Standby Mode:
Quick Reference Guidelines for Infusomat® Space
Infusomat® Space Volumetric - Quick Reference Guidelines
1. Stop the infusion by pressing . 2. Press once to access the Standby Function.
Clear button Parameter Setting Keys
3. Answer YES with arrow key to “Use standby?” • Standby will now be activated with previously used standby time.
4. To change the desired duration the pump goes into standby press
Bolus button
Scroll menus, change/enter values
On/Off button
arrow key.
Ok button Door Open button
To confirm settings
5. Use the directional keypad
to change the time or press
to clear and start from zero.
Start/Stop Space Connect button To link the pump to space control/ barcode scanner
6. Once desired time has been chosen press to confirm. • The pump will now start to count down the desired standby time. • Standby mode can be cancelled at any time by pressing
.
! If the giving set is in-situ the pump cannot be switched off. Í
Standby is activated after pressing
for 3 seconds.
Software F/G/H/J
BOLUS/ STANDBY MODE
Data Lock:
Quick Reference Guidelines for Infusomat® Space
! If your pump is configured to prime the line press Í
to prime and repeat as necessary. Ensure the pump is disconnected from the patient before priming.
1. From the Main Menu use the to scroll down to the Options Menu. 2. Once the Options Menu is highlighted press to access. 3. Use the
arrow keys to highlight Data Lock and press
4.
Use the and then press
Starting an infusion:
Quick Reference Guidelines for Infusomat® Space
to access.
1. *Enter the Volume To Be Infused “VTBI” by using 2. In the Main Menu
to set Data Lock at Level 1 or Level 2
, and confirm with highlight rate with
.
and select with
.
.
LEVEL ONE will allow the user to start and stop the infusion, change the disposable, and turn the pump off. All other actions will require the code. LEVEL TWO will only allow the user to stop the infusion. All other actions require the code. 5. Activate Data Lock by using the directional keypad to input the 4 digit code: ! Once activated any changes need to be made within 20 seconds. After this, a key symbol appears on the screen. At this point any further changes to infusion parameters will require the code.
Í
3. Select the correct rate with . 4. Press to commence infusion. • Once VTBI and Rate parameters are entered, the Time is automatically calculated. • Moving arrows on the display plus a green operating light indicate the pump is infusing. ! Check the displayed values with the values intended. Start only if values correspond. Í
• * You do not need to enter a Volume To Be Infused if you have a drop sensor connected.
6. De-activate the Data Lock by following steps 1 to 5 and returning the Data Lock level to OFF. PRESSURE SETTINGS/
STARTING THE
DATA LOCK
INFUSION
Quick Reference Guidelines for Infusomat® Space
Change infusion set:
• Interrupt connection to the patient to avoid dosage errors during line change. • Never leave the pump unattended during a line change. • Pump must be powered ON to change the giving set. 1. Press to stop infusion. • Ensure the roller clamp is closed on the giving set; disconnect from the patient.
19981_Infuso_SETUP_Layout 1 19/02/2013 11:12 Page 2
2. Press and to open the pump door. 3. Pull door down to open. 4. Press the green opening lever of the anti-free flow device within the pump.
5. Remove set and replace with a new line. 6. Press pump door firmly to close. ! Check the displayed values with the values intended. Start only if values correspond. Í
This booklet is provided for reference only. Prior to operating this equipment all users should consult the full Instructions For Use.
B. Braun Medical Ltd | Hospital Care | Thorncliffe Park | Sheffield | S35 2PW Tel: 0114 225 9000 | Fax: 0114 225 9111 | www.bbraun.co.uk
XX-SIIFUM-09-14
CHANGING RATES/ SYRINGE
TROUBLE SHOOTING
“Battery empty” The battery pack is discharged, you have 3 minutes before the pump will automatically turn off. Connect to mains or replace battery pack.
Enter new VTBI and continue therapy 1. Press to mute alarm. 2. Use to scroll down to VTBI and press . 3. Using enter new VTBI and press . 4. Press to recommence infusion. Or Continue without VTBI (only if drop sensor connected) Press and Press to resume. Or To end infusion and turn off Press and hold .
“Time expired” The pre selected time has expired. To continue without preset time 1. Press to mute alarm. 2. Press to resume. Or to enter new time 1. Press to mute alarm. 2. Use to select time and press . 3. Use and to enter a time. 4. Press . Or to end infusion Remove disposable and turn the pump off.
• Red LED Alarm “VTBI infused” Pre selected volume infused and KVO inactive.
Trouble shooting RED alarms:
Quick Reference Guidelines for Infusomat® Space
Stop infusion / turn off pump:
Quick Reference Guidelines for Infusomat® Space
! The giving set must be removed in order to switch the pump off. Í
1. Stop the infusion at any time by pressing
Bolusing:
Quick Reference Guidelines for Infusomat® Space
! The pump must be infusing in order to deliver a bolus. Í
• To deliver a bolus with a preset bolus limit: 1. Press 2. Press 3. Using 4. Press
to access the bolus screen. to access the bolus limit screen. set a bolus dose. to deliver the bolus.
! Remember that the bolus can be stopped at any time by pressing Í
• To deliver a manual bolus: 1. Press to access the bolus screen. 2. Press and hold the to deliver a manual bolus. • The pump will only allow a manual bolus to be delivered for a maximum of 10 seconds. ! Ensure you check the delivery rate on the bolus screen before delivering the bolus. Í
Change rate of infusion:
Quick Reference Guidelines for Infusomat® Space
During Infusion: The infusion will NOT be interrupted 1. Whilst the pump is running, press
.
.
to open rate setting.
• The green operating light will go out. 2. Turn the pump off by pressing
for 3 seconds.
! If the giving set is in-situ the pump cannot be switched off. Í
Standby is activated after pressing
2. Select the new rate with . 3. Press to confirm. If you do not press within 20 seconds a “value not accepted” alarm will sound.
for 3 seconds. Stopping the Infusion to change rate: 1. Press to stop infusion. 2. Press to open rate setting. 3. Select the correct rate with . 4. Press to commence infusion with modified rate. ! Always check the displayed value with the value intended. Í
Quick Reference Guidelines for Infusomat® Space
Adjusting pressure settings:
1. If the infusion is running press to return to the Main Menu. When the infusion is stopped it will automatically be on the Main Menu screen.
2. Use to highlight the Options Menu. 3. Once Options is highlighted, press .
4. Press
to access the Pressure settings.
Quick Reference Guidelines for Infusomat® Space
Turn pump on and load infusion set:
1. Press to switch on the pump. The pump will now go through a series of self checks. 2. Press and to open the pump door. • If your pump is not configured to do so, manually prime the line. Prior to loading giving set, ensure roller clamp is closed. 3. Insert the infusion line from right to left. • First ensure the opening lever is pressed down and the yellow hazard light is flashing. A. Insert the first white two holed clip. B. Insert the second white clip.
A
19981_Infuso_SETUP_Layout 1 19/02/2013 11:12 Page 1
B 5. Use to adjust the pressure alarm setting. Adjust in accordance with local policy. The higher the pressure alarm level is set, the higher the resistance must rise before triggering an occlusion alarm. Increasing the pressure alarm setting may lead to a prolonged alarm reaction time. 6. Press to confirm the new pressure setting. • If the pump is running it will automatically default back to the running display. If the pump is not infusing, press to return to the Main Menu.
C. Insert the green clamp into the anti-free flow slot.
C • When correctly inserted the yellow hazard light will stop flashing. 4. Push and hold the door firmly for a few seconds to close. 5. Confirm giving set type with . 6. You may now open the roller clamp on the giving set.
Standby Mode:
Quick Reference Guidelines for Infusomat® Space
Infusomat® Space Volumetric - Quick Reference Guidelines
1. Stop the infusion by pressing . 2. Press once to access the Standby Function.
Clear button Parameter Setting Keys
3. Answer YES with arrow key to “Use standby?” • Standby will now be activated with previously used standby time.
4. To change the desired duration the pump goes into standby press
Bolus button
Scroll menus, change/enter values
On/Off button
arrow key.
Ok button Door Open button
To confirm settings
5. Use the directional keypad
to change the time or press
to clear and start from zero.
Start/Stop Space Connect button To link the pump to space control/ barcode scanner
6. Once desired time has been chosen press to confirm. • The pump will now start to count down the desired standby time. • Standby mode can be cancelled at any time by pressing
.
! If the giving set is in-situ the pump cannot be switched off. Í
Standby is activated after pressing
for 3 seconds.
Software F/G/H/J
BOLUS/ STANDBY MODE
Data Lock:
Quick Reference Guidelines for Infusomat® Space
! If your pump is configured to prime the line press Í
to prime and repeat as necessary. Ensure the pump is disconnected from the patient before priming.
1. From the Main Menu use the to scroll down to the Options Menu. 2. Once the Options Menu is highlighted press to access. 3. Use the
arrow keys to highlight Data Lock and press
4.
Use the and then press
Starting an infusion:
Quick Reference Guidelines for Infusomat® Space
to access.
1. *Enter the Volume To Be Infused “VTBI” by using 2. In the Main Menu
to set Data Lock at Level 1 or Level 2
, and confirm with highlight rate with
.
and select with
.
.
LEVEL ONE will allow the user to start and stop the infusion, change the disposable, and turn the pump off. All other actions will require the code. LEVEL TWO will only allow the user to stop the infusion. All other actions require the code. 5. Activate Data Lock by using the directional keypad to input the 4 digit code: ! Once activated any changes need to be made within 20 seconds. After this, a key symbol appears on the screen. At this point any further changes to infusion parameters will require the code.
Í
3. Select the correct rate with . 4. Press to commence infusion. • Once VTBI and Rate parameters are entered, the Time is automatically calculated. • Moving arrows on the display plus a green operating light indicate the pump is infusing. ! Check the displayed values with the values intended. Start only if values correspond. Í
• * You do not need to enter a Volume To Be Infused if you have a drop sensor connected.
6. De-activate the Data Lock by following steps 1 to 5 and returning the Data Lock level to OFF. PRESSURE SETTINGS/
STARTING THE
DATA LOCK
INFUSION
Quick Reference Guidelines for Infusomat® Space
Change infusion set:
• Interrupt connection to the patient to avoid dosage errors during line change. • Never leave the pump unattended during a line change. • Pump must be powered ON to change the giving set. 1. Press to stop infusion. • Ensure the roller clamp is closed on the giving set; disconnect from the patient.
19981_Infuso_SETUP_Layout 1 19/02/2013 11:12 Page 2
2. Press and to open the pump door. 3. Pull door down to open. 4. Press the green opening lever of the anti-free flow device within the pump.
5. Remove set and replace with a new line. 6. Press pump door firmly to close. ! Check the displayed values with the values intended. Start only if values correspond. Í
This booklet is provided for reference only. Prior to operating this equipment all users should consult the full Instructions For Use.
B. Braun Medical Ltd | Hospital Care | Thorncliffe Park | Sheffield | S35 2PW Tel: 0114 225 9000 | Fax: 0114 225 9111 | www.bbraun.co.uk
XX-SIIFUM-09-14
CHANGING RATES/ SYRINGE
TROUBLE SHOOTING
“Battery empty” The battery pack is discharged, you have 3 minutes before the pump will automatically turn off. Connect to mains or replace battery pack.
Enter new VTBI and continue therapy 1. Press to mute alarm. 2. Use to scroll down to VTBI and press . 3. Using enter new VTBI and press . 4. Press to recommence infusion. Or Continue without VTBI (only if drop sensor connected) Press and Press to resume. Or To end infusion and turn off Press and hold .
“Time expired” The pre selected time has expired. To continue without preset time 1. Press to mute alarm. 2. Press to resume. Or to enter new time 1. Press to mute alarm. 2. Use to select time and press . 3. Use and to enter a time. 4. Press . Or to end infusion Remove disposable and turn the pump off.
• Red LED Alarm “VTBI infused” Pre selected volume infused and KVO inactive.
Trouble shooting RED alarms:
Quick Reference Guidelines for Infusomat® Space
Stop infusion / turn off pump:
Quick Reference Guidelines for Infusomat® Space
! The giving set must be removed in order to switch the pump off. Í
1. Stop the infusion at any time by pressing
Bolusing:
Quick Reference Guidelines for Infusomat® Space
! The pump must be infusing in order to deliver a bolus. Í
• To deliver a bolus with a preset bolus limit: 1. Press 2. Press 3. Using 4. Press
to access the bolus screen. to access the bolus limit screen. set a bolus dose. to deliver the bolus.
! Remember that the bolus can be stopped at any time by pressing Í
• To deliver a manual bolus: 1. Press to access the bolus screen. 2. Press and hold the to deliver a manual bolus. • The pump will only allow a manual bolus to be delivered for a maximum of 10 seconds. ! Ensure you check the delivery rate on the bolus screen before delivering the bolus. Í
Change rate of infusion:
Quick Reference Guidelines for Infusomat® Space
During Infusion: The infusion will NOT be interrupted 1. Whilst the pump is running, press
.
.
to open rate setting.
• The green operating light will go out. 2. Turn the pump off by pressing
for 3 seconds.
! If the giving set is in-situ the pump cannot be switched off. Í
Standby is activated after pressing
2. Select the new rate with . 3. Press to confirm. If you do not press within 20 seconds a “value not accepted” alarm will sound.
for 3 seconds. Stopping the Infusion to change rate: 1. Press to stop infusion. 2. Press to open rate setting. 3. Select the correct rate with . 4. Press to commence infusion with modified rate. ! Always check the displayed value with the value intended. Í
Quick Reference Guidelines for Infusomat® Space
Adjusting pressure settings:
1. If the infusion is running press to return to the Main Menu. When the infusion is stopped it will automatically be on the Main Menu screen.
2. Use to highlight the Options Menu. 3. Once Options is highlighted, press .
4. Press
to access the Pressure settings.
Quick Reference Guidelines for Infusomat® Space
Turn pump on and load infusion set:
1. Press to switch on the pump. The pump will now go through a series of self checks. 2. Press and to open the pump door. • If your pump is not configured to do so, manually prime the line. Prior to loading giving set, ensure roller clamp is closed. 3. Insert the infusion line from right to left. • First ensure the opening lever is pressed down and the yellow hazard light is flashing. A. Insert the first white two holed clip. B. Insert the second white clip.
A
19981_Infuso_SETUP_Layout 1 19/02/2013 11:12 Page 1
B 5. Use to adjust the pressure alarm setting. Adjust in accordance with local policy. The higher the pressure alarm level is set, the higher the resistance must rise before triggering an occlusion alarm. Increasing the pressure alarm setting may lead to a prolonged alarm reaction time. 6. Press to confirm the new pressure setting. • If the pump is running it will automatically default back to the running display. If the pump is not infusing, press to return to the Main Menu.
C. Insert the green clamp into the anti-free flow slot.
C • When correctly inserted the yellow hazard light will stop flashing. 4. Push and hold the door firmly for a few seconds to close. 5. Confirm giving set type with . 6. You may now open the roller clamp on the giving set.
Standby Mode:
Quick Reference Guidelines for Infusomat® Space
Infusomat® Space Volumetric - Quick Reference Guidelines
1. Stop the infusion by pressing . 2. Press once to access the Standby Function.
Clear button Parameter Setting Keys
3. Answer YES with arrow key to “Use standby?” • Standby will now be activated with previously used standby time.
4. To change the desired duration the pump goes into standby press
Bolus button
Scroll menus, change/enter values
On/Off button
arrow key.
Ok button Door Open button
To confirm settings
5. Use the directional keypad
to change the time or press
to clear and start from zero.
Start/Stop Space Connect button To link the pump to space control/ barcode scanner
6. Once desired time has been chosen press to confirm. • The pump will now start to count down the desired standby time. • Standby mode can be cancelled at any time by pressing
.
! If the giving set is in-situ the pump cannot be switched off. Í
Standby is activated after pressing
for 3 seconds.
Software F/G/H/J
BOLUS/ STANDBY MODE
Data Lock:
Quick Reference Guidelines for Infusomat® Space
! If your pump is configured to prime the line press Í
to prime and repeat as necessary. Ensure the pump is disconnected from the patient before priming.
1. From the Main Menu use the to scroll down to the Options Menu. 2. Once the Options Menu is highlighted press to access. 3. Use the
arrow keys to highlight Data Lock and press
4.
Use the and then press
Starting an infusion:
Quick Reference Guidelines for Infusomat® Space
to access.
1. *Enter the Volume To Be Infused “VTBI” by using 2. In the Main Menu
to set Data Lock at Level 1 or Level 2
, and confirm with highlight rate with
.
and select with
.
.
LEVEL ONE will allow the user to start and stop the infusion, change the disposable, and turn the pump off. All other actions will require the code. LEVEL TWO will only allow the user to stop the infusion. All other actions require the code. 5. Activate Data Lock by using the directional keypad to input the 4 digit code: ! Once activated any changes need to be made within 20 seconds. After this, a key symbol appears on the screen. At this point any further changes to infusion parameters will require the code.
Í
3. Select the correct rate with . 4. Press to commence infusion. • Once VTBI and Rate parameters are entered, the Time is automatically calculated. • Moving arrows on the display plus a green operating light indicate the pump is infusing. ! Check the displayed values with the values intended. Start only if values correspond. Í
• * You do not need to enter a Volume To Be Infused if you have a drop sensor connected.
6. De-activate the Data Lock by following steps 1 to 5 and returning the Data Lock level to OFF. PRESSURE SETTINGS/
STARTING THE
DATA LOCK
INFUSION
Quick Reference Guidelines for Infusomat® Space
Change infusion set:
• Interrupt connection to the patient to avoid dosage errors during line change. • Never leave the pump unattended during a line change. • Pump must be powered ON to change the giving set. 1. Press to stop infusion. • Ensure the roller clamp is closed on the giving set; disconnect from the patient.
19981_Infuso_SETUP_Layout 1 19/02/2013 11:12 Page 2
2. Press and to open the pump door. 3. Pull door down to open. 4. Press the green opening lever of the anti-free flow device within the pump.
5. Remove set and replace with a new line. 6. Press pump door firmly to close. ! Check the displayed values with the values intended. Start only if values correspond. Í
This booklet is provided for reference only. Prior to operating this equipment all users should consult the full Instructions For Use.
B. Braun Medical Ltd | Hospital Care | Thorncliffe Park | Sheffield | S35 2PW Tel: 0114 225 9000 | Fax: 0114 225 9111 | www.bbraun.co.uk
XX-SIIFUM-09-14
CHANGING RATES/ SYRINGE
TROUBLE SHOOTING
“Battery empty” The battery pack is discharged, you have 3 minutes before the pump will automatically turn off. Connect to mains or replace battery pack.
Enter new VTBI and continue therapy 1. Press to mute alarm. 2. Use to scroll down to VTBI and press . 3. Using enter new VTBI and press . 4. Press to recommence infusion. Or Continue without VTBI (only if drop sensor connected) Press and Press to resume. Or To end infusion and turn off Press and hold .
“Time expired” The pre selected time has expired. To continue without preset time 1. Press to mute alarm. 2. Press to resume. Or to enter new time 1. Press to mute alarm. 2. Use to select time and press . 3. Use and to enter a time. 4. Press . Or to end infusion Remove disposable and turn the pump off.
• Red LED Alarm “VTBI infused” Pre selected volume infused and KVO inactive.
Trouble shooting RED alarms:
Quick Reference Guidelines for Infusomat® Space
Stop infusion / turn off pump:
Quick Reference Guidelines for Infusomat® Space
! The giving set must be removed in order to switch the pump off. Í
1. Stop the infusion at any time by pressing
Bolusing:
Quick Reference Guidelines for Infusomat® Space
! The pump must be infusing in order to deliver a bolus. Í
• To deliver a bolus with a preset bolus limit: 1. Press 2. Press 3. Using 4. Press
to access the bolus screen. to access the bolus limit screen. set a bolus dose. to deliver the bolus.
! Remember that the bolus can be stopped at any time by pressing Í
• To deliver a manual bolus: 1. Press to access the bolus screen. 2. Press and hold the to deliver a manual bolus. • The pump will only allow a manual bolus to be delivered for a maximum of 10 seconds. ! Ensure you check the delivery rate on the bolus screen before delivering the bolus. Í
Change rate of infusion:
Quick Reference Guidelines for Infusomat® Space
During Infusion: The infusion will NOT be interrupted 1. Whilst the pump is running, press
.
.
to open rate setting.
• The green operating light will go out. 2. Turn the pump off by pressing
for 3 seconds.
! If the giving set is in-situ the pump cannot be switched off. Í
Standby is activated after pressing
2. Select the new rate with . 3. Press to confirm. If you do not press within 20 seconds a “value not accepted” alarm will sound.
for 3 seconds. Stopping the Infusion to change rate: 1. Press to stop infusion. 2. Press to open rate setting. 3. Select the correct rate with . 4. Press to commence infusion with modified rate. ! Always check the displayed value with the value intended. Í
Quick Reference Guidelines for Infusomat® Space
Adjusting pressure settings:
1. If the infusion is running press to return to the Main Menu. When the infusion is stopped it will automatically be on the Main Menu screen.
2. Use to highlight the Options Menu. 3. Once Options is highlighted, press .
4. Press
to access the Pressure settings.
Quick Reference Guidelines for Infusomat® Space
Turn pump on and load infusion set:
1. Press to switch on the pump. The pump will now go through a series of self checks. 2. Press and to open the pump door. • If your pump is not configured to do so, manually prime the line. Prior to loading giving set, ensure roller clamp is closed. 3. Insert the infusion line from right to left. • First ensure the opening lever is pressed down and the yellow hazard light is flashing. A. Insert the first white two holed clip. B. Insert the second white clip.
A
19981_Infuso_SETUP_Layout 1 19/02/2013 11:12 Page 1
B 5. Use to adjust the pressure alarm setting. Adjust in accordance with local policy. The higher the pressure alarm level is set, the higher the resistance must rise before triggering an occlusion alarm. Increasing the pressure alarm setting may lead to a prolonged alarm reaction time. 6. Press to confirm the new pressure setting. • If the pump is running it will automatically default back to the running display. If the pump is not infusing, press to return to the Main Menu.
C. Insert the green clamp into the anti-free flow slot.
C • When correctly inserted the yellow hazard light will stop flashing. 4. Push and hold the door firmly for a few seconds to close. 5. Confirm giving set type with . 6. You may now open the roller clamp on the giving set.
Standby Mode:
Quick Reference Guidelines for Infusomat® Space
Infusomat® Space Volumetric - Quick Reference Guidelines
1. Stop the infusion by pressing . 2. Press once to access the Standby Function.
Clear button Parameter Setting Keys
3. Answer YES with arrow key to “Use standby?” • Standby will now be activated with previously used standby time.
4. To change the desired duration the pump goes into standby press
Bolus button
Scroll menus, change/enter values
On/Off button
arrow key.
Ok button Door Open button
To confirm settings
5. Use the directional keypad
to change the time or press
to clear and start from zero.
Start/Stop Space Connect button To link the pump to space control/ barcode scanner
6. Once desired time has been chosen press to confirm. • The pump will now start to count down the desired standby time. • Standby mode can be cancelled at any time by pressing
.
! If the giving set is in-situ the pump cannot be switched off. Í
Standby is activated after pressing
for 3 seconds.
Software F/G/H/J
BOLUS/ STANDBY MODE
Data Lock:
Quick Reference Guidelines for Infusomat® Space
! If your pump is configured to prime the line press Í
to prime and repeat as necessary. Ensure the pump is disconnected from the patient before priming.
1. From the Main Menu use the to scroll down to the Options Menu. 2. Once the Options Menu is highlighted press to access. 3. Use the
arrow keys to highlight Data Lock and press
4.
Use the and then press
Starting an infusion:
Quick Reference Guidelines for Infusomat® Space
to access.
1. *Enter the Volume To Be Infused “VTBI” by using 2. In the Main Menu
to set Data Lock at Level 1 or Level 2
, and confirm with highlight rate with
.
and select with
.
.
LEVEL ONE will allow the user to start and stop the infusion, change the disposable, and turn the pump off. All other actions will require the code. LEVEL TWO will only allow the user to stop the infusion. All other actions require the code. 5. Activate Data Lock by using the directional keypad to input the 4 digit code: ! Once activated any changes need to be made within 20 seconds. After this, a key symbol appears on the screen. At this point any further changes to infusion parameters will require the code.
Í
3. Select the correct rate with . 4. Press to commence infusion. • Once VTBI and Rate parameters are entered, the Time is automatically calculated. • Moving arrows on the display plus a green operating light indicate the pump is infusing. ! Check the displayed values with the values intended. Start only if values correspond. Í
• * You do not need to enter a Volume To Be Infused if you have a drop sensor connected.
6. De-activate the Data Lock by following steps 1 to 5 and returning the Data Lock level to OFF. PRESSURE SETTINGS/
STARTING THE
DATA LOCK
INFUSION
Quick Reference Guidelines for Infusomat® Space
Change infusion set:
• Interrupt connection to the patient to avoid dosage errors during line change. • Never leave the pump unattended during a line change. • Pump must be powered ON to change the giving set. 1. Press to stop infusion. • Ensure the roller clamp is closed on the giving set; disconnect from the patient.
19981_Infuso_SETUP_Layout 1 19/02/2013 11:12 Page 2
2. Press and to open the pump door. 3. Pull door down to open. 4. Press the green opening lever of the anti-free flow device within the pump.
5. Remove set and replace with a new line. 6. Press pump door firmly to close. ! Check the displayed values with the values intended. Start only if values correspond. Í
This booklet is provided for reference only. Prior to operating this equipment all users should consult the full Instructions For Use.
B. Braun Medical Ltd | Hospital Care | Thorncliffe Park | Sheffield | S35 2PW Tel: 0114 225 9000 | Fax: 0114 225 9111 | www.bbraun.co.uk
XX-SIIFUM-09-14
CHANGING RATES/ SYRINGE
TROUBLE SHOOTING
“Battery empty” The battery pack is discharged, you have 3 minutes before the pump will automatically turn off. Connect to mains or replace battery pack.
Enter new VTBI and continue therapy 1. Press to mute alarm. 2. Use to scroll down to VTBI and press . 3. Using enter new VTBI and press . 4. Press to recommence infusion. Or Continue without VTBI (only if drop sensor connected) Press and Press to resume. Or To end infusion and turn off Press and hold .
“Time expired” The pre selected time has expired. To continue without preset time 1. Press to mute alarm. 2. Press to resume. Or to enter new time 1. Press to mute alarm. 2. Use to select time and press . 3. Use and to enter a time. 4. Press . Or to end infusion Remove disposable and turn the pump off.
• Red LED Alarm “VTBI infused” Pre selected volume infused and KVO inactive.
Trouble shooting RED alarms:
Quick Reference Guidelines for Infusomat® Space
Stop infusion / turn off pump:
Quick Reference Guidelines for Infusomat® Space
! The giving set must be removed in order to switch the pump off. Í
1. Stop the infusion at any time by pressing
Bolusing:
Quick Reference Guidelines for Infusomat® Space
! The pump must be infusing in order to deliver a bolus. Í
• To deliver a bolus with a preset bolus limit: 1. Press 2. Press 3. Using 4. Press
to access the bolus screen. to access the bolus limit screen. set a bolus dose. to deliver the bolus.
! Remember that the bolus can be stopped at any time by pressing Í
• To deliver a manual bolus: 1. Press to access the bolus screen. 2. Press and hold the to deliver a manual bolus. • The pump will only allow a manual bolus to be delivered for a maximum of 10 seconds. ! Ensure you check the delivery rate on the bolus screen before delivering the bolus. Í
Change rate of infusion:
Quick Reference Guidelines for Infusomat® Space
During Infusion: The infusion will NOT be interrupted 1. Whilst the pump is running, press
.
.
to open rate setting.
• The green operating light will go out. 2. Turn the pump off by pressing
for 3 seconds.
! If the giving set is in-situ the pump cannot be switched off. Í
Standby is activated after pressing
2. Select the new rate with . 3. Press to confirm. If you do not press within 20 seconds a “value not accepted” alarm will sound.
for 3 seconds. Stopping the Infusion to change rate: 1. Press to stop infusion. 2. Press to open rate setting. 3. Select the correct rate with . 4. Press to commence infusion with modified rate. ! Always check the displayed value with the value intended. Í
Quick Reference Guidelines for Infusomat® Space
Adjusting pressure settings:
1. If the infusion is running press to return to the Main Menu. When the infusion is stopped it will automatically be on the Main Menu screen.
2. Use to highlight the Options Menu. 3. Once Options is highlighted, press .
4. Press
to access the Pressure settings.
Quick Reference Guidelines for Infusomat® Space
Turn pump on and load infusion set:
1. Press to switch on the pump. The pump will now go through a series of self checks. 2. Press and to open the pump door. • If your pump is not configured to do so, manually prime the line. Prior to loading giving set, ensure roller clamp is closed. 3. Insert the infusion line from right to left. • First ensure the opening lever is pressed down and the yellow hazard light is flashing. A. Insert the first white two holed clip. B. Insert the second white clip.
A
19981_Infuso_SETUP_Layout 1 19/02/2013 11:12 Page 1
B 5. Use to adjust the pressure alarm setting. Adjust in accordance with local policy. The higher the pressure alarm level is set, the higher the resistance must rise before triggering an occlusion alarm. Increasing the pressure alarm setting may lead to a prolonged alarm reaction time. 6. Press to confirm the new pressure setting. • If the pump is running it will automatically default back to the running display. If the pump is not infusing, press to return to the Main Menu.
C. Insert the green clamp into the anti-free flow slot.
C • When correctly inserted the yellow hazard light will stop flashing. 4. Push and hold the door firmly for a few seconds to close. 5. Confirm giving set type with . 6. You may now open the roller clamp on the giving set.
Standby Mode:
Quick Reference Guidelines for Infusomat® Space
Infusomat® Space Volumetric - Quick Reference Guidelines
1. Stop the infusion by pressing . 2. Press once to access the Standby Function.
Clear button Parameter Setting Keys
3. Answer YES with arrow key to “Use standby?” • Standby will now be activated with previously used standby time.
4. To change the desired duration the pump goes into standby press
Bolus button
Scroll menus, change/enter values
On/Off button
arrow key.
Ok button Door Open button
To confirm settings
5. Use the directional keypad
to change the time or press
to clear and start from zero.
Start/Stop Space Connect button To link the pump to space control/ barcode scanner
6. Once desired time has been chosen press to confirm. • The pump will now start to count down the desired standby time. • Standby mode can be cancelled at any time by pressing
.
! If the giving set is in-situ the pump cannot be switched off. Í
Standby is activated after pressing
for 3 seconds.
Software F/G/H/J
BOLUS/ STANDBY MODE
Data Lock:
Quick Reference Guidelines for Infusomat® Space
! If your pump is configured to prime the line press Í
to prime and repeat as necessary. Ensure the pump is disconnected from the patient before priming.
1. From the Main Menu use the to scroll down to the Options Menu. 2. Once the Options Menu is highlighted press to access. 3. Use the
arrow keys to highlight Data Lock and press
4.
Use the and then press
Starting an infusion:
Quick Reference Guidelines for Infusomat® Space
to access.
1. *Enter the Volume To Be Infused “VTBI” by using 2. In the Main Menu
to set Data Lock at Level 1 or Level 2
, and confirm with highlight rate with
.
and select with
.
.
LEVEL ONE will allow the user to start and stop the infusion, change the disposable, and turn the pump off. All other actions will require the code. LEVEL TWO will only allow the user to stop the infusion. All other actions require the code. 5. Activate Data Lock by using the directional keypad to input the 4 digit code: ! Once activated any changes need to be made within 20 seconds. After this, a key symbol appears on the screen. At this point any further changes to infusion parameters will require the code.
Í
3. Select the correct rate with . 4. Press to commence infusion. • Once VTBI and Rate parameters are entered, the Time is automatically calculated. • Moving arrows on the display plus a green operating light indicate the pump is infusing. ! Check the displayed values with the values intended. Start only if values correspond. Í
• * You do not need to enter a Volume To Be Infused if you have a drop sensor connected.
6. De-activate the Data Lock by following steps 1 to 5 and returning the Data Lock level to OFF. PRESSURE SETTINGS/
STARTING THE
DATA LOCK
INFUSION
Quick Reference Guidelines for Infusomat® Space
Change infusion set:
• Interrupt connection to the patient to avoid dosage errors during line change. • Never leave the pump unattended during a line change. • Pump must be powered ON to change the giving set. 1. Press to stop infusion. • Ensure the roller clamp is closed on the giving set; disconnect from the patient.
19981_Infuso_SETUP_Layout 1 19/02/2013 11:12 Page 2
2. Press and to open the pump door. 3. Pull door down to open. 4. Press the green opening lever of the anti-free flow device within the pump.
5. Remove set and replace with a new line. 6. Press pump door firmly to close. ! Check the displayed values with the values intended. Start only if values correspond. Í
This booklet is provided for reference only. Prior to operating this equipment all users should consult the full Instructions For Use.
B. Braun Medical Ltd | Hospital Care | Thorncliffe Park | Sheffield | S35 2PW Tel: 0114 225 9000 | Fax: 0114 225 9111 | www.bbraun.co.uk
XX-SIIFUM-09-14
CHANGING RATES/ SYRINGE
TROUBLE SHOOTING
“Battery empty” The battery pack is discharged, you have 3 minutes before the pump will automatically turn off. Connect to mains or replace battery pack.
Enter new VTBI and continue therapy 1. Press to mute alarm. 2. Use to scroll down to VTBI and press . 3. Using enter new VTBI and press . 4. Press to recommence infusion. Or Continue without VTBI (only if drop sensor connected) Press and Press to resume. Or To end infusion and turn off Press and hold .
“Time expired” The pre selected time has expired. To continue without preset time 1. Press to mute alarm. 2. Press to resume. Or to enter new time 1. Press to mute alarm. 2. Use to select time and press . 3. Use and to enter a time. 4. Press . Or to end infusion Remove disposable and turn the pump off.
• Red LED Alarm “VTBI infused” Pre selected volume infused and KVO inactive.
Trouble shooting RED alarms:
Quick Reference Guidelines for Infusomat® Space
Stop infusion / turn off pump:
Quick Reference Guidelines for Infusomat® Space
! The giving set must be removed in order to switch the pump off. Í
1. Stop the infusion at any time by pressing
Bolusing:
Quick Reference Guidelines for Infusomat® Space
! The pump must be infusing in order to deliver a bolus. Í
• To deliver a bolus with a preset bolus limit: 1. Press 2. Press 3. Using 4. Press
to access the bolus screen. to access the bolus limit screen. set a bolus dose. to deliver the bolus.
! Remember that the bolus can be stopped at any time by pressing Í
• To deliver a manual bolus: 1. Press to access the bolus screen. 2. Press and hold the to deliver a manual bolus. • The pump will only allow a manual bolus to be delivered for a maximum of 10 seconds. ! Ensure you check the delivery rate on the bolus screen before delivering the bolus. Í
Change rate of infusion:
Quick Reference Guidelines for Infusomat® Space
During Infusion: The infusion will NOT be interrupted 1. Whilst the pump is running, press
.
.
to open rate setting.
• The green operating light will go out. 2. Turn the pump off by pressing
for 3 seconds.
! If the giving set is in-situ the pump cannot be switched off. Í
Standby is activated after pressing
2. Select the new rate with . 3. Press to confirm. If you do not press within 20 seconds a “value not accepted” alarm will sound.
for 3 seconds. Stopping the Infusion to change rate: 1. Press to stop infusion. 2. Press to open rate setting. 3. Select the correct rate with . 4. Press to commence infusion with modified rate. ! Always check the displayed value with the value intended. Í
Quick Reference Guidelines for Infusomat® Space
Adjusting pressure settings:
1. If the infusion is running press to return to the Main Menu. When the infusion is stopped it will automatically be on the Main Menu screen.
2. Use to highlight the Options Menu. 3. Once Options is highlighted, press .
4. Press
to access the Pressure settings.
Quick Reference Guidelines for Infusomat® Space
Turn pump on and load infusion set:
1. Press to switch on the pump. The pump will now go through a series of self checks. 2. Press and to open the pump door. • If your pump is not configured to do so, manually prime the line. Prior to loading giving set, ensure roller clamp is closed. 3. Insert the infusion line from right to left. • First ensure the opening lever is pressed down and the yellow hazard light is flashing. A. Insert the first white two holed clip. B. Insert the second white clip.
A
19981_Infuso_SETUP_Layout 1 19/02/2013 11:12 Page 1
B 5. Use to adjust the pressure alarm setting. Adjust in accordance with local policy. The higher the pressure alarm level is set, the higher the resistance must rise before triggering an occlusion alarm. Increasing the pressure alarm setting may lead to a prolonged alarm reaction time. 6. Press to confirm the new pressure setting. • If the pump is running it will automatically default back to the running display. If the pump is not infusing, press to return to the Main Menu.
C. Insert the green clamp into the anti-free flow slot.
C • When correctly inserted the yellow hazard light will stop flashing. 4. Push and hold the door firmly for a few seconds to close. 5. Confirm giving set type with . 6. You may now open the roller clamp on the giving set.
Quick Reference Guidelines for Infusomat® Space
Trouble shooting YELLOW alarms:
• Yellow LED Alarm “VTBI near end” The pre selected volume is nearly infused. Press to mute alarm, infusion will continue and the time remaining until end of VTBI will flash on the display.
“Battery nearly empty” The battery is almost discharged. You have 30 minutes before the pump will automatically turn off. Press to mute alarm, connect to mains or replace battery pack.
“VTBI near end” The pre selected volume is nearly infused. Press to mute alarm, infusion will continue and the time remaining until end of VTBI will flash on the display. “Time near end” Pre selected time has nearly finished.
Press to mute alarm, infusion will continue and the time remaining will flash on the display.
Press to mute alarm, infusion will continue and the time remaining will flash on the display.
STOPPING AN INFUSION
STOPPING AN INFUSION
TROUBLE SHOOTING
TROUBLE SHOOTING
Trouble shooting YELLOW alarms:
Press to mute alarm, infusion will continue and the time remaining until end of VTBI will flash on the display.
Quick Reference Guidelines for Infusomat® Space
“Battery nearly empty” The battery is almost discharged. You have 30 minutes before the pump will automatically turn off. Press to mute alarm, connect to mains or replace battery pack.
“VTBI near end” The pre selected volume is nearly infused. Press to mute alarm, infusion will continue and the time remaining until end of VTBI will flash on the display.
“Time near end” Pre selected time has nearly finished.
“Time near end” Pre selected time has nearly finished.
Press to mute alarm, infusion will continue and the time remaining will flash on the display.
Press to mute alarm, infusion will continue and the time remaining will flash on the display.
STOPPING AN INFUSION
STOPPING AN INFUSION
TROUBLE SHOOTING
TROUBLE SHOOTING
Quick Reference Guidelines for Infusomat® Space
Trouble shooting YELLOW alarms:
• Yellow LED Alarm “VTBI near end” The pre selected volume is nearly infused. Press to mute alarm, infusion will continue and the time remaining until end of VTBI will flash on the display.
“Battery nearly empty” The battery is almost discharged. You have 30 minutes before the pump will automatically turn off. Press to mute alarm, connect to mains or replace battery pack.
Trouble shooting YELLOW alarms:
• Yellow LED Alarm
• Yellow LED Alarm “VTBI near end” The pre selected volume is nearly infused.
Trouble shooting YELLOW alarms:
• Yellow LED Alarm
“Time near end” Pre selected time has nearly finished.
Quick Reference Guidelines for Infusomat® Space
19981_Infuso_SETUP_Layout 1 19/02/2013 11:12 Page 4
Quick Reference Guidelines for Infusomat® Space
Quick Reference Guidelines for Infusomat® Space
“Battery nearly empty” The battery is almost discharged. You have 30 minutes before the pump will automatically turn off. Press to mute alarm, connect to mains or replace battery pack.
Trouble shooting YELLOW alarms:
• Yellow LED Alarm “Battery nearly empty” The battery is almost discharged. You have 30 minutes before the pump will automatically turn off. Press to mute alarm, connect to mains or replace battery pack.
“VTBI near end” The pre selected volume is nearly infused. Press to mute alarm, infusion will continue and the time remaining until end of VTBI will flash on the display.
“Time near end” Pre selected time has nearly finished.
“Time near end” Pre selected time has nearly finished.
Press to mute alarm, infusion will continue and the time remaining will flash on the display.
Press to mute alarm, infusion will continue and the time remaining will flash on the display.
STOPPING AN INFUSION
STOPPING AN INFUSION
TROUBLE SHOOTING
TROUBLE SHOOTING
“Battery nearly empty” The battery is almost discharged. You have 30 minutes before the pump will automatically turn off. Press to mute alarm, connect to mains or replace battery pack.
Quick Reference Guidelines for Infusomat® Space
• Red LED Alarm “Device Alarm” Press to switch the pump off, then immediately switch back on. If self check is then completed, continue to use the device. Should alarm re-occur, note the error code, withdraw pump from use and report fault as per local policy. “Pressure high” Occlusion in the system, set pressure level was exceeded. 1. Press to mute alarm. 2. Check and act accordingly. Is the tubing kinked or damaged? Is the patient access patent? If there is an inline filter, is that patent? 3. Increase occlusion pressure if necessary (see Adjusting Pressure Settings). 4. Press to recommence infusion.
Quick Reference Guidelines for Infusomat® Space
• Red LED Alarm “Device Alarm” Press to switch the pump off, then immediately switch back on. If self check is then completed, continue to use the device. Should alarm re-occur, note the error code, withdraw pump from use and report fault as per local policy. “Pressure high” Occlusion in the system, set pressure level was exceeded. 1. Press to mute alarm. 2. Check and act accordingly. Is the tubing kinked or damaged? Is the patient access patent? If there is an inline filter, is that patent? 3. Increase occlusion pressure if necessary (see Adjusting Pressure Settings). 4. Press to recommence infusion.
Quick Reference Guidelines for Infusomat® Space
• Red LED Alarm “Device Alarm” Press to switch the pump off, then immediately switch back on. If self check is then completed, continue to use the device. Should alarm re-occur, note the error code, withdraw pump from use and report fault as per local policy.
19981_Infuso_SETUP_Layout 1 19/02/2013 11:12 Page 3
“Pressure high” Occlusion in the system, set pressure level was exceeded. 1. Press to mute alarm. 2. Check and act accordingly. Is the tubing kinked or damaged? Is the patient access patent? If there is an inline filter, is that patent? 3. Increase occlusion pressure if necessary (see Adjusting Pressure Settings). 4. Press to recommence infusion.
Trouble shooting RED alarms: “Check upstream” The upstream sensor triggers an alarm. Check if roller clamp is closed or if infusion line is kinked. “Air bubble” / “Accumulated air” Air detected in the line. Press to mute alarm. To purge the line disconnect from the patient and press . Repeat until the air is removed. Press to recommence the infusion. Or Press to cancel prime (consider removing the line and manually prime to remove the air) press to recommence the infusion.
Trouble shooting RED alarms: “Check upstream” The upstream sensor triggers an alarm. Check if roller clamp is closed or if infusion line is kinked. “Air bubble” / “Accumulated air” Air detected in the line. Press to mute alarm. To purge the line disconnect from the patient and press . Repeat until the air is removed. Press to recommence the infusion. Or Press to cancel prime (consider removing the line and manually prime to remove the air) press to recommence the infusion.
Trouble shooting RED alarms: “Check upstream” The upstream sensor triggers an alarm. Check if roller clamp is closed or if infusion line is kinked. “Air bubble” / “Accumulated air” Air detected in the line. Press to mute alarm. To purge the line disconnect from the patient and press . Repeat until the air is removed. Press to recommence the infusion. Or Press to cancel prime (consider removing the line and manually prime to remove the air) press to recommence the infusion.
Quick Reference Guidelines for Infusomat® Space
• Red LED Alarm “Device Alarm” Press to switch the pump off, then immediately switch back on. If self check is then completed, continue to use the device. Should alarm re-occur, note the error code, withdraw pump from use and report fault as per local policy. “Pressure high” Occlusion in the system, set pressure level was exceeded. 1. Press to mute alarm. 2. Check and act accordingly. Is the tubing kinked or damaged? Is the patient access patent? If there is an inline filter, is that patent? 3. Increase occlusion pressure if necessary (see Adjusting Pressure Settings). 4. Press to recommence infusion.
Quick Reference Guidelines for Infusomat® Space
• Red LED Alarm “Device Alarm” Press to switch the pump off, then immediately switch back on. If self check is then completed, continue to use the device. Should alarm re-occur, note the error code, withdraw pump from use and report fault as per local policy. “Pressure high” Occlusion in the system, set pressure level was exceeded. 1. Press to mute alarm. 2. Check and act accordingly. Is the tubing kinked or damaged? Is the patient access patent? If there is an inline filter, is that patent? 3. Increase occlusion pressure if necessary (see Adjusting Pressure Settings). 4. Press to recommence infusion.
Quick Reference Guidelines for Infusomat® Space
• Red LED Alarm “Device Alarm” Press to switch the pump off, then immediately switch back on. If self check is then completed, continue to use the device. Should alarm re-occur, note the error code, withdraw pump from use and report fault as per local policy. “Pressure high” Occlusion in the system, set pressure level was exceeded. 1. Press to mute alarm. 2. Check and act accordingly. Is the tubing kinked or damaged? Is the patient access patent? If there is an inline filter, is that patent? 3. Increase occlusion pressure if necessary (see Adjusting Pressure Settings). 4. Press to recommence infusion.
Trouble shooting RED alarms: “Check upstream” The upstream sensor triggers an alarm. Check if roller clamp is closed or if infusion line is kinked. “Air bubble” / “Accumulated air” Air detected in the line. Press to mute alarm. To purge the line disconnect from the patient and press . Repeat until the air is removed. Press to recommence the infusion. Or Press to cancel prime (consider removing the line and manually prime to remove the air) press to recommence the infusion.
Trouble shooting RED alarms: “Check upstream” The upstream sensor triggers an alarm. Check if roller clamp is closed or if infusion line is kinked. “Air bubble” / “Accumulated air” Air detected in the line. Press to mute alarm. To purge the line disconnect from the patient and press . Repeat until the air is removed. Press to recommence the infusion. Or Press to cancel prime (consider removing the line and manually prime to remove the air) press to recommence the infusion.
Trouble shooting RED alarms: “Check upstream” The upstream sensor triggers an alarm. Check if roller clamp is closed or if infusion line is kinked. “Air bubble” / “Accumulated air” Air detected in the line. Press to mute alarm. To purge the line disconnect from the patient and press . Repeat until the air is removed. Press to recommence the infusion. Or Press to cancel prime (consider removing the line and manually prime to remove the air) press to recommence the infusion.
“Battery empty” The battery pack is discharged, you have 3 minutes before the pump will automatically turn off. Connect to mains or replace battery pack. “Time expired” The pre selected time has expired. To continue without preset time 1. Press to mute alarm. 2. Press to resume. Or to enter new time 1. Press to mute alarm. 2. Use to select time and press . 3. Use and to enter a time. 4. Press . Or to end infusion Remove disposable and turn the pump off.
Trouble shooting RED alarms: 19981_Infuso_SETUP_Layout 1 19/02/2013 11:12 Page 1
Quick Reference Guidelines for Infusomat® Space
TROUBLE SHOOTING Quick Reference Guidelines for Infusomat® Space
Enter new VTBI and continue therapy 1. Press to mute alarm. 2. Use to scroll down to VTBI and press . 3. Using enter new VTBI and press . 4. Press to recommence infusion. Or Continue without VTBI (only if drop sensor connected) Press and Press to resume. Or To end infusion and turn off Press and hold . • Red LED Alarm “VTBI infused” Pre selected volume infused and KVO inactive. Quick Reference Guidelines for Infusomat® Space
Stop infusion / turn off pump:
! The giving set must be removed in order to switch the pump off. Í
1. Stop the infusion at any time by pressing
Bolusing:
! The pump must be infusing in order to deliver a bolus. Í
• To deliver a bolus with a preset bolus limit: 1. Press 2. Press 3. Using 4. Press
to access the bolus screen. to access the bolus limit screen. set a bolus dose. to deliver the bolus.
! Remember that the bolus can be stopped at any time by pressing Í
.
• To deliver a manual bolus: 1. Press to access the bolus screen. 2. Press and hold the to deliver a manual bolus. • The pump will only allow a manual bolus to be delivered for a maximum of 10 seconds. ! Ensure you check the delivery rate on the bolus screen before delivering the bolus. Í
Quick Reference Guidelines for Infusomat® Space
Change rate of infusion:
During Infusion: The infusion will NOT be interrupted 1. Whilst the pump is running, press
.
to open rate setting.
• The green operating light will go out. 2. Turn the pump off by pressing
2. Select the new rate with . 3. Press to confirm. If you do not press within 20 seconds a “value not accepted” alarm will sound.
for 3 seconds.
Í
! If the giving set is in-situ the pump cannot be switched off. Standby is activated after pressing for 3 seconds.
Stopping the Infusion to change rate: 1. Press to stop infusion. 2. Press to open rate setting. 3. Select the correct rate with . 4. Press to commence infusion with modified rate. ! Always check the displayed value with the value intended. Í
Adjusting pressure settings:
Quick Reference Guidelines for Infusomat® Space
1. If the infusion is running press to return to the Main Menu. When the infusion is stopped it will automatically be on the Main Menu screen.
2. Use to highlight the Options Menu. 3. Once Options is highlighted, press .
4. Press
Turn pump on and load infusion set:
Quick Reference Guidelines for Infusomat® Space
1. Press to switch on the pump. The pump will now go through a series of self checks. 2. Press and to open the pump door. • If your pump is not configured to do so, manually prime the line. Prior to loading giving set, ensure roller clamp is closed. 3. Insert the infusion line from right to left. • First ensure the opening lever is pressed down and the yellow hazard light is flashing. A. Insert the first white two holed clip. B. Insert the second white clip.
A
to access the Pressure settings.
C. Insert the green clamp into the anti-free flow slot.
B 5. Use to adjust the pressure alarm setting. Adjust in accordance with local policy. The higher the pressure alarm level is set, the higher the resistance must rise before triggering an occlusion alarm. Increasing the pressure alarm setting may lead to a prolonged alarm reaction time. 6. Press to confirm the new pressure setting. • If the pump is running it will automatically default back to the running display. If the pump is not infusing, press to return to the Main Menu.
C • When correctly inserted the yellow hazard light will stop flashing. 4. Push and hold the door firmly for a few seconds to close. 5. Confirm giving set type with . 6. You may now open the roller clamp on the giving set.
Standby Mode:
Quick Reference Guidelines for Infusomat® Space
Infusomat® Space Volumetric - Quick Reference Guidelines
1. Stop the infusion by pressing . 2. Press once to access the Standby Function.
Clear button Parameter Setting Keys
3. Answer YES with arrow key to “Use standby?” • Standby will now be activated with previously used standby time.
4. To change the desired duration the pump goes into standby press
Bolus button
Scroll menus, change/enter values
On/Off button
arrow key.
Ok button Door Open button
To confirm settings
5. Use the directional keypad
to change the time or press
to clear and start from zero.
Start/Stop Space Connect button To link the pump to space control/ barcode scanner
6. Once desired time has been chosen press to confirm. • The pump will now start to count down the desired standby time. • Standby mode can be cancelled at any time by pressing
.
! If the giving set is in-situ the pump cannot be switched off. Í
Standby is activated after pressing
for 3 seconds.
Software F/G/H/J
BOLUS/ STANDBY MODE
Data Lock:
Quick Reference Guidelines for Infusomat® Space
! If your pump is configured to prime the line press Í
to prime and repeat as necessary. Ensure the pump is disconnected from the patient before priming.
1. From the Main Menu use the to scroll down to the Options Menu. 2. Once the Options Menu is highlighted press to access. 3. Use the
arrow keys to highlight Data Lock and press
4.
Use the and then press
Starting an infusion:
Quick Reference Guidelines for Infusomat® Space
to access.
1. *Enter the Volume To Be Infused “VTBI” by using 2. In the Main Menu
to set Data Lock at Level 1 or Level 2
, and confirm with highlight rate with
.
and select with
.
.
LEVEL ONE will allow the user to start and stop the infusion, change the disposable, and turn the pump off. All other actions will require the code. LEVEL TWO will only allow the user to stop the infusion. All other actions require the code. 5. Activate Data Lock by using the directional keypad to input the 4 digit code: ! Once activated any changes need to be made within 20 seconds. After this, a key symbol appears on the screen. At this point any further changes to infusion parameters will require the code.
Í
3. Select the correct rate with . 4. Press to commence infusion. • Once VTBI and Rate parameters are entered, the Time is automatically calculated. • Moving arrows on the display plus a green operating light indicate the pump is infusing. ! Check the displayed values with the values intended. Start only if values correspond. Í
• * You do not need to enter a Volume To Be Infused if you have a drop sensor connected.
6. De-activate the Data Lock by following steps 1 to 5 and returning the Data Lock level to OFF. PRESSURE SETTINGS/
STARTING THE
DATA LOCK
INFUSION
Quick Reference Guidelines for Infusomat® Space
Change infusion set:
• Interrupt connection to the patient to avoid dosage errors during line change. • Never leave the pump unattended during a line change. • Pump must be powered ON to change the giving set. 1. Press to stop infusion. • Ensure the roller clamp is closed on the giving set; disconnect from the patient.
19981_Infuso_SETUP_Layout 1 19/02/2013 11:12 Page 2
2. Press and to open the pump door. 3. Pull door down to open. 4. Press the green opening lever of the anti-free flow device within the pump.
5. Remove set and replace with a new line. 6. Press pump door firmly to close. ! Check the displayed values with the values intended. Start only if values correspond. Í