Operating Manual
60 Pages
Preview
Page 1
Operating Manual BaehrTec A2000
BaehrTec A2000 Dry technology pedicure device with hand piece
Operating Manual
English– V104 – Status 02/2016
Page 1/60
Operating Manual BaehrTec A2000
1
Table of Contents Page Table of Contents ... 2 Note to users ... 4
1 2 2.1
Symbols ... 5 2.1.1 2.1.2 2.1.3
2.2 2.3
Foreword ... 8 General product description and application purpose ... 9 2.3.1 2.3.2 2.3.3 2.3.4
3
2.1.1 Symbols in these operating instructions ... 5 2.1.2 Nameplate with performance figures... 5 2.1.3 Symbols on the packaging ... 7
Operator requirements... 9 Staff and Patient Protection ... 10 Information on electromagnetic compatibility ... 11 Safety notices ... 11
Before using the device for the first time ... 12 3.1 3.2
4
Scope of delivery ... 12 What to observe before every use! ... 13 Description of device ... 16
4.1
Description control device ... 16 4.1.1 4.1.2 4.1.3 4.1.4 4.1.5 4.1.6 4.1.7
4.2 4.3 4.4 4.5
Front view ... 16 Side view with handpiece holder ... 17 Side view with dust bag lid (closed) ... 18 Side view without dust bag lid and turbine protection filter (open)... 19 Dust bag lid (interior) ... 20 Rear view ... 21 Bottom view ... 22 Description handpiece ... 23 Description non-heating device connecting cable ... 25 Description of suction system ... 26 Colour display ... 28
4.5.1 4.5.2 4.6 4.7 4.8
Work mode... 28 Adjustment mode (Options Menu) ... 29 Home Function (button (17)) ... 31 Memory buttons (buttons (29) - (31)) ... 31 Disconnecting/connecting the dust bag lid to the controller ... 32
4.8.1 4.8.2 4.9
Disconnecting ... 32 Disconnecting ... 32 Changing the filter ... 33
4.9.1 4.9.2 4.9.3 4.10 4.11 4.11.1
Page 2/60
Changing the dust bag... 34 Changing the turbine protection filter ... 35 Changing the filter cassette ... 36 Getting started ... 39 Accessories ... 42 Foot pedal ... 42
English– V104 – Status 02/2016
Operating Manual BaehrTec A2000
5
Servicing and care ... 43 5.1 5.2 5.3 5.4 5.5
Safety notices ... 43 Care (Disinfection)... 44 Guarantee... 44 Recycling/disposal ... 45 Self-help in the event of malfunctions ... 46 5.5.1 5.5.2 5.5.3 5.5.4 5.5.5 instruments)
6 7
Change fuse... 48 Cleaning the handpiece button ... 48 Changing the dampers for the handpiece case ... 50 Easy-Clean clamping mechanism ... 51 Instrument changing aid (insertion and removal aid for very small 54
Technical Data ... 57 Spare parts/accessories ... 60
English– V104 – Status 02/2016
Page 3/60
Operating Manual BaehrTec A2000
2
Note to users
This operating manual must be read through thoroughly before using the device for the first time and the contained instructions and rules must be precisely observed. These operating instructions form part of the device user agreement. By thoroughly reading them, you will be familiarised fully with the functioning and operation of the device and will therefore be able to recognise and avoid operating errors, dangers and damage. Please retain these operating instructions in a safe place with the device.
Please read these operating instructions carefully. Please keep these operating instructions for reference purposes – also in case you wish to clean the device. Please observe all warnings and instructions in these operating instructions and on the device. If you ever clean the device, the power supply must be disconnected fully. Remove the power plug from the safety outlet. When cleaning / disinfecting, please observe the notes described in chapter 5 and its subchapters. Do not place the device near heat sources, such as radiators, air-conditioning units, refrigerators and the like. Please also avoid positioning near to water sources (for example sinks) and / or chemicals. Ensure an appropriately hygienic environment. Place the device on a firm and non-slip base. Avoid positioning on unstable tables, carts or the like. If the device falls, this can cause severe damage and injuries. To guarantee that the device functions reliably, protect it from cold, and also from overheating. Therefore, avoid temperatures below +10° C and above +35° C. Should you use an extension cable, ensure that the overall power supply is not higher than the capacity of the extension cable. Please understand that we cannot accept any liability for accessories of any kind not included in the scope of delivery. This also applies to any consequential damages that may occur. Please avoid:
Touching plug contacts with sharp and / or metallic objects. Placing water, beverages and other liquids close to the device. Leaving children unsupervised with the device. Touching the mains plug with wet and / or damp hands.
Please do not carry out any repairs on the device yourself, as this will result in the cancellation of the guarantee claim. For all repairs, please contact qualified experts who are authorised to carry these out. If necessary, please ask the manufacturer or the distributor of the device (see label on the device).
Page 4/60
English– V104 – Status 02/2016
Operating Manual BaehrTec A2000
On the basis of the risk assessment already implemented we have noticed that (electro) magnetic fields can lead to interference. Therefore, when using the device, please switch off all devices and equipment (mobile telephones, WLAN, etc.) completely that generate or could generate such fields. Should this not be feasible, then the distance between the A2000 and these devices must be at least 50 cm in order to exclude any malfunction form occurring.
Please do not use the device in the following cases, and contact the manufacturer:
If the power cable / insulation shows signs of damage. If the device was exposed to moisture and / or wetness. If the device has been dropped and / or if the device housing is damaged.
In the event of lighting strike and / or surge, the device may be damaged. For this reason, we also recommend installing surge protection and removing the plug during a storm and / or after a long period of non-use in order to protect the device from voltage peaks. Please observe the customary power supply specifications before connecting the device to the mains. 2.1
Symbols
2.1.1
2.1.1 Symbols in these operating instructions
Attention! This symbol indicates a danger to humans or the device. This symbol must always be given the utmost attention. Read the corresponding sections especially carefully and adhere very strictly to the specifications.
This symbol provides especially useful advice and gives additional information on operating the device.
CE sign (Communauté Européenne) with the number of the certification authority. A product bearing this label fulfils the requirements of the corresponding EU guideline (the applicable European Standard). 2.1.2
2.1.2 Nameplate with performance figures
Fig. 1 BaehrTec A2000
English– V104 – Status 02/2016
Page 5/60
Operating Manual BaehrTec A2000
CE sign (Communauté Européenne) with the number of the certification authority. A product bearing this label fulfils the requirements of the corresponding EU guideline (the applicable European Standard).
Application part of type B This application part guarantees protection against electric shock due to the compliance of the leakage currents with standards (Type B).
It is a mandatory requirement that these operating instructions are read and observed before using the device.
Electrical/electronic waste. Devices with this label must be disposed of properly and must not be put in household waste.
This symbol indicates which fuse(s) is/are used in the device.
Protection class II This is a device of protection class II with functional earthing. ON (max) / OFF (min) Indicates how the device should be operated. The following applies to the device: Operating time: 15 minutes (maximum) pause time: 10 min (minimum) These approved operating times correspond to the common work procedure in podiatry/pedicure.
Page 6/60
English– V104 – Status 02/2016
Operating Manual BaehrTec A2000
Functional earthing This symbol indicates that the power supply is earthed (the labelling is directly on the power supply unit).
Manufacturer
2.1.3
2.1.3 Symbols on the packaging
Transport upright (up = in direction of arrow)
Protect from impacts!
Protect from wetness!
Permitted temperature range: -10°C to +40°C
Permitted humidity range: 30% to 85%
800 hPa - 1060 hPa
Permitted air pressure: 800 hPa - 1060 hPa
English– V104 – Status 02/2016
Page 7/60
Operating Manual BaehrTec A2000
2.2
Foreword
Dear customer! We are delighted that you have chosen to purchase this foot care dry technology device. The BaehrTec A2000 boasts technical features that will help to enhance your work. The BaehrTec A2000 is made from many high-quality aluminium, stainless-steel and plastic parts which are also used in sports car and aircraft construction and thus guarantee the ultimate stability and quality. Furthermore, the micro-processor-controlled electronics ensure maximum power and performance from the electronic components. In addition, the BaehrTec A2000 has an electronic readjustment for the handpiece motor, which provides power and performance even in the lowest rotation speed range. Try it for yourself – you'll be amazed. Another highlight is the Easy-Speed concept. Thanks to the instrument illustrations on the controller, setting the engine speed for the instrument currently in use is guaranteed to be child’s play (however, this is no substitute for the user checking that the maximum speed for the instrument currently in use is not exceeded). In addition to the proven Easy-Speed concept, the BaehrTec A2000 also has a colour display which shows you additional information and allows you to change the device settings (e.g. colour background). Contemporary electronic devices are characterised by their energy-saving features. For this reason, we have consciously decided against a standby mode for the BaehrTec A2000. Therefore, when your device is not required, please turn it off using the main switch, which is located in a user-friendly position on the front of the device. Think about the environment and your wallet. The BaehrTec A2000 gives you the following outstanding advantages: - very low weight - very low noise level - simple operating with a high degree of operating safety - high performance, perfectly adjusted to the working conditions - high robustness (for mobile use) - high and long reliability - high energy-saving potential (no standby mode) - Button on handpiece The BaehrTec A2000 foot care device has been made and tested in accordance with strict quality criteria, and it complies with Directive 93/42 EEC for medical devices. We hope your new device brings you plenty of enjoyment and we wish you every success in operating it. Your Gustav Baehr GmbH
Page 8/60
English– V104 – Status 02/2016
Operating Manual BaehrTec A2000
2.3
General product description and application purpose
The BaehrTec A2000 foot care device is intended for use in medical foot care. It must only be used by trained professionals. The BaehrTec A2000 foot care device moves rotating instruments (cutters, grinders etc.). These can be used to strip away hard skin, calluses, nails etc. and remove corns. In detail, the A2000 is intended for the following: • • • • • • •
for cleaning and grinding the nail fold and removing ingrown nails to smooth and strip away mycotic and non-mycotic nails to polish non-mycotic nails where necessary to remove deep callosities or clavi using the hollow cutter grinding and smoothing the areas around the digit if these are macerated or calloused to smooth plantar soles with the twister or the cutting grinder Drill through the nail with a round or hollow drill with grinders for clavi or subungual haematoma to roughen the nail in preparation for brace correction for preparation in the event of panaritium to remove severe plate-like callosities
• • •
Other types and fields of application are carried out at your own risk, and may conceal dangers. No form of misappropriation is permitted. Improper use may lead to damage to persons or objects. The manufacturer cannot be held responsible for damages caused by improper use, unqualified staff or incorrect operation. All warranty claims are void following improper use or opening the device.
WARNING: This device may not be altered without the permission of the manufacturer.
2.3.1
Operator requirements
This device must only be used by trained and instructed podiatrists, medical chiropodists, doctors or persons in related occupational categories. They must be familiar with the appropriate working method and have a relevant qualification. The operator is obliged to/must ensure that
• • •
only fault-free and flawless work equipment is used protects himself, the patient and others from dangers contamination through the device is avoided
English– V104 – Status 02/2016
Page 9/60
Operating Manual BaehrTec A2000
2.3.2
Staff and Patient Protection It is essential that you read this section with the utmost care! It contains important information on protecting yourself, others and the device from damage!
• • • • • • • • •
• • •
Only use high-quality rotating instruments with standardised shaft (diameter 2.35 mm) from the Baehr product range. Please observe the specific instructions of rotating instruments when using them. Above all, observe the manufacturer's information on maximum speeds, cleaning, disinfection and sterilisation. Disinfect, clean and sterilise the instruments after each use. Only use cleaned, disinfected and sterilised instruments for each change of patient to avoid a possible transfer of germs to the next patient. Disinfect the handpiece after each use and before each change of patient. (Please ensure that no disinfecting agent or other liquid can enter the device). Disinfect all parts of the device that could have come into contact with contaminated patients after each use and before each change of patient. (Please ensure that no disinfecting agent or other liquid can enter the device). The operating staff must wear protective gloves as well as eye, mouth and nose protection when using the device. When using the device, the staff must take care that neither hair nor any other loose objects such as wipes, cotton wool or the like can enter the area of the rotating tools. A hair net must be worn where necessary. The operating staff must bear in mind that the particles that are removed when working with the rotating tools may chip. Open and untreated wounds on the patient which are in the direct vicinity of the working area should therefore be covered in a sterile way in order to protect them from any splintering particles. The device must be maintained and cleaned according to the instructions before and after long pauses in use. Only accessories authorised for use with the device may be used. The national statutory provisions must be observed during use, in particular:
the currently applicable work regulations the currently applicable accident prevention measures To guarantee constant readiness for operation and preservation of value, the prescribed care work and maintenance services must be performed. The device must only be repaired using replacement parts approved by the manufacturer and in accordance with the manufacturer’s instructions. The recommended maintenance services (after notification, but at the latest within 24 months) and inspection and repair work must only be performed by the manufacturer. This device must not be modified without the permission of the manufacturer.
Page 10/60
English– V104 – Status 02/2016
Operating Manual BaehrTec A2000
2.3.3
Information on electromagnetic compatibility
We would like to point out that due to EN 60601-1 on the electromagnetic compatibility of electromedical devices that:
•
medical, electric devices are subject to particular precautionary measures and so must be operated according to the requirements of these operating instructions.
•
portable and mobile high-frequency communication facilities may affect the functionality of electrical devices.
•
in order to comply with the EMV requirements of EN 60601-1, only original feed lines, accessories and spare parts may be used. Only use manufacturer approved mains cables to operate the device. If you require a new cable, please contact the manufacturer. Operating the device using a different cable is not permitted.
2.3.4
Safety notices
The device is not authorised for operation in potentially explosive areas. Before every application, the operator must make sure of the functional safety and proper condition of the device. Improper handling, maintenance and care may lead to premature deterioration and malfunctions. This can result in a reduced product life. Therefore, please clean and care for the device regularly and properly and have it sent for servicing regularly (observe service indicator or at the latest within 24 months)! Damaged functional parts can cause damage or injuries to persons or objects. Furthermore this may result in (even greater) damage to your device. Stop working immediately and disconnect the device from the mains when functional parts are damaged and contact customer service. Electromagnetic fields may affect the functionality of implanted systems (e.g. pacemakers). Please ask your patients before beginning the treatment whether they have such a system. Due to the complex interactions between electric devices and mobile telephones, it is possible that mobile phones that are switched on may affect the device, even though the device fulfils the applicable requirements relating to electromagnetic fields. Do not use your mobile phone while working and also inform your patients that their mobile phones should be switched off during treatment. Remove electronic devices, which could cause interaction (e.g. hearing aids, etc.) while using the device. Otherwise, the distance between the device and the upper part of the body of the person to be treated must be at least 50 cm, in order to rule out any malfunctions. If you put the handpiece down there is risk of injury when reaching for the handpiece. In the event of injuries from used instruments, this may lead to infections. Please take care when setting down the handpiece so that you do not injure yourself.
English– V104 – Status 02/2016
Page 11/60
Operating Manual BaehrTec A2000
3
Before using the device for the first time
3.1
Scope of delivery
Before first use, you should check that all items have been delivered. Items delivered: 1 x A2000 controller incl. handpiece (these are firmly connected to each other) 1 x operating instructions 1 pc. dust bag microfibre (1 bag already installed) Art-No.: 21185 1 x turbine protection filter (1 filter is already inserted for you) Dimensions: (~ 93 x 78 mm) 1 x sound-insulating fleece (1 filter is already inserted for you) Dimensions: (~ 48 x 57.5 mm) 1 x coarse filter fleece (1 filter is already inserted for you) Dimensions: (~ 48 x 57.5 mm) 1 x carbon filter fleece (1 filter is already inserted for you) Dimensions: (~ 48 x 57.5 mm) 1 x ultra-fine filter fleece (1 filter is already inserted for you) Dimensions: (~ 48 x 57.5 mm) 2 x carbon filter foam (2 filters are already inserted for you) Dimensions: (~ 48 x 57.5 mm) 1 x power cable with straight connector 20970006 1 x Easy-Clean tool kit for BaehrTec A2000 Art No.40285 1 x damper for handpiece case (set) Art-No.: 40286 For more information, please see the section on "Changing the dampers for the handpiece case". If your delivery is incomplete, please inform us of this immediately. Please keep the delivery box along with any packaging accessories. The packaging was developed for this device and provides the best possible protection during transportation. Therefore, please use the original packaging should you send your device in for servicing. There shall be no entitlement to guarantee for any damages that are caused due to inadequate packaging during transportation.
Page 12/60
English– V104 – Status 02/2016
Operating Manual BaehrTec A2000
3.2
What to observe before every use! It is essential that you read this section with the utmost care! It contains important information on protecting yourself, others and the device from damage!
The description used in this section BaehrTec A2000 refers both to the control device and the handpiece. Before use, please check whether the type of current and the mains voltage of the power source are suitable for using the device. Information on the type of current and the mains voltage can be found on the nameplate on the controller. When setting up the device, ensure that it is placed on a level base, that it cannot fall down and that the extracted air can escape easily. Ensure that the BaehrTec A2000 remains out of reach of children. Do not expose the device to direct heat sources (heaters, strong sunlight etc.). Ensure that the power cable is not damaged due to squeezing, snapping or rubbing on sharp edges. If you notice any damage to the mains cable, please immediately stop working with your device, turn off the main switch and immediately remove the safety plug from the safety socket. To be able to work with the device again, please order a new power cable. To be able to work with the device again, please order a new power cable. Only use manufacturer approved mains cables to operate the device. If you require a new cable, please contact the manufacturer. Operating the device using different cables is not permitted. Never operate the device with a damaged power cable. Your device has a detachable connector plug mechanism (see section "rear view") Please ensure that the device is positioned so that a disconnection is unlikely at all times. Do not work with the BaehrTec A2000 in damp areas such as saunas or swimming complexes. Wetness and moisture on the controller can lead to dangerous current leaks, which poses the risk of an electric shock. Disconnect the device immediately from the socket. Please send the device in for service with a description of the defect. Avoid significant differences in temperature. This can cause dampness (condensation). Protect the BaehrTec A2000 from frost. The device must be switched off and disconnected from the mains before carrying out any cleaning/maintenance work. Disconnect the safety plug from the mains immediately should you discover any damage or malfunctions of the device. The manufacturer accepts no liability for damages to objects, animals or persons which are caused by incorrect operation of the BaehrTec A2000.
English– V104 – Status 02/2016
Page 13/60
Operating Manual BaehrTec A2000
Please ensure that the BaehrTec A2000 and the instruments are always in an immaculate hygienic condition, in order that they do put your own health or that of others at risk. Find out more in the section "Care (disinfection)" of the BaehrTec A2000. Rings or jewellery worn while working with the device may cause scratches on the handpiece. Such damage is excluded from the guarantee. Whilst working with the device, you should avoid wearing jewellery. Please only work with the suction turned on in order to suction off any pathogen-containing dust and to prevent the handpiece motor or the handpiece from becoming warm. Never submerge the device in liquids, and do not suck up any liquids. If you suck up any cotton wool, paper or the like, this may cause the suction openings in the handpiece to become blocked. This may heavily affect the suction power. If the symbol for a filter change appears on the display, it is essential that you change the dust bag (the display can only function reliably if all filters are correctly applied. Therefore, only work with the device if all filters are applied). If you do not do this, the device will switch automatically after approximately half a minute to suction level 3 to prevent any damage to your device (please refer to the section "Filter change" for more information on how to change the dust bag and the filter). Never work with the device if the dust bag or the filter are not inserted, as in such cases, the device may be damaged and the guarantee shall expire.
Page 14/60
English– V104 – Status 02/2016
Operating Manual BaehrTec A2000
Warning – risk of injury! As shown in the illustration, you should avoid "pulling motions" during your work, as this may cause the instruments to slip out. Please take care not to exert too much pressure when using the instrument (burns on skin).
Fig. 2 Incorrect operation
Fig. 3 Correct operation
English– V104 – Status 02/2016
Page 15/60
Operating Manual BaehrTec A2000
4
Description of device
4.1
Description control device
4.1.1
Front view
Fig. 4 (1) Controller housing Main switch: (2) "ON" and (3) "OFF" with corresponding (4) LED (5) – (12) Push buttons for instrument speeds (6,000 – 40,000 rpm) with the corresponding (5a) – (12a) LEDs (13) Push button to switch handpiece motor on and off Push buttons handpiece speed: (14) + and (15) - (1,000 rpm) (16) Push button for instrument direction right/left (17) Push button "Home" (18) Push button "Options menu" (19) Colour display (20) Push button for operational readiness of suction (21) - (26) Push buttons for "Suction level" (Levels 1-6) Push buttons for "Change of suction level": (27) + and (28) (29) - (31) Push buttons "Memory"
Page 16/60
English– V104 – Status 02/2016
Operating Manual BaehrTec A2000
4.1.2
Side view with handpiece holder
Fig. 5 (32) Handpiece holder
English– V104 – Status 02/2016
Page 17/60
Operating Manual BaehrTec A2000
4.1.3
Side view with dust bag lid (closed)
Fig. 6 (33) Dust bag lid with handle (34) Handpiece outlet (35) Screws for handpiece outlet Must only be unfastened by the manufacturer (36) Suction hose
Page 18/60
English– V104 – Status 02/2016
Operating Manual BaehrTec A2000
4.1.4
Side view without dust bag lid and turbine protection filter (open)
Fig. 7 (37) 4x magnet holder (38) Filter grille
English– V104 – Status 02/2016
Page 19/60
Operating Manual BaehrTec A2000
4.1.5
Dust bag lid (interior)
Fig. 8 (39) Receptacle for dust bags (40) Mounting bolts for magnet holder (41) Rubber seal for dust bag lid (42) Dust bag
Page 20/60
English– V104 – Status 02/2016