CASPAR Bipolar Forceps, straight, 7 3/4", (195 mm), working length: 4", (100 mm), bayonet-shaped, insulated, width: 2 mm
Haemostatic forceps Instructions for Use
5 Pages
Preview
Page 1
AESCULAP® en
Effective
Instructions for use/Technical description Haemostatic forceps
Aesculap AG | Am Aesculap-Platz | 78532 Tuttlingen | Germany Phone +49 (0) 7461 95-0 | Fax +49 (0) 7461 95-26 00 | www.bbraun.com AESCULAP® – a B. Braun brand
TA016590-AU
2024-11
eifu.bbraun.com
Effective Name: TA016590-AU - Data ID: 282984237 - Document No.: NA - Version: 1.0 Document ID: IFU-AE-003689 - Effective Date: 2024-11-18 - Page: 1 of 5 - Print date: 2024-11-19 14:39 (CET)
en AESCULAP® Haemostatic forceps 1
About this document
Note General risk factors associated with surgical procedures are not described in this document.
1.1
Scope
These instructions for use apply to Haemostatic forceps. Note Instructions for use and further information about B. Braun/AESCULAP products can be found on the B. Braun eIFU website at eifu.bbraun.com
1.2
Safety messages
Safety messages make clear the dangers to patient, user and/or product that could arise during the use of the product. Safety messages are labeled as follows: WARNING Indicates a possible threat of danger. If not avoided, minor or moderate injury may result. CAUTION Indicates a possible threat of material damage. If not avoided, the product may be damaged.
2
Clinical use
2.1
Areas of use and limitations of use
2.1.1
Intended purpose
Haemostatic clamps are used for clamping of tissue and small vessels.
2.1.2
Intended use
Haemostatic clamps are used for clamping of tissue and small vessels.
2.1.3
Indications
Note The manufacturer is not responsible for any use of the product against the specified indications or the described applications. For indications, see Intended use.
2.1.4
Contraindications
3.2
Service life
Effective
Materials for reusable surgical instruments are generally chosen to be suitable for repeated processing. However, it should be noted that each mechanical, chemical and thermal treatment can lead to stress and thus to aging of the material. The service life of the product is limited by damage, normal wear, type and duration of use, as well as by handling, storage and transport of the product. End-of-life indicators for these products are signs of corrosion and cracks as well as deformation in the jaw area and loss of tension. Influences of processing using the validated procedure that lead to damage to the product are not known. Careful visual and functional inspection before each use is the best way to detect a product that is no longer functional, see Visual inspection and see Functional test.
3.3
General information
Dried or affixed surgical residues can make cleaning more difficult or ineffective and lead to corrosion. Therefore the time interval between application and processing should not exceed 6 h; also, neither fixating precleaning temperatures >45 °C nor fixating disinfecting agents (active ingredient: aldehydes/alcohols) should be used. Excessive measures of neutralizing agents or basic cleaners may result in a chemical attack and/or to fading and the laser marking becoming unreadable visually or by machine for stainless steel. Residues containing chlorine or chlorides e.g. in surgical residues, medicines, saline solutions and in the service water used for cleaning, disinfection and sterilization will cause corrosion damage (pitting, stress corrosion) and result in the destruction of stainless steel products. These must be removed by rinsing thoroughly with demineralized water and then drying. Additional drying, if necessary. Only process chemicals that have been tested and approved (e.g. VAH or FDA approval or CE mark) and which are compatible with the product’s materials according to the chemical manufacturers’ recommendations may be used for processing the product. All the chemical manufacturer's application specifications must be strictly observed. Failure to do so can result in the following problems: ■ Optical changes of materials, e.g. fading or discoloration of titanium or aluminum. For aluminum, the application/process solution only needs to be of pH >8 to cause visible surface changes. ■ Material damage such as corrosion, cracks, fracturing, premature aging or swelling. ► Do not use metal cleaning brushes or other abrasives that would damage the product surfaces and could cause corrosion. ► Further detailed advice on hygienically safe and material-/value-preserving reprocessing can be found at www.a-k-i.org, link to “AKI-Brochures”, “Red brochure”.
3.4
Initial treatment and disposal at the point of use
► If applicable, rinse non-visible surfaces preferably with deionized water, with a disposable syringe for ex-
ample. ► Remove any visible surgical residues to the extent possible with a damp, lint-free cloth. ► Transport the dry product in a sealed waste container for cleaning and disinfection within 6 hours.
No known contraindications.
2.1.5
Intended patient population
There are no general gender, age or ethnic limitations on patient population for the use of the product when used within its intended use. Restrictions are defined by the contraindications.
3.5
Preparation before cleaning
► Remove coarse contamination by rinsing and flushing with cold, clean water.
3.6
Cleaning, disinfecting and drying
2.2
Safety information
3.6.1
Product-specific safety information on the processing procedure
2.2.1
Clinical user
Damage to or destruction of the product due to inappropriate cleaning/disinfecting agents and/or excessive temperatures! ► Use cleaning and disinfecting agents according to the manufacturer's instructions. ► Observe specifications regarding concentration, temperature and exposure time. ► Do not exceed the maximum allowable disinfection temperature of 95 °C.
General safety information To prevent damage caused by improper setup or use and to not compromise the manufacturer warranty and liability: ► Use the product only according to these instructions for use. ► Follow the safety and maintenance instructions. ► Ensure that the product and its accessories are only operated and used by qualified personnel. ► Store any new or unused products in a dry, clean and safe place. ► Prior to use, check that the product is in good working order. ► Keep the instructions for use accessible for the user. Note The user is obligated to report all severe events in connection with the product to the manufacturer and the responsible authorities of the state in which the user is located. Notes on surgical procedures It is the user's responsibility to ensure that the surgical procedure is performed correctly. Appropriate clinical training as well as a theoretical and practical proficiency of all the required operating techniques, including the use of this product, are prerequisites for the successful use of this product. The user is required to obtain information from the manufacturer if there is an unclear preoperative situation regarding the use of the product.
2.2.2
3.6.2
Validated cleaning and disinfection procedure
Note Processing may only take place in accordance with the following listed procedures in version V6. These are documented in the brochure “Validated Reprocessing Procedures” (AVA-V6) C63402. You will also find this brochure on the B. Braun eIFU site at eifu.bbraun.com Validated procedure
Short description
Specific requirements
Manual cleaning with immersion disinfection
► Use a suitable cleaning brush. ► Use a disposable syringe
see Manual cleaning and disinfection and subsection: ■ see Manual cleaning with immersion disinfection
20 ml. ► Drying phase: Use a lint-free
cloth or medical compressed air. ► Clean the product having movable hinges in the open position or while moving the joints.
Sterility
The product is supplied non-sterile and intended to be used in sterile condition. ► Clean the new product after removing its transport packaging and prior to its initial sterilization.
2.3
Application
► Place the product in a tray
Mechanical alkaline cleaning and thermal disinfection
that is suitable for cleaning (avoiding rinsing blind spots). ► Place the product on the tray with all product links and joints open.
WARNING Risk of injury and/or malfunction! ► Prior to each use, inspect the product for loose, bent, broken, cracked, worn or fractured components. ► Always carry out a function test prior to each use of the product.
3
Validated processing procedure
Note Note the separate brochure “Validated Reprocessing Procedures” (AVA-V6) C63402, see Validated cleaning and disinfection procedure
3.1
3.7
Manual cleaning and disinfection
3.7.1
Manual cleaning with immersion disinfection
Phase
Step
T [°C/ °F]
t [min]
Conc. [%]
Water quality
Chemical/Note
I
Disinfecting cleaning
RT (cold)
≥ 15
2
D-W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
II
Intermediate rinse
RT (cold)
1
–
D-W
–
III
Disinfection
RT (cold)
15
2
D-W
Aldehyde-free, phenol-free, and QUAT-free concentrate, pH ~ 9*
IV
Final rinse
RT (cold)
1
–
FD-W
–
V
Drying
RT
–
–
–
–
Safety information
Note Adhere to national statutory regulations, national and international standards and directives, and local, clinical hygiene instructions for sterile processing. Note For patients with Creutzfeldt-Jakob disease (CJD), suspected CJD or possible variants of CJD, observe the relevant national regulations concerning the processing of products. Note Mechanical processing should be favored over manual cleaning as it gives better and more reliable results. Note Successful processing of this medical device can only be ensured if the processing method is first validated. The operator/sterile processing technician is responsible for this. Note Up-to-date information about processing and material compatibility can be found on the B. Braun eIFU site at eifu.bbraun.com The validated steam sterilization procedure was carried out in the AESCULAP sterile container system.
see Mechanical cleaning/disinfection and subsections: ■ see Mechanical alkaline cleaning and thermal disinfecting
D-W FD-W RT *Recommended
Effective
Drinking water Fully demineralized water (low-germ, max. 10 CFU / 100 ml, as well as low endotoxin contamination, max. 0 25 endotoxin units/ml) Room temperature B. Braun Stabimed fresh
Name: TA016590-AU - Data ID: 282984237 - Document No.: NA - Version: 1.0 Document ID: IFU-AE-003689 - Effective Date: 2024-11-18 - Page: 2 of 5 - Print date: 2024-11-19 14:39 (CET)
Phase I ► Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are moistened. ► Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed from the surface. ► If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min. ► Mobilize non-rigid components, such as set screws, links, etc. during cleaning. ► Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a disposable syringe. Phase II ► Rinse/flush the product thoroughly (all accessible surfaces) under running water. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Drain any remaining water fully. Phase III ► Fully immerse the product in the disinfectant solution. ► Mobilize non-rigid components, such as set screws, joints, etc. during rinsing. ► Rinse lumens at least 5 times at the beginning of the exposure time using an appropriate disposable syringe. Ensure that all accessible surfaces are moistened. Phase IV ► Rinse/flush the product thoroughly (all accessible surfaces). ► Mobilize non-rigid components, such as set screws, joints, etc. during final rinse. ► Rinse lumens with an appropriate disposable syringe at least five times. ► Drain any remaining water fully. Phase V ► Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning and disinfection procedure.
3.8
Mechanical cleaning/disinfection
Note The cleaning and disinfection device must be of tested and approved effectiveness (e.g. compliance with EN ISO 15883). Note The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.
3.8.1
Mechanical alkaline cleaning and thermal disinfecting
Phase
Step
T [°C/ °F]
t [min]
Water quality
Chemical/Note
I
Prerinse
< 25/ 77
3
D-W
–
II
Cleaning
55/131
10
FD-W
■ Concentrate, alkaline: – pH ≈ 13 – less than 5 % anionic surfactant ■ 0,5% working solution – pH = 11*
III
Intermediate rinse
> 10/ 50
>1
FD-W
–
IV
Thermal disinfection
90/194
5
FD-W
–
V
Drying
–
–
–
According to the program for cleaning and disinfection device
D-W FD-W *Recommended
Drinking water Fully demineralized water (low-germ, max. 10 CFU / 100 ml, as well as low endotoxin contamination, max. 0 25 endotoxin units/ml) B. Braun Helimatic Cleaner alcaline
► Check visible surfaces for residues after mechanical cleaning/disinfecting. ► Repeat the cleaning/disinfecting process if necessary.
3.9
Inspection
► Allow the product to cool down to room temperature. ► Dry the product if it is wet or damp. ► Any damages, signs of corrosion, cracks, missing labels, loose or missing parts, functional impairments like
loss of tension are end-of-life indicators.
3.9.1
Visual inspection
► Make sure all contaminants have been removed. Pay particular attention to mating surfaces, hinges, shafts,
recessed areas, drill grooves and the sides of teeth on rasps. ► If the product is still dirty: Repeat the cleaning and disinfection procedure. ► Inspect the product for damage, e.g., damaged insulation or corroded, loose, bent, broken, cracked, worn
or severely scratched and fractured components.
3.10 Packaging
Effective
► Appropriately protect products with fine working tips. ► Store products with ratchet locks fully opened or locked no further than in the first notch. ► Place the product in its holder or on a suitable tray, making sure it is positioned to prevent damage. ► Ensure that any fine working tips, blades and/or sharp edges are covered. ► Pack trays appropriately for the sterilization process (e.g., in AESCULAP sterile containers). ► Use sterile barrier packaging system in accordance with ISO 11607-1. ► Ensure that the packaging provides sufficient protection against contamination of the product during stor-
age.
3.11
Steam sterilization
Note To avoid breakage due to stress crack corrosion, sterilize the instruments with the lock fully open or locked no further than on the first ratchet tooth. ► Check to ensure that the sterilizing agent comes into contact with all external and internal surfaces (e.g., by opening any valves and faucets). ► Use validated sterilization process. – Steam sterilization using fractional vacuum process – Steam sterilizer according to EN 285 and validated according to ISO 17665 – Allowed sterilization parameters, see table below ► If several devices are sterilized simultaneously in the same steam sterilizer: Make sure that the maximum allowed load according to the specifications of the manufacturer is not exceeded. Allowed sterilization parameters Sterilization process
T [°C]
Holding time [min]
Drying time (at least recommended) [min]
Steam sterilization (fractionated vacuum process)
134
3 - 18
20
The sterilization of products approved for 134 °C is permissible in the temperature range from 134 °C to 137 °C.
3.12 Storage Shelf life depends on the quality of the packaging system or material, the tightness of the seals, and the storage conditions. ► Store sterile products at room temperature in a dust-free, clean, dry, and pest-free environment. ► Follow the storage instructions provided by the sterile barrier system manufacturer.
3.13 Transport Transport and storage must not adversely affect the characteristics of the processed medical device. ► Use appropriate transport systems and aids to prevent damage or recontamination.
4
Technical service
CAUTION Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and forfeiture of applicable licenses. ► Do not modify the product. ► For service and repairs, contact the national B. Braun/AESCULAP agency. Service address Aesculap Technischer Service Am Aesculap-Platz 78532 Tuttlingen / Germany Phone: +49 7461 95-1601 Fax: +49 7461 16-2887 E-Mail: [email protected]
5
Disposal
WARNING Risk of infection due to contaminated products! ► Adhere to national regulations when disposing of or recycling the product, its components and its packaging. WARNING Risk of injury due to sharp-edged and/or pointed products! ► Ensure that the packaging prevents injury by the product when disposing of or recycling the product. Note The user institution is obliged to process the product before its disposal, see Validated processing procedure. ► Detailed information concerning the disposal of the product is available through the national B. Braun / AESCULAP agency, see Technical service. TA016590-AU
2024-11
► Inspect the product for missing or faded labels. ► Check the surfaces for rough spots. ► Check the product for burrs that could damage tissue or surgical gloves. ► Check the product for loose or missing parts. ► Inspect the product carefully: The surface of the product shall be clean and free of any signs of corrosion
and cracks. ► Especially check the area of the joint and the jaw in open position from both sides. If in doubt, use mag-
nification. ► Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service,
see Technical service.
3.9.2
Functional test
CAUTION Damage (metal cold welding / friction corrosion) to the product caused by insufficient lubrication! ► Prior to function checks, lubricate moving parts (e.g., joints, pusher components and threaded rods) with maintenance oil suitable for the respective sterilization process (e.g., for steam sterilization: STERILIT® I oil spray JG600 or STERILIT® I drip lubricator JG598). ► Check that the product functions correctly. ► Check that all moving parts are working properly (e.g. hinges, locks/latches, sliding parts etc.). ► Check for compatibility with associated products. ► Check the jaw area for deformation and loss of tension. ► Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service,
see Technical service.
Effective
Name: TA016590-AU - Data ID: 282984237 - Document No.: NA - Version: 1.0 Document ID: IFU-AE-003689 - Effective Date: 2024-11-18 - Page: 3 of 5 - Print date: 2024-11-19 14:39 (CET)
Document Control
B|BRAUN
& Signature Page
Effective
This document is signed electronically in compliance with the B. Braun electronic signature policies and procedures by following persons:
UserName: Mouafo Tchio, Armand Joel (mouaarde) Title: Technischer Redakteur Date: Friday, 15 November 2024, 11:21 W. Europe Daylight Time Meaning: Document signed as Author ================================================
UserName: Vohl, Georg (vohlgede) Title: Senior Project Manager - Medical Scientific Affairs Date: Friday, 15 November 2024, 11:24 W. Europe Daylight Time Meaning: Approve Document ================================================
UserName: Bonno, Katharina (gfelkade) Title: Lifetime Evaluation and Processing Date: Friday, 15 November 2024, 13:10 W. Europe Daylight Time Meaning: Approve Document ================================================
UserName: Bellikli, Serkan (bellsrde) Title: R&D Expert Engineer Date: Monday, 18 November 2024, 07:22 W. Europe Daylight Time Meaning: Approve Document ================================================
UserName: Willemsen, Stefan (willstde) Title: Regulatory Affairs Date: Monday, 18 November 2024, 13:30 W. Europe Daylight Time Meaning: Approve Document ================================================
Effective Name: TA016590-AU - Data ID: 282984237 - Document No.: NA - Version: 1.0 Document ID: IFU-AE-003689 - Effective Date: 2024-11-18 - Page: 4 of 5 - Print date: 2024-11-19 14:39 (CET)
Document Control
B|BRAUN
& Signature Page
Effective
Title: TA016590-AU Initiator: Armand Joel ? Mouafo Tchio
This document is signed electronically in compliance with the B. Braun electronic signature policies and procedures by following persons:
UserName: Mouafo Tchio, Armand Joel (mouaarde) Title: Technischer Redakteur Date: Monday, 18 November 2024, 13:54 W. Europe Daylight Time Meaning: Final Release of the Document ================================================
Effective Name: TA016590-AU - Data ID: 282984237 - Document No.: NA - Version: 1.0 Document ID: IFU-AE-003689 - Effective Date: 2024-11-18 - Page: 5 of 5 - Print date: 2024-11-19 14:39 (CET)