BIS VIEW™ Monitoring System OPERATING MANUAL
Rx only
EC REP
Aspect Medical Systems, Inc. One Upland Road Norwood, MA 02062 U.S.A. (Tel) 617-559-7000 (Tel) 888-BIS INDE(X) (U.S. only) (Fax) 617-559-7400 [email protected] www.aspectmedical.com
Aspect Medical Systems International B.V. Rijnzathe 7d2 3454 PV De Meern The Netherlands Tel: +31.30.662.9140 0123 Fax: +31.30.662.9150 [email protected] 070-0089
1.02
BIS VIEW™ Monitoring System OPERATING MANUAL
© Copyright, 2008, Aspect Medical Systems, Inc. All rights reserved. Copying or other reproduction of this document is prohibited without prior written consent of Aspect Medical Systems, Inc. BIS VIEW is a trademark of Aspect Medical Systems, Inc. Aspect, Bispectral Index, BIS, the BIS logo, BISx, the BISx logo and Zipprep are trademarks of Aspect Medical Systems, Inc. and are registered in the U.S.A., E.U. and other countries.
TABLE OF CONTENTS ABOUT THIS MANUAL... i INTRODUCING THE BIS VIEW MONITORING SYSTEM... ii 1
SAFETY PRECAUTIONS... 1-1
1.1
Warnings... 1-1
1.2
Cautions ... 1-3
1.3
Key to Symbols ... 1-6
2
SYSTEM SETUP AND PREPARATION FOR USE ... 2-1
2.1
BIS VIEW Monitor Setup and Checkout ... 2-1
2.2 Environment... 2-2 2.2.1 Shipping and Storage Environment... 2-2 2.2.2 Operating Environment ... 2-2 2.2.3 Power Requirements and System Grounding ... 2-3 2.2.4 Electromagnetic Compatibility Requirements... 2-4 2.2.5 Site Preparation: Mounting the Monitor ... 2-4 2.2.5.1 Mounting the Monitor using the Pole Clamp ... 2-4 2.2.5.2 Optional Mounting Accessories ... 2-5 2.3 The BIS VIEW Monitoring System – Equipment and Supplies ... 2-6 2.3.1 The BIS VIEW Monitor... 2-7 2.3.1.1 Front Panel... 2-7 2.3.1.2 Soft Keys ... 2-7 2.3.1.3 Alarm Key ... 2-7 2.3.1.4 ON/Standby button ... 2-7 2.3.1.5 Rear Panel ... 2-8 2.3.1.6 Integral Battery ... 2-9 2.3.2 BISx ... 2-10 2.3.3 Patient Interface Cable (PIC)... 2-11 2.3.4 BIS Sensor... 2-11 2.4
Cable Connections... 2-11
2.5 2.5.1 2.5.2
Start Procedure/Standby Mode ... 2-12 Starting the Monitor for the First Time ... 2-12 Starting the Monitor from Standby Mode... 2-12
2.6 2.6.1 2.6.2 2.6.3
Initial Menu Settings... 2-13 Language Selection... 2-13 Date and Time ... 2-13 Save Settings... 2-14
3
OPERATING THE BIS VIEW MONITORING SYSTEM ... 3-1
3.1
Preparing for Operation ... 3-1
3.2
Sensor Check... 3-4
3.3 3.3.1 3.3.2 3.3.3 3.3.4 3.3.5 3.3.6 3.3.7 3.3.8
BIS Number Display Screen ... 3-6 BIS (Bispectral Index) Value ... 3-6 Battery Icon... 3-7 Alarm Icon ... 3-7 Signal Quality Indicator ... 3-8 Electromyograph (EMG) Indicator ... 3-8 Case ID... 3-8 Message Region ... 3-8 Soft Key Screen Selections... 3-9
3.4 3.4.1 3.4.2
BIS Trend Display Screen with Sensor Status... 3-9 BIS Trend Graph ... 3-10 Target Range ... 3-10
3.5
BIS Trend Display Screen with EEG... 3-10
3.6 Menu Selections ... 3-11 3.6.1 Chart Data... 3-11 3.6.2 Export Data... 3-12 3.6.3 Setup ... 3-13 3.6.3.1 Screen Options ... 3-13 3.6.3.2 Alarms: The Alarms Menu... 3-14 3.6.3.2.1 Target Range ... 3-15 3.6.3.2.2 Alarm Volume Menu / Test Alarm Volume ... 3-16 3.6.3.3 Smoothing Rate: The Smoothing Rate Menu... 3-16 3.6.3.4 Date/Time... 3-17 3.6.3.5 Settings: Active and Saved Monitor Settings... 3-17 3.6.4 Maintenance ... 3-18 3.6.5 Diagnostics ... 3-18 3.6.6 Demo Case... 3-18 3.7
Ending a Case ... 3-19
3.8
Data Transfer ... 3-19
3.9 3.9.1 3.9.2 3.9.3 3.9.4
How the BIS VIEW Monitoring System Works ... 3-20 Bispectral Index (BIS) ... 3-21 System Self-Checks... 3-22 Data Memory ... 3-23 Battery Operation... 3-23
4
QUICK REFERENCE GUIDE ... 4-1
5
PREVENTIVE MAINTENANCE, CARE AND CLEANING... 5-1
5.1
Care and Cleaning ... 5-1
5.2 5.2.1 5.2.2 5.2.3 5.2.4 5.2.5 5.2.6 5.2.7
Maintenance ... 5-2 Replace the PIC (Patient Interface Cable) ... 5-2 Checking Cable Integrity ... 5-3 System Checkout ... 5-3 Checking the Battery... 5-3 Replacing the Battery ... 5-5 Replacing the Power Supply... 5-5 Checking Leakage Current... 5-6
5.3
Technical Documentation ... 5-7
5.4
Instrument Identification ... 5-8
6
DIAGNOSTICS AND TROUBLESHOOTING ... 6-1
6.1 6.1.1 6.1.2 6.1.3
Maintenance ... 6-1 Software Update... 6-1 Default Settings... 6-2 Language... 6-2
6.2 6.2.1 6.2.2 6.2.3 6.2.4
Diagnostics... 6-3 Impedance Checking... 6-3 Diagnostic Codes ... 6-4 System Configuration Information ... 6-4 Advanced Diagnostics ... 6-4
6.3
BIS VIEW System Messages and Corrective Actions... 6-5
6.4
Using the Reset button ... 6-11
6.5
What to do if the BIS VIEW Monitoring System Requires Service... 6-11
7
APPENDIX I: MENUS, PROCESSED VARIABLES AND GLOSSARY. 7-1
7.1
Menu Map ... 7-1
7.2
Menu Listing ... 7-2
7.3
Processed EEG Variables ... 7-3
7.4
Glossary... 7-4
8 APPENDIX II: SPECIFICATIONS, WARRANTY AND SOFTWARE LICENSE AGREEMENT... 8-1 8.1
Specifications... 8-1
8.2 8.2.1 8.2.2
Electromagnetic Compatibility Specifications ... 8-5 Accessories... 8-5 IEC 60601-1-2:2001 Electromagnetic Compatibility Guidance ... 8-6
8.3
Warranty ... 8-11
8.4
Software License Agreement... 8-13
TABLE OF FIGURES Figure 1 - Symbol Key (page 1 of 3) ... 1-6 Figure 2 - Pole Clamp... 2-5 Figure 3 - The BIS VIEW Monitoring System... 2-6 Figure 4 - Rear Panel ... 2-8 Figure 5 - BISx and PIC... 2-10 Figure 6 - Connecting the PIC... 3-3 Figure 7 - Sensor Check Graphic Screen... 3-4 Figure 8 - BIS Number Display ... 3-6 Figure 9 - Screen Features – BIS Trend Display Screen with Sensor Status... 3-9 Figure 10 - BIS Trend Display Screen with EEG... 3-10 Figure 11 - Trend Display with Target Range ... 3-14 Figure 12 - BIS Range Guidelines ... 3-21 Figure 13 - Replacing the Power Supply... 5-5 Figure 14 - BIS VIEW Menu Map ... 7-1
ABOUT THIS MANUAL This Operating Manual contains all of the information you need to set up and operate the Aspect™ Medical Systems’ BIS VIEW™ Monitoring System (Figure 3). It also includes specific cleaning and test procedures you may occasionally be required to perform. Although this manual is intended for trained medical personnel, it does not assume prior knowledge or experience with operator-programmable medical electronics devices. Keep this Operating Manual with the BIS VIEW monitor for use by the operator. This manual is also intended to be a service information manual for service technicians or biomedical engineering personnel. Before attempting to set up or use the BIS VIEW system, please familiarize yourself with the safety information provided in this section.
i
INTRODUCING THE BIS VIEW MONITORING SYSTEM Introduction The BIS VIEW Monitoring System is a user-configurable patient monitoring system designed to monitor the hypnotic state of the brain based on acquisition and processing of EEG signals. The BIS VIEW system processes raw EEG signals to produce a single number, called the Bispectral Index™, or BIS, which correlates with the patient's level of hypnosis. The BIS VIEW monitor display consists of: • • • • •
The current BIS number Trend graph of BIS values over time Raw EEG waveforms in real time Various signal quality indicators (EMG, SQI, SR) Alarm Indicator and Messages
The system performs self-tests to ensure that the monitor and its components are functioning properly and that impedance levels of patient sensors are within acceptable limits. Easy-to-use menus allow the user to change the data display and review stored data.
ii
Important Information about Using BIS Monitoring The BIS VIEW Monitoring System is intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in their proper use. The system, and all its associated parameters, is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals. The BIS Index, one of the BIS VIEW monitor’s output parameters, may be used as an aid in monitoring the effects of certain anesthetic agents; and its usage with certain anesthetic agents may be associated with a reduction in primary anesthetic use and a reduction in emergence and recovery time. Use of the BIS Index for monitoring to help guide anesthetic administration may be associated with the reduction of incidence of awareness with recall in adults during general anesthesia and sedation. BIS is a complex monitoring technology intended for use as an adjunct to clinical judgment and training. Clinical judgment should always be used when interpreting the BIS in conjunction with other available clinical signs. Reliance on the BIS alone for intraoperative anesthetic management is not recommended. As with any monitored parameter, artifacts and poor signal quality may lead to inappropriate BIS values. Potential artifacts may be caused by poor skin contact (high impedance), muscle activity or rigidity, head and body motion, sustained eye movements, improper sensor placement and unusual or excessive electrical interference. BIS values should also be interpreted cautiously with certain anesthetic combinations, such as those relying primarily on either ketamine or nitrous oxide/narcotics to produce unconsciousness. Due to limited clinical experience in the following applications, BIS values should be interpreted cautiously in patients with known neurological disorders and those taking other psychoactive medications. The BIS education site, www.biseducation.com, offers relevant information and published articles on the clinical use of BIS. In addition, there is a “Monitoring Consciousness Using the Bispectral Index During Anesthesia” Clinician’s Pocket Guide available on the website and through your local Aspect Representative. For more information, please contact Aspect Medical Systems at (800) 442-2051. If you require additional information on the use of BIS, please contact Aspect Medical Systems Clinical Support at 800-442-8655 or 617-559-7655 if calling from outside of the USA.
iii
SECTION 1 SAFETY PRECAUTIONS ______________________________________________________________________ SECTION 1
1 SAFETY PRECAUTIONS INTRODUCTION Caution: Carefully read this entire manual before using the monitor in a clinical setting.
WARNINGS, CAUTIONS, AND NOTES The terms warning, caution, and note have specific meanings in this manual. • • •
A WARNING advises against certain actions or situations that could result in personal injury or death. A CAUTION advises against actions or situations that could damage equipment, produce inaccurate data, or invalidate a procedure, although personal injury is unlikely. A NOTE provides useful information regarding a function or procedure.
KEY TO SYMBOLS A key to the symbols that may appear on the BIS VIEW system appears at the end of this section.
1.1
Warnings
EXPLOSION HAZARD: DO NOT USE THE BIS VIEW SYSTEM IN A FLAMMABLE ATMOSPHERE OR WHERE CONCENTRATIONS OF FLAMMABLE ANESTHETICS MAY OCCUR. MONITOR IS NOT DESIGNED FOR USE IN MRI ENVIRONMENT. USE ONLY THE POWER CORD SUPPLIED BY THE MANUFACTURER. NEVER ADAPT THE PLUG FROM THE MONITOR TO FIT A NONSTANDARD OUTLET. U.S.A. REQUIREMENT: FOR PROPER GROUNDING, THE POWER RECEPTACLE MUST BE A THREE-WIRE GROUNDED OUTLET. A HOSPITAL GRADE OUTLET IS REQUIRED. NEVER ADAPT THE THREEPRONG PLUG FROM THE MONITOR TO FIT A TWO-SLOT OUTLET. IF THE OUTLET HAS ONLY TWO SLOTS, MAKE SURE THAT IT IS REPLACED WITH A THREE-SLOT GROUNDED OUTLET BEFORE ATTEMPTING TO OPERATE THE MONITOR.
1-1
SECTION 1 SAFETY PRECAUTIONS ______________________________________________________________________ IF THE INTEGRITY OF THE EXTERNAL PROTECTIVE EARTH GROUND IS IN DOUBT, THE BIS VIEW SYSTEM SHALL BE OPERATED FROM ITS INTERNAL BATTERY POWER SOURCE ONLY. BE SURE THE MONITOR IS MOUNTED SECURELY IN PLACE TO AVOID PERSONAL OR PATIENT INJURY. WHEN CONNECTING EXTERNAL EQUIPMENT (e.g., DATA CAPTURE COMPUTER), THE SYSTEM LEAKAGE CURRENT MUST BE CHECKED AND MUST BE LESS THAN THE IEC 60601-1-1 LIMIT. THE USE OF ACCESSORY EQUIPMENT NOT COMPLYING WITH THE EQUIVALENT SAFETY REQUIREMENTS OF THIS EQUIPMENT MAY LEAD TO A REDUCED LEVEL OF SAFETY OF THE RESULTING SYSTEM. CONSIDERATION RELATING TO THE CHOICE SHALL INCLUDE: • USE OF THE ACCESSORY IN THE PATIENT VICINITY • EVIDENCE THAT THE SAFETY CERTIFICATION OF THE ACCESSORY HAS BEEN PERFORMED IN ACCORDANCE TO THE APPROPRIATE IEC 60601-1 AND/OR IEC 60601-1-1 HARMONIZED NATIONAL STANDARD. DUE TO ELEVATED SURFACE TEMPERATURE, DO NOT PLACE BISx™ IN PROLONGED DIRECT CONTACT WITH PATIENT’S SKIN, AS IT MAY CAUSE DISCOMFORT. THE CONDUCTIVE PARTS OF ELECTRODES OR SENSOR AND CONNECTORS SHOULD NOT CONTACT OTHER CONDUCTIVE PARTS, INCLUDING EARTH. TO REDUCE THE HAZARD OF BURNS DURING USE OF HIGHFREQUENCY SURGICAL EQUIPMENT, THE SENSOR OR ELECTRODES SHOULD NOT BE LOCATED BETWEEN THE SURGICAL SITE AND THE ELECTRO-SURGICAL UNIT RETURN ELECTRODE. TO REDUCE THE HAZARD OF BURNS DURING USE OF BRAINSTIMULATING DEVICES (e.g., TRANSCRANIAL ELECTRICAL MOTOR EVOKED POTENTIAL), PLACE STIMULATING ELECTRODES AS FAR AS POSSIBLE FROM THE BIS SENSOR AND MAKE CERTAIN THAT SENSOR IS PLACED ACCORDING TO PACKAGE INSTRUCTIONS. THE SENSOR MUST NOT BE LOCATED BETWEEN DEFIBRILLATOR PADS WHEN A DEFIBRILLATOR IS USED ON A PATIENT CONNECTED TO THE BIS VIEW SYSTEM. TO MINIMIZE THE RISK OF PATIENT STRANGULATION, THE PATIENT INTERFACE CABLE (PIC) MUST BE CAREFULLY PLACED AND SECURED.
1-2
SECTION 1 SAFETY PRECAUTIONS ______________________________________________________________________ DO NOT EXPORT BISx HISTORY DATA WHILE A CASE IS IN PROGRESS. SHOCK HAZARD: DO NOT ATTEMPT TO DISCONNECT THE POWER CORD WITH WET HANDS. MAKE CERTAIN THAT YOUR HANDS ARE CLEAN AND DRY BEFORE TOUCHING THE POWER CORD. UNIVERSAL PRECAUTIONS SHALL BE OBSERVED TO PREVENT CONTACT WITH BLOOD OR OTHER POTENTIALLY INFECTIOUS MATERIALS. PLACE CONTAMINATED MATERIALS IN REGULATED WASTE CONTAINER. DO NOT MIX DISINFECTING SOLUTIONS (e.g., BLEACH AND AMMONIA), AS HAZARDOUS GASES MAY RESULT. ELECTRICAL SHOCK HAZARD: DO NOT REMOVE MONITOR COVERS DURING OPERATION OR WHILE POWER IS CONNECTED TO MONITOR. ELECTRICAL SHOCK HAZARD: THE MANUFACTURER'S INSPECTION OF THIS APPARATUS VERIFIED THAT THE GROUND LEAKAGE CURRENT AND THE PATIENT SAFETY CURRENT WERE LESS THAN THE SPECIFIED LIMITS ESTABLISHED BY THE APPLICABLE SAFETY STANDARDS. AS A MATTER OF SAFE PRACTICE, THE INSTITUTION SHOULD CONDUCT PERIODIC TESTS TO VERIFY THESE CURRENTS. WHENEVER AN EVENT SUCH AS SPILLAGE OF BLOOD OR SOLUTIONS OCCURS, RE-TEST LEAKAGE CURRENT BEFORE FURTHER USE. LEAKAGE CURRENT MUST BE CHECKED BY A QUALIFIED BIOMEDICAL ENGINEERING TECHNICIAN WHENEVER INSTRUMENT CASE IS OPENED. POWER SUPPLY IS INTERNALLY FUSED. REPLACE POWER SUPPLY ONLY WITH ASPECT MEDICAL SYSTEMS BIS VIEW POWER SUPPLY.
1.2
Cautions
Read this entire manual carefully before using the monitor in a clinical setting. Do not autoclave the BISx or Monitor. Autoclaving will seriously damage both components. Do not block ventilation inlet holes on the underside of monitor. Do not open the BISx for any reason. The seal to prevent liquids from entering the BISx may be damaged if opened.
1-3
SECTION 1 SAFETY PRECAUTIONS ______________________________________________________________________ The BIS VIEW system has been designed to operate with a BIS sensor. The sensor is a silver/silver chloride electrode array that utilizes Aspect's patented Zipprep™ technology and uses a proprietary connector. Use of other electrodes is not recommended. Do not remove drive while export is in progress. Continuous impedance checking may need to be disabled if the 1 nanoampere 128 Hz impedance check signal interferes with other equipment (e.g., evoked potential monitors). Considerations when using Electro-Convulsive Therapy (ECT) equipment during BIS monitoring: Place ECT electrodes as far as possible from the BIS sensor to minimize the effect of interference. Certain ECT equipment may interfere with the proper function of the BIS monitoring system. Check for compatibility of equipment during patient setup. Check the battery periodically by operating a BIS VIEW monitor that has been disconnected from the wall socket and that has been charged to full capacity (at least 6 hours of charge time). After long periods of storage, charge the battery for 6 hours to assure full capacity. If the BIS VIEW monitor fails to operate reliably from the battery for approximately 45 minutes, battery replacement is required. The BIS VIEW monitor contains an internal Lithium ion battery. The battery must be removed by a qualified service technician and disposed of or recycled in accordance with the national laws of the country. Contact Aspect Medical Systems, Inc. or the local distributor for a replacement battery: Aspect part number 186-0208. Avoid liquid ingress to the Patient Interface Cable. Contact of fluids with the PIC sensor connector can interfere with PIC performance. The BIS VIEW system complies with the electromagnetic compatibility requirements of IEC 60601-1-2. Operation of this device may affect or be affected by other equipment in the vicinity due to electromagnetic interference (EMI). If this occurs: • Increase separation between devices • Re-orient device cabling • Plug devices into separate outlet circuit branches Refer to Section 8.2 “Electromagnetic Compatibility Specifications.” Do not disconnect the BISx during the software update. When connecting or disconnecting BISx, take care not to touch the exposed contacts of either connector. Damage due to electrostatic discharge may result.
1-4
SECTION 1 SAFETY PRECAUTIONS ______________________________________________________________________ Using accessories other than those specified may result in increased electromagnetic emissions or decreased electromagnetic immunity of the BIS VIEW Monitoring System. The BIS VIEW Monitor should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the BIS VIEW Monitor should be observed to verify normal operation in the configuration in which it will be used. To completely remove power from the unit: put the monitor in standby mode, disconnect power cord from the power cord receptacle of the monitor, then remove the battery from the monitor. The BIS VIEW monitor may not power up entirely if battery power is low. If that should occur, connect unit to wall power and press the Reset button. (Refer to Section 6.4 “Using the Reset Button”). Service or repairs must be performed only by qualified biomedical technicians. Important: The BIS VIEW systems comply with the European Medical Device Directive (MDD) and applicable regulatory requirements of the country distributed to and carry the CEXXXX Marking. Declarations of Conformity provided upon request where appropriate.
1-5
SECTION 1 SAFETY PRECAUTIONS ______________________________________________________________________
1.3
Key to Symbols Caution: Consult Accompanying Documents
USB-A
Universal Serial Bus: Type A
Data I/O, RS-232 Serial Port USB-B
Universal Serial Bus: Type B
Caution: Hot Surface
Storage Temperature Limits
Type BF Equipment
Type BF Equipment Defibrillator-proof
Alternating Current (A/C)
Direct Current (D/C)
Battery Location
Reset Button
Monitor Power ON
Monitor Power OFF or Standby Mode
Do not Reuse
Use by YYYY-MM-DD or YYYY-MM
Latex-free product
PVC-free product
Manufacturer
Date of Manufacture
Authorized Representative in the European Community
Catalog Number
Batch Code
Serial Number
Conformité Européenne (CE) Marking of Conformity to European Medical Device Directive. CEXXXX represents the Notified Body number Recognized under the Component Recognition Program of Underwriters Laboratories Inc.
Classified by Underwriters Laboratories Inc.® with respect to electric shock, fire and mechanical hazards only, in accordance with UL 60601-1 and IEC60601-2-26 Packaging Labeling: Fragile, Keep Dry, and This Side Up
Figure 1 - Symbol Key (page 1 of 3) 1-6
SECTION 1 SAFETY PRECAUTIONS ______________________________________________________________________
Crossed out wheelie bin indicates separate treatment from general waste at end of life *
Recyclable
Product marked with a number contains certain toxic or hazardous substances or elements, and can be used safely during its Environment-Friendly Use Period (EFUP). The product should be recycled. The Environment-Friendly Use Period is valid only when the product is operated under the conditions defined in the product manual. **
Product marked with the “e” does not contain any toxic or hazardous substances or elements, and is green and environmental. The product can be recycled. **
*Contact Aspect Medical Systems International B.V. for a Return Materials Authorization (RMA) number or contact your municipality or the nearest collection site to dispose of waste equipment. According to the WEEE Directive 2002/96/EC, all waste electrical and electronic equipment (EEE) should be disposed of and collected separately and treated according to the best available and environmentally friendly techniques. EEE contains hazardous substances to the (human) environment but also EEE is a valuable resource of new raw materials. Therefore it is important to collect WEEE separately from other waste. Aspect Medical products are subject to the Directive and we therefore urge you to dispose of the equipment separately from ‘normal’ household waste and make sure that it is treated at an electronics recycler. ** Refer to www.aspectmedical.com for Material Declaration Data Sheets. Figure 1 - Symbol Key (Page 2 of 3)
1-7
SECTION 1 SAFETY PRECAUTIONS ______________________________________________________________________
Operating on Battery
No Battery is Installed in Monitor
Ringing Bell Icon – High Priority Alarm Condition exists. The alarm sounds unless the alarm key has been pressed to silence it. Yellow Bell with dotted ‘X’ - Alarms Paused
Red Bell with ‘X’ - Alarms Silenced
Figure 1 - Symbol Key (page 3 of 3)
1-8
SECTION 2 SYSTEM SETUP AND PREPARATION FOR USE ______________________________________________________________________ SECTION 2
2 SYSTEM SETUP AND PREPARATION FOR USE ____________________________________________________
INTRODUCTION This section provides setup instructions for the Aspect BIS VIEW Monitor, BISx, and accessories. It includes: • • • • • •
2.1
Setup checklist Proper environment Required equipment and supplies Cable connections Start and shutdown procedures Initial menu settings
BIS VIEW Monitor Setup and Checkout
1. Open packages and inspect for all components: • Monitor (P/N 185-0205) • Power cord • Pole clamp • BISx (P/N 185-0145-AMS) • PIC (Patient interface cable, connects BISx to patient) Sensors are sold separately. For a list of available sensors please contact Aspect Medical Systems, Inc. or your local distributor. 2. Connect power cable to monitor, plug power plug into appropriate wall outlet. • Verify that light to right of ON/Standby button is yellow. 3. Start up monitor by pressing the ON/Standby button (lower right corner). • Verify that light to right of ON/Standby button is green. • Verify all self-tests complete successfully. (Do not connect or disconnect equipment, or press keys until the monitor has completed its tests.) • Verify next screen says “Connect BISx.” 4. Connect BISx with PIC to monitor. • Verify that screen says “Connect sensor or cable.” 5. Connect PIC and sensor. • Verify SENSOR CHECK begins. 6. Disconnect power cord from monitor.
2-1
SECTION 2 SYSTEM SETUP AND PREPARATION FOR USE ______________________________________________________________________ • •
Verify ‘OPERATING ON BATTERY BACKUP’ is displayed Verify battery icon displays.
7. Reconnect power cord. • Verify battery icon is not displayed. • Verify “OPERATING ON BATTERY BACKUP” is not displayed. 8. End of checkout.
2.2
Environment
2.2.1 Shipping and Storage Environment The monitor and its accessories can be stored or shipped within the following environmental limits. Note that these limits apply to non-operational storage and shipping situations. Temperature Humidity Pressure
-10°C to +60°C 15% to 95% (non-condensing) 800 mm Hg (1500 feet below sea level) to 360mm Hg (20,000 feet above sea level)
Protect the monitor from sudden temperature changes that can lead to condensation within the instrument. To minimize condensation, avoid moving the system between heated buildings and outside storage. Once moved inside, allow the monitor to stabilize in the unopened shipping container at the inside ambient temperature before unpacking and placing into service. Before operation, wipe down all visible condensation and allow the system to reach equilibrium at room temperature.
2.2.2 Operating Environment The BIS VIEW Monitoring System is not designed for use in areas containing flammable gases or vapors. WARNING: EXPLOSION HAZARD: DO NOT USE THE BIS VIEW SYSTEM IN A FLAMMABLE ATMOSPHERE OR WHERE CONCENTRATIONS OF FLAMMABLE ANESTHETICS MAY OCCUR. MONITOR IS NOT DESIGNED FOR USE IN MRI ENVIRONMENT. The BIS VIEW monitor is designed to operate safely under the following conditions. Conditions outside these ranges could affect reliability. Temperature Humidity
0°C to +40°C 15% to 95% (non-condensing)
2-2