Tender LiV™ Care (TLC)
Catalog Number 340-8230, 340-8250, 340-7450, 340-7451, 340-7470, 340-7498, 340-7499, 340-7495, 340-7850, 340-7475, 340-7480, 340-7484, 340-7485, 340-7486, 340-7800, 340-7800S, 340-7800L, 340-7802, 340-7802S, 340-7802L, 340-7803, 340-7803S, 340-7803L, 340-7805, 340-7805S, 340-7805L, 340-7806, 340-7806S, 340-7806L, 340-7820, 340-7820S, 340-7820L, 340-7821, 340-7821S, 340-7821L, 340-7824, 340-7824S, 340-7824L, 340-7824NM, 340-7825, 340-7825S, 340-7825L, 340-7830, 340-7830S, 340-7830L, 340-7831, 340-7831S, 340-7831L, 340-7835, 340-7835S, 340-7835L, 340-7840, 340-7840S, 340-7840L, 340-7862, 340-7862S, 340-7862L
BEFORE THE PROCEDURE
Storage Transport
DURING THE PROCEDURE
Preparation Storage
AFTER THE PROCEDURE
Cleaning and Sterilization
Before the Procedure Reference: Instructions for Reprocessing of Reusable Devices, P/N 440-11975-01 • Store instruments in the proper tray at all times • Transport instruments in tray to avoid damage • Test function prior to use Contact Customer Service for replacements European Representative: Köntges SPRL Avenue Hellevelt 35 B-1180 Brussels Belgium Tel: +32 (0) 2 375 51 63 FAX: +32 (0) 2 375 89 06 e-mail: [email protected]
Manufactured by: 7555 Innovation Way Mason, Ohio 45040 USA Customer Service: 1-866-349-2342 (toll free) 1-513-755-4100 (phone)
During the Procedure • Flush instruments with saline periodically • DO NOT ALLOW BLOOD TO DRY ON INSTRUMENTS After the Procedure CAUTION! • Avoid caustic cleaners Cleaning solutions should be pH neutral • Avoid automated washers Instruments should be hand washed • Avoid tap water Use DI or RO water only • Avoid wet loads Instruments must be dry when stored Prior to Sterile Processing Department (SPD) • Hand rinse and flush with sterile water • Send to Sterile Processing Department (SPD) immersed in sterile water Cleaning at Sterile Processing Department (SPD) • Fill sink with Enzymatic Cleaner and de-ionized (DI) or reverse osmosis filtered (RO) water to manufacturer specifications • Immerse entire instrument in the sink • Using syringe with flush port adapter attached, flush instrument 2x or until no debris is seen exiting the instrument • Using soft cleaning brush, scrub jaws of instrument until clean LiV™ Care and Handling Guide
Page 1 of 2
P/N 490-13978-01 Rev G
Tender LiV™ Care (TLC) • • • •
Actuate instrument to assure free movement and that all spots are clean Rinse with DI or RO water Flush with DI or RO water 2x or until no cleaner is present Using air hose and compressed air, blow instrument dry (especially down the flush port)
Steam Sterilization • Please be sure to use DI or RO water • Pre-Vacuum, Steam Sterilization – 132C (270F) to 137C (278.6F) 4 Minutes (recommended) • Pre-Vacuum, Steam Sterilization – 134C (275F) to 137C (278.6F) 3 Minutes Keep instruments in the proper tray at all times for transport to avoid damage - DO NOT STACK INSTRUMENTS ON EACH OTHER Disposal Method •
Follow local governing ordinances and recycling plans regarding disposal or recycling of device components.
European Representative: Köntges SPRL Avenue Hellevelt 35 B-1180 Brussels Belgium Tel: +32 (0) 2 375 51 63 FAX: +32 (0) 2 375 89 06 e-mail: [email protected]
Manufactured by: 7555 Innovation Way Mason, Ohio 45040 USA Customer Service: 1-866-349-2342 (toll free) 1-513-755-4100 (phone)
Graphic Symbols for Device Labeling Catalogue Number
Serial Number
Quantity
Date of Manufacture
Manufacturer/Company Address
Authorized Representative in the European Union
Instrument supplied NonSterile
CE mark, Product conforms with the essential requirements in the European Medical Devices Directive 93/42/EEC.
Does not contain Natural Rubber Latex
Contains no di (2-ethylhexyl) Consult Instructions For Use phthalate (DEHP) released from polyvinyl chloride (PVC)
LiV™ Care and Handling Guide
Page 2 of 2
Caution: U.S. Federal law restricts this device to sale by or on the order of a physician or other licensed practitioner.
P/N 490-13978-01 Rev G