-2
1
2
3
Pull the collar back to open the attachment.
Insert the rotating instrument and keep turning it until it snaps into place. Release the collar and check that the connection is secure by pulling in an axial direction.
Burring attachment AR-300B
1
2
Open the attachment chuck mechanism and insert the rotary instrument.
Close the chucking system and note the corresponding marks on the attachment. Check that the connection is secure by pulling in an axial direction.
Removing the battery
1
Press
2
Open the battery housing by pressing the locking device.
Remove the battery by pulling on the handle in an axial direction.
Unsterile
4
3
5
Sterilization
1
2
3
4 V-300T
Do not wash or sterilize the batteries.
Only wash or sterilize the battery housing when open and empty.
Only use the Arthrex autoclave basket (V-300T) or instrument case (AR-300C) for processing.
134˚C Arthrex recommends sterilization in accordance with EN 285/EN 13060, class B sterilization holding time a minimum of 4 minutes at 273°F (134°C) (drying time of at least 15 minutes). No lubrication necessary after processing.
AR-300C AR-300C
General tips Li-Ion Battery The batteries can remain in the charging station until the next surgery. They can also be stored fully charged next to the charging station.
No self-discharge!
Maintenance-free – DO NOT OIL! Cleaning, disinfection, and drying • All system components with this symbol can be cleaned and disinfected in a thermo washer disinfector. • Only place the attachment in the thermo washer disinfector in an upright position. • Do not place the handpiece, the battery housing, or the attachments in the ultrasonic bath. • Arrange for sufficient drying after the thermo disinfection.
This description of technique is provided as an educational tool and clinical aid to assist properly licensed medical professionals in the usage of specific Arthrex products. As part of this professional usage, the medical professional must use their professional judgment in making any final determinations in product usage and technique. In doing so, the medical professional should rely on their own training and experience and should conduct a thorough review of pertinent medical literature and the product’s Directions For Use. Postoperative management is patient specific and dependent on the treating professional’s assessment. Individual results will vary and not all patients will experience the same postoperative activity level or outcomes.
View U.S. Patent information at www.arthrex.com/corporate/virtual-patent-marking © 2018 Arthrex, Inc. All rights reserved. LI1-00048-EN_B