B.Braun Medical
Bipolar Forceps, straight, 7", (180 mm), working length: 2 3/4", (70 mm), bayonet-shaped, insulated, width: 0,70 mm
Bipolar Coagulation Forceps Instructions for Use
3 Pages
Preview
Page 1
Aesculap®
Aesculap Neurosurgery
Instructions for use Bipolar coagulation forceps Gebrauchsanweisung Bipolare Koagulations-Pinzetten Mode d'emploi Pinces de coagulation bipolaires Instrucciones de manejo Pinzas de coagulación bipolares Istruzioni per l'uso Pinzette per coagulazione bipolari Instruções de utilização Pinças de coagulação bipolares Gebruiksaanwijzing Bipolaire coagulatiepincetten Bruksanvisning Bipolära koagulations-pincetter Инструкция по применению Биполярные коагуляционные пинцеты Návod k použití Bipolární koagulační pinzety Instrukcja użytkowania Bipolarne pęsety koagulacyjne Návod na používanie Bipolárne koagulačné pinzety Kullanım Kılavuzu Bipolar koagülasyon pensleri
Aesculap AG | Am Aesculap-Platz | 78532 Tuttlingen | Germany Phone +49 (0) 7461 95-0 | Fax +49 (0) 7461 95-26 00 | www.aesculap.com Aesculap – a B. Braun company TA-Nr. 013772
11/12
- DIR 93/42/EEC Technical alterations reserved
V6
Änd.-Nr. 46258
General information
Aesculap® Bipolar coagulation forceps Symbols on product and packages Caution, general warning symbol Caution, see documentation supplied with the product
Intended use Bipolar coagulation forceps are used for hemostatic coagulation in surgical procedures.
Safe handling and preparation CAUTION Federal law restricts this device to sale by, or on oder of a physician! ► Ensure that the product and its accessories are operated and used only by persons with the requisite training, knowledge, or experience. ► Read, follow, and keep the instructions for use. ► Use the product only in accordance with its intended use, see Intended use. ► Remove the transport packaging and clean the new product, either manually or mechanically, prior to its initial sterilization. ► Store any new or unused products in a dry, clean, and safe place. ► Prior to each use, inspect the product for loose, bent, broken, cracked, worn, or fractured components. ► Do not use the product if it is damaged or defective. Set aside the product if it is damaged. ► Replace any damaged components immediately with original spare parts.
Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore the time interval between application and processing should not exceed 6 h; also, neither fixating pre-cleaning temperatures >45 °C nor fixating disinfecting agents (active ingredient: aldehydes/alcohols) should be used. Excessive measures of neutralizing agents or basic cleaners may result in a chemical attack and/or to fading and the laser marking becoming unreadable visually or by machine for stainless steel. Residues containing chlorine or chlorides e.g. in surgical residues, medicines, saline solutions and in the service water used for cleaning, disinfection and sterilization will cause corrosion damage (pitting, stress corrosion) and result in the destruction of stainless steel products. These must be removed by rinsing thoroughly with demineralized water and then drying. Additional drying, if necessary. Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems: ■ Optical changes of materials, e.g. fading or discoloration of titanium or aluminum. For aluminum, the application/process solution only needs to be of pH >8 to cause visible surface changes. ■ Material damage such as corrosion, cracks, fracturing, premature aging or swelling. ► Do not use metal cleaning brushes or other abrasives that would damage the product surfaces and could cause corrosion. ► Further detailed advice on hygienically safe and material-/value-preserving reprocessing can be found at www.ak-i.org, link to Publications, Red Brochure – Proper maintenance of instruments.
Preparations at the place of use ► If applicable, rinse non-visible surfaces preferably with deionized water, with a disposable syringe for example. ► Remove any visible surgical residues to the extent possible with a damp, lint-free cloth. ► Transport the dry product in a sealed waste container for cleaning and disinfection within 6 hours.
Cleaning/disinfection Product-specific safety notes on the reprocessing procedure
Risk of injury from ignition or explosion of flammable gases! Sparks may occur when using the HF device as directed. ► Observe the safety guidelines in the instructions for use of the HF device.
CAUTION
WARNING
WARNING
Thermal injuries to patients/users due to insufficient insulation of leads in active accessories! ► Adjust the HF device to an appropriate setting to ensure that the peak output voltage does match or not exceed the accessory voltage rating specified for the product.
The manufacturer has tested the product and verified that its insulation can withstand 20 reprocessing cycles. In clinical practice, the service life will depend on the individual intraoperative usage and the hospital's specific reprocessing conditions. ► Prior to each use, inspect products for: damage or surface changes to the insulation. ► Immediately sort out damaged or inoperative products and have them sent to Aesculap Technical Service, see Technical Service. ► Adjust the HF power output to the intended surgical intervention. Take into account clinical experience or reference values. ► Select the lowest possible HF power output. ► Keep the product’s contact surfaces clean during surgery. Remove encrusted tissue residues or body fluids with a moistened swab. The accessory voltage rating of the product is 600 Vp. The accessory voltage rating must exceed or match the peak output voltage with which the product is operated in combination with a suitable HF device at an appropriate operating mode/setting (see IEC/DIN EN 60601-2-2). To avoid HF burns: ► Always keep the working end of the product in the user’s field of vision whenever the HF power is activated. ► Prior to activating the HF device, check that the working end of the product is not touching any electrically conductive accessories. ► Prior to each use, visually inspect the product for: damage or surface changes to the insulation. ► Never place the product on or next to the patient. ► When using accessories for endoscopy or laparoscopy, deactivate the automatic switch-on mode of the HF device. ► Follow the instructions for use of the HF device.
Damage to the product due to inappropriate cleaning/disinfecting agents and/or excessive temperatures! ► Use cleaning and disinfecting agents according to the manufacturer’s instructions which – are approved for use, for example, on aluminum, plastic materials, and high-grade steel, – do not attack softeners (e.g. in silicone). ► Observe specifications regarding concentration, temperature and exposure time.
Immersion treatment in a 3 % H2O2 solution for approx. 5 minutes is a particularly effective and gentle method to dissolve encrustations from HF instruments. Subsequently, the debris can be removed by hand, with a medium-hard brush and or in an ultrasonic bath. This is followed by the conventional reprocessing steps. ► Carry out ultrasound cleaning: – as an effective mechanical supplement to manual cleaning/disinfection. – as a pre-cleaning procedure for products with encrusted residues, in preparation for mechanical cleaning/ disinfection. – as an integrated mechanical support measure for mechanical cleaning/disinfection. – for additional cleaning of products with residues left after mechanical cleaning/disinfection.
Validated cleaning and disinfection procedure Validated procedure
Specific requirements
Reference
Manual cleaning with ultrasound and immersion disinfection
■ 20 ml disposable syringe ■ Keep working ends open for
Chapter Manual cleaning/disinfection and sub-chapter:
cleaning.
■ Chapter Manual cleaning with ultrasound and immersion disinfection
■ Drying phase: Use a lint-free cloth or medical compressed air Manual pre-cleaning with ultrasound and brush, and subsequent mechanical alkaline cleaning and thermal disinfection
■ Place the instrument in a tray that is suitable for cleaning (avoiding rinsing blind spots).
■ Keep working ends open for cleaning.
■ Place instruments in the tray
Chapter Mechanical cleaning/disinfection with manual pre-cleaning and sub-chapter:
■ Chapter Manual pre-cleaning with ultrasound and brush
■ Chapter Mechanical alkaline
with their hinges open.
Safe operation Risk of injury and/or malfunction! ► Always carry out a function check prior to using the product. WARNING
Validated reprocessing procedure
cleaning and thermal disinfecting
Manual cleaning/disinfection ► Prior to manual disinfecting, allow water to drip off for a sufficient length of time to prevent dilution of the dis-
infecting solution. ► After manual cleaning/disinfection, check visible surfaces visually for residues. ► Repeat the cleaning /disinfection process if necessary.
Manual cleaning with ultrasound and immersion disinfection Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
Note Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for sterile processing.
I
Ultrasonic cleaning
RT (cold)
>15
2
D–W
Note For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possible variants of CJD, observe the relevant national regulations concerning the reprocessing of products.
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Intermediate rinse
RT (cold)
1
-
D–W
-
Note Mechanical reprocessing should be favored over manual cleaning as it gives better and more reliable results.
III
Disinfection
RT (cold)
15
2
D–W
Note Successful processing of this medical device can only be ensured if the processing method is first validated. The operator/sterile processing technician is responsible for this. The recommended chemistry was used for validation.
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
IV
Final rinse
RT (cold)
1
-
FD-W
-
V
Drying
RT
-
-
-
-
General safety instructions
Note If there is no final sterilization, then a virucidal disinfectant must be used. Note For the latest information on reprocessing and material compatibility see also the Aesculap extranet at www.extranet.bbraun.com The validated steam sterilization procedure was carried out in the Aesculap sterile container system.
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) RT: Room temperature *Recommended: BBraun Stabimed
► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure. Phase I ► Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for at least 15 min. Ensure that all accessible surfaces are immersed and acoustic shadows are avoided. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Drain any remaining water fully. Phase III ► Fully immerse the product in the disinfectant solution. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Rinse lumens at least five times at the beginning of the exposure time with an appropriate disposable syringe. Ensure that all accessible surfaces are moistened. Phase IV ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during final rinse. ► Rinse lumens with an appropriate disposable syringe at least five times. ► Drain any remaining water fully.
Inspection, maintenance and checks Damage (metal seizure/friction corrosion) to the product caused by insufficient lubrication! ► Prior to function checks, lubricate moving parts (e.g. joints, pusher components and threaded rods) with maintenance oil suitable for the respective sterilization process (e.g. for steam sterilization: Aesculap STERILIT® I oil spray JG600 or STERILIT® I drip lubricator JG598).
CAUTION
► Allow the product to cool down to room temperature. ► After each complete cleaning, disinfecting and drying cycle, check that the instrument is dry, clean, operational,
and free of damage (e.g. broken insulation or corroded, loose, bent, broken, cracked, worn, or fractured components). ► Dry the product if it is wet or damp. ► Repeat cleaning and disinfection of products that still show impurities or contamination. ► Check that the product functions correctly. ► Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service, see Technical Service. ► Check for compatibility with associated products.
Packaging ► Appropriately protect products with fine working tips. ► Place the product in its holder or on a suitable tray. Ensure that all cutting edges are protected. ► Pack trays appropriately for the intended sterilization process (e.g. in sterile Aesculap containers). ► Ensure that the packaging provides sufficient protection against recontamination of the product during storage.
Steam sterilization
Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure.
Mechanical cleaning/disinfection with manual pre-cleaning Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark according to DIN EN ISO 15883). Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
Manual pre-cleaning with ultrasound and brush
► Check to ensure that the sterilizing agent will come into contact with all external and internal surfaces (e.g. by
opening any valves and faucets). ► Validated sterilization process
– Steam sterilization using fractionated vacuum process – Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665 – Sterilization using fractionated vacuum process at 134 °C/holding time 5 min ► When sterilizing several instruments at the same time in a steam sterilizer, ensure that the maximum load capacity of the steam sterilizer specified by the manufacturer is not exceeded.
Sterilization for the US market ■ Aesculap advises against sterilizing the device by flash sterilization or chemical sterilization. ■ Sterilization may be accomplished by a standard prevacuum cycle in a steam autoclave. To achieve a sterility assurance level of 10-6 , Aesculap recommends the following parameters:
Phase
Step
T [°C/°F]
t [min]
Conc. [%]
Water quality
Chemical
I
Ultrasonic cleaning
RT (cold)
>15
2
D–W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Rinsing
RT (cold)
1
-
D–W
-
D–W: Drinking water RT: Room temperature *Recommended: BBraun Stabimed ► Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure. Phase I ► Clean the product in an ultrasonic cleaning bath (frequency 35 kHz) for at least 15 min. Ensure that all accessible surfaces are immersed and acoustic shadows are avoided. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.
Mechanical alkaline cleaning and thermal disinfecting Machine type: single-chamber cleaning/disinfection device without ultrasound Phase
Step
T [°C/°F]
t [min]
Water quality
Chemical
I
Prerinse
<25/77
3
D–W
-
II
Cleaning
55/131
10
FD-W
■ Concentrate, alkaline: – pH = 13 – <5 % anionic surfactant
■ 0.5 % working solution – pH = 11* III
Intermediate rinse
>10/50
1
FD-W
-
IV
Thermal disinfecting
90/194
5
FD-W
-
V
Drying
-
-
-
According to the program for cleaning and disinfection device
D–W: FD–W:
Drinking water Fully desalinated water (demineralized, low microbiological contamination: drinking water quality at least) *Recommended: BBraun Helimatic Cleaner alkaline ► Check visible surfaces for residues after mechanical cleaning/disinfecting.
Aesculap Orga Tray/Sterile container (perforated bottom) Minimum cycle parameters* Sterilization method
Temp.
Time
Minimum drying time
Prevacuum
270 °F/275 °F
4 min
20 min
*Aesculap has validated the above sterilization cycle and has the data on file. The validation was accomplished in an Aesculap sterile container cleared by FDA for the sterilization and storage of these products. Other sterilization cycles may also be suitable, however individuals or hospitals not using the recommended method are advised to validate any alternative method using appropriate laboratory techniques. Use an FDA cleared accessory to maintain sterility after processing, such as a wrap, pouch, etc. WARNING for the US market If this device is/was used in a patient with, or suspected of having Creutzfeldt-Jakob Disease (CJD), the device cannot be reused and must be destroyed due to the inability to reprocess or sterilize to eliminate the risk of crosscontamination.
Storage ► Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area.
Technical Service Risk of injury and/or malfunction! ► Do not modify the product. WARNING ► For service and repairs, please contact your national B. Braun/Aesculap agency.
Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and forfeiture of applicable licenses. Service addresses Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 (7461) 95-1602 Fax: +49 (7461) 16-5621 E-Mail: [email protected] Or in the US: Attn. Aesculap Technical Services 615 Lambert Pointe Drive Hazelwood MO, 63042 Aesculap Repair Hotline Phone: +1 (800) 214-3392 Fax: +1 (314) 895-4420 Other service addresses can be obtained from the address indicated above.
Disposal ► Adhere to national regulations when disposing of or recycling the product, its components and its packaging!